HEINE® NT300 Charger Please read and follow these instructions for use and keep them for future reference.
General Conditions of Warranty Instead of the statutory warranty time period of 2 years, HEINE will grant a guarantee of 5 years from the date of the consignment of the goods ex works, concerning its equipment (excluding disposables, e.g. bulbs, singleuse articles, and rechargeable batteries). Date of consignment means that HEINE hands the goods over to the transport carrier, freight forwarder or any other person designated by the Customer for the transport of the goods without loading the collecting vehicle. The guarantee covers irreproachable workmanship, on condition of the proper use of the equipment and the observation of the operating instructions. During the warranty- and guarantee time period, errors and deficiencies arising on the equipment will be rectified free of charge, in so far as such are evidenced by defective materials, processing and/or constructional errors. Should buyer complain of a material deficiency during the warranty time period, then the onus of proof is always to be on the orderer, that the product was defective already upon receipt of the goods. The statutory warranty and the guarantee do not apply to loss or damage caused by wear and tear, negligent use, the non-employment of original HEINE components and/or spares (in particular bulbs, as these have been especially developed for HEINE instruments in accordance with the following criterions: colour temperature, useful service life, safety, optical quality and performance. The statutory warranty and the guarantee do not apply to interventions by persons not authorised by HEINE or when the operating instructions are not observed by the customer. Any modification of a HEINE product with parts or additional parts which do not conform to the original HEINE specification will invalidate the warranty for the correct function of the product and further invalidate any warranty claims which result from such a change or modification. Further claims, in particular claims for replacement of loss or damage, which are experienced otherwise than directly on the HEINE product itself, are hereby excluded. For U.S. only: Caution: Federal law restricts this device to sale by or on the order of a Physician or Practitioner! Intended Use The HEINE® NT300 charger is designed exclusively for charging HEINE BETA® rechargeable handles, HEINE BETA® Slim rechargeable handles and HEINE Laryngoscope rechargeable handles that must contain a HEINE rechargeable battery. Warnings and Safety Information CAUTION! Indicates potential hazardous situations. Ignoring the corresponding instructions may lead to dangerous situations of mild to moderate extent. (Background color yellow; foreground color black). NOTE! Indicates valuable advice in terms of installation, operation, maintenance or repair. Notes are important, but not related to hazardous situations.
Product overview
Primary plug:
EU
10/76
med 0613
USA / Japan
2014-03-10
UK
Australia
The primary plug is interchangeable. This allows a world wide use. The procedure of changing the plugs is as per the pictures. The HEINE® NT300 is intended for installation in medical areas outside the patient environment (at least 1.5 metres from the patient or patient support pursuant to IEC 60601-1, see figure). To prevent mechanical and electrical breakdown, never install or store the charger in the following areas: - Areas exposed to high moisture or condensation - Areas exposed to extreme environmental conditions - Areas subject to constant vibrations - Areas subject to severe temperature fluctuations - Outdoors Installation Do not place the NT300 on soft furniture, cloth or carpet. Do not insert any objects in any opening in the housing.
Fig1: The above figure defines the patient environment (minimum radius around the patient). Installation, operation and use The HEINE® NT300 is compatible with the following rechargeable handles: - BETA L rechargeable handle - BETA NT rechargeable handle - BETA Slim rechargeable handle - Standard F.O. laryngoscope handle - Short F.O. laryngoscope handle - F.O. angled handle - Small F.O. laryngoscope handle
med 0613
2014-03-10
11/76
The NT300 may only be used at the mains voltage specified on the identification label. Unplug the unit from the mains if not in use over long periods. In the event of fluid ingress in the housing, the charger should be taken out of service and checked by an authorised person. The NT300 must not be used in rooms where an explosion hazard is present. Follow the instructions for the charger and rechargeable handles. The following handles are not compatible with the NT300 charger: BETA TR, BETA R, BETA cord handle, mini3000 battery handle. Plug the mains transformer into the mains and the NT300 is ready for use. NT300 ensures that batteries are kept in good condition irrespective of handle size. When replacing the handles in the charging compartments, ensure that the internal contacts in the compartment are clean and that the instruments are switched off. Instruments placed next to each other must not touch each other. If different-sized handles are used, it is possible to fit adaptors so that each handle is held in place correctly. Reduction rings for:
BETA Handle F.O. angled handle Standard F.O. Small F.O. laryngoscope handle Short F.O. laryngoscope handle (optional)
BETA Slim rechargeable handle (optional)
Ideally, the rechargeable handle should always be replaced and stored in the charger. Handles cannot be overcharged. The rechargeable battery can only be charged when fitted to the handle. The NT 300 must be disconnected from the mains supply when the adaptors are being fitted. If the second charging compartment is not in use it can be sealed off by the dust cap which is supplied with the unit. The pulsating light around the charging compartment shows that current is flowing and that the battery is being recharged. When the light is permanently lit, the charging sequence is completed. The battery in the handle is fully-charged and is ready to supply maximum power. If the light is switched off, this indicates that the charging compartment is empty or the unit is defective. Hygienic Cleaning / Processing: The device must be cleaned by sufficiently qualified personnel. We recommend that external surfaces of the NT300 be wiped clean with a disinfectant cleaning wipe. The external surfaces may also be wiped clean with a cloth moistened with water or with a neutral pH or slightly alkaline cleansing agent. Any disinfectant used must be approved by the disinfectant manufacturer for use on plastics. The manufacturer’s instructions must be followed. Spray disinfection or immersion disinfection is not permitted since there is a risk of moisture ingress, which would damage the device. Disinfection: The article can be neither machine-cleaned, nor thermal disinfected or sterilized. These processes lead to irreparable damage to the product.
12/76
med 0613
2014-03-10
Maintenance The charger is maintenance free. However, it must be tested by a qualified electrician at periodic intervals of 24 months. Do not open the power adapter (risk of electric shock). Service This product does not require regular service. Disposal The product must be recycled as separated electrical and electronic devices. Please observe the relevant state-specific disposal regulations. Used or defect batteries must be disposed adequately. Avoid skin contact with defect batteries. Technical specification Mains supply: Current consumption: Output: Fuse: Charging current: Charging time: Safety class: Protection class: Environmental conditions for operation Temperature: Humidity: Air pressure: Environmental conditions for transport and storage Temperature: Humidity: Air pressure:
100 – 240 V AC / 50 – 60 Hz max. 30 VA 15Vdc, 2A Integral overload protection Max. 1.5 A for BETA L Max. 0.5 A for other HEINE rechargeable batteries BETA L rechargeable battery -> approx. 2 hrs Other HEINE rechargeable batteries -> up to 4 hrs II IP 20
+10°C to +40°C 10% to 75% r.h. 700 hPa to 1060 hPa
-20°C to +50°C 10% to 95% r.h. 500 hPa to 1060 hPa
General Notes and Warnings The HEINE® NT300 charger is a high-quality piece of equipment. Always handle with care. Check the correct operation of the product at regular intervals. Do not use the equipment if it shows visible signs of damage. Do not use the device in hyperbaric chambers, in explosive or oxygen loaded environments! Do not use the device near strong magnetic fields like MRI scanners. We can only guarantee the performance of this device when fitted and used with original HEINE accessories and spare parts. Otherwise the HEINE Warranty is terminated. The NT300 must not be charged using rechargeable batteries from other manufacturers. Do not manipulate or modify the product! The LED indicates that the power adapter is ready for use. Do not pull the cable to disconnect the power adapter from the mains. Do not store or use the charger outdoors.
med 0613
2014-03-10
13/76
Explanation of utilized symbols The following symbols are used on the device or on the packaging: The CE mark indicates that the product complies with the European medical device directive 93/42/EEC.
Fragile, handle with care!
REF
Catalogue- or reference number
For indoor use only
SN
Serial number
“Grüner Punkt” (country-specific)
Manufacturer
Follow instructions for use! (Background: blue, foreground: white)
Keep dry!
Date of manufacture Product bearing this symbol may not be disposed of together with general household waste, but instead requires separate disposal according to local provisions. (European Waste Electrical and Electronic Equipment Directive, WEEE)
Approval mark of SIQ (for Slovenia) Trade mark of the manufacturer
ta40/F
UL approval mark. Valid in the US and Canada.
Temperature limits in °C for storage and transport Temperature limits in °F for storage and transport
Rated ambient temperature/converter temperature class
Class II equipment
IPx0
Humidity limitation for storage and transport
IP code (Ingress Protection Rating) Approval mark of Australia
Pressure limitation for storage and transport Instructions for use Alternating current (AC) Fuse Interference may occur in the vicinity of equipment marked with the following symbol.
Electromagnetic Compatibility Medical electric devices are subject to special precautionary measures with regard to electromagnetic compatibility (EMC). Portable and mobile high frequency communication equipment can affect medical electric devices. This device is intended for use by medical professionals in the electromagnetic environment specified below. The user of this device should assure that it is used in such an environment. The use of accessories, converters or cables other than the ones specified by HEINE might lead to increased emission reduced electrical immunity of the medical equipment. The device may not be stacked directly near or used directly beside other devices. If the device is to be operated in a stack or with other devices, the device should be watched to ensure it operates properly in this location.
14/76
med 0613
2014-03-10
Guidance and manufacturer’s declaration – electromagnetic emissions The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such environment. Emission test
Compliance
Electromagnetic environment – Guidelines
RF emissions CISPR11
Group 1
The device uses RF energy only for its internal function. Therefore, RF-emission is very low and it is unlikely that any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Warning: This device is intended only for use by medical professionals. This is a device of class A CISPR 11 in the domestic environment, this device may cause radio interference, so that it may be necessary in this case, to take appropriate remedial measures, as e.g. orientation, new arrangement or shielding of the device or restrict the connection to the site.
Harmonic Emissions IEC 61000-3-2
Class A
Voltage Fluctuations/ Flicker Emissions IEC 61000-3-3
Passed
Guidance and manufacturer declaration - Electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such environment. Electromagnetic environment – Guideline
Immunity test
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD) acc. to IEC 61000-4-2
± 6 kV contact discharge ± 8 kV air discharge
± 6 kV contact discharge ± 8 kV air discharge
Floors should be wood, concrete or covered with ceramic tiles. If floors are covered with synthetic material, the relative humidity should be at least 30 %.
Electrical fast transient/ burst IEC 61000-4-4
± 2 kV for mains cables ± 1 kV for input and output lines
± 2 kV for mains cables ± 1 kV for input and output lines
The supply voltage quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
± 1 kV voltage phase – phase, ± 2 kV voltage phase – earth
± 1 kV voltage phase – phase ± 2 kV voltage phase – earth
Mains power quality should be that of a typical commercial or hospital environment.
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
< 5% UT, (>95% dip in UT) < 5% UT, (>95% dip in UT) Mains power quality should be that for 1/2 period for 1/2 period of a typical commercial or hospital 40% UT, (60% dip in UT) 40% UT, (60% dip in UT) environment. If the user of the device for 5 periods for 5 periods device requires continued operation 70% UT, (30% dip in UT) 70% UT, (30% dip in UT) during power mains interruptions, it for 25 periods for 25 periods is recommended that the device be <5% UT, (>95% dip in UT) <5% UT, (>95% dip in UT) powered by a UPS (uninterruptible for 5 seconds for 5 seconds power supply) or a battery.
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Comment: UT is the a.c. supply voltage prior to application of the test level.
med 0613
2014-03-10
15/76
Guidance and manufacturer’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such environment. Immunity test
IEC 60601 test level Compliance level
Electromagnetic environment Guidelines
Conducted RF IEC 61000-4-6
3 Veff 150 kHz to 80 MHz
3 V eff
Portable and mobile RF communication equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Radiated HF IEC 61000-4-3
3 V/m 80MHz to 2,5GHz
3 V/m
Recommended separation distance: d = 3,5/3 * SQRT (P/W) d = 3,5/3 * SQRT (P/W) 80 MHz to 800 MHz d = 7/3 * SQRT (P/W) 800 MHz to 2,5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site surveyaa, should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
Note 1: At 80Hz and 800MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V / m. Recommended separation distances for portable and mobile RF communication equipment and the device The device is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled. The customer or user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter (W)
Separation distance according to frequency of transmitter (m) 150 kHz to 80 MHz d = 3,5/3 * SQRT (P)
80 MHz to 800 MHz d = 3,5/3 * SQRT (P)
800 MHz to 2,5 GHz d = 7/3 * SQRT (P)
0.01
0.1
0.1
0.2
0.1
0.4
0.4
0.7
1
1.2
1.2
2.3
10
3.7
3.7
7.4
100
11.7
11.7
23.3
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
16/76
med 0613
2014-03-10
Manufacturer: HEINE Optotechnik GmbH & Co. KG Kientalstr. 7 · 82211 Herrsching · Germany www.heine.com Patents: www.heine.com