Camera settings for photographic documentation with the HEINE DELTA® 20 T / HEINE DELTA® 20 PLUS / DELTA® 20 Dermatoscope and HEINE SLR Photo adaptor Digital high-resolution photography is the method of choice for optimum image documentation in dermatology. The combination of the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 and a digital SLR camera allows for consistent high quality photographs. This step-by-step guide will show you how to connect a DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 to a Digital SLR camera, and which settings are optimal for your camera. For proper use please read the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 user manual carefully. 1. The following current digital SLR cameras are recommended in connection with the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20.
Manufacturer
Type
HEINE SLR Photo Adaptor
EOS 7D Mark II EOS 60D / 80D
Canon
EOS 100D / 700D
K-000.34.185
EOS 750D / 760D / 1300D D500 / D3200 / D3300 Nikon
D3400 / D5200 / D5300
K-000.34.186
D5500 / D7100 / D7200 Other currently available models from Canon and Nikon are not supported or deliver photographic material of insufficient quality. The current Olympus camera portfolio is not compatible any more with HEINE SLR adaptors. Some of the older Canon, Nikon or Olympus models are compatible with the SLR adaptor. If you wish to use an older camera, please contact [email protected] to get the compatibility checked.
2. F or best results, please assemble the following components: DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 Dermatoscope with contact plate and scale, BETA handle as power source, HEINE Photo Accessory Set consisting of HEINE SLR Photo adaptor (Canon, Nikon or Olympus) adaptor cord X-000.99.231 , and belt clip . The distance ring is part of the adaptor set.
1
3. Connect one end of the adaptor cord to the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 Dermatoscope head and the other end to the BETA handle by slightly turning the head of the handle. (Put the handle into your belt clip, it will give you more freedom of movement).
4. C onnect the HEINE SLR Photo adaptor to the bayonet lock of the camera by placing the red or white spots on top of each other and twist to lock.
5. C onnect the distance ring onto the line marked “Camera” on the focusing ring of the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 Dermatoscope head. Once attached, tighten the focus ring down until it makes contact with the distance ring.
6. Connect the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 to the SLR photo adaptor. Push the sliding ring of the adaptor with one hand down towards the camera and hold it in this position. With your free hand, insert the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 Dermatoscope head into the adaptor until it clicks in the housing. Now release the sliding ring of the adaptor to lock the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 onto the SLR photo adaptor. Check whether the dermatoscope can easily rotate on the adaptor and if it is locked in by gently pulling the head away from the camera body.
"click"
7. Adjust the settings on your camera to optimize image quality. Following are instructions on how to make the proper adjustments for Nikon and Canon. (Please note that “auto” settings do not lead to good results.)
2
Example: Nikon (pictures © Nikon)
1. Please turn the mode dial to program “M”.
2. Set the ISO sensitivity to 400 by pressing the ISO - button while turning the rear main command dial until the correct value is displayed. The correct value appears on the bottom corner of the upper control panel.
400
ISO button
Main command dial
Control panel
3. S et the exposure time to 1 / 40 s. Press the shutter-release briefly until the exposure time appears on the upper control panel. Adjust the value by turning the rear main command dial.
40
40 Main command dial
3
4. T he white balance needs to be set once. To do this, press the WB - button and turn the rear main command dial until ”PRE“ is displayed on the upper control panel. 40
Control panel
View finder
button
Main command dial
Control panel
Turn the WB button until the “PRE” symbol starts to blink. Please point the camera with the adapted DELTA® 20 T / DELTA® 20 Plus / DELTA 20 head (switched on) on a white or neutral gray object like paper and press the shutter-release button. The display will show “good” once the measuring was accurate. In all other cases please repeat the process.
button
Control panel
View finder
The manual white balance picture can be found in the menu. For using a previously set white balance, please also use the menu. Please choose a white balance picture in the menu. We recommend naming the white balance picture taken with the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 accordingly.
4
Example: EOS
(pictures © Canon) 1. Please turn the mode dial to the program Tv.
2. Set the ISO sensitivity to 400. To do so, press the ISO- button and select the value 400 on the display.
3. Adjust the exposure time to 1 / 40 s by pushing the menu button until the following menu appears:
The value can be set by turning the main dial (in the front). 4. The white balance needs to be set once. To do so, please press the menu-button until the following menu is displayed:
Press the WB button, select the customer white balance on the display and confirm with the “set” button.
5
Point the camera with the adapted DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 head (switched-on) on a white or neutral grey object like paper and press the shutter button. To accept the white balance picture, please press the menu-button until the following screen is displayed.
Please select the custom WB with the set-button and confirm the chosen picture with the set-button. We recommend naming the white balance picture taken with the DELTA® 20 T / DELTA® 20 Plus / DELTA® 20 accordingly.
Canon Camera Name Chart USA EOS 7D Mark II EOS 60D EOS 80D Rebel SL 1 Rebel T5i Rebel T6i Rebel T6s Rebel T6
Japan EOS 7D Mark II EOS 60D EOS 80D Kiss X7 Kiss X7i Kiss X8i 8000 D Kiss X80
med 0112e 12.16
Europe EOS 7D Mark II EOS 60D EOS 80D EOS 100D EOS 700D EOS 750D EOS 760D EOS 1300D
HEINE Optotechnik GmbH & Co.KG Kientalstr. 7 · 82211 Herrsching · Germany · Tel +49 (0) 8152 / 38 0 Fax +49 (0) 8152 / 38 - 202 · E-Mail: [email protected] · www.heine.com
6