HELMER
Platelet Systems- Incubators and Agitators
Platelet Storage Pro Line Incubator PC100-Pro, PC900-Pro, PC1200-Pro and Agitator PF15Pro,PF48-Pro, PF96-Pro Rev D Operation Manual Oct 2021
Use and Operation Manual
33 Pages
Preview
Page 1
Platelet Storage Instructions for Use and Operation Manual Pro Line Incubator - Pro Line Agitator
Countertop Incubator PC100-Pro PC900-Pro PC1200-Pro
Countertop Agitator PF15-Pro PF48-Pro PF96-Pro
360415/D
Document History Revision
Date
CO
Supersession
Revision Description
A
6 JAN 2020*
15008
n/a
Initial release.
B
19 MAY 2020
15365
B supersedes A
• Added detail regarding agitator speed to Enable or Disable Motion section. • Updated Notified Body information in the Compliance section.
C
16 MAR 2021
15943
C supersedes B
• Added CE mark for agitators to Section 9.1
D
4 OCT 2021
16380
D supersedes C
• Updated Safety Compliance Section
* Date submitted for Change Order review. Actual release date may vary.
Document Updates The document is furnished for information use only, is subject to change without notice and should not be construed as a commitment by Helmer Scientific. Helmer Scientific assumes no responsibility or liability for any errors or inaccuracies that may appear in the informational content contained in this material. For the purpose of clarity, Helmer Scientific considers only the most recent revision of this document to be valid.
Notices and Disclaimers Confidential / Proprietary Notices Use of any portion(s) of this document to copy, translate, disassemble or decompile, or create or attempt to create by reverse engineering or otherwise replicate the information from Helmer Scientific products is expressly prohibited.
Copyright and Trademark Helmer®, i.Series®, i.Center®, AgiTrak™, and Rel.i™ are registered trademarks or trademarks of Helmer, Inc. in the United States of America. Copyright © 2021 Helmer, Inc. All other trademarks and registered trademarks are the property of their respective owners. Helmer, Inc., doing business as (DBA) Helmer Scientific and Helmer.
Disclaimer This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain Helmer Scientific products. Any failure to follow the instructions as described could result in impaired product function, injury to the operator or others, or void applicable product warranties. Helmer Scientific accepts no responsibility or liability resulting from improper use or maintenance of its products. The screenshots and component images appearing in this guide are provided for illustrative purposes only, and may vary slightly from the actual software screens and/or product components. Helmer Scientific 14400 Bergen Boulevard Noblesville, IN 46060 USA www.helmerinc.com
Part No. 360415/Rev D
Helmer Scientific Pro Line Platelet Storage
Instructions for Use
Contents 1
About this Manual... 1 1.1
Intended Audience... 1
1.2
Model Reference... 1
1.3
Intended Use... 1
1.4
Safety Precautions and Symbols... 1
1.5
Avoiding Injury... 2
1.6
General Recommendations... 3
Platelet Incubator (stand-alone) 2
3
Installation... 4 2.1
Location... 4
2.2
Placement and Leveling... 4
2.3
Chart Recorder... 5
Pro Line Platelet Incubator Operation... 6 3.1
Initial Start-Up... 6
3.2
Operation... 7
3.3
Users and Passwords... 7
3.4
Change Temperature Setpoint... 8
3.5
Set Alarm Parameters... 8
3.6
Active Alarms... 9
3.7
Mute and Disable Active Alarms... 9
3.8
Min/Max Temperature Monitoring... 9
3.9
Electronic Access Control (Optional)... 10
Pro Line Platelet Agitator (stand-alone) 4
5
Installation... 11 4.1
Location... 11
4.2
Power Connection... 11
4.3
Mounting Brackets... 12
4.4
Placement, Leveling and Setup... 12
4.5
Storage Configuration... 12
4.6
Load the Platelet Agitator... 13
Pro Line Platelet Agitator Operation... 14 5.1
Initial Start-Up... 14
5.2
Motion Alarm... 14
5.3
Motion Alarm Controls... 14
5.4
Enable or Disable Motion... 15
360415/D
ii
Helmer Scientific Pro Line Platelet Storage
Instructions for Use
Pro Line Platelet Agitator (installed in Platelet Incubator) 6
7
8
Installation... 16 6.1
Location... 16
6.2
Storage Configuration... 16
6.3
Power and Communication Connection... 17
6.4
Agitator Mounting Brackets... 17
6.5
Placement and Leveling... 18
6.6
Load the Platelet Agitator... 18
Pro Line Platelet Agitator Operation... 19 7.1
Initial Start-Up... 19
7.2
Motion Alarm... 19
7.3
Motion Alarm Controls... 19
7.4
Enable or Disable Motion... 21
7.5
AgiTrak Setup and Information... 21
Product Specifications... 23 8.1
9
Operating Standards... 23
Compliance... 25 9.1
Safety Compliance... 25
9.2
Environmental Compliance... 25
9.3
Electromagnetic Compliance... 25
Appendix A: i.Series Active Alarms... 27 Appendix B: i.Series Application Icons... 28
360415/D
iii
Helmer Scientific Pro Line Platelet Storage
1
About this Manual
1.1
Intended Audience
Instructions for Use
This manual provides information on how to use the Pro Line platelet incubator and platelet agitator. It is intended for use by end users of the platelet incubator and platelet agitator and authorized service technicians.
1.2
Model Reference
This manual covers all Pro Line platelet incubators and platelet agitators which may be identified by size or model number.
1.3
Intended Use Note This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Helmer platelet incubators are intended to provide the controlled temperature environment required for the storage of platelet products. Helmer platelet agitators are intended to provide controlled agitation required for the storage of platelet products. The devices are intended to be operated by personnel who have procedures in place for meeting FDA, AABB, EU or any other applicable regulations for the processing and storage of platelet products.
1.4
Safety Precautions and Symbols Symbols found in this document The following symbols are used in this manual to emphasize certain details for the user: Task Indicates procedures which need to be followed. Note Provides useful information regarding a procedure or operating technique when using Helmer Scientific products. NOTICE Advises the user against initiating an action or creating a situation which could result in damage to equipment; personal injury is unlikely.
Symbols found on the units The following symbols may be found on the incubator or agitator or incubator packaging or agitator packaging: Caution: Risk of damage to equipment or danger to operator
EU Authorized Representative
Refer to documentation
360415/D
1
Helmer Scientific Pro Line Platelet Storage
1.5
Instructions for Use
Avoiding Injury
Review safety instructions before installing, using, or maintaining the equipment. Platelet Incubator
Platelet Agitator
• Before moving unit, remove the installed agitator (if applicable). • Before moving unit, ensure door is closed. • Before moving unit, disconnect the AC power cord and secure the cord. • When moving unit, use assistance from a second person. • Never physically restrict any moving component. • Avoid removing electrical service panels and access panels unless so instructed. • Do not store or place objects or liquid containers on top of the incubator. • Keep hands away from pinch points when closing the door or when agitation motion is enabled (if applicable). • Avoid sharp edges when working inside the electrical compartment. • Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good laboratory practices. • Proceed with caution when adding and removing product from the platelet incubator. • Use manufacturer supplied power cord only. • Using the equipment in a manner not specified by Helmer Scientific may impair the protection provided by the equipment. • The platelet incubator is not considered to be a storage cabinet for flammable or hazardous materials.
• Before moving unit, remove contents from the drawers. • Do not open multiple drawers at the same time. • Before moving unit, disconnect the DC power cord and secure the cord. • When moving unit, use assistance from a second person. • Never physically restrict any moving component. • Avoid removing electrical service panels and access panels unless so instructed. • Keep hands away from pinch points when agitation motion is enabled. • Ensure both left and right side drawer stop panels are fully installed prior to operating the agitator. • Avoid sharp edges when working inside the electrical compartment. • Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good laboratory practices. • Proceed with caution when adding and removing products from the platelet agitator. • Use only manufacturer supplied power supply/cord when operating stand-alone or within incubator. • Using the equipment in a manner not specified by Helmer Scientific may impair the protection provided by the equipment. • The platelet agitator is not considered to be a storage cabinet for flammable or hazardous materials.
REQUIRED: Decontaminate parts prior to sending for service or repair. Contact Helmer or your distributor for decontamination
instructions and a Return Authorization Number.
360415/D
2
Helmer Scientific Pro Line Platelet Storage
1.6
Instructions for Use
General Recommendations
General Use Platelet Incubator
Platelet Agitator
Allow platelet incubator or platelet agitator to come to room temperature before switching power on. During initial startup, motion alarm may sound if the motion is disabled. The low temperature alarm may sound while the platelet incubator reaches operating temperature.
Allow platelet agitator to come to room temperature before switching power on. During initial startup, motion alarm may sound if the motion is disabled. During initial startup for stand alone operation, device selection of “X” is required for motion to occur.
Initial Loading After the platelet incubator reaches room temperature, allow chamber temperature to stabilize at the setpoint before storing product. After the platelet agitator reaches room temperature, begin storing product.
360415/D
3
Helmer Scientific Pro Line Platelet Storage
Instructions for Use
Platelet Incubator (stand-alone) 2
Installation
2.1
Location Note Hot ambient temperatures with high humidity may cause condensation on the outside of the incubator.
♦ Has a sturdy, level surface. ♦ Has a grounded outlet meeting national electric code (NEC) and local electrical requirements. ♦ Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents. ♦ Meets limits specified for ambient temperature (15 °C to 35 °C) and relative humidity. ♦ Minimum 24” (610 mm) above for ambient temperatures of 28 °C to 35 °C. ♦ Minimum 4” (102 mm) above for ambient temperatures of 15 °C to 28 °C. ♦ Minimum 12” (305 mm) behind for ambient temperatures of 28 °C to 35 °C. ♦ Minimum 4” (102 mm) behind for ambient temperatures of 15 °C to 28 °C.
2.2
Placement and Leveling • The product should not be used adjacent to other equipment. If adjacent use is necessary, the product should be observed to verify normal operation in the configuration in which it will be used. • The use of accessories other than those specified for the product by Helmer is not recommended. They may result in increased emissions or decreased immunity of the device. • Refer to the Electromagnetic Compliance section for additional information. Notes • Ensure the AC power and backup battery power are turned off on the incubator prior to connecting an agitator. • Rear stand-off brackets are provided with the platelet incubator and should be installed prior to placing the incubator in its location.
1. Align keyhole openings in stand-off bracket with screws on back of incubator and slide down to engage. 2. Tighten screws using a #2 Phillips screwdriver to secure. 3. Place platelet incubator on sturdy surface. 4. Ensure platelet incubator is level
360415/D
4
Helmer Scientific Pro Line Platelet Storage
2.3
Instructions for Use
Chart Recorder Note For complete information, refer to the Temperature Chart Recorder Operation and Service Manual.
A B
F
C D E
Chart recorder with paper and battery installed. Table 1. Chart Recorder Label
Description
Function
A
Left and Right Arrow buttons
Adjust settings and stylus position.
B
LED
Indicates status of chart recorder in operating mode, or selected temperature range in paper change mode.
C
Chart change button
Adjust position of stylus when changing chart paper, or run a test pattern.
D
Stylus
Mark temperature line on paper.
E
Reset button
Restart chart recorder.
F
Backup battery
Provides power during AC power failure. Connect prior to use.
Install / Replace Chart Paper Note For accurate temperature reading, ensure the current time is aligned with the time line groove when the chart knob is fully tightened. 1. Press and hold C button. When stylus begins to move left, release button. The LED flashes. 2. When stylus stops moving, remove chart knob then move knob up and away. 3. Place chart paper on chart recorder. 4. Gently lift stylus and rotate paper so current time line corresponds to time line groove. 5. Hold chart paper in place while making sure the chart knob is fully tightened. (Failure to fully tighten the knob can result in paper slipping and losing time.)
6. Press and hold C button. When stylus begins to move right, release button. 7. Confirm stylus is marking on paper and stops at the correct temperature. 8. Calibrate chart recorder to match primary temperature if needed and close recorder door.
360415/D
Stylus and time line groove
5
Helmer Scientific Pro Line Platelet Storage
3
Pro Line Platelet Incubator Operation
3.1
Initial Start-Up
Instructions for Use
1. Plug the power cord into a grounded outlet meeting the electrical requirements on the product specification label. 2. Turn the AC power switch ON. 3. Turn the backup battery switch ON. 4. The Start screen is displayed.
Start screen The language screen is displayed when the i.C3 is powered on. Use the Language screen to select the i.C3 display language.
Language screens Note English is the default language. If an alarm sounds, temporarily mute the alarm by touching the Mute button.
Home screen Mute icon Note Active alarms are displayed on the Home screen. If an alarm condition other than High Temperature occurs, refer to the service manual for troubleshooting.
360415/D
6
Helmer Scientific Pro Line Platelet Storage
3.2
Instructions for Use
Operation Notes • Refer to the i.C3 User Guide for complete information regarding the i.C3 User Interface. • The i.C³ Home screen displays temperature and alarm information, and provides icons to gain access to other functions of the i.C³. • After two minutes of inactivity, the screensaver will be displayed. To return to the Home screen, touch the screensaver.
Home screen
3.3
Home screensaver (touch to return to Home screen)
Users and Passwords
The Users and Passwords screen provides a way for the administrator-level user to limit access to certain screens. The administrator-level password can be changed by selecting Change Password. The Access Control Setup screen can be opened by choosing the Access Setup button from the Users and Passwords screen as well as from the Access Log screen.
> Enter settings password. Select Users and Passwords.
Users and Passwords screen
Change Password keypad
Note Default Settings password is 1234 Change Password 1. Select the Change Password button. A numeric keypad is displayed. 2. Enter a unique 4-digit code and select the √. A numeric keypad appears 3. Re-enter the 4-digit code to confirm and select the √. 4. Select the Back arrow icon to return to the previous screen or the Home icon to return to the Home screen.
360415/D
7
Helmer Scientific Pro Line Platelet Storage
3.4
Instructions for Use
Change Temperature Setpoint
> Enter Settings password. Select Temperature Setpoints. Touch minus (-) or plus (+) on spin box to change value.
Settings screen Notes • Default Settings password is 1234. • Default setpoint is 22.0 °C.
3.5
Set Alarm Parameters > Enter Settings password. Select Alarm Settings. Touch minus (-) or plus (+) on spin box to set each alarm parameter.
Settings screen
Alarm Settings screens Alarm settings control the circumstances and timing of alarm condition indicators displayed on the i.C3 Home screen.
360415/D
8
Helmer Scientific Pro Line Platelet Storage
3.6
Instructions for Use
Active Alarms
Active alarms are displayed on the Home screen. Refer to Appendix A for a list of potential active alarms.
Home screen with active alarm Notes • When the door switch is bypassed, the incubator and door open alarm continue to operate as if the door is closed. • The door switch may be bypassed by opening the door and pulling the switch cylinder. • The number of agitators, internal and external fans, and heat pumps will vary based on incubator model. • Fan failure alarm may occur if facility power is lost.
3.7
Mute and Disable Active Alarms
Audible alarms may be temporarily muted by touching the Mute icon. The delay duration can be set and changed by selecting Sound Settings from the Settings screen. The duration may be set to any value from 1 - 60 minutes. The delay time remaining will be displayed in the bottom right corner of the icon. If the alarm is still active after the mute delay has ended, the audible alarm will sound.
Unmuted
Muted
> Enter the Settings password. Scroll down to select Sound Settings. Touch minus (-) or plus (+) on spin box to set the mute duration.
3.8
Min/Max Temperature Monitoring
The Min/Max temperature display provides the highest and lowest Primary Monitor probe temperature reading since the last system reset (power-on event) or manually-initiated reset. Touch the Reset icon to the right of the display to manually reset.
Notes • The Min/Max temperature display can be turned on or off through Display Settings. • Once the time reaches the maximum display of 999 hours and 60 minutes, the message will display “>999:60”, but minimum and maximum temperatures will continue to be tracked. 360415/D
9
Helmer Scientific Pro Line Platelet Storage
3.9
Instructions for Use
Electronic Access Control (Optional)
Allows user-specific secure access to the incubator. The Access Control Setup screen can be opened from the Access Log screen or by choosing the Access Setup button from the Users and Passwords screen. Notes • During a power failure, the optional Access Control lock will remain locked. It can be unlocked using backup battery power until battery power is depleted or until the backup battery switch is switched OFF. • Switching the back-up battery key switch OFF will disable the monitoring system during a power failure. • During a power failure, use the mechanical door key to provide secure storage for incubator contents. • Refer to i.C3 User Guide for complete information regarding Access Control. Setup Configure and manage user-specific accounts to allow controlled access to the incubator.
Access Control Setup password screen
Access Control Setup screen
Enter the supervisor PIN to set up Access Control and follow the on-screen prompts to set up users. Notes • Initial factory supervisor PIN is 5625 • The supervisor PIN cannot be deleted, and should be changed to prevent unauthorized user ID setup. The supervisor PIN does not allow access to the unit. At least one user ID must be set up to gain access to the unit. Open Incubator with Access Control
Access Control Home screen Enter a valid PIN using the keypad.
360415/D
10
Helmer Scientific Pro Line Platelet Storage
Instructions for Use
Pro Line Platelet Agitator (stand-alone) 4
Installation
4.1
Location • The product should not be used adjacent to other equipment. If adjacent use is necessary, the product should be observed to verify normal operation in the configuration in which it will be used. • The use of accessories other than those specified for the product by Helmer is not recommended. They may result in increased emissions or decreased immunity of the device. • Refer to the Electromagnetic Compliance section for additional information. Note Add 1.5” (38 mm) to the agitator width to accommodate the trolley frame when agitation motion is enabled.
♦ To ensure continuous operation of linearly shifting loads, the location surface must be level and adequately accommodate the full weight of the agitator when loaded with product. ♦ Has a grounded outlet meeting the electrical requirements listed on the product specification label. ♦ Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents. ♦ Minimum 0.5” (13 mm) behind. ♦ Minimum 0.75” (20 mm) on left and right sides. ♦ Meets limits specified for ambient temperature (15˚C to 35˚C) and relative humidity.
4.2
Power Connection Note Use only the optional Helmer AC/DC power supply for stand-alone configuration.
The platelet agitator may be used in a stand-alone configuration. A power supply with adaptable plugs is available for stand-alone use. Select and install the desired plug prior to attaching the power supply to the agitator. The power supply is not used when configured inside a Pro Line platelet incubator.
Configure and Attach Power Supply 1. Remove the cover plate by using the thumb to push and hold the spring loaded locking key downward while sliding the plate forward. Retain the cover plate in secure location for future use. 2. Select the desired plug and slide in place until it locks (a clicking sound will occur).Make sure the plug is firmly attached. 3. Attach the power supply to the platelet agitator, and ensure the rotating lock is finger tight prior to plugging power supply into facility AC.
360415/D
11
Helmer Scientific Pro Line Platelet Storage
4.3
Instructions for Use
Mounting Brackets
Mounting brackets are included and may be installed for use inside the incubator or in stand-alone configuration. Install Mounting Brackets 1. Carefully place the agitator on its back on a solid surface allowing access to the bottom of the unit. 2. Locate the two nutserts in the right or left side toward the front of the unit and align with the two holes in the mounting bracket. 3. Hand thread the screws through each hole, and secure using a #2 Phillips screwdriver. 4. Repeat steps 2 and 3 for the opposite side. 5. Return the agitator to the upright position. 6. Carefully place the agitator in the desired location aligning the hole in each bracket with the holes in the mounting surface. 7. Hand thread the screws through the bracket and into the mounting surface. Secure using a #2 Phillips screwdriver.
4.4
Placement, Leveling and Setup NOTICE • To prevent damage to the platelet agitator, lift using the ends of the base. Do not use the platelet agitator storage frame, trolley or trolley drawer to lift agitator. If the platelet agitator base is not accessible, lift using the ends of the storage frame. • The platelet agitator communication switch is fragile, do not use excessive force when changing the setting.
1. Place platelet agitator on sturdy surface. 2. Ensure platelet agitator is level. 3. Using a small flathead screwdriver, turn the communication switch to the X position. Ensure the arrow (shown in red for visibility in the adjacent picture) is pointing to the X.
4.5
Storage Configuration
Drawers can be removed or moved to create additional storage space. Label holders are available and may be installed on drawers.
To avoid injury, ensure both left and right side drawer stop panels are fully installed prior to operating the agitator. Thumb Screws
Remove and Replace Drawers 1. Remove the thumb screws securing the drawer stop panels to the left and right sides of the agitator. (Note the orientation of each panel) 2. Carefully pull each panel from the agitator and set panels and thumb screws aside. 3. Slide the drawer(s) out and remove. 4. Reinstall drawer in desired location by aligning the outer edges of the drawer with the slots in the drawer guides and push inward. 5. Reinstall the drawer stop panels in the same orientation as removed, and secure with thumb screws ensuring they are finger tight.
Install Label Holders (optional)
1. Insert the tabs on the label holder into the slots on the drawer. 2. Pivot the holder around the drawer handle and align the hole on the label holder with the corresponding hole on the drawer. 3. Push thumb screw through the hole in the label holder and through the hole in the drawer to secure.
360415/D
12
Helmer Scientific Pro Line Platelet Storage
4.6
Instructions for Use
Load the Platelet Agitator
Table 2. Agitator Capacity Model
Capacity WBD/Random Bags
SDP/Apheresis Bags
PF15-Pro
15 (2 per drawer; 3 per shelf)
7 (1 per drawer/shelf)
PF48-Pro
48 (6 per drawer/shelf)
16 (2 per drawer/shelf)
PF96-Pro
96 (12 per drawer/shelf)
32 (4 per drawer/shelf)
PF15-Pro
NOTICE When opening drawer, grasp the center of the handle (not label holder). Open one drawer at a time. Open the drawer to be loaded and lay the platelet bags flat. The top of the storage frame may also be used for bag storage. Avoid stacking bags. Maintain enough space around each bag for air circulation. For thicker bags, remove and/or adjust drawers. Place the bag tubing under or around the bag.
360415/D
13
Helmer Scientific Pro Line Platelet Storage
5
Pro Line Platelet Agitator Operation
5.1
Initial Start-Up
Instructions for Use
Use only rechargeable 9V NiMH batteries (1 included) for backup power to the motion alarm. Notes • Backup battery may need charging for 24 to 48 hours prior to use. Charging occurs when system is on facility power. • Turning the alarm ON/OFF switch ON turns the motion alarm on and further allows the backup battery to recharge. When in the OFF position, the Alarm is not activated and the battery will not recharge. 1. Plug the power supply cord into a grounded outlet that meets the electrical requirements on the product specification label. 2. Switch the alarm ON/OFF switch ON. 3. Select alarm volume and alarm delay settings. 4. Set the communication switch to “X” when in stand-alone configuration using a small flathead screwdriver. 5. Switch power ON/OFF switch ON.
5.2
Motion Alarm
♦ Enable the motion alarm when using the platelet agitator. The red LED will illuminate. ♦ Disable the motion alarm when not using the platelet agitator. Lack of motion triggers the alarm.
5.3
Motion Alarm Controls
Motion alarm switch.
Alarm volume and delay controls
When platelet agitator motion stops the motion alarm is activated. The alarm condition is communicated as follows: ♦ Red alarm LED on motion alarm switch flashes. ♦ Audible alarm buzzer sounds when in stand-alone mode, the motion alarm switch is turned on, adjustable alarm delay time has been exceeded, and volume has been turned up. ♦ Through a dry (no voltage) connection to an external monitoring device (if connected). ♦ Through a 9V connection to an external monitoring device (if connected). Enable and Disable Motion Alarm 1. Switch the motion alarm ON/OFF switch ON. 2. Set volume to desired level. 3. Switch the motion alarm ON/OFF switch OFF. NOTICE If motion stops while the motion alarm is switched OFF, communication of the alarm (visual, audible, and signals to external devices) is suppressed.
360415/D
14
Helmer Scientific Pro Line Platelet Storage
Instructions for Use
Alarm Delay The duration of time between when agitation stops and when the alarm sounds is the alarm delay. The alarm delay is set using the alarm delay control. Notes • The minimum alarm delay that can be set is approximately 10 seconds. • Maximum alarm delay is approximately 10 minutes. • The default motion alarm delay is set at the halfway point (approximately 4 to 5 minutes). Set Alarm Delay 1. Using a small flat-head screwdriver, rotate the control to the left (counterclockwise) to shorten the motion alarm delay. 2. Using a small flat-head screwdriver, rotate the control to the right (clockwise) to extend the motion alarm delay. Alarm Volume The motion alarm volume has a variable setting. Notes • Rotating the volume control fully counter-clockwise will silence the audible alarm. If the motion alarm switch is turned ON, the LED will blink providing a visual alarm when the agitator is in an alarm condition. • Alarm volume is applicable only when the agitator alarm switch is in the ON position and the platelet agitator is in stand-alone mode, or when communication is lost when configured in a platelet incubator. Set Alarm Volume Rotate the motion alarm dial to the appropriate position for the desired volume level.
5.4
Enable or Disable Motion
Agitation switch.
Agitator Speed Control
Notes • The agitator speed is factory set to 72 CPM (as displayed in green) and is applicable when used in stand-alone operation. • For a setpoint of 72 CPM, rotate the arrow into the center of the green zone. The green zone represents an expanded 72 CPM area. Any point to the right of the green zone is a setpoint greater than 72 CPM. • Contact Helmer Technical Service for instructions regarding Agitator Speed Verification (T3I468). • For a platelet agitator in stand-alone configuration, switching the agitation ON/OFF switch ON will start the agitation motion. • For a platelet agitator in stand-alone configuration or installed in a platelet incubator, switching the agitation ON/OFF switch OFF will stop the agitation motion. Start/Stop Agitation 1. Select the desired speed from 40 to 80 CPM. 2. Load items into the platelet agitator. 3. Switch the agitation ON/OFF switch ON to begin movement. 4. Switch the motion alarm switch ON/OFF switch ON. 5. Switch the motion alarm switch ON/OFF switch OFF. 6. Switch the agitation ON/OFF switch OFF to stop movement. 7. Remove items from the platelet agitator. 360415/D
15
Helmer Scientific Pro Line Platelet Storage
Instructions for Use
Pro Line Platelet Agitator (installed in Platelet Incubator) 6
Installation
6.1
Location Notes • Only Helmer Pro Line platelet agitator may be used with Pro Line platelet incubator. • Hot ambient temperatures with high humidity may cause condensation on the outside of the incubator. • When placing a Pro Line agitator in the PC900-Pro or PC1200-Pro, ensure agitator placement allows the roll-top door to open. • Add 1.5” (38 mm) to the agitator width to accommodate the trolley frame when agitation motion is enabled.
To ensure continuous operation of linearly shifting loads, the location surface must be level and adequately accommodate the full weight of the incubator with installed agitator when loaded with product.
6.2
Storage Configuration
Drawers can be removed or moved to create additional storage space. Label holders are available and may be installed on drawers.
To avoid injury, ensure both left and right side drawer stop panels are fully installed prior to operating the agitator. Thumb Screws
Remove and Replace Drawers 1. Remove the thumb screws securing the drawer stop panels to the left and right sides of the agitator. (Note the orientation of each panel) 2. Carefully pull each panel from the agitator and set panels and thumb screws aside. 3. Slide the drawer(s) out and remove. 4. Reinstall drawer in desired location by aligning the outer edges of the drawer with the slots in the drawer guides and push inward. 5. Reinstall the drawer stop panels in the same orientation as removed, and secure with thumb screws ensuring they are finger tight.
Install Label Holders (optional)
1. Insert the tabs on the label holder into the slots on the drawer. 2. Pivot the holder around the drawer handle and align the hole on the label holder with the corresponding hole on the drawer. 3. Push thumb screw through the hole in the label holder and through the hole in the drawer to secure.
360415/D
16