Hall Surgical
Hall PowerPro Pneumatic Handpieces and Attachments
MicroFree Battery Handpiece Instruction Manual Rev AE
Instruction Manual
29 Pages
Preview
Page 1
Hall® MicroFree™ Battery Handpiece Instruction Manual High Speed Drill: PRO8000SB Medium Speed Drill: PRO8100SB Sagittal Saw: PRO8200SB Oscillating Saw: PRO8300SB Reciprocating Saw: PRO8400SB
Proprietary Information This manual contains information deemed proprietary to CONMED Corporation. The information contained herein, including all of the designs and related materials, is the sole property of CONMED and/or its licensors. CONMED and/or its licensors reserve all patent, copyright and other proprietary rights to this document, including all design, manufacturing methodology and reproduction. This document, and any related materials, is confidential and is protected by copyright laws and shall not be duplicated, transmitted, transcribed, stored in a retrieval system, or translated into any human or computer language in any form or by any means, electronic, mechanical, magnetic, manual or otherwise, or disclosed to third parties, in whole or in part, without the prior express written consent of CONMED. CONMED reserves the right to revise this publication and to make changes from time to time in the contents hereof without obligation to notify any person of such revision or changes, unless otherwise required by law.
Linvatec, Hall, and SmartGuard are trademarks or registered trademarks of CONMED Corporation.
© CONMED Corporation 2018. All Rights Reserved.
Record the Model and Serial Numbers of the handpieces, and date received. Retain for future reference.
Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No. Handpiece Model No.
Serial No. Serial No. Serial No. Serial No. Serial No. Serial No.
Date Date Date Date Date Date
Table of Contents 1.0
INTRODUCTION 1.1 Operating Principle 1.2 Indications for Use 1.3 Intended Use 1.4 Contraindications 1.5 Warnings and Precautions 1.5.1 Warnings 1.5.2 Precautions 1.6 Environmental Directives 1.7 Product Photographs and Drawings 1.8 Symbol Definitions 1.8.1 Product Symbols 1.8.2 Warnings and Information Symbols
1 1 1 1 2 2 3 3 3 4 4 4
2.0
SYSTEM INSTALLATION AND OPERATION 2.1 Product Description 2.1.1 Handpiece Descriptions 2.2 Assembly/Installation Instructions 2.2.1 HALL Small Bone Lithium Power Unit Installation 2.2.2 Hall Bur Guard Assembly Instructions 2.2.3 Bur Installation and Removal 2.2.4 Reciprocating Saw Blade and Rasp Installation and Removal 2.2.5 Sagittal Saw Blade Installation and Removal 2.2.6 Oscillating Saw Blade Installation and Removal 2.3 Operating Instructions 2.3.1 Drill Operation 2.3.2 Reciprocating Saw Operation 2.3.3 Sagittal Saw Operation 2.3.4 Oscillating Saw Operation 2.4 Preoperative Functional Test 2.4.1 Bur Guard Functional Test 2.4.2 Medium Bur Guard (1375-012/1375-012P) Functional Test 2.4.3 Handpiece Functional Test
6 6 7 7 7 8 9 9 10 11 11 11 11 12 12 12 12 13
3.0
MAINTENANCE 3.1 Cleaning Information 3.1.1 Warnings, Precautions and Notes 3.1.2 Manual Cleaning Instructions 3.1.3 Automated Cleaning Instructions 3.2 Disinfection Information 3.2.1 Manual Disinfection Instuctions 3.2.2 Automated Disinfection Instuctions 3.3 Sterilization Information 3.3.1 Warnings, Precautions and Notes 3.3.2 Inspection Recommendations 3.3.3 Sterilization Instructions 3.4 Troubleshooting 3.5 Maintenance Schedule
14 14 14 14 15 15 15 16 16 16 16 17 18
4.0
TECHNICAL SPECIFICATIONS 4.1 Product Technical Specifications 4.1.1 Handpieces 4.2 Product Environmental Requirements 4.2.1 Environmental Technical Specifications 4.2.2 Electromagnetic Requirements 4.3 Handpieces, Attachments and Accessories
19 19 21 21 22 24
5.0
CUSTOMER SERVICE 5.1 Assistance and Repair
25 I
1.0
INTRODUCTION
It is recommended that personnel study this manual before attempting to operate, clean, or sterilize this or associated equipment. The safe and effective use of this equipment requires the understanding of and compliance with all warnings, precautionary notices, and instructions marked on the product, and included in this manual. This equipment is designed for use by medical professionals completely familiar with the required techniques and instructions for use of the equipment.
1.1
Operating Principle
The MicroFree Battery Handpieces are powered by the Hall® Small Bone Lithium Power Unit (L3500SB) to provide force to the accessory (burs, blades, drill bits or attachment) for bone cutting, shaping, drilling and soft tissue resection with the drills, or bone cutting, shaping, or resection with the saws. The MicroFree Battery Handpieces are controlled by an activation lever on the handpiece. Consult the associated HALL Small Bone Lithium Power Unit (L3500SB) instruction insert prior to operating this equipment.
1.2
Indications for Use
The MicroFree Battery Handpieces (PRO8000SB, PRO8100SB, PRO8200SB, PRO8300SB and PRO8400SB) and their accessories perform cutting of soft tissue and bone. The fields of application include: orthopedic, arthroscopic, and plastic/reconstructive procedures.
1.3
Intended Use
Same as Indications for Use above.
1.4
Contraindications
None
1
1.5
Warnings and Precautions
Do not bypass this section. It contains warnings and precautions that must be thoroughly understood before operating any of the equipment. Lack of understanding or adherence to these warnings and precautions may result in injury or even death to the patient. The words WARNING, PRECAUTION, and NOTE carry special meanings and they must be read carefully. WARNING: A warning contains critical information regarding serious adverse reactions and potential safety hazards that can occur in proper use or misuse of the equipment. Failure to observe the information or procedures presented in a Warning may result in injury or other serious adverse reactions to the patient and/or surgical staff. PRECAUTION: A precaution contains instructions for any special care to be exercised by the practitioner for the safe and effective use of the equipment. Failure to observe the information or procedures presented in a Precaution may result in damage to the equipment. NOTE: A note is added to provide additional focused information. This information has no critical effect on the patient or equipment.
1.5.1 Warnings 1. Eye protection is recommended when operating equipment. Eye injury may result. 2. It is the surgeon’s responsibility to be familiar with the appropriate surgical techniques prior to use of the equipment and its associated accessories. 3. Do not use equipment if, upon receipt, package is opened, damaged, or shows any signs of tampering. 4. Do not use equipment in the presence of flammable anesthetics, gases, disinfecting agents, cleaning solutions, or any material susceptible to ignition due to electrical sparking. 5. Do not use sterile equipment beyond the expiration date listed on the label. Sterility of the product cannot be assured beyond the expiration date. 6. Handpieces are supplied non-sterile. Clean and sterilize prior to each use. 7. Do not contact the moving parts on the handpieces. Injury to the operator may occur. 8. Continually check handpiece for overheating. If overheating is sensed, immediately discontinue use and return equipment for service. Overheating of the blade or bur may cause damage to the blade or bur and may cause burns or thermal necrosis. 9. While handpiece is not in use do not place on patient/surgical drapes. Place handpiece on mayo stand. 10. Do not immerse the equipment in fluids. Immersion may render the device inoperable. 11. Failure to follow the specified service interval could result in reduced instrument performance or overheating of the handpiece. Overheating can lead to possible burn injury to the patient or medical personnel. Rotation of handpiece usage per day will assist with proper performance. (Refer to section “3.5 Maintenance Schedule”). 12. Do not attach, insert or remove accessories or attachments while the handpiece is operating. Injury to the operator and/or damage to the equipment may occur. Place the HALL Small Bone Lithium Power Unit safety mechanism in the “safe” position prior to installation or removal of items. 13. Avoid contact with cutting tip of blade or bur when locking into handpiece. Tips are sharp and may cause injury. 14. After use, blades, burs and tubing sets may be a potential biohazard and should be handled and disposed in accordance with acceptable medical practice and applicable local and national requirements. 15. Do not use burs for plunge cutting. Injury or damage may occur. 16. Disposable blades and burs are supplied sterile and are for single-use only. Do not re-sterilize or reuse. The ability to effectively clean and re-sterilize these single use devices has not been established and subsequent re-use may adversely affect the performance, safety and/or sterility of the devices. 17. When using the MicroFree Sagittal Saw (PRO8200SB), placing excessive bending or twisting force on the sagittal saw blade may cause the collet to open and release the saw blade. Do not use saw blades to pry, remove bone grafts, or as a leverage point. Patient or user injury could occur. 18. Bur guards should always be checked before use. To reduce the risk of injury, prior to surgery, spin the bur guard on a bur. If the bur guard spins freely, the bearing is still good. Otherwise, the bur guard must be sent for repair immediately. DO NOT USE. Overheating can occur if bur guard bearings are worn or not kept clean. 19. Do not operate the drills without the appropriate bur guard and collet locked. Always use a bur of the proper length. The tip of the bur guard should cover the safe line on the bur, if applicable. Without the stabilization that the proper guard provides, the bur can break and be propelled with great force. 20. Loss of use of the MicroFree Handpiece due to battery depletion does not create a hazardous situation. 2
1.5.2 Precautions 1. 2. 3. 4.
United States Federal law restricts sale of this device to or on the order of a physician. This device should only be used in compliance with its intended use. Handle all equipment carefully. If any equipment is dropped or damaged in any way, return it immediately for service. Use only associated CONMED approved equipment and accessories. Using unapproved accessories may result in improper operation, and may result in non-compliance to medical standards. 5. The warranty becomes void and the manufacturer is not liable for direct or resulting damage if: • The device or the accessories are improperly used, prepared or maintained; • The instructions in the manual are not adhered to; • Non-authorized persons perform repairs, adjustments or alterations to the device or accessories. 6. There are no user-serviceable parts inside. No modification of this equipment is allowed. 7. Prior to each use, perform the following: • Ensure all accessories are correctly and completely attached. (Refer to Section “2.2 Assembly/Installation Instructions”). • Perform the required Preoperative Functional Tests for the equipment and accessories. (Refer to Section “2.4 Preoperative Functional Test”). 8. Clean and sterilize all equipment and associated accessories according to instructions for use. (Refer to section “3.1 Cleaning Information”, section “3.2 Disinfection Information” and section “3.3 Sterilization Information”). 9. Handpieces are factory sealed. Do not disassemble as this may void the warranty. 10. Always inspect for bent, dull or damaged blades or burs before each use. Do not attempt to straighten or sharpen. Do not use if damaged. 11. After each use, thoroughly clean the handpiece, attachments and accessories. (Refer to Section “3.0 MAINTENANCE”). 12. Do not stall handpieces, damage can occur. 13. Do not operate the oscillating saw or reciprocating saw without a blade locked securely in place. Damage to the handpiece may occur. 14. Never lock the drill collet without a bur inserted. Damage to the collet may result. 15. Different saw blade designs and blade features can produce varying cutting efficiencies. For example: shorter finetoothed blades will cut hard bone more efficiently than shorter coarse-toothed blades.
1.6
Environmental Directives
WEEE Directive [2002/96/EC] on Waste Electrical and Electronic Equipment. This statement only applies to European countries with regard to the Waste Electrical and Electric Equipment (WEEE) European Directive. The WEEE symbol on the product or its packaging indicates that this product must not be disposed of with other waste. Instead, it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for the recycling of Waste Electrical and Electronic Equipment. The separate collection and recycling of your waste equipment at the time of disposal will help conserve natural resources and ensure that it is recycled in a manner that protects human health and the environment. For more information about where you can drop off your medical equipment at the end of its useful life for recycling, please contact CONMED.
1.7
Product Photographs and Drawings
The pictures in this manual are for reference only. Items shown may not represent the actual product. However, procedural steps are identical, unless otherwise specified. When necessary, the actual pictures will be represented.
3
1.8
Symbol Definitions
1.8.1 Product Symbols SAFE RUN
Safe Mode
LOCK
Run Mode
Collet Lock Position Indicates Handpiece should not be Immersed in any fluid.
1.8.2 Warnings and Information Symbols
REF
Catalog Number
SN
Serial Number
Manufacturer
Date of Manufacture
Consult Instructions for Use
Refer to Instruction Manual/Booklet (for critical safety instruction)
Caution
DEHP Symbol DEHP
Authorized Representative in the European Community
CE Mark of Conformity
Prescription Only: Federal Law restricts this device to sale by or on the order of a physician
No User Service Recommended. Refer servicing to qualified CONMED service personnel
Non Sterile
STERILE EO
2
STERILE
Sterile - Sterilized Using EO
STERILE
R
Sterile Sterile - Sterilized Using Irradiation
Do Not Steam Sterilize
Do Not Sterilize
Do Not Resterilize
Do Not Reuse (for Single Use Only)
Do Not Use Oil
Do Not Use for Plunge Cutting
Eye Protection Required
Biohazard Risk
STERILIZE
Do Not Immerse
QTY
Quantity
Type B Applied Part
Type BF Applied Part
UL Classification Mark
UL Recognized Components
4
~
Rating Fuse
Fuse Location
Alternating Current
Protective Earth Ground
Equipotentiality (Equipment Potential)
Non-Ionizing Electromagnetic Radiation (RF Symbol)
Temperature Limitation
Humidity Limitation
Atmospheric Pressure Limitation
Use by Date
Fragile
This Side Up
Do Not Use if Package is Damaged
Keep Dry
Warning: Corrosive Substance
Warning: Electrical Hazard/High Voltage
Waste Electrical and Electronic Equipment (WEEE) Symbol. Regarding European Union end-of-life of product, indicating separate collection for electrical and electronic equipment
Recycle. Batteries contain materials which must be recycled or disposed of properly. The disposal of batteries as municipal waste is prohibited. Dispose or recycle in accordance with your local, state and governmental regulations. In the U.S. call 1-800-237-0169, or outside the U.S. contact your local CONMED representative for additional information on battery disposal or recycling.
5
2.0
SYSTEM INSTALLATION AND OPERATION
2.1
Product Description
2.1.1 Handpiece Descriptions 1. Bur Lock Collar - Rotate to lock or unlock a bur. 2. Blade Collet - Holds the blade in place. 3. Blade Lock Collar - Rotate to open or lock the blade collet. 4. Collet Lock Mechanism - Locks a blade into place. 5. Activation Lever - Depress to operate the handpiece. 6. HALL Small Bone Lithium Power Unit Connector - The HALL Small Bone Lithium Power Unit (REF L3500SB) connects here to provide power to the handpiece. 7. Collet Indexing Lock - locks collet at 45° increments.
NOTE: Do not allow saline to enter into the connector and HALL Small Bone Lithium Power Unit. Damage may occur to the handpiece.
6
2.2
Assembly/Installation Instructions
2.2.1 HALL Small Bone Lithium Power Unit Installation To attach the HALL Small Bone Lithium Power Unit: 1. Put HALL Small Bone Lithium Power Unit in “SAFE” mode. 2. Ensure pins and sockets on HALL Small Bone Lithium Power Unit and handpiece are completely dry prior to connecting. 3. Insert the handpiece into the HALL Small Bone Lithium Power Unit. 4. Push together until fully seated and a click is heard and/or felt. NOTE: Do not force the handpiece into the HALL Small Bone Lithium Power Unit. This may bend the pins and damage the handpiece. To remove the HALL Small Bone Lithium Power Unit: 1. Put HALL Small Bone Lithium Power Unit in “SAFE” mode. 2. Pull HALL Small Bone Lithium Power Unit and handpiece apart.
2.2.2 Hall Bur Guard Assembly Instructions 2.2.2.1
Bur Guards
These instructions apply to following Guards: • Medium Bur Guard (1375-012/1375-012P) • Long Bur Guard (1375-011/1375-011P) • Extra-Long Bur Guard (1375-023/1375-023P) To install a bur guard: 1. Slide the appropriate bur guard over the end of the drill. 2. Ensure it is seated completely.
Prior to installing a bur, but only while using the 1375-012/1375-012P Medium Bur Guard, it is recommended that a SmartGuard® Protector Sleeve (1375-112) be attached. Reference the associated insert that comes packaged with the SmartGuard for additional information. NOTES: • SmartGuard Protector Sleeves are designed to be attached to the Medium Bur Guard during surgical applications. They are intended to be used as a heat-sensing and insulating device that changes color from PURPLE (SAFE to use bur guard) to PINK (REPLACE bur guard) to indicate over-heating of the attached bur guard. • Overheating might occur if bearings are worn or are not kept clean. If overheating occurs while a SmartGuard Protector Sleeve is attached to the Medium Bur Guard, the SmartGuard Protector Sleeve will change color from PURPLE to PINK. Immediately discontinue use of the bur guard to prevent possible injury to the patient and/or operator. Replace the bur guard and return original bur guard for service. • Under prolonged use or intense cutting, the tip of the SmartGuard Protector Sleeve may turn pink while the rest of the sleeve retains its purple color. If more than 50% of the SmartGuard Protector Sleeve turns pink, discontinue use immediately to prevent possible injury to the patient and/or operator. Replace the bur guard and return original bur guard for service.
7
To attach a SmartGuard Protector Sleeve: 1. Remove the SmartGuard Protector Sleeve from its sterile packaging within the sterile field. 2. Press-fit the SmartGuard Protector Sleeve onto the end of the associated bur guard until fully seated (see figure below).
3. After the SmartGuard Protector Sleeve is attached, install a bur. To remove the SmartGuard Protector Sleeve: 1. Remove the bur.
2. Grasp the tabs of the SmartGuard Protector Sleeve between the thumb and forefinger and pull. 3. Discard the used SmartGuard Protector Sleeve properly.
2.2.3 Bur Installation and Removal Note: Always ensure HALL Small Bone Lithium Power Unit is in “SAFE” mode prior to installing or removing bur from drill. To install a bur: 1. Select a proper length bur for the guard. 2. Twist the bur lock counter-clockwise to the unlocked position.
3. Insert the bur to the safe line or until the bur seats completely if no safe line is observed. 4. Lock the bur in place by twisting the bur lock clockwise until the red indicator dots are aligned.
5. Always pull on the bur to ensure bur is fully locked in position. To remove bur: Twist the bur lock counter-clockwise to the unlocked position and remove bur.
8
2.2.4 Reciprocating Saw Blade and Rasp Installation and Removal Note: Always ensure HALL Small Bone Lithium Power Unit is in “SAFE” mode prior to installing or removing blade from Reciprocating Saw. To insert a blade: 1. Twist the blade locking collar counter-clockwise and insert the shank of the blade or rasp into the collet. Ensure the blade or rasp is fully seated.
2. Twist the blade lock collar clockwise and tighten securely.
Round shank blades or rasps can be seated at any position within a 360° range. Flat blades must be aligned with the slot in the collet.
To remove blades or rasps: Twist the blade lock collar counter-clockwise and remove the blade.
2.2.5 Sagittal Saw Blade Installation and Removal Note: Always ensure HALL Small Bone Lithium Power Unit is in “SAFE” mode prior to installing or removing blade from Sagittal Saw. To Attach Blades: 1. Depress the collet lock mechanism to open the collet. Position the blade on the pins inside the collet.
9
2. Depress the opposite side of the handpiece collet area to lock the blade into position. An audible click should be heard.
3. Blades may be positioned at 45° intervals within a 180° arc.
4. Verify the blade is securely attached by activating the handpiece. To remove saw blades: Depress the collet lock mechanism and remove the blade.
2.2.6 Oscillating Saw Blade Installation and Removal Note: Always ensure HALL Small Bone Lithium Power Unit is in “SAFE” mode prior to installing or removing blade from Oscillating Saw. To attach saw blades: 1. Ensure the blade lock collar is in the open position until the arrows are no longer aligned.
2. Check collet pins for debris, damage, or wear. Do not use if noted. 3. Position the blade on the pins inside the collet.
10
4. Twist the blade lock collar and align the arrows in the “LOCK” position to secure the blade.
5. Blades may be positioned at 45° intervals within a 360° circumference of the collet.
6. Verify the blade is securely attached by briefly activating the handpiece. To remove saw blades: Twist the blade lock collar until the arrows are no longer aligned in the “LOCK” position and remove the blade.
2.3
Operating Instructions
PRECAUTION: Discontinue use if overheating is sensed in handpiece during operation.
2.3.1 Drill Operation NOTE: Operation of a drill (PRO8000SB, PRO8100SB) without a bur locked in place is considered a failure mode. If the handpiece is activated for several seconds without a bur in place, the handpiece will not run. Operation cannot be initiated until a bur is locked in place. To operate the handpiece after the guard and bur are attached: 1. Place the “SAFE/RUN” Slide on the HALL Small Bone Lithium Power Unit in the “RUN” position. 2. Depress the handpiece Activation Lever.
2.3.2 Reciprocating Saw Operation To activate the handpiece: 1. Place the “SAFE/RUN” Slide on the HALL Small Bone Lithium Power Unit in the “RUN” position. 2. Depress the handpiece Activation Lever. 3. Place the HALL Small Bone Lithium Power Unit back in the “SAFE” position. 4. Retighten the blade locking collar.
2.3.3 Sagittal Saw Operation To activate the handpiece: 1. Place the “SAFE/RUN” Slide on the HALL Small Bone Lithium Power Unit in the “RUN” position. 2. Depress the handpiece Activation Lever. To index the head: 1. The sagittal saw head can be indexed at 45 degree increments. 2. Pull the collet indexing lock back. 3. Rotate the head to the desired 45 degree increment. 4. Release the collet indexing lock. 11
2.3.4 Oscillating Saw Operation To activate the handpiece: 1. Place the “SAFE/RUN” Slide on the HALL Small Bone Lithium Power Unit in the “RUN” position. 2. Depress the handpiece Activation Lever.
2.4
Preoperative Functional Test
Prior to operating the MicroFree handpieces, perform the following preoperative tests to verify proper functioning. Any operating difficulties should be reported to your local Sales Representative or CONMED Customer Service.
2.4.1 Bur Guard Functional Test NOTE: This test excludes the Medium Bur Guard, 1375-012/1375-012P. 1. Remove the guard from the handpiece and insert a bur into the nose of the guard. 2. While holding the bur, spin the guard. The guard should spin freely around the bur shaft without resistance.
3. Attach the guard and bur to the handpiece. 4. Operate the handpiece for approximately 30 seconds, then stop the handpiece. Once the bur has stopped moving, carefully feel the end of the guard where the guard encases the shaft of the bur for overheating. If overheating is noticed, return guard for service.
2.4.2 Medium Bur Guard (1375-012/1375-012P) Functional Test Prior to testing, remove the guard from the handpiece. Check for worn bearings: 1. Insert a bur into the nose of the guard. 2. While holding the bur, spin the guard. The guard should spin freely around the bur shaft without resistance.
3. Attach the bur guard to the drill. 4. Attach a SmartGuard Protector Sleeve (1375-112). Reference the associated insert available on the CONMED website. 5. Select a proper length bur for the guard and install a bur. 6. Operate the handpiece for at least 30 seconds. Stop the handpiece and carefully feel the end of the guard where the guard encases the shaft of the bur for overheating. If overheating is noticed, return the guard for service. Or, if the SmartGuard Protector Sleeve changes color from PURPLE to PINK, immediately discontinue use of the bur guard. Replace the bur guard and return original bur guard for service.
12
2.4.3 Handpiece Functional Test Before operating the handpiece, check for: • any loose or missing parts • any physical damage • movable parts that do not move freely Performance Testing: 1. Assemble the appropriate guard, bur and/or blade according to the assembly instructions. 2. Operate the handpiece for one minute according to the operating instructions. Monitor the handpiece for any of the following: • excessive noise • excessive vibration • handpiece or attachment are hot to the touch during the test procedure or during surgical use.
13
3.0
MAINTENANCE
3.1
Cleaning Information
3.1.1 Warnings, Precautions and Notes 1. Follow universal precautions for protective apparel when handling and cleaning contaminated instruments. 2. After use (post-operatively) do not allow soiled/contaminated devices to dry prior to pre-cleaning and disinfecting. Clean instruments within 30 minutes after use to minimize the potential of blood and debris drying. Consider covering the re-usable device with a cloth saturated with distilled water to avoid drying of soil. 3. Never clean equipment in an ultrasonic cleaner. 4. Always disconnect the handpiece from the HALL Small Bone Lithium Power Unit prior to cleaning. 5. Always detach accessories from equipment prior to cleaning. 6. Never clean handpieces with bleach, chlorine-based detergents, liquid or chemical disinfectants, or any products containing sodium hydroxide (such as, INSTRU-KLENZ or Buell Cleaner). These products degrade the anodized aluminum coating and may result in reduced handpiece reliability. 7. To prevent corrosion, avoid contact with agents containing iodine or chlorine.
3.1.2 Manual Cleaning Instructions 1. Remove the attachments, HALL Small Bone Lithium Power Unit and disposables from the handpieces. 2. Thoroughly scrub the handpiece and attachments with a clean, soft brush dampened with a mild, pH-balanced detergent. Ensure the blade locking collet on the oscillator and reciprocator handpieces is in the open position. Remove all traces of blood, debris and stains. 3. To clean the cannulated section of the handpiece and attachment, feed the wire end of a cleaning brush through the cannulation of the handpiece or attachment. Pull the brush completely through and repeat until all debris is removed. 4. Manipulate all moving parts of the handpiece and attachments to ensure all debris is removed. If not, clean again until all debris is removed. 5. Keeping the nose of the handpiece pointed downward, rinse under tap water (minimum temperature of 25°C/77°F) for a minimum of 30 seconds using a minimum of 6 liters to remove all traces of soap. Rinse all attachments likewise. 6. Flush all surfaces free of tap water with distilled/deionized water (minimum temperature of 25°C/77°F) for a minimum of 20 seconds using a minimum of 4 liters of distilled/deionized water. Ensure handpiece is visibly free of detergent or cleaning residues. 7. Gently shake the equipment free of water and wipe the surfaces with a clean, lint-free towel. 8. For cleaning instructions of the Hall Small Bone Lithium power unit, refer its instructions of use.
3.1.3 Automated Cleaning Instructions MicroFree handpieces are suitable for use in washer/sanitizers. They are designed with advanced materials and sealing technologies and are individually tested to ensure the prevention of ingress of fluids during use and cleaning. MicroFree handpieces have an Ingress Protection rating of IPX6 and IPX8, ensuring they are suitably protected against water intrusion. 1. Under running water, remove all traces of blood and debris. 2. Load handpiece in washer/sanitizer and operate according to machine manufacturer’s instructions. It is recommended to use a neutral pH cleaner. Only use approved solutions for cleaning powered instruments according to the manufacturer’s recommendations. Ensure all collet mechanisms are in the open position. 3. Run washer/sanitizer according to the manufacturer’s specifications. A dry cycle is recommended but not required.
14
3.2
Disinfection Information
3.2.1 Manual Disinfection Instructions NOTE: CaviCide® Spray Ready-to-use or a comparable disinfection solution of PH between 11.0 and 12.49 are highly recommended by CONMED. However, the use of alternative methods of disinfection or disinfection agents outside the scope of this document may be suitable for disinfecting the devices. Alternate disinfection methods must be validated by the end user. 1. Spray the device thoroughly with CaviCide® Spray Ready-to-use disinfectant. 2. Ensure the device remains wet for a minimum of ten (10) minutes by spraying the device as needed with the CaviCide® Spray Ready-to-use. 3. Rinse under running deionized water for a minimum of one (1) minute. 4. Thoroughly dry all surfaces of the device using a sterile, lint free wipe, changing wipes when necessary to ensure the device is completely dry.
3.2.2 Automated Disinfection Instructions 1. 2. 3. 4.
Place device in an automated washer disinfector filled with purified water. Run thermal rinse cycle at 194˚F (90 ˚C) for five (5) minutes. Let device cool in washer for a minimum of 60 minutes. Removed device from washer and thoroughly dry the device with a sterile lint free wipe.
15
3.3
Sterilization Information
3.3.1 Warnings, Precautions and Notes 1. The use of disinfecting solutions for an exterior instrument wipe will not sterilize the equipment and is not recommended. 2. Do not sterilize equipment or accessories using Ethylene Oxide (EtO). 3. Do not sterilize equipment or accessories using a STERIS system or by comparable sterilization methods. 4. Do not sterilize equipment or accessories in cold sterilants like CIDEX. 5. Always detach accessories from equipment prior to sterilization. 6. Do not use equipment while warm. Allow adequate time for cooling prior to use. Cool by exposure to room temperature. Operation of equipment that is not completely cool or dry may decrease performance and/or reliability. 7. When sterilizing equipment and attachments in a system sterilization tray, use the recommended sterilization parameters in its instruction manual. 8. Do not “Peel Pack” handpieces for sterilization. Sterilization in a sealed pouch traps moisture which can cause damage. 9. Attachments or handpieces with collet mechanisms must be sterilized with the collet fully open. 10. A minimum dry cycle according to the Sterilization Instructions must be run on all equipment and attachments every time the product is sterilized. Failure to use a dry cycle on the products may lead to reduced product performance or premature product failure. 11. In instances where a facility does not have the ability to perform Pre-vacuum sterilization with dry cycles that correspond for the recommendations above, we recommend that the handpiece be dried in a dry heat sterilizer at 250°F (121°C) after sterilization, once a day for two hours. This process will eliminate moisture that may have built up during previous gravity sterilization cycles. Following this process should ensure optimum performance of the handpiece. Refer also to Section “1.5 Warnings and Precautions”.
3.3.2 Inspection Recommendations 1. Inspect device prior to sterilization. 2. Generally, unmagnified visual inspection under good light conditions is sufficient. All parts of the device should be checked for visible soil and/or corrosion. 3. Functional checks should be performed where possible. 4. Mating devices should be checked for proper assembly. 5. Ensure moving parts move freely. When necessary, lubricate moving parts with Hall Blitz® Instrument Cleaner (M105A) or equivalent. 6. Remove and replace damaged instruments/containers.
3.3.3 Sterilization Instructions Steam sterilization is safe and effective and has no contraindications for sterilizing this equipment. Handpieces and attachments may be processed in a pre-vacuum steam sterilizer (Steam Pre-vacuum) or in a gravity (downward) displacement sterilizer (Steam Gravity). 1. Individually wrap handpiece and accessories. In the United States, use FDA-cleared sterilization wrap. 2. Follow the recommended minimum sterilization exposure times listed below. 3. Use the recommended tray parameters in its instruction manual when sterilizing handpieces and attachments in a system sterilization tray. 4. The instructions provided have been validated as being capable of preparing the MicroFree handpiece for re-use. It remains the responsibility of the processor to ensure that the processing is actually performed using equipment, materials and personnel in the processing facility to achieve the desired result. This requires validation and routine monitoring of the process. Likewise, any deviation by the processor from the instructions provided should be properly evaluated for effectiveness and potential adverse consequences.
16
Recommended sterilization exposure times are as follows:
Table 1: Sterilization Parameters of MicroFree Battery Handpieces without a System Sterilization Tray in the United States Method
Temperature
Exposure
Dry Cycle
Steam Pre-Vacuum
270°F (132°C)
3 minutes
8 minutes
Steam Pre-Vacuum
270°F (132°C)
4 minutes
8 minutes
Steam Pre-Vacuum
273°F (134°C)
3 minutes
8 minutes
Steam Gravity
250°F (121°C)
30 minutes
8 minutes
Table 2: Sterilization Parameters of MicroFree Battery Handpieces without a System Sterilization Tray outside the United States Method
Minimum Temperature
Maximum Temperature
Minimum Exposure
Maximum Exposure
Minimum Dry Cycle
Maximum Dry Cycle
Steam PreVacuum
273°F (134°C)
278°F (137°C)
3 minutes
18 minutes
8 minutes
No maximum
Steam Gravity
250°F (121°C)
-
30 minutes
-
8 minutes
No maximum
3.4
Troubleshooting
Table 3: Troubleshooting Guide
Symptom
Possible Cause
Corrective Action
Handpiece does not operate.
• •
Handpiece faulty. Power unit safety is in the safe, or off position. Twist collet not in the fully locked position. Handpiece not fully seated in power unit.
• •
Worn bearings in the bur guard, or collet. Moisture in the handpiece.
• •
• • Handpiece Overheats
• •
• •
•
17
Return for service. Move safety to the appropriate operating position. Twist collet to the fully locked position. Securely connect handpiece to power unit Refer to section “2.4.1 Bur Guard Functional Test”. If resistance is noticed, return bur guard for service. If no resistance is noticed, return for service. Resterilize the handpiece according to sterilization parameters and drying times. (Refer to section “3.3.1 Warnings, Precautions and Notes”).