HELMER
Platelet Systems- Incubators and Agitators
i.Series and Horizon Series Platelet Incubator various models Operation Manual Rev P May 2014
Operation Manual
28 Pages
Preview
Page 1
Platelet Incubator Operation Manual i.Series® and Horizon Series™
Model Group
i.Series
Horizon Series
Countertop
PC100i, PC900i, PC1200i (Version A)
PC100h, PC900h, PC1200h (Version A)
Floor
PC2200i, PC3200i, PC4200i (Version A)
PC2200h, PC3200h, PC4200h (Version A)
HELMER SCIENTIFIC 14400 Bergen Boulevard Noblesville, IN 46060 USA PH +1.317.773.9073 FAX +1.317.773.9082 USA and Canada 800.743.5637 360093-1/P
0086
ISO 13485:2003 CERTIFIED
Document History Revision
Date
CO
Supersession
Revision Description
L
28 JUN 2013
8414
Supersedes A, B, C, D, E, F, G, Revised layout for ease of navigation and locating information. H, I, J, K
M
23 DEC 2013
8965
► Added 100 V information. M supersedes L ► Added caution note for remote alarm interface. ► Revised section I for consistency with existing manuals.
N
31 JAN 3014
9113
N supersedes M Changed voltage for remote alarm contacts.
O
22 APR 2014* 9386
O supersedes N Revised voltage specification for remote alarm contacts.
P
22 MAY 2014* 9497
P supersedes O
► Added information to environmental/application specifications from technical audit. ► Added 0086 to CE certification.
* Date submitted for Change Order review. Actual release date may vary.
360093-1/P
i
Contents Section I: General Information... 4 1 About this Manual... 4 1.1 1.2 1.3
Intended Audience... 4 Model References... 4 Copyright and Trademark... 4
2 Safety... 4 2.1 2.2 2.3
Safety Definitions... 4 Product Labels... 5 Avoiding Injury... 5
3 General Recommendations... 6 3.1 3.2 3.3
Intended Use... 6 General Use... 6 Initial Loading... 6
4 Specifications... 6 5 Compliance... 7 5.1 5.2
Regulatory Compliance... 7 WEEE Compliance... 7
6 Installation... 8 6.1 6.2
Location Requirements... 8 6.1.1 Placement... 8 Chart Recorder... 8 6.2.1 Install and Change Chart Paper... 9
7 Maintenance Schedule... 10
Section II: i.Series® Models... 11 8 Operation... 11 8.1 8.2 8.3 8.4 8.5 8.6
Initial Start Up... 11 Configure a Platelet Agitator for Use in an i.Series Incubator (Optional)... 11 Load the PC4200h Built-In Agitator... 12 Main Screen Functions... 13 Chamber Setpoint... 14 Temperature Monitor Setpoints... 14 8.6.1 Change System Password... 15 8.6.2 Temperature Alarm Setpoints... 15 8.6.3 Active Alarms... 15 8.6.4 Mute an Active Alarm... 16
9 Components... 16 9.1 9.2 9.3 9.4
Control Door... 16 Control Panel... 16 Alarm Panel... 17 Chamber... 17
360093-1/P
ii
Section III: Horizon Series™ Models... 18 10 Operation... 18 10.1 10.2 10.3 10.4 10.5
10.6 10.7 10.8 10.9
Initial Start Up... 18 Configure a Platelet Agitator for Use in a Horizon Series Incubator (Optional)... 18 Load the PC4200h Built-In Agitator... 18 Chamber Setpoint... 19 Temperature Monitor Setpoints... 19 10.5.1 High Alarm Setpoint... 19 10.5.2 Low Alarm Setpoint... 20 Alarm Volume Settings... 20 10.6.1 Temperature and Power Failure Alarm Volume... 20 10.6.2 Motion Alarm Volume (PC4200h)... 21 Alarm Delay Settings... 22 10.7.1 Temperature Alarm Delay... 22 10.7.2 Motion Alarm Delay (PC4200h)... 23 Mute Audible Alarms... 23 Enable or Disable Audible Alarms with the Alarm Key Switch... 24
11 Components... 25 11.1 11.2 11.3 11.4
Control Door... 25 Control Panel... 25 Alarm Panel... 26 Chamber... 26
360093-1/P
iii
General Information
Section I: General Information 1
About this Manual
1.1
Intended Audience This manual is intended for use by end users of the platelet incubator and authorized service technicians.
1.2
Model References Generic references are used throughout this manual to group models that contain similar features. For example, “PC100 models” refers to all models of that size (PC100i, PC100h). This manual covers all platelet incubators, which may be identified singly, by their size, or by their respective “Series.”
1.3
Copyright and Trademark Helmer®, i.Series®, Horizon Series™, AgiTrak™, and Rel.i™ are registered trademarks or trademarks of Helmer, Inc. in the United States of America. Copyright © 2014 Helmer, Inc. All other trademarks and registered trademarks are the property of their respective owners. Helmer, Inc., doing business as (DBA) Helmer Scientific and Helmer.
2
Safety The operator or technician performing maintenance or service on Helmer Scientific products must (a) inspect the product for abnormal wear and damage, (b) choose a repair procedure which will not endanger his/her safety, the safety of others, the product, or the safe operation of the product, and (c) fully inspect and test the product to ensure the maintenance or service has been performed properly.
2.1
Safety Definitions The following general safety alerts appear with all safety statements within this manual. Read and abide by the safety statement that accompanies the safety alert symbol.
360093-1/P
WARNING
The safety statement that follows this safety alert symbol indicates a hazardous situation which, if not avoided, could result in serious injury.
CAUTION
The safety statement that follows this safety alert symbol indicates a hazardous situation which, if not avoided, could result in minor or moderate injury.
NOTICE
The safety statement that follows this safety alert symbol indicates a situation which, if not avoided, could result in damage to the product or stored inventory.
4
General Information
2.2
Product Labels The following general safety and information alerts appear on the product to identify potential hazards to the operator or service technician.
2.3
360093-1/P
Caution: Risk of damage to equipment or danger to operator
Caution: Unlock all casters
Caution: Hot surface
Earth / ground terminal
Caution: Shock/electrical hazard
Protective earth / ground terminal
Avoiding Injury ► Review safety instructions before installing, using, or maintaining the equipment. ► Before moving unit, remove contents from the drawers (if applicable). ► Do not open multiple drawers at the same time (if applicable). ► Before moving unit, ensure door(s) are closed and casters (if applicable) are unlocked and free of debris. ► Before moving unit, disconnect the AC power cord and secure the cord. ► When moving unit, use assistance from a second person. ► Never physically restrict any moving component. ► Avoid removing electrical service panels and access panels unless so instructed. ► Keep hands away from pinch points when closing the door or when agitation motion is enabled (if applicable). ► Avoid sharp edges when working inside the electrical compartment. ► Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good laboratory practices. ► Proceed with caution when adding and removing samples from the platelet incubator. ► Use supplied power cord only. ► Using the equipment in a manner not specified by Helmer Scientific may impair the protection provided by the equipment. ► Decontaminate parts prior to sending for service or repair. Contact Helmer Scientific or your distributor for decontamination instructions and a Return Authorization Number. ► Ensure biological materials are stored safely, in accordance with all applicable organizational, regulatory, and legal requirements. ► The platelet incubator is not considered to be a storage cabinet for flammable or hazardous materials.
5
General Information
3
General Recommendations
3.1
Intended Use Helmer platelet incubators are intended to provide the controlled temperature environment required for the storage of platelet products. The devices are intended to be operated by personnel who have procedures in place for meeting FDA, AABB, EU or any other applicable regulations for the processing and storage of platelet products.
3.2
General Use Allow platelet incubator to come to room temperature before switching power on. During initial startup, motion alarm may sound if the motion is disabled, and low temperature alarm may sound while platelet incubator reaches operating temperature.
3.3
Initial Loading After platelet incubator reaches room temperature, allow chamber temperature to stabilize at the setpoint before storing product.
4
Specifications PC100
PC900
PC1200
PC2200
PC3200
PC4200
Height
25.00” (635 mm)
30.25” (768 mm)
30.25” (768 mm)
60.00” (1524 mm)
75.50” (1918 mm)
75.50” (1918 mm)
Width
21.25” (540 mm)
26.50” (673 mm)
40.75” (1035 mm)
40.25” (1022 mm)
40.25” (1022 mm)
40.25” (1022 mm)
Depth
23.50” (597 mm)
27.75” (705 mm)
27.75” (705 mm)
30.25” (768 mm)
30.25” (768 mm)
30.25” (768 mm)
Weight
105 lbs (48 kg)
159 lbs (72 kg)
208 lbs (94 kg)
363 lbs (165 kg)
431 lbs (196 kg)
663 lbs (301 kg)
Physical
Electrical Input Voltage and Frequency
100 V, 50/60 Hz / 115 V, 50/60 Hz / 230 V, 50/60 Hz
Voltage Tolerance
115 V, 50/60 Hz / 230 V, 50/60 Hz
±10%
Circuit Breakers
12.0 A (100 V, 115 V) 6.0 A (230 V, quantity 2)
15.0 A, 2 A (115 V)
20.0 A, 5.0 A (115 V)
Fuses
n/a
10.0 A (230 V, quantity 2)
10.0 A (230 V, quantity 2)
8.0 A (100 V, 115 V) Power 5.0 A (230 V) Consumption (1) Power Source (2)
8.0 A (100 V) 9.0 A (115 V) 4.5 A (230 V)
8.0 A (100 V) 9.0 A (115 V) 4.5 A (230 V)
11.5 A (115 V) 7.0 A (230 V)
12.0 A (115 V) 7.0 A (230 V)
14.5 A (115 V) 8.3 A (230 V)
Varies (refer to product specification label)
Agitation Speed (3) (cycles/minute)
n/a
Remote Alarm Capacity
60 (115 V) 60 (230 V, 50 Hz) 72 (230 V, 60 Hz)
i.Series: 0.5 A at 30 V (RMS); 1.0 A at 24 V (DC) Horizon Series: 0.25 A at 30 V (RMS); 0.25 A at 60 V (DC)
Internal Outlet Maximum Current Draw (4)
0.5 A
n/a
Control and Monitoring
360093-1/P
Interface
i.Series: Monitoring and display system and separate temperature control system Horizon Series: Temperature control and display system
Alarms
i.Series: High, low, and condenser temperature; door open; low battery; no battery; AC power failure; change chart paper; agitator 1, 2, and 3 motion Horizon Series: High and low temperature; AC power failure; agitator motion (PC4200h)
6
General Information PC100
PC900
PC1200
PC2200
PC3200
PC4200
Environmental
Application
► ► ► ► ► ► ► ►
Indoor use only Altitude (maximum): 2000 m Ambient temperature range: 15 °C to 35 °C Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C, decreasing linearly to 50% at 40 °C Temperature control range: 20 °C to 35 °C Overvoltage category: II Pollution degree: 2 Mains supply voltage: ±10% of nominal voltage
(1) Power consumption is measured in full-load Amperes. (2) Product specification label is located on the back of the platelet incubator. (3) Agitation speed is ±10% of nominal speed. (4) PC4200 platelet incubators do not include an internal outlet(s). CAUTION
► The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm system(s) that uses normally-open or normally-closed dry contacts. ► If an external power supply exceeding 30 V (RMS) or 60 V (DC) is connected to the remote alarm monitoring system’s circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
5
Compliance
5.1
Regulatory Compliance This device complies with the requirements of directive 93/42/EEC concerning Medical Devices, as amended by 2007/47/EC.
0086
Sound level is less than 70 dB(A).
EC
REP
5.2
Emergo Europe Molenstraat 15 2513 BH The Hague, Netherlands
WEEE Compliance The WEEE (waste electrical and electronic equipment) symbol (right) indicates compliance with European Union Directive WEEE 2002/96/EC and applicable provisions. The directive sets requirements for the labeling and disposal of certain products in affected countries. When disposing of this product in countries affected by this directive: ► Do not dispose of this product as unsorted municipal waste. ► Collect this product separately. ► Use the collection and return systems available locally. For more information on the return, recovery, or recycling of this product, contact your local distributor.
360093-1/P
7
General Information
6
Installation
6.1
Location Requirements ► Has a sturdy, level surface. ► Has a grounded outlet meeting national electric code (NEC) and local electrical requirements. ► Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents. ► Countertop models: Minimum 4” (102 mm) above and behind. ► Floor models: Minimum 4” (102 mm) on left and right sides. ► Meets limits specified for ambient temperature and relative humidity.
6.1.1
Placement WARNING 1 2 3
6.2
To prevent tipping, ensure the casters are unlocked (floor models).
Ensure all casters are unlocked (floor models). Place platelet incubator on study surface or roll platelet incubator into place and lock casters (floor models). Ensure platelet incubator is level.
Chart Recorder A B
F
C D E
Chart recorder with paper and battery installed. Label
360093-1/P
Description
Function
A
Left and Right Arrow Adjust settings and stylus position buttons
B
LED
C
Chart change button Adjust position of stylus when changing chart paper, or run a test pattern
Indicates status of chart recorder in operating mode, or selected temperature range in paper change mode
D
Stylus
Mark temperature line on paper
E
Reset button
Restart chart recorder
F
Backup battery
Provides power during AC power failure. Connect prior to use.
8
General Information
6.2.1
Install and Change Chart Paper 1 Press and hold C button. When stylus begins to move left, release button. The LED flashes to indicate current temperature range. 2 When stylus stops moving, remove chart knob then move knob up and away from chart paper. 3 Place new chart paper on chart recorder. 4 Gently lift stylus and rotate paper so current time line corresponds to time line groove.
5
Hold chart paper and reinstall chart knob.
NOTE 6 7 8
360093-1/P
For accurate temperature reading, ensure that current time is aligned with time line groove when chart knob is tightened.
Confirm the temperature range is set to the correct value. Press and hold button C. When the stylus begins to move right, release the button. Confirm the stylus is marking the temperature correctly.
9
General Information
7
Maintenance Schedule Maintenance tasks should be completed according to the following schedule. Refer to the service manual for more detail on the various tasks. NOTE
These are recommended minimum requirements. Regulations for your organization or physical conditions at your organization may require maintenance items to be performed more frequently, or only by designated service personnel.
Task
Frequency Quarterly
Test the high and low temperature alarms.
Test the power failure alarm (as required by your organization’s protocols).
Test the door open alarm.
(i.Series)
Test the no battery alarm.
(i.Series)
Test the motion alarm (i.Series platelet incubators with platelet agitators installed, and PC4200 platelet incubator).
Check the temperature calibration for the temperature monitor and change it if necessary.
Check the temperature calibration for the temperature controller and change it if necessary.
Check the backup battery for the motion alarm system after an extended power failure and change it if necessary, or change the battery if it has been in service for one year (PC4200).
Annually
(Models with chart recorders) Check the backup battery for the chart recorder after an extended power failure and change it if necessary, or change the battery if it has been in service for one year. Refer to the Temperature Chart Recorder Operation and Service Manual. Clean the condenser grill.
Clean the exterior and interior.
Clean the door gaskets.
(PC4200) Check agitation wheels, wheel bearings, and drive rod assembly for wear. Clean and lubricate moving parts. Replace moving parts if worn or when prompted by the agitation maintenance reminder (i.Series).
NOTE
360093-1/P
As Needed
► i.Series: During a power failure, the backup battery provides power to the monitoring system and the power failure alarm. If the backup battery is not functioning, the power failure alarm will not be activated. ► If the backup battery does not provide power to the monitoring system during the power failure alarm test, replace the battery. ► If batteries have been in service for one year, replace batteries.
10
i.Series® Models
Section II: i.Series® Models 8
Operation
8.1
Initial Start Up 1 Plug the power cord into a grounded outlet that meets the electrical requirements on the product specification label. 2 Switch the AC ON/OFF switch ON. 3 Connect the backup batteries for the monitoring system and alarms. 4 Connect the backup battery for the temperature chart recorder. 5 Install the alarm check tube (PC100i). Position the tube below the probe tip. 6 Select the display language. ► When the platelet incubator is powered on, the System Options is displayed. ► Press the INC or DEC buttons to select the language. ► Press the HOME button. 7 Press the MUTE button if the Low Temperature alarm sounds. NOTE
8.2
► When the door switch is bypassed, the incubator and door open alarm continue to operate as if the door is closed. ► The door switch controls power to the built-in agitator on PC4200i platelet incubator, and controls power to the internal outlets on all other models. ► The door switch may be bypassed by opening the door and pulling the switch cylinder.
Configure a Platelet Agitator for Use in an i.Series Incubator (Optional) Helmer i.Series platelet agitators may be installed in a stand-alone configuration, or in Helmer i.Series platelet incubators. ► Motion data is transmitted from the platelet agitator, through the data cable, to the platelet incubator. ► The platelet incubator interprets the motion data and generates its own motion alarm, based on its own alarm delay period. ► If the motion alarm is not disabled on the platelet agitator, the motion alarms on both devices will sound. NOTE
360093-1/P
► Disable the platelet agitator motion alarm when installing the platelet agitator in a Helmer i.Series platelet incubator. ► Refer to the platelet agitator service manual for information regarding the installation of a platelet agitator in a platelet incubator.
11
i.Series® Models
8.3
Load the PC4200h Built-In Agitator NOTE
Model PF4200i CAUTION
► This chapter applies to PC4200h models only. ► For instsructions in loading all other platelet agitator models, refer to the platelet agitator operation manual. Capacity Random Bags Apheresis Bags 396
132
When opening drawer, grasp handle (not label holder). Open one drawer at a time.
Open the drawer to be loaded and lay the platelet bags flat. ► Avoid stacking bags. ► Maintain enough space around each bag for air circulation. For thicker bags, remove drawers. ► Place the tubing under or around the bag.
360093-1/P
12
i.Series® Models
8.4
Main Screen Functions All screens on the monitoring system are accessible from the Main screen. Press the UP or DOWN buttons to highlight a menu option, then press the SELECT button to display the selected option.
Left: Temperature graph screen (displayed by default). Center: Home screen. Right: Main screen. Option
360093-1/P
Function
Event Log
View historical information: alarms and operational events
System alarm test and status
► ► ► ►
Edit configuration settings
Access the Configuration screen (with password) and edit: ► Text language ► Date and time and/or format ► Temperature units ► Volume and pattern of audible alarms ► Enable or disable chart paper timer ► Enable or disable temperature graph display ► Change alarm-related setpoints and timers ► Calibrate temperature probe monitor reading ► Change some settings to factory default values ► Change password ► Configure AgiTrak system to monitor agitation
View configuration
► ► ► ► ► ►
Product/company information
► Software versions for the control and display components of the monitoring system ► Helmer contact information
i.Help
► View real-time agitation speed and cycle counts ► View historical agitation event logs and details ► Reset agitator cycle counters
NOTE
Refer to the platelet incubator service manual for a complete list of i.Center monitoring system functions, settings, and screens.
Start or stop automatic alarm test View number of days left before TCR paper change View door status (open or closed) View condenser temperature
Date and time formats Alarm-related setpoints and timers Volume and pattern for audible alarms Setting for the chart paper timer Setting for the temperature graph display Settings for agitator alarms
13
i.Series® Models
8.5
Chamber Setpoint
Temperature monitor/controller display and buttons. NOTE
1 2 3 8.6
► Default setpoint is 22.0 °C ► Displayed chamber temperature may be different than setpoint. ► Refer to the service manual for program parameters and default parameter values.
*
On the temperature controller, press and hold . ► The temperature controller displays the current setpoint. While holding press Up Arrow or Down Arrow to change the temperature setpoint. ► Setpoint changes in 0.1 ºC increments. Release all buttons; the temperature setpoint is changed.
*
Temperature Monitor Setpoints
Left: Home screen. Right: Main screen. Information displayed on the Home screen: ► Current readings from temperature probe(s) ► Remaining battery charge ► Current date and time ► Active alarm status Features accessible from the Homescreen: ► Access the Main screen to view and change settings ► View active alarms ► Mute audible alarms ► View a graph of the chamber temperature (past 24 hours of operation) ► Adjust the screen contrast (center button)
360093-1/P
14
i.Series® Models
Features accessible from the Main screen: ► View Event Log data ► Start or stop system alarm tests and view incubator status ► View and edit configuration settings ► View product and company information ► Access the i.Help system 8.6.1
Change System Password The default password is 1234. Passwords must be four digits, ranging from 1 to 5. Change the password: 1 On the Home screen, press the MAIN button. 2 Press the DOWN button to select Edit Configuration. Press the SELECT button. 3 Enter the password when prompted. 4 Press the DOWN button to select Change Password. Press the SELECT button. 5 Enter the new password, then re-enter the new password when prompted. ► If password entries match, the “update” message is displayed. ► If password entries do not match, the “incorrect match” message is displayed. Repeat the procedure to change the password.
8.6.2
Temperature Alarm Setpoints 1 Press the MAIN button. 2 Press the DOWN button to highlight Edit Configuration. Press the SELECT button. 4 Enter the password when prompted. 5 Press the DOWN button to highlight Alarm Setpoints. Press the SELECT button. 6 Press the DOWN button to highlight the desired temperature alarm setting. 7 Press the INC or DEC buttons to set the temperature alarm setpoint. 8 Press the BACK button to return to the Edit Configuration screen, or press the HOME button to exit. The new settings are saved.
8.6.3
Active Alarms The Home screen identifies active alarms. If multiple alarms are active, the display will indicate the number of alarms and will cycle through multiple alarms.
Home screen with one active alarm (Low Temp), highlighted.
360093-1/P
15
i.Series® Models
8.6.4
Mute an Active Alarm ► By default audible alarms are set to ON. ► Alarms that are activated while existing alarms are muted are audible. ► The mute timer may be reset with each new active alarm. ► Muting an alarm does not clear the visual alarm. NOTE
Muting an alarm clears the signal sent through the remote alarm interface.
► If an alarm is still active when the mute timer expires, the active alarm is sent through the remote alarm interface. Mute an active alarm: 1 On the Home screen, press the MUTE button once to mute the alarm for five minutes. 2 Press the MUTE button multiple times to extend the mute timer (up to 60 minutes).
9
Components
9.1
Control Door
Control door (circled). 9.2
Control Panel
A B C
Control panel, behind control door. Label
360093-1/P
Description
A
Main power switch
B
Temperature controller
C
Backup battery key switch
16
i.Series® Models
9.3
Alarm Panel A B
C
G H
I
D E F E Left: PC1200i (230 V) alarm panel. Right: PC2200i alarm panel.
9.4
Label
Description
A
Flash port
B
RS-232 COM port (optional)
C
Remote alarm interface
D
Alarm buzzer
E
Circuit breakers
F
Power connector
G
Flash port
H
Alarm buzzer
I
Remote alarm interface
Chamber A
B A B
D
C
Chamber probes (multiple models shown). Label
360093-1/P
Description
A
Temperature control and temperature chart recorder probe
B
Temperature monitoring probe
C
Tube for alarm checks (PC100i)
D
Lower probe (PC2200i, 3200i, and 4200i)
17
Horizon Series™ Models
Section III: Horizon Series™ Models 10
Operation
10.1
Initial Start Up 1 Plug the power cord into a grounded outlet that meets the electrical requirements on the product specification label. 2 Switch the AC ON/OFF switch ON. 3 Connect the backup battery for the Power Failure alarm. 4 Connect the backup battery for the Motion alarm (PC4200h only). 5 Connect the backup battery for the temperature chart recorder. 6 Install the alarm check tube (PC100h). Position the tube below the probe tip. 7 Press the MUTE button if the Low Temperature alarm sounds. NOTE
10.2
► When the door switch is bypassed, the incubator will continue to operate as if the door is closed. ► The door switch controls power to the built-in agitator on PC4200h, and controls power to the internal outlets on all other models. ► The door switch may be bypassed by opening the door and pulling the switch cylinder.
Configure a Platelet Agitator for Use in a Horizon Series Incubator (Optional) Helmer Horizon Series platelet agitators may be installed in a stand-alone configuration, or in Helmer Horizon Series platelet incubators. NOTE
10.3
Refer to the platelet agitator service manual for information regarding the installation of a platelet agitator in a platelet incubator.
Load the PC4200h Built-In Agitator NOTE
Model PF4200h CAUTION
► This chapter applies to PC4200h models only. ► For instsructions in loading all other platelet agitator models, refer to the platelet agitator operation manual. Capacity Random Bags Apheresis Bags 396
132
When opening drawer, grasp handle (not label holder). Open one drawer at a time.
Open the drawer to be loaded and lay the platelet bags flat. ► Avoid stacking bags. ► Maintain enough space around each bag for air circulation. For thicker bags, remove drawers. ► Place the tubing under or around the bag.
360093-1/P
18
Horizon Series™ Models
10.4
Chamber Setpoint
Temperature monitor/controller display and buttons. NOTE
1 2 3 10.5
► Default setpoint is 22.0 °C ► Displayed chamber temperature may be different than setpoint. ► Refer to the service manual for program parameters and default parameter values.
*
On the temperature controller, press and hold . ► The temperature controller displays the current setpoint. While holding press Up Arrow or Down Arrow to change the temperature setpoint. ► Setpoint changes in 0.1 ºC increments. Release all buttons; the temperature setpoint is changed.
*
Temperature Monitor Setpoints
Temperature monitor/controller display and buttons. Features accessible on the monitoring and control system: ► View current chamber temperature and active alarms ► View and change monitor temperature offset ► View and change temperature alarm setpoints ► View and change alarm delay settings 10.5.1
High Alarm Setpoint NOTE
► Default setpoint is 24.0 °C. ► Alarm activates if the chamber temperature is greater than the alarm setpoint ► Refer to the service manual for program parameters and default parameter values.
Change the setpoint: 1 On the temperature monitor/controller, press and hold Up Arrow and Down Arrow. “tunE” and “oFF” will flash on the display. ► The temperature monitor/controller is now in Level 1 program mode. 2 Press Up Arrow or Down Arrow as necessary to select the “AL.HI” parameter. 3 Hold and press Up Arrow or Down Arrow to change the parameter. 4 Release all buttons; the parameter value is changed.
*
360093-1/P
19