ClinActiv + MCM Instructions for Use Rev 001
Instructions for Use
2 Pages
Preview
Page 1
Installation procedure
1 Instructions for use
ClinActiv® + MCM™ ClinActiv® + >
>
Before placing the patient on the bed and therapeutic mattress
5
Carry out individual risk assessments including but not limited to: •
risk of entrapment,
•
potential falls from the bed,
•
patient in state of confusion,
•
patient’s learning ability,
•
children (aged less than 12 or under 1.46 m tall),
•
persons lacking the mental capacity to recognize risky actions,
•
unauthorized persons,
•
bed/mattress combination.
3
4 = Click
7
Installation Place the rolled up ClinActiv® +/ClinActiv® + MCM™ therapeutic mattress on the top of the bed and unroll it. Check that the foot symbol on the cover is at the foot of the bed. The CPR valve must be positioned at the foot end of the bed and must remain accessible for use in the event of emergency deflation. The mattress must be correctly attached and centered on the sleep surface. The straps must be attached to the mobile parts of the bed and the attachments and power cord must be secured so as not be become entangled in the mobile parts.
>
2
Mobilizing the patient Optimal therapeutic benefits: Adjustable head section raised to between 0° and 45°. The performance of the mattress diminishes at angles greater than 45°.
*166438*
6 10
8 9
Use Indication
Contraindication • • •
atypical anatomy unstable fracture of the spine cervical or skeletal traction.
*166438*
>
CPR
Keys and indicator lights
Place the bed chassis in the CPR position before using the emergency CPR function.
5
Continuous Low Pressure Module (CLP)
7
4
6
3
8
2
To activate the CPR
1
and turn 1 Push clockwise the
B
yellow CPR valve.
2 To cancel the CPR and turn 1 Push anticlockwise the
2
5
>
>
Disinfection
6
3
8
2
To transfer the ClinActiv® + from one bed to another: • Never leave the patient on the mattress. • Make sure that the bed chassis brakes are applied to prevent any accidental movements. • Do not walk on or run over the power cord.
The ClinActiv® + and ClinActiv® + MCM™ mattresses (203 x 92 x 18 cm) are compatible with a number of beds including: • Evolution, • Hill-Rom AvantGuard 1200, • Hill-Rom AvantGuard 1600.
7
4
Transferring a mattress
Compatible mattresses and beds
D
A
Alternated Low Pressure Module (ALP)
yellow CPR valve.
>
C
1 E
B
A
D
MCM™ module remote control
B Keys A - On/Off B - Maximum inflation (P-Max) C - Bed exit alarm D - Alarm silence E - MCM™ mode On/Off*
>
C
Indicator lights 1 - CPR 2 - Microclimate management (MCM™) 3 - Seat cushion inflation 4 - Safety lock* 5 - Continuous/alternated low pressure 6 - Internal pressure indicator 7 - Mains power fault 8 - Mattress malfunction * See user manual, Ref. 166182
Disinfection
ZONES TO BE CLEANED and/or DISINFECTED according to INSTRUCTIONS opposite.
1 Parts to be kept clean at all times. 2 After a patient’s departure. Whenever the bed linen is changed, if the inspection reveals ingress of fluids.
1
Insufficient cleaning = risk of infection (biological danger)!
2
Manufactured by Hill-Rom Inc. Distributed by Hill-Rom Inc. Hill-Rom reserves the right to make changes to the design, characteristics and models without prior notice. The only warranty Hill-Rom makes is the express written warranty extended on the sale or rental of its products. © 2010 Hill-Rom Services, Inc. ALL RIGHTS RESERVED.
w w w. h i l l - r o m . c o m
Ref : 166438 rev.001