Hill-Rom

Hill-Rom 900 Beds

Hill-Rom 900 Basic Initialisation Instruction

Basic Initialisation Instruction

1 Page

Product Information: HR900 [LI900Bx] BASIC HR900 INITIALISATION INSTRUCTION Product Code Description  HR900 [LI900Bx] This instruction is to be used in the event your HR900 bed functions do not work or you are experiencing flashing lights on the Lateral Caregiver Unit [LCU]  Process  If you are experiencing a flashing LCU or audible alarm, please proceed with Initialisation instructions, below. Firstly, attempt to disable the attendant lockout controls on the LCU. If unsuccessful, initialise the system by 1 simultaneously pressing the HEAD DOWN and THIGH DOWN keys, until the system stops beeping. Please note that the two buttons must be activated at exactly the same time, held for approximately 5 seconds or until audible beep ceases. If audible beep ceases, test functionality by 2 running HILO UP until both actuators are fully extended (bed is at its highest point). All sensors and limit switches will now be Initialised (synchronised)  or
File Type: PDF
File Size: 286 KB
File Name: Hill-Rom - Hill-Rom 900 Basic Initialisation Instruction.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions