HELMER
Platelet Systems- Incubators and Agitators
Pro Line Incubator PC100-Pro, PC900-Pro, PC1200-Pro Rev C Service Manual Sept 2021
Service Manual
39 Pages
Preview
Page 1
Pro Line Platelet Incubator Service Manual
Countertop PC100-Pro PC900-Pro PC1200-Pro
360389/C
Document History Revision
Date
CO
Supersession
Revision Description
A
5 JUL 2019
14829
N/A
Initial Release
B
19 MAY 2020
15444
B supersedes A
• Updated Location Requirements to include reference to Product Specifications. • Updated the use of caution symbols and statements throughout the document.
C
08 SEP 2021
16200
C supersedes B
• Added Battery Alarm Test to Alarm Tests section.
* Date submitted for Change Order review. Actual release date may vary.
Document Updates The document is furnished for information use only, is subject to change without notice and should not be construed as a commitment by Helmer Scientific. Helmer Scientific assumes no responsibility or liability for any errors or inaccuracies that may appear in the informational content contained in this material. For the purpose of clarity, Helmer Scientific considers only the most recent revision of this document to be valid.
Notices and Disclaimers Confidential / Proprietary Notices Use of any portion(s) of this document to copy, translate, disassemble or decompile, or create or attempt to create by reverse engineering or otherwise the information from Helmer Scientific products is expressly prohibited. Copyright and Trademark Helmer®, i.Series®, i.Center®, i.C3®, AgiTrak™, and Rel.i™ are registered trademarks or trademarks of Helmer, Inc. in the United States of America. Copyright © 2021 Helmer, Inc. All other trademarks and registered trademarks are the property of their respective owners. Helmer, Inc., doing business as (DBA) Helmer Scientific and Helmer. Disclaimer This manual is intended as a guide to provide the operator with necessary instructions on the proper use and maintenance of certain Helmer Scientific products. Any failure to follow the instructions as described could result in impaired product function, injury to the operator or others, or void applicable product warranties. Helmer Scientific accepts no responsibility for liability resulting from improper use or maintenance of its products. The screenshots and component images appearing in this guide are provided for illustrative purposes only, and may vary slightly from the actual software screens and/or product components.
Helmer Scientific 14400 Bergen Boulevard Noblesville, IN 46060 USA www.helmerinc.com
Part No. 360389/ Rev C
Helmer Scientific Platelet Incubator Pro
Service Manual
Contents 1
2
3
4
5
6
About this Manual... 3 1.1
Intended Audience... 3
1.2
Model Reference... 3
1.3
Intended Use... 3
1.4
Safety Symbols and Precautions... 3
1.5
Avoiding Injury... 4
1.6
Product Labels... 5
Installation and Configuration... 6 2.1
Location Requirements... 6
2.2
Placement and Leveling... 6
2.3
Connect Back-Up Power... 6
2.4
Platelet Agitator and Platelet Incubator Compatibility... 7
2.5
Install Pro Series Platelet Agitator (Optional)... 7
2.6
Temperature Probes... 8
2.7
Chart Recorder... 9
2.8
External Monitoring Devices... 10
Controls... 11 3.1
Home Screen and Screensaver... 11
3.2
Home Screen Functions... 11
3.3
Event Log... 12
3.4
Alarm Reference... 13
3.5
Settings... 14
3.6
Sensor Calibration... 17
3.7
AgiTrak™ System... 20
Maintenance... 21 4.1
Alarm Tests... 22
4.2
Test and Replace Back-up Battery... 25
4.4
Upgrade System Firmware... 27
4.5
Clean the Platelet Incubator... 27
Service... 28 5.1
Condensate Tray... 28
5.2
CSV Download... 28
Troubleshooting... 30 6.1
General Operation Problems... 30
6.2
Chamber Temperature Problems... 31
6.3
Alarm Activation Problems... 31
7
Parts... 32
8
Schematics... 34 8.1
PC100-Pro, PC900-Pro, PC1200-Pro... 34
Appendix A: Warranty... 37
360389/C
ii
Helmer Scientific Platelet Incubator Pro
1
About this Manual
1.1
Intended Audience
Service Manual
This manual provides information on how to use the platelet incubator Pro. It is intended for use by end users of the platelet incubator and authorized service technicians.
1.2
Model Reference
This manual covers all Pro Series platelet incubators which may be identified by size or model number.
1.3
Intended Use Note This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Helmer platelet incubators are intended to provide the controlled temperature environment required for the storage of platelet products. The devices are intended to be operated by personnel who have procedures in place for meeting FDA, AABB, EU or any other applicable regulations for the processing and storage of platelet products.
1.4
Safety Symbols and Precautions Symbols and precautions found in this document The following symbols are used in this manual to emphasize certain details for the user: Task Indicates procedures which need to be followed. Note Provides useful information regarding a procedure or operating technique when using Helmer Scientific products. NOTICE Advises the user against initiating an action or creating a situation which could result in damage to equipment; personal injury is unlikely.
Symbols found on the units The following symbols may be found on the incubator or incubator packaging: Caution: Risk of damage to equipment or danger to operator
360389/C
Refer to documentation
3
Helmer Scientific Platelet Incubator Pro
1.5
360389/C
Service Manual
Avoiding Injury Review safety instructions before installing, using, or maintaining the equipment. ♦ Before moving unit, remove the installed agitator (if applicable). ♦ Before moving unit, ensure door is closed. ♦ Before moving unit, disconnect the AC power cord and secure the cord. ♦ When moving unit, use assistance from a second person. ♦ Never physically restrict any moving component. ♦ Avoid removing electrical service panels and access panels unless so instructed. ♦ Do not store or place objects or liquid containers on top of the incubator. ♦ Keep hands away from pinch points when closing the door or when agitation motion is enabled (if applicable). ♦ Avoid sharp edges when working inside the electrical compartment. ♦ Ensure biological materials are stored at recommended temperatures determined by standards, literature, or good laboratory practices. ♦ Proceed with caution when adding and removing product from the platelet incubator. ♦ Use manufacturer supplied power cord only. ♦ Using the equipment in a manner not specified by Helmer Scientific may impair the protection provided by the equipment. ♦ The platelet incubator is not considered to be a storage cabinet for flammable or hazardous materials. ♦ REQUIRED: Decontaminate parts prior to sending for service or repair. Contact Helmer or your distributor for decontamination instructions and a Return Authorization Number.
4
Helmer Scientific Platelet Incubator Pro
1.6
Service Manual
Product Labels Note Service information varies depending on the model and power requirements.
This information appears on the product specification label, on the back of the platelet incubator. A B
C
Sample Product Specification label.
Label
Description
A
Model
B
Serial number (SN)
C
Power requirements
360389/C
5
Helmer Scientific Platelet Incubator Pro
2
Installation and Configuration
2.1
Location Requirements
Service Manual
Note Hot ambient temperatures with high humidity may cause condensation on the outside of the unit. ♦ Has a sturdy, level surface. ♦ Has a grounded outlet meeting national electric code (NEC) and local electrical requirements. ♦ Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents. ♦ Meets specified limits for ambient temperature and relative humidity as stated in the Product Specification section of the Platelet Storage Instructions for Use and Operation manual. ♦ Minimum 24” (610 mm) above for ambient temperatures of 28 °C to 35 °C. ♦ Minimum 4” (102 mm) above for ambient temperatures of 15 °C to 28 °C. ♦ Minimum 12” (305 mm) behind for ambient temperatures of 28 °C to 35 °C. ♦ Minimum 4” (102 mm) behind for ambient temperatures of 15 °C to 28 °C.
2.2
Placement and Leveling Note Rear stand-off brackets are provided with the unit and should be installed prior to placing the incubator in its location. 1. Align keyhole openings in stand-off bracket with screws on back of incubator and slide down to engage. 2. Tighten screws using a #2 Phillips screwdriver to secure. 3. Place platelet incubator on sturdy surface. 4. Ensure platelet incubator is level.
2.3
Rear Stand-off Bracket
Connect Back-Up Power
The monitoring and alarm system, and chart recorder each have a back-up battery, enabling a period of continuous operation if power is lost. The PC100-Pro has an optional additional full system back-up battery enabling a period of continuous operation of the incubator and agitator if power is lost. Battery life varies by manufacturer as well as voltage level remaining. The battery must have 48 consecutive hours of power to be fully charged. Providing full power is available and no battery-related alarms are active, back-up power for the monitoring system is available for up to 18 hours. The batteries are located on the top of the platelet incubator, under a removable access panel. The back-up battery switch is OFF for shipping. Turn the back-up battery switch ON to provide the monitoring system with back-up power in the event of AC power failure. Notes • The monitoring system will start on battery power alone. If the platelet incubator was previously not connected to AC power and the batteries are connected, the monitoring system will begin running on battery power. • If AC power is lost, the monitoring system will continue to collect data until battery power is depleted. • Use only a 12V 7AHr SLA battery for the standard monitoring and alarm system battery back-up. • Use only a 12V 18AHr SLA battery (quantity 2) for the PC100-Pro integrated full system battery back-up.
Before installing or replacing batteries, switch the main power switch to OFF and the back-up battery switch to OFF. Disconnect the power cord from the power receptacle.
360389/C
6
Helmer Scientific Platelet Incubator Pro
2.4
Service Manual
Platelet Agitator and Platelet Incubator Compatibility
Pro Series platelet incubators may only be fitted with Pro Series platelet agitators. When installed in a Pro Series platelet incubator, the Pro Series platelet agitator can be monitored by the platelet incubator’s AgiTrak™ system. The AgiTrak™ system receives alarm information through a direct connection from the platelet agitator’s alarm system to the platelet incubator’s alarm system. Note Only Pro Series platelet incubators can monitor Pro Series platelet agitators. Platelet Incubator
Compatible Helmer Platelet Agitator
PC100-Pro
PF15-Pro
PC900-Pro
PF48-Pro
PC1200-Pro
PF96-Pro
2.5
Install Pro Series Platelet Agitator (Optional)
Helmer Pro Series platelet agitators can be installed inside Helmer Pro Series platelet incubators. The agitator can be connected to the incubator through a data cable. The connection allows the platelet incubator to monitor the platelet agitator motion, and to sound an alarm when agitation motion stops. Power is supplied to the agitator through the coiled DC power cable. Connect the data cable and DC power cable supplied with the incubator prior to placing the agitator inside the incubator. The agitator communication switch must be set to position 1 for use in the incubator. The power cable has a rotating locking feature on each end to ensure a secure connection. When installing the agitator inside the incubator, finger tighten each rotating lock. For additional information, refer to the Platelet Agitator Manuals. Notes • Only Helmer Pro series platelet agitator may be used with Pro series platelet incubator. • Use only the manufacturer supplied power cord when installing the Pro series agitator in a Pro series Incubator. • Retain the agitator AC/DC power supply if shipped with the agitator for potential use in stand-alone configuration. • Ensure the AC power and back-up battery power switches are turned off on the incubator prior to connecting an agitator. • Ensure data cable is carefully positioned to the right of the agitator to prevent damage caused by agitation motion. • The communication switch is fragile, do not use excessive force when changing the setting. • To ensure continuous operation of linearly shifting loads, the location surface must adequately accommodate the full weight of the incubator with installed agitator when loaded with product. • Incubator restraining brackets are included in packaging and recommended when configured with an agitator operating at a setpoint speed greater than 75 cycles per minute (CPM), or when located on a slick surface.
Agitator Power Connection
PC900-Pro with PF48-Pro installed
360389/C
Agitator communication switch
7
Helmer Scientific Platelet Incubator Pro
2.6
Service Manual
Temperature Probes Notes • Temperature probes are fragile; handle with care. • Remote probes may also be introduced through the rear port.
The temperature monitor and control probes are located on the right side of the unit. NOTICE • The incubator power switch and battery back-up switch must be turned OFF and the AC power cord disconnected from the power receptacle prior to installing an external probe. • DO NOT attach remote probe wires to the AC Terminal block wires or the AC power supply. AC Power Supply
Remote probe routing path Mounting clip
Install Remote Probe 1. Switch the AC power switch to OFF; switch the back-up battery switch to OFF; disconnect the power cord from the power receptacle. 2. Open incubator door and locate the access port plug at the top of the chamber. Use a flat blade screwdriver to pry off the access port plug. 3. Use a #2 Phillips screwdriver to remove the screws securing the top cover, and set top cover and screws aside. 4. Locate the access port on the top of the unit, and peel back the putty. 5. Install the remote probe through the rear panel of the incubator, down through the hole in the top of the unit, and into the chamber. 6. Attach the remote probe to the plastic mounting clip inside the chamber. 7. Secure the remote probe wires to the wire bundle in the top of the unit using wire ties. 8. Reseal the top access port with putty. 9. Reinstall the top cover and secure with screws using a #2 Phillips screwdriver. 10. Reconnect the power cord to the wall receptacle. Switch the main power switch to ON; switch the back-up battery switch to ON. 11. Allow the temperature to stabilize at the setpoint before moving inventory back into the incubator.
360389/C
8
Helmer Scientific Platelet Incubator Pro
2.7
Service Manual
Chart Recorder Note For complete information, refer to the Temperature Chart Recorder Operation and Service Manual.
The chart recorder has a battery system, enabling a period of continuous operation if power is lost. Battery life varies by manufacturer as well as voltage level remaining. Providing full power is available, back-up power for the temperature chart recorder is available for up to 14 hours. Prior to use: ♦ Install battery. ♦ Add paper. ♦ Calibrate chart recorder to match upper chamber temperature. Set up and Operation Open the door by pressing and releasing.
Chart recorder
Chart recorder (door open)
Install Battery Connect the leads to the battery to provide back-up power to the chart recorder. Install or Replace Chart Paper Note For accurate temperature reading, ensure that current time is aligned with time line groove when chart knob is tightened.
Chart recorder stylus and time line groove
1. Press and hold C button. When stylus begins to move left, release button. The LED flashes to indicate current temperature range. 2. When stylus stops moving, remove chart knob then move knob up and away. 3. Place chart paper on chart recorder. 4. Gently lift stylus and rotate paper so current time line corresponds to time line groove. 5. Hold chart paper in place while making sure the chart knob is fully tightened. (Failure to fully tighten the knob can result in paper slipping and losing time.)
6. Confirm temperature range is set to the correct value. 7. Press and hold C button. When stylus begins to move right, release button. 8. Confirm stylus is marking temperature correctly.
360389/C
9
Helmer Scientific Platelet Incubator Pro
2.8
Service Manual
External Monitoring Devices
The remote alarm interface is a relay switch with three terminals: ♦ Common (COM) ♦ Normally Open (NO) ♦ Normally Closed (NC) Terminals are dry contacts and do not supply voltage. Interface circuit is either normally open or normally closed, depending on terminals used. Requirements for your alarm system determine which alarm wires must connect to terminals. NOTICE • The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm system(s) that uses normally-open or normally-closed dry contacts. • If an external power supply exceeding 33V (AC) RMS or 30V (DC) is connected to the remote alarm monitoring system’s circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user. • For any platelet incubator without a platelet agitator installed, a bypass plug (PN 402149-1) is needed for a normallyclosed interface circuit. The terminals on the remote alarm interface have the following maximum load capacity: ♦ 1A at 33V (AC) RMS or 30V (DC) Connect to Remote Alarm Interface 1. Switch the Battery ON/OFF key switch OFF. Switch the AC ON/OFF switch OFF. 2. On the alarm panel, locate the remote alarm terminals. 3. Connect remote alarm wires to appropriate terminals, according to requirements for your alarm system. 4. Use a cable tie to relieve strain on alarm wires (as necessary). 5. Switch the Battery ON/OFF switch ON. Switch the AC ON/OFF switch ON. 6. Touch MUTE to disable the high temperature alarm while platelet incubator reaches operating temperature.
360389/C
Remote alarm interface.
10
Helmer Scientific Platelet Incubator Pro
3
Service Manual
Controls
i.Series -Pro platelet incubators are equipped with the i.C3 monitoring and control system. The i.C3 system combines temperature control and monitoring into a single user interface. Note Please refer to the i.C3 User Guide for complete information regarding the i.C3 User Interface.
3.1
Home Screen and Screensaver
The Home Screen is the default screen and is displayed when: ♦ The Home icon is touched from any other screen ♦ There is no interaction for two minutes on any screen other than those used to enter a password.
Home Screen Screensaver
3.2
Home Screen Functions Note Refer to the i.C3 User Guide for options available on all i.C3 screens.
♦ View current interior temperature readings ♦ View current system time and date ♦ Access any of five home screen applications (touch i.C³ APPS for additional applications) ♦ View information about current alarm events ♦ View whether the monitoring system is running on battery power ♦ Mute audible alarms ♦ View a graph of the interior temperature ♦ View unit ID ♦ Shortcut to Event Log
360389/C
11
Helmer Scientific Platelet Incubator Pro
3.3
Service Manual
Event Log
Event Log screen
Event Log Detail screen
View Event Log Detail 1. From the Home screen, touch the Event Log icon. The Event Log screen appears. 2. Touch an Event. The Event Log Detail screen appears 3. Select the Back arrow to return to the previous screen or the Home icon to return to the Home screen. Acknowledge an Event 1. From the Home screen, touch the Event Log icon. The Event Log screen appears. 2. Touch an Event. The Event Log Detail screen appears 3. Touch the Event Cause button at the bottom of the screen. A possible cause menu appears. 4. Select the appropriate cause for the event or select Other and use the alphanumeric keyboard to make an entry. 5. Touch the Action Taken button at the bottom of the screen. A possible actions menu appears. 6. Select the appropriate action taken or select Other and use the alphanumeric keyboard to make an entry. 7. Touch the Signature button at the bottom of the screen. An alphanumeric keyboard appears. 8. Enter your signature. 9. Touch to save or to cancel.
360389/C
12
Helmer Scientific Platelet Incubator Pro
3.4
Service Manual
Alarm Reference
If an alarm condition is met, an alarm activates. Some alarms are visual only; others are visual and audible. Some Alarms are sent through the remote alarm interface. The table below indicates if an alarm is audible (A), visual (V), or sent through the remote alarm interface (R). Notes • Agitator Alarms (shaded) are configured through AgiTrak™ and appear only when a platelet agitator is installed inside a platelet incubator and the agitator power switch is ON. • Disconnecting the agitator will not clear active agitator alarms. All active agitator alarms will remain active on the monitor until disabled in the AgiTrak™ system or the agitator is reconnected and the problem is corrected. Table 1. i.Series Alarm Reference Alarm
Alarm Type
Alarm
Alarm Type
Primary Monitor Probe High Temperature
A, V, R
Internal Fan Failure
A, V, R
Primary Monitor Probe Low Temperature
A, V, R
Heat Pump 1 Failure
A, V, R
Heat Pump 1 High Temp
A, V, R
Heat Pump 1 Connection Error
A, V, R
Door Open
A, V, R
Communication Failure (1,2 or 3)
A, V, R
Low Battery
V
MPB Communication Failure
A, V, R
No Battery
A, V, R
Drive Space is Low
A, V, R
Control Probe Failure
A, V, R
Drive Space is Full
A, V, R
Primary Monitor Probe Failure
A, V, R
Agitator High Speed
A, V, R
Power Failure
A, V, R
Agitator Low Speed
A, V, R
Condensate Tray Full
A, V, R
Agitator Communication Failure
A, V, R
External Fan Failure (1 or 2)
A, V, R
Agitator Maintenance Required
A, V, R
360389/C
13
Helmer Scientific Platelet Incubator Pro
3.5
Service Manual
Settings
> Through the i.C3 monitoring and control system, current settings may be viewed and changed. To view settings, touch i.C3 APPS, Settings. Use a touch-drag motion to scroll up or down to select the desired setting.
Settings screen
Notes • If the Settings screen is password protected, enter the appropriate password. If viewing settings for the first time, enter the factory default password of “1234”. • Default values for general settings, alarm settings, and display settings are available in the i.C3 User Guide. • Changing temperature settings affects operation of the incubator. Do not change settings unless instructed in product documentation or by Helmer Technical Service. The i.C3 temperature and controller is programmed at the factory. To change a setting, first enter the Settings screen, then select the setting. The method for accessing the Settings mode for each setting varies. Temperature Setpoint The setpoint is the temperature at which the incubator operates. The factory default setting for the primary monitor probe is 22.0 °C. Change Temperature Setpoint 1. Touch i.C³ APPS, Settings. 2. Enter the Settings password (default password is “1234”). 3. Touch minus (–) or plus (+) on the Temperature Setpoint spin box to select the desired value.
360389/C
14
Helmer Scientific Platelet Incubator Pro
Service Manual
User Configurable Alarm Settings The following alarm settings may be changed by the operator. The setpoint for temperature alarms may be changed (where applicable), as well as the time delay between when the alarm condition commences and when the visual and audible alarms are initiated. Note Agitator Alarms (shaded) are configured through AgiTrak™ and appear only when a platelet agitator is installed inside a platelet incubator and the agitator power switch is ON. Table 2. User Configurable Alarms Alarm
Description
Default Setting
Default Time Delay
Primary Monitor Probe High Temp
High temperature at which alarm condition occurs
24 °C
0 minutes
Primary Monitor Probe Low Temp
Low temperature at which alarm condition occurs
20 °C
0 minutes
Power Failure
Time after power failure occurs until alarm sounds
-
1 minute
Probe Failure
Time after probe failure occurs until alarm sounds
-
0 minutes
Door Open (Time)
Time door remains open until alarm sounds
-
3 minutes
Heat Pump 1 High Temp
High temperature at which alarm condition occurs
50 °C
1 minute
Agitator High Speed
High speed at which alarm condition occurs
75 CPM
1 minute
Agitator Low Speed
Low speed at which alarm condition occurs
65 CPM
1 minute
Alarm Settings screens
Change Alarm Setting 1. Touch i.C3 APPS, Settings. 2. Enter the Settings password (default password is “1234”). 3. Touch Alarm Settings. 4. Touch minus (-) or plus (+) on the spin box corresponding to the alarm setting to be changed. 5. Touch Home to exit the Alarm Settings screen.
360389/C
15
Helmer Scientific Platelet Incubator Pro
Service Manual
Non-Configurable Alarms The following alarms indicate operation conditions which require the attention of the operator or a qualified service technician. Note Agitator Alarms (shaded) are configured through AgiTrak™ and appear only when a platelet agitator is installed inside a platelet incubator and the agitator alarm switch is ON. Table 3. User Non-Configurable Alarms Alarm
Description
Agitator Communication Failure
Occurs if Agitator Alarm Monitoring Enabled is set to YES in the Agitator Setup screen and the Incubator cannot communicate with that agitator
Agitator Maintenance Required
• Occurs if Agitator Alarm Monitoring Enabled is set to YES in the Agitator Setup screen and the cycle count for that agitator reaches 75,686,400 counts (the approximate number of cycles when operating at 72 CPM over a 2 year period) • Resetting the count in the Agitator Status screen clears the alarm
Communication Failure
Communication Failure 1 • Triggered if communication is lost between i.C³ display board and control board. • Unit will continue to run with previously saved settings • Screen will not display temperature changes or alarm conditions • i.C³ system will continue to reset until connection is re-established Communication Failure 2 • Triggered if communication is lost between i.C³ display board and internal system memory • Unit will continue to run with previously saved settings Communication Failure 3 • Triggered if the database is corrupted • The database is archived and a new database is automatically created • Unit will continue to run with previously saved settings
Condensate Tray Full
Triggered when condensate tray is full of water.
Drive Space Low
• Triggered if SD card containing downloadable historical date is approaching capacity • New data will continue to be saved for up to 3 more months • Data can be downloaded, but doing so will not free up capacity (SD card replacement recommended)
Drive Space Full
• Triggered if the SD card containing downloadable historical data has reached its capacity • No new data will be saved • Data can be downloaded, but doing so will not free up capacity (SD card replacement required)
External Fan Failure
Triggered when an external fan speed drops below the acceptable low speed limit
Internal Fan Failure
Triggered when the internal fan has fallen below the acceptable low speed limit
Heat Pump Failure
Triggered when electrical reading indicate the heat pump is not heating/cooling properly
Heat Pump Connection Error
Triggered when temperature readings indicate the heat pump is not heating/cooling properly
Low Battery
Triggered when monitoring system rechargeable battery voltage is low
MPB Communication Failure
Triggered when the i.C³ is not able to communicate with the Incubator specific Master Power Board.
360389/C
16
Helmer Scientific Platelet Incubator Pro
3.6
Service Manual
Sensor Calibration
> Sensor calibration values are programmed at the factory. Calibration values can be viewed and changed through the i.C³ monitoring and control system. To view calibration settings, touch i.C³ APPS, Settings, and select Sensor Calibration.
Settings screen Sensor Calibration screen
Notes • If the Settings screen is password protected enter appropriate password. If viewing settings for the first time, enter factory default password of “1234”. • Offset values are adjusted to the tenth of a degree (0.1). • After one hour of no interaction, the Home screen or Temperature Graph screensaver (if enabled) is displayed. View Sensor Calibration Values 1. Touch the Settings icon. 2. Enter the Settings password. 3. Touch Sensor Calibration. Sensor offset values and their current temperature readings are displayed. 4. Touch Home to return to the Home screen. Verify primary monitor probe is reading chamber temperature correctly by comparing the probe reading to the temperature measured by a calibrated reference thermometer. If the probe is not reading correctly, change the value displayed on the monitor. Notes • If the variance is within acceptable limits for your organization, changing probe settings is optional. • Default setting for chamber temperature is 22.0 ºC. Value is factory-preset • Value can be changed from 20.0 ºC to 35.0 ºC
360389/C
17
Helmer Scientific Platelet Incubator Pro
Service Manual
Calibrate Primary Monitor Probe Model
Location
PC100-Pro
Top right corner
PC900-Pro, PC1200-Pro
Bottom right corner
1. Using the table above, place a calibrated independent reference thermometer traceable per national standards in the specified location. 2. Close the door and allow the chamber temperature to stabilize for 30 minutes. 3. Observe and note the thermometer temperature. If the independent thermometer corresponds to the displayed temperature, proceed to Step 9. 4. Subtract the displayed temperature reading from the independent thermometer reading to determine the offset value. 5. Touch, i.C³ APPS, Settings 6. Enter the Settings password (default password is “1234”). 7. Touch Sensor Calibration. 8. Touch minus (-) or plus (+) on the corresponding spin box to increase or decrease the value by the value calculated in Step 4. The message “New Setting Saved” appears next to the spin box. 9. Remove the thermometer from the chamber and close the door. Control Probe The temperature controller senses chamber temperature through the control probe. Chamber temperature will be controlled as required to match the temperature set point based on the control probe reading. NOTICE • Control Probe Offset is factory-preset and should not be changed. Contact Helmer Technical Service for instructions regarding changing the Control Sensor Offset. • Monitor temperature must be verified and accurate prior to adjusting the Control Sensor Offset.
360389/C
18
Helmer Scientific Platelet Incubator Pro
Service Manual
Factory Default Settings Settings listed below may be simultaneously returned to factory default values. Note The factory default settings may not be the same as the settings that were factory-calibrated before the refrigerator was shipped. Table 4. Default Settings Setting
Restored Value
Home Screen Application Icons
i.C³ APPS, Settings, Temperature Graph, Temperature Alarm Test, AgiTrak
Display Brightness
High (level 3)
Password (for Settings screen)
1234
Sounds
On
Alarm Volume
9
Alarm Tone
3
Mute Duration
15
Temperature Calibration Values
Not affected (values previously entered during setup)
Unit ID
Not affected (previously selected during setup)
Date Format
MM/DD/YY
Day
Not affected (maintained in real-time clock)
Month Year Time Format
12-hour
Minute
Not affected (maintained in real-time clock)
Hour AM/PM Language
Not affected (Language previously selected during setup)
Temperature Units
°C
Password Protection (for Settings screen)
On
Temperature Graph Screensaver
On
Access Control (optional) as Home Page
On (When Access Control is enabled in the Edit Factory Settings screen)
High Temperature Alarm Setpoint
24.0 °C
High Temperature Alarm Time Delay
0 minutes
Low Temperature Alarm Setpoint
20.0 °C
Low Temperature Alarm Time Delay
0 minutes
Power Failure Alarm Time Delay
1 minute
Probe Failure Alarm Time Delay
0 minutes
Door Open (Time) Alarm Time Delay
3 minutes
Heat Pump 1 High Temperature Alarm Setpoint
50.0 °C
Chamber Setpoint
22.0 °C
USB Port
On
Agitator 1 Speed Setpoint
72
Agitator 1 High Speed Alarm
75
Agitator 1 Low Speed Alarm
65
Agitator Motion Alarm Delay
1 minute
Restore Settings 1. Touch the Settings icon. 2. Scroll down and touch the Restore Factory Settings button. The Restore Factory Settings confirmation box appears. 3. Touch to restore the factory settings or to maintain the current settings and clear the message. 360389/C
19