Hills Healthcare
KH 400 GE QUIK-STAND ECONOMY HOIST Operating, Maintenance and Cleaning Manual KEIM-035
Operating, Maintenance and Cleaning Manual
8 Pages
Preview
Page 1
KH 400 GE QUIK- STAND ECONOMY HOIST
OPERATING, MAINTENANCE & CLEANING MANUAL
KEIM-035
INDEX
PAGE
SPECIFICATIONS
2
SPARE PARTS
3
SAFETY PRECAUTIONS
4
OPERATING INSTRUCTIONS
4
BATTERY CHARGING AND CARE
6
MAINTENANCE INSTRUCTIONS
7
SPECIFICATIONS DIMENSIONS:
Length Width Height Effective lift range Leg gating (inside) Under base clearance Base height Leg height from floor
LIFT SPEED: LIFT DRY WEIGHT: LIFT CAPACITY: CASTORS:
35mm per second +/- 36 Kg (with battery) Max load - 150kg Ø125mm Total lock rear polyurethane castor Ø75mm Swivel grey rubber castors – front. Note: Larger castors available on request. (Easier to move) Epoxy powder coated Padded kneepad vinyl Tested by Linak Australia to standards:- AS 3108-1990 - AS 3200.1-1990
FINISH: ELECTRICAL ITEMS:
KH 400GE
2
1100mm 670mm 970mm 950mm – 1700mm 510mm – 900mm 25mm 100mm 110mm
QUIK STAND ECONOMY HOIST
KH 400GE
3
QUIK STAND ECONOMY HOIST
SAFETY PRECAUTIONS. 1. The hoist is manufactured and tested with safety and function for the carer and patients as the utmost importance. 2. The MAXIMUM safe working load is 150 Kg’s. 3. EMERGENCY LOWERING: The emergency lowering button can be used to lower the hoist if there is a handset failure. The switch is located on the controller under the big red emergency stop switch. Simply press using a ballpoint pen or similar object, DO NOT use a sharp object such as a compass point. 4. Always use the correct slings on the hoist.
5. For safety precautions only properly trained personal should use the hoist.
OPERATING INSTRUCTIONS BATTERY PACK INSTALLATION Simply locate the battery on the mounting bracket and clip the release lever into the secure position. The Hoist is now ready for use.
HANDSET: - RAISE/LOWER: The hoist operates via a hand controller. To raise or lower the hoist, simply press the required switch on the touch pad.
KH 400GE
4
QUIK STAND ECONOMY HOIST
LEG GATING: To OPEN or CLOSE the “Gate” on the legs, simply press the foot pedal to open or close the legs.
SLINGS AVAILABLE KH400SF - PREMIUM STANDING SLING (Medium) with Velcro waist strap. KH400SFB - PREMIUM STANDING SLING (Medium) with buckle waist strap. These slings are used for patients that are up to 110 kg's. KH400SL - PREMIUM STANDING SLING (Large) with Velcro waist strap. KH400SLB - PREMIUM STANDING SLING (Large) with buckle waist strap. These slings are used for patients that are between 100 - 150 kg's. SLING FITTING INSTRUCTION When using a seating sling for example, the front mounting points are used for the chest straps and the back for the leg straps. Please refer to the fitting instructions relevant to your choice of sling. BACK BACK
FRONT
For your convenience Kerry Equipment has supplied a “User Records Chart” (located at the back of this manual). It is used to record patient lifting preferences to which mounting points and sling colours best suit them. (Hint: Copy the chart and keep the original with this manual) KNOB RECOMMENDED USAGE When using the standing hoist there are two recommended mounting points relevant to patient size. The front knobs are used for patients over 5’6” (165cm) and the back knobs are used for shorter patients ie: below 5’6” (165cm).
KH 400GE
5
QUIK STAND ECONOMY HOIST
KNOB 1 FRONT
KNOB 2 BACK
BATTERY CHARGING AND CARE CHARGING: Before the hoist can be operated the battery pack will need to be charged. Minimum 4 hours for flat battery. Please follow the instructions below. Plug the charger unit into the 240 Volt supply outlet. The unit is now ready to accept a battery that requires charging. To charge the battery, simply locate the charging plug securely into the female socket on the base of the battery. If a two-battery system is going to be used, replace the first battery once charged, with the second battery. Charge as above. NOTE: A battery pack may remain on charge indefinitely without any detrimental effects to the battery or charging unit.
BATTERY CARE: The hoist uses fully sealed lead acid batteries. They are made to be recharged on a regular basis. To care for your batteries and obtain the maximum life DO NOT allow the batteries to run totally flat before recharging. The controller is fitted with an audio low battery warning alarm. When this warning alarm sounds complete the lift process, then change the battery and recharge. Kerry Equipment recommends that a 2-battery system be used, this allows one battery to always be on charge while the other is in use, and hence the hoist may never be out of service with a flat battery.
KH 400GE
6
QUIK STAND ECONOMY HOIST
MAINTENANCE INSTRUCTIONS The hoist requires periodical lubrication. These points are: CASTORS. It is always best practice to remove all hair and lint from the castors as soon as possible, always allowing free movement of all the wheels. The head bearing will need lubricating monthly. Simply spray TAC-2 up into the bearing while rotating the castor. The wheel also requires lubricating every month. To do this the wheel must be unbolted and removed from the castor frame. TAC-2 can then be sprayed into the hub centre to lubricate the internal bearings.
BASE PIVOT POINTS and FOOT PEDAL All pivot points will need to be lubricated regularly with TAC-2 or similar lubricant. Simply spray the points for a second or two on either side. The hoist has to be turned onto its side for access to the following lubrication points. (To do this 2 people are required.) Both people are to stand on one side of the hoist. One near the push handle, the other by the side of the hoist. The person near the push handle locks the total lock castors. In conjunction with one another, pull the hoist towards the side, lowering the hoist onto the floor. You now have access to all the lubrication points under the base. i.e.: pivot points, leg lever and leg pivot. Simply spray TAC-2 on all the pivot points.
KH 400GE
7
QUIK STAND ECONOMY HOIST
ARM AND ACTUATOR PIVOT POINTS The arm and actuator pivot points will need to be lubricated regularly with TAC-2 or similar lubricant. Simply spray on each side of the recommended points for a second or two.
TAC-2 is a CRC product and is available from most automotive outlets or from Kerry Equipment. KERRY EQUIPMENT SERVICE VAN READY FOR ALL YOU NEEDS CALL:08 9240 2944
CLEANING INSTRUCTIONS 1.
The hoist can be cleaned with a low strength detergent, warm water and a soft textured cloth. This includes all Epoxy powder coated framework and yoke.
2.
CAUTION Do not immerse any parts in water. Do not allow water/cleaning liquid to enter the battery or controller. Clean up spills immediately.
3.
SLING CARE Protect from direct sunlight. Repair any cuts or fraying immediately. Slings are machine washable. Please refer to care instructions on the sling.
Distributed by: -
KH 400GE
8
QUIK STAND ECONOMY HOIST