ENGLISH
MyoSure Rod Lens Hysteroscope ®
Instructions for Use
U.S. FEDERAL LAW RESTRICTS THIS DEVICE TO SALE BY OR ON THE ORDER OF A PHYSICIAN Read these instructions completely prior to using the MyoSure Rod Lens Hysteroscope. These instructions describe two MyoSure Rod Lens Hysteroscopes: Catalogue number
40-250
Product
Hysteroscope working channel diameter
Outflow channel part number
MyoSure 3 mm Rod Lens Hysteroscope
40-201
Compatible MyoSure Tissue Removal Devices
Note: All Models may not be available in all regions. Contact your Hologic Representative for a list of models available in your region.
MyoSure Tissue Removal Device
10-401
MyoSure LITE Tissue Removal Device 30-401LITE MyoSure REACH Tissue Removal Device 10-401FC
50-250XL
MyoSure XL 4 mm Rod Lens Hysteroscope
50-201XL MyoSure Tissue Removal Device
10-401
MyoSure LITE Tissue Removal Device 30-401LITE MyoSure XL Tissue Removal Device 50-501XL MyoSure REACH Tissue Removal Device 10-401FC
TABLE 1. Device Description
The Operative Hysteroscopy System can be combined with a hysteroscopic fluid management system to provide continuous flow hysteroscopy capability. The hysteroscope is normally coupled to a camera and video display unit for visualization.
The MyoSure Rod Lens Hysteroscope is intended for use in visualizing the uterine cavity and performing operative hysteroscopy procedures including use with the MyoSure Tissue Removal Device. The MyoSure Rod Lens Hysteroscope system includes a removable outflow channel and the hysteroscope. The removable outflow channel is intended to be used to provide a fluid outflow lumen when the hysteroscope is being used in a diagnostic mode. The removable outflow channel includes a sealed entry port to permit the introduction of instrumentation.
The Myosure Rod Lens Hysteroscope is a type BF applied part per IEC 60601-1. Performance Parameters
The reusable rod lens hysteroscope utilizes rod lenses for visualization and fibers for illumination. The hysteroscope includes a working channel designed to accommodate the respective MyoSure Tissue Removal Device. (See Table 1.) 1
Scope Working Length
184 mm
Angle of View
0°
Field of View
80°
ENGLISH
Indications for Use
• When using HF surgical equipment, keep the working part of the active electrode in the field of view to avoid accidental burns. • The hysteroscope and outflow channel are shipped non-sterile. They must be thoroughly cleaned and sterilized before use. • If scope light post adapters have been used, they need to be disassembled, cleaned, and sterilized before every subsequent use. • Uterine perforation can result in possible injury to bowel, bladder, major blood vessels, and ureter. • High energy radiated light emitted from illuminating fiber at the distal end of the scope may give rise to temperatures exceeding 106°F/41°C (within 8 mm in front of the scope). Do not leave tip of scope in direct contact with the patient tissue or combustible materials, as burns may result. Lower the light source output when working in close proximity to the object. • The hysteroscope light post and adapter may exceed temperatures of 41°C. Hysteroscopes should not be placed on the patient or on combustible materials, as burns may result. • To prevent potential safety hazard to the patient caused by accidental loss of function of the device (i.e., front end damage by surgical instruments) it is recommend to have an additional sterile “stand-by” device during surgical procedures. • When scopes are used with laser equipment, appropriate filtering spectacles must be worn by the operating team. In some cases, a specific filter must be put between the scope and camera head to prevent camera damage by high-power laser radiation. Contact your laser supplier for details. To prevent scope damage by high-power laser radiation, always ensure that the laser delivery fiber is seen through the scope and not directed at the scope before energizing the laser. For Continuous Flow Hysteroscopy:
The MyoSure Rod Lens Hysteroscope is used to provide viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. Diagnostic Hysteroscopy • Abnormal uterine bleeding • Infertility and pregnancy wastage • Evaluation of abnormal hysterosalpingogram • Intrauterine foreign body • Amenorrhea • Pelvic Pain Operative Hysteroscopy • Directed biopsy • Removal of submucous fibroids and large polyps • Submucous Myomectomy (see Contraindications) • Transection of intrauterine adhesions • Transection of intrauterine septa Contraindications • Acute pelvic inflammatory disease Hysteroscopy may also be contraindicated by the following conditions, depending on their severity or extent: • Inability to distend uterus • Cervical stenosis • Cervical/vaginal infection • Uterine bleeding or menses • Known pregnancy • Invasive carcinoma of the cervix • Recent uterine perforation • Medical contraindication or intolerance to anesthesia Contraindications to Hysteroscopic Myomectomy Hysteroscopic myomectomy should not be undertaken without adequate training, preceptorship, and clinical experience. The following are clinical conditions that can significantly complicate hysteroscopic myomectomy:
• If liquid distension medium is used, strict fluid intake and output surveillance should be maintained. Intrauterine instillation exceeding 1 liter should be followed with care due to the possibility of fluid overload. Potential Complications of Continuous Flow Hysteroscopy:
• Severe anemia • Inability to circumnavigate a myoma due to myoma size (e.g., predominantly intramural myomas with small submucous components).
• Hyponatremia • Hypothermia • Uterine perforation resulting in possible injury to adjacent anatomy • Pulmonary edema • Cerebral edema
Warnings • For use only by physicians trained in hysteroscopy • Suspicion of pregnancy should suggest a pregnancy test before the performance of diagnostic hysteroscopy. • The MyoSure hysteroscope is only to be used in conjunction with accessories that comply with the following safety standards: National/Regional versions of IEC 60601-1, the general safety requirements for medical devices; IEC 606011-2, collateral requirements for electromagnetic disturbances; and, as applicable, IEC 60601-2-18, particular safety requirements for endoscope equipment and accessories; and IEC 60601-2-2, particular safety requirements for High Frequency (HF) surgical equipment and accessories. Before using any accessory, be sure to follow the instructions provided with the accessory, including in the case of a HF electrode, the maximum recurring peak voltage rating.
Precautions • Vaginal ultrasonography before hysteroscopy may identify clinical conditions that will alter patient management. • Intrauterine distension can usually be accomplished with pressures in the range of 35–75 mm Hg. Unless the systemic blood pressure is excessive, it is seldom necessary to use pressures greater than 75–80 mm Hg. • Do not use the seals if the sterile package is open or appears compromised. Do not use the device if damage is observed. • Avoid exposing the scope to sudden temperature changes. Do not immerse hot scopes into cold water or liquid.
2
ENGLISH
• Any mechanical manipulation of the eyepiece may result in seal breakage, therefore do not attempt to remove the eyepiece. • Avoid contact with metal parts of the scope and other conductive accessories by ensuring before activation of the HF output that the active electrode is at a sufficient distance from the tip of the scope. • To avoid perforation, do not use the scope tip as a probe and exercise caution when the scope is being inserted through the cervix and when the scope tip is near the uterine wall. These devices contain the hazardous substance Lead (Pb) in excess of the allowable amount within the REACh Regulation (EC 1907/2006). Lead may damage fertility or the unborn child, causes damage to organs through prolonged or repeated exposure, is very toxic to aquatic life with long lasting effects, may cause cancer, is very toxic to aquatic life and may cause harm to breast-fed children.
Seal Cap
Seal FIGURE 2. SEAL INSTALLATION
To Attach Fluid Connections: The hysteroscope accepts a standard male luer connection for fluid inflow. The Removable Outflow Channel includes a universal stopcock for both luer and friction connections.
Inspection Prior to Use Prior to each use, the outer surface of the insertion portion of the hysteroscope should be inspected to ensure there are no unintended rough surfaces, sharp edges or protrusions. Check that both the hysteroscope and outflow channel contain seals.
Needed Equipment for Indicated Procedures Fiber optic light source, fiber optic light guide (not supplied with this product)
Hysteroscope System Set-up Instructions
Hysteroscope Assembly/Disassembly Instructions
The MyoSure Rod Lens Hysteroscope consists of a removable outflow channel (40-201) and hysteroscope (40-200) as shown in Figure 1.
The MyoSure Rod Lens Hysteroscope is compatible with MetalHalide and Xenon light sources with up to 300 watts of power. Place the correct adapter on the light post of the fiber optic scope and on the instrument end of the light guide. Adapters are available for connection to Storz, Olympus, Wolf, and ACMI light sources as shown in Figure 3.
The MyoSure XL Rod Lens Hysteroscope consists of a removable outflow channel (50-201XL) and hysteroscope (50-200XL).
Removable Outflow Channel
ACMI Dyonics/Wolf Storz/Olympus
Hysteroscope
FIGURE 3. LIGHT POST ADAPTERS The light post threads may be lubricated as needed, being sure to remove any excess lubricant as required. Make sure that the fiber optic surface remains free of foreign matter. Do not use tools to tighten the adapters – hand tighten only.
FIGURE 1. REPRESENTATIVE HYSTEROSCOPE & OUTFLOW CHANNEL To place removable outflow channel into hysteroscope:
Directions for Use
Insert the removable outflow channel into the proximal seal of the hysteroscope working channel. Reverse this process to remove the outflow channel.
The surgeon may look through the direct-view hysteroscope (with eyepiece) directly with his or her eye. If a video system is being connected to the scope, thread a camera coupler onto the camera head and then insert the eyepiece into the camera coupler.
To Insert Sterile Single-Use Seal (40-902): Both the hysteroscope and the outflow channel contain single-use seals for their working channels. Figure 2 below illustrates the installation of the seals.
Plug the video cable into the camera control unit (CCU). Turn on the power to the monitor, CCU, and light source. Adjust the video system components per the manufacturer’s instructions. The system is now ready to use.
Caution: To ensure proper performance of the system and prevent leaks, install new seals in the hysteroscope and the outflow channel prior to use.
3
ENGLISH
Hysteroscope Cleaning Instructions-General The device should not be allowed to dry after procedure before cleaning to ensure effective removal of contaminant material.
• If still inserted, separate the removable outflow channel from the hysteroscope. • Light post adaptors must be removed prior to cleaning and sterilization. • Remove single-use seals from hysteroscope and removable outflow channel • Warning: Failure to remove the single-use seals from hysteroscope and removable outflow channel will affect proper cleaning and sterilization of the product. • Open the stopcocks on the hysteroscope and removable outflow channel. • Flush all lumens of the hysteroscope and removable outflow channel with warm tap water • Scrub the hysteroscope and removable outflow channel using a nylon-bristled brush that is suitable to contact the full interior dimensions (diameter and length) of the lumens. Scrub all surfaces, crevices, interior cavities of the stopcock and lumens to remove any visible debris. Do not scratch any of the optical surfaces. • The following brushes dimensions are recommended • For the main lumen, a nylon-bristled brush with bristle area length of 2” (50mm), bristle diameter of .236” (6mm) and an overall length of 14” (35cm) is recommended • For all other lumens, a nylon-bristled brush with bristle area length of 2” (50mm), bristle diameter of .197” (5mm) and an overall length of 14” (35cm) is recommended • Utilizing the stopcocks, flush the lumens with an enzymatic, neutral pH cleaner a minimum of three (3) times ensuring that no air remains within the lumen. • Hysteroscope, removable outflow channel and accessory components should be soaked in an enzymatic, neutral pH cleaner in accordance with cleaning solution instructions. • Thoroughly rinse the hysteroscope, including flushing all lumens and accessory components to completely remove the cleaning solution. • Dry the hysteroscope with a lint free soft cloth or filtered compressed air. • Check visually for cleanliness to ensure that all debris have been removed. If not visually clean, repeat reprocessing steps until the device is visually clean.
To remove these deposits, a tube of bio-compatible polishing paste is enclosed with each hysteroscope. To remove the deposits, dab some polishing paste onto a clean, cotton-tipped swab. Gently press the swab onto the optical surface to be cleaned and scrub the surface with a circular motion. Rinse the optical surface with water to remove any remaining polishing paste. NOTE: Cleaning should only be performed when the image as viewed through the scope is cloudy, and not as part of your routine cleaning procedures. NOTE: Do not use any ultrasonic cleaning methods. The energy transmitted through fluid cavitations will damage seals and optical surfaces and will void the warranty. NOTE: Foreign matter remaining on the fiber surface of the light post after cleaning may tend to burn and discolor the surface when exposed to a high intensity light source. Sterilization Hysteroscope and outflow channel should be sterilized in a container which secures the instruments in place. Be sure the distal working length does not experience any undue forces or stress which can damage the delicate internal optics. Sterilize the hysteroscope and outflow channel with the following validated steam autoclave, ethylene oxide (Et0) gas, Sterrad® or STERIS Amsco® V-PRO® methods. Sterrad®-Devices meet guidelines for Sterrad 100S, NX, and 100NX systems and require the use of a Sterrad-compatible tray or container systems. Refer to manufacturer’s Instructions for Use for more information. STERIS® Amsco® V-PRO® Devices meet guidelines for Amsco® V-PRO® 1, V-PRO 1 Plus and V-PRO maX Low Temperature Sterilization Systems and should use the STERIS compatible tray or container systems. Refer to manufacturer’s Instructions for Use for more information.
Hysteroscope Cleaning Instructions- Optical Surfaces
Steam Autoclave Wrapped
Due to insufficient cleaning or foreign matter contamination, deposits may develop on the three optical surfaces of the hysteroscope as shown below.
Follow standard hospital procedures: Pre-vacuum method:
270°F (132°C) for 4 minutes and 30 minutes dry time Gravity method: 270°F (132°C) for 15 minutes and 30 minutes dry time IMPORTANT: It is recommended that the institution employ procedures which include the use of biological indicators in order to determine the effectiveness of the sterilization process.
These are: • The distal tip • The proximal window or eyepiece • The fiber optic light post
Maintenance We recommend that you inspect the hysteroscope carefully before and after the procedure for possible signs of damage. 4
ENGLISH
First, check the image quality of the scope by viewing the monitor. If image quality is impaired:
on their respective packages; v) licensed Software is warranted to operate in accordance with published specifications; vi) Services are warranted to be supplied in a workman-like manner; vii) non-Hologic Manufactured Equipment is warranted through its manufacturer and such manufacturer’s warranties shall extend to Hologic’s customers, to the extent permitted by the manufacturer of such non-Hologic Manufactured Equipment. Hologic does not warrant that use of Products will be uninterrupted or error-free, or that Products will operate with non-Hologic authorized third-party products.
• Check the distal and proximal lenses of the hysteroscope for cracked or scratched lenses. • Check the surface cleanliness of the distal and proximal lenses. A foggy or cloudy image can be the result of moisture entering the optical system or lack of cleanliness of exterior surfaces. When viewing reflected light, the surfaces should appear smooth and shiny. As a second step, check the illumination system of the scope. Reduced brightness can result from fiber damage:
These warranties do not apply to any item that is: (a) repaired, moved, or altered other than by Hologic authorized service personnel; (b) subjected to physical (including thermal or electrical) abuse, stress, or misuse; (c) stored, maintained, or operated in any manner inconsistent with applicable Hologic specifications or instructions, including Customer’s refusal to allow Hologic recommended Software upgrades; or (d) designated as supplied subject to a non-Hologic warranty or on a pre-release or “as-is” basis.
• Check for fiber optic damage in the scope by holding the distal end of the scope toward a low power light and observing the light post on the hub. The center of the light post should appear clear or white. Noticeable black spots indicate serious damage to the fiber illumination bundle in the scope. This will affect light transmission and the brightness of the image viewed on the monitor. • Check the light cable for damaged fibers by holding one end of the cable toward a low power light and observing the other end. Broken fiber will appear as black spots in the light field. A damaged light cable will affect its ability to transmit light and the brightness of the image viewed on the monitor.
TECHNICAL SUPPORT AND PRODUCT RETURN INFORMATION Contact Hologic Technical Support for proper disposal of any part of the MyoSure Hysteroscopic Tissue Removal System or if the MyoSure Hysteroscopic Tissue Removal System fails to operate as intended. If product is to be returned to Hologic for any reason, Technical Support will issue a Returned Materials Authorization (RMA) number and biohazard kit if applicable. Return the MyoSure Hysteroscopic Tissue Removal System according to the instructions provided by Technical Support. Be sure to clean and sterilize the product before returning it and include all accessories in the box with the returned unit.
Storage The MyoSure Rod Lens Hysteroscope and Removable Outflow Channel should be stored either in their shipping box or in a sterilization tray. In either case, proper care should be taken to ensure that the hysteroscope and outflow channel are immobile to prevent any damage.
Return used or opened product according to the instructions provided with the Hologic-supplied biohazard kit. For More Information
Service - Accessories
For technical support or reorder information in the United States, please contact:
The following are replacement/service parts for the MyoSure Rod Lens Hysteroscopes and Removable Outflow Channels: REF
Description
40-201
Replacement MyoSure Outflow Channel
50-201XL
Replacement MyoSure XL Outflow Channel
ASY-04996
Hysteroscope Light Source Adapters 1 each: Wolf and Storz
40-902
MyoSure Single Use Seal Set - 10 per box
40-904
MyoSure Hysterscope and Outflow Channel Seal Cap
Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 USA Phone: 1.800.442.9892 (toll-free) www.hologic.com International customers, contact your distributor or local Hologic Sales Representative: European Representative Hologic BV Da Vincilaan 5 1930 Zaventem Belgium Phone: +32 2 711 46 80
WARRANTY, SERVICE, AND REPAIR WARRANTIES Except as otherwise expressly stated in the Agreement: i) Equipment manufactured by Hologic is warranted to the original Customer to perform substantially in accordance with published product specifications for one (1) year starting from the date of shipment, or if Installation is required, from the date of Installation (“Warranty Period”); ii) digital imaging mammography x-ray tubes are warranted for twenty-four (24) months, during which the x-ray tubes are fully warranted for the first twelve (12) months and are warranted on a straight-line prorated basis during months 13-24; iii) replacement parts and remanufactured items are warranted for the remainder of the Warranty Period or ninety (90) days from shipment, whichever is longer; iv) consumable Supplies are warranted to conform to published specifications for a period ending on the expiration date shown
Hologic, MyoSure, and associated logos are registered trademarks of Hologic, Inc. and/or its subsidiaries in the United States and other countries. All other trademarks, registered trademarks, and product names are the property of their respective owners.
© 2014-2021 Hologic, Inc.
AW-07813-001 Rev. 012 5