System Configuration Guide Overview
Accessing Custom Mode
The Hospira GemStar Pump is available in the following factory configurations: 6-Therapy (List 13086) 7-Therapy (List 13087) Pain Management (List 13088) The Configuration Menu allows the biomedical technician to configure several of the pump's functions. The Configuration Menu varies depending on the factory configuration of the pump. Note: If the current configuration is the same as the factory defaults, the pump displays CURRENT CONFIG: DEFAULT. (See Figure 1 for a sample printout of a 7-therapy pump with factory default settings) If any factory defaults have been changed, the pump displays CURRENT CONFIG: CUSTOM. (See Figure 2 for a sample printout of a 7-therapy pump with custom settings) If an item is configured that makes other items unavailable, the pump displays DISABLED in the line for those items. Note: From the submenus, press [BACK-UP] to return to the Custom Mode main menu.
Note: Keypad lock levels do not affect access to the Custom Mode.
1.
From STOP or PROGRAMMING mode, press [OPTIONS].
2.
Select 2 HISTORIES.
3.
Select 4 CONFIGURATION. The current configuration displays.
4.
Press [CHANGE]. The pump displays CUSTOM MODE ACCESS CODE.
5.
To enter the CUSTOM mode press [PURGE], [NO], [OPTIONS], [BACK-UP], [START]. The Custom Mode Main Menu displays. Note: You must enter the code within 30 seconds or the pump will return to the previous display. Tip: If you press an invalid key, press [CHANGE] to zero out the display and start again.
Frequently Used Keys Press the up and down arrows to scroll through the configuration menu and display messages. Press the up arrow to insert a decimal point in a numeric value.
0
Press the number keys to select menu items and to enter numeric values. If the desired option is 01, you must press [0] and [1]. Use the associated number key to select or deselect settings. If you enter an incorrect numeric value, press [CHANGE] before pressing [YES/ENTER] to reset the value to zero. Press [YES/ENTER] to: Answer YES to (Yes or No) display messages. Accept selected settings and return to the Configuration Menu. Press [NO] to answer NO to (Yes or No) display messages.
Press [BACK-UP] to return to a previous display without accepting changes (when in sub-menu).
GEMSTAR SYSTEM CONFIGURATION GUIDE
430-10991-002 (Rev. 06/06)
GEMSTAR SYSTEM CONFIGURATION GUIDE
2
430-10991-002 (Rev. 06/06)
Configuring the Pump 1.
To configure the pump, press [ENTER] to confirm selection in submenu. Press [BACKUP] to return to configuration menu. If configuration is complete, press YES to save all configurations.
2.
Select menu item. Use up and down arrows to view menu and submenu items. Note: press both
3.
Select submenu. NOTE: DO NOT PRINT WITHOUT BEING CONNECTED. If you do not have a cable, select No. NOTE: Press both keys (e.g., [0] and [1]) to access sub-options, such as 01 THERAPIES. From the submenus, press [BACK-UP] to return to the Custom Mode main menu.
The following table lists which functions are available under each of the Configuration submenus: Select this menu item:
To configure the
Default Settings
Available Setting
Description
YES OR NO
Select YES to reset the pump to factory configured settings.
following functions: 1 RESET DEFAULTS
NA
YES
2 GENERAL PUMP
01 THERAPIES
T P I C W mL V
(Continued on next page)
Any combination of the (TPN) (Pain Management) seven therapies (Intermittent) (Continuous) (Weight Dosed) (mL/hr Only) (Variable)
Use the associated number key to select or deselect therapies. For example, press [2] to select or deselect Pain Management. The pump will only display selected therapies during programming. This item is only available if the pump is configured for 6 or 7 Therapies.
02 FORCED REVIEW
YES
YES OR NO
Select NO to turn OFF FORCED REVIEW. The pump will not prompt for a Program Review after programming.
03 SHIFT TOTALS
YES
YES OR NO
Select NO to turn OFF SHIFT TOTALS. The pump will not print or display SHIFT TOTALS in the Histories Menu.
04 DISPLAYED TOTALS
CONTAINER
SHOW CONTAINER SHOW SHIFT
Select SHOW SHIFT to display current shift totals in the Run Mode and Stop Mode displays. The pump shows the first line of the display as: SHIFT XXXX.X mL
05 CONTACT PHONE NUMBER
blank
Any combination of digits and/or decimal points up to 16 characters
Enter up to 16 characters using the number keys and [UP ARROW] (e.g., 888.555.1234). This number will display when the Operation Test Call Back message displays and when a service alarm occurs.
Select this menu item:
To configure the
Default Settings
Available Setting
Description
ON (0.5 mL) 2 mL ALL SELECTIONS (ON, 2 mL, OFF)
Setting the Air Sensitivity to 2 mL will not affect the ON default setting in mL/hr Only and Weight Dosed programs.
following functions: 2 GENERAL PUMP
06 AIR SENSITIVITY ALL SELECTIONS
(Continued)
If the Air Sensitivity is configured to ON or 2 mL, the AIR SENSITIVITY message will not display when entering a new program or changing a program. The user will not be able to change the Air Sensitivity setting through the Options Menu.
07 DISP OT WARN
NO
YES OR NO
Select YES to display the Operation Test Warning at power-on when the call back date is reached.
365 DAYS
1 - 365 DAYS
Enter the number of days between Operation Tests.
DISABLED
1 - 364 DAYS
The Operation Test Call Back displays a message to the user when the Operation Test is due. Enter the number of days prior to the Operation Test due date for the call back message to display. If DISP OT WARN is enabled, the default OT callback is 7 days.
13000
Any combination of 3-5 The Lock/Unlock Access Code must consist of 3 to 5 digits. The digits default access code for Rate Change, Container Lock, and Full Lock is 13000.
(Display Operation Test Warning) 08 OT FREQUENCY
(Operation Test Frequency) 09 OT CALL BACK
(Operation Test Call Back) 3 KEYPAD LOCK
01 LOCK CODE
The Therapy Lock Access Code is the reverse of the Lock/Unlock Access Code. The default access code for Therapy Lock is 00031. 02 LOCK REMINDER
NO
YES OR NO
Select YES to display a Lock Reminder when the pump is unlocked or locked only under Therapy Lock and is placed in Run Mode. Select NO to disable the Lock Reminder when the pump is unlocked or locked only under Therapy Lock and is placed in Run Mode.
(Continued on next page)
GEMSTAR SYSTEM CONFIGURATION GUIDE
3
430-10991-002 (Rev. 06/06)
GEMSTAR SYSTEM CONFIGURATION GUIDE
Select this menu item:
To configure the
4
430-10991-002 (Rev. 06/06)
Default Settings
Available Setting
Description
03 UNLOCK PROMPT
NO
YES OR NO
Select YES to display the LOCK SEQUENCE NUMBER screen when the pump is locked under a Full, Container or Rate lock and 2 or 3 is selected on the RESUME PROGRAM screen.
04 AUTO LOCK
1 NO AUTO LOCK
1
NO AUTO LOCK
Select one of the lock level settings to force the pump to lock the keypad under the selected lock level when the user presses [START].
2
RATE CHANGE
3
CONTAINER
4
FULL
For example, select CONTAINER auto lock and the pump will automatically apply the Container lock level when the user presses [START].
following functions: 3 KEYPAD LOCK (Continued)
4 BOLUS
01 BOLUS PLACEBO
YES
YES OR NO
This setting determines how the pump responds when the user requests a bolus dose and a bolus dose in NOT available. If YES is selected, the pump sounds 3 beeps (same as in a regular bolus delivery); however, the pump does not display the asterisks to indicate bolus delivery. If NO is selected, the pump does not sound any beeps or display asterisks.
02 PM MAX BOLUS
25 mL
0.1 - 25 mL
Enter the maximum bolus amount for a Pain Management program.
125 mL/hr
1 - 125 mL/hr
Enter the bolus delivery rate for a Pain Management program.
400 mL/hr
1 – 1000 mL/hr
Enter the bolus delivery rate for a Variable Time program.
1 - 5 minutes
Enter the minimum lockout interval for bolus delivery.
(Pain Management Maximum Bolus) 03 PM BOLUS RATE
(Pain Management Bolus Rate) 04 VT BOLUS RATE
(Variable Time Bolus Rate) 05 MINIMUM LOCKOUT 5 minutes (Domestic) 3 minutes (International)
Select this menu item:
To configure the
Default Settings
Available Setting
Description
following functions: 5 PM (SPECIFIC)
01 PAIN MGMT (PM) CONT (continuous) Any combination of the The pump will only display selected delivery types during three delivery types programming. BOLUS (bolus) (Pain Management) C+B 02 PM UNITS
(continuous + bolus)
mg, mcg, mL
Any combination of the The pump will only display selected units during programming. three units
25 mL/hr
0.1 - 25 mL/hr
Enter the maximum delivery rate for a Pain Management program.
25 mL
0.1 - 25 mL
Enter the maximum loading dose for a Pain Management program.
125 mL/hr
1 – 125 mL/hr
Enter the loading dose rate for a Pain Management program.
ALL LIMITS
1 HOUR
Select 1-Hour, 4-Hour, or # Boluses/Hour to disable PAIN MANAGEMENT DOSE LIMIT SELECTION when entering a new program or when changing a program.
(Pain Management Units) 03 PM MAX RATE
(Pain Management Maximum Rate) 04 PM MAX LD
(Pain Management Maximum Loading Dose) 05 PM LD RATE
(Pain Management Loading Dose Rate) 06 PM DOSE LIMIT
(Pain Management Dose Limit)
4 HOUR # BOLUSES/HOUR ALL LIMITS
07 PM MAX PRIME
8.3 mL
4.6 – 8.3 mL
(Pain Management Maximum Prime Volume)
GEMSTAR SYSTEM CONFIGURATION GUIDE
5
The pump will display only the selected dose limit screen. For example, if 4 HOUR is selected, the SET 4-HOUR DOSE LIMIT screen will display when a new program is entered or a program is changed. No other dose limit options will be available. Enter the maximum pain management priming volume, which determines the total volume that can be primed during a Pain Management program between an event that allows priming and the time [START] is pressed.
430-10991-002 (Rev. 06/06)
GEMSTAR SYSTEM CONFIGURATION GUIDE
Select this menu item:
To configure the
6
430-10991-002 (Rev. 06/06)
Default Settings
Available Setting
Description
01 PIGGYBACKING
YES
YES OR NO
Select NO to disable Piggybacking options.
02 WD UNITS
01 mcg/kg/min
mcg/kg/min
(Weight Dosed)
02 mcg/kg/hr
mcg/kg/hr
03 mcg/min
mcg/min
This setting allows you to configure from 1 - 16 different units of measure for weight dosed therapies. Press the associated number keys to add or change a setting.
04 mg/kg/min
mcg/hr
05 mg/kg/hr
mg/kg/min
06 mg/min
mg/kg/hr
07 mL/hr
mg/min
08 NOT USED
mg/hr
following functions: 6 OTHER
09 NOT USED 10 NOT USED 11 NOT USED 12 NOT USED 13 NOT USED 14 NOT USED 15 NOT USED 16 NOT USED 03 WD PRECEDEX
NO
For example, press [0], [8] to add a unit of measure. The first available unit of measure will display. Use the arrow keys to scroll through the available units of measure. Press [YES/ENTER] to select the displayed unit of measure.
grams/hr ng/kg/min U/kg/hr U/min U/hr mUn/min mEq/hr mL/hr NOT USED YES OR NO
Precedex is not available unless mcg/kg/hr is selected. Press Yes to enable Precedex delivery while programming a Weight Dosed therapy.
60 minutes
1 - 60 minutes
Enter the maximum delivery time for a Weight Dosed program.
05)WD LD MAX RATE 1000 mL/hr
1 - 1000 mL/hr
Enter the maximum delivery rate for a Weight Dosed program.
04)WD LD MAX TIME
4.
To exit CUSTOM mode after all changes have been made, press [BACKUP]. Then press [YES/ENTER]. Save changes. Press [YES].
5.
[BACKUP] to return to STOP or PROGRAMMING screen.
***************************************** * HOSPIRA GEMSTAR * * DEFAULT MODE CONFIGURATION * *****************************************
***************************************** * HOSPIRA GEMSTAR * * CUSTOM MODE CONFIGURATION * *****************************************
Pump Serial Number: XXXXXXXX
Pump Serial Number: XXXXXXXX
10:45AM JUN 11, 05 VERSION X.X.X
11:15AM JUN 11, 05 VERSION X.X.X
CONFIGURATION CHANGED hh:mmXM MMM DD, YY PUMP SETTINGS DEFAULT THERAPIES T P I C W mL V PROGRAM REVIEW YES SHIFT TOTALS YES SHIFT OR CONT CONTAINER CONTACT PHONE AIR SENSITIVITY ALL DISPLAY OPERATION TEST WARNING NO OPERATION TEST FREQUENCY 365 DAYS OPERATION TEST CALL BACK TIME DISABLED LOCK CODE 13000 LOCK REMINDER NO UNLOCK PROMPT NO AUTO LOCK NO AUTO LOCK BOLUS PLACEBO YES PAIN MANAGEMENT MAX BOLUS 25 mL PAIN MANAGEMENT BOLUS RATE 125 mL/hr VARIABLE TIME BOLUS RATE 400 mL/hr MINIMUM LOCKOUT 5 MINUTES PAIN MANAGEMENT PROG. TYPES C B C+B PAIN MANAGEMENT UNITS mL mg mcg PAIN MANAGEMENT MAX RATE 25 mL/hr PAIN MANAGEMENT MAX LD 25 mL PAIN MANAGEMENT LD RATE 125 mL/hr PAIN MANAGEMENT DOSE LIMIT ALL LIMITS PAIN MANAGEMENT MAX PRIME 8.3 mL PIGGYBACKING YES WEIGHT DOSED UNITS DEFAULT ENABLE PRECEDEX NO WEIGHT DOSED LD MAX TIME 60 MINUTES WEIGHT DOSED LD MAX RATE 1000 mL/hr
CONFIGURATION CHANGED hh:mmXM MMM DD, YY PUMP SETTINGS CUSTOM THERAPIES T P mL PROGRAM REVIEW YES SHIFT TOTALS NO SHIFT OR CONT SHIFT CONTACT PHONE AIR SENSITIVITY ALL DISPLAY OPERATION TEST WARNING NO OPERATION TEST FREQUENCY 365 DAYS OPERATION TEST CALL BACK TIME 7 DAYS LOCK CODE 12345 LOCK REMINDER NO UNLOCK PROMPT YES AUTO LOCK NO AUTO LOCK BOLUS PLACEBO YES PAIN MANAGEMENT MAX BOLUS 25 mL PAIN MANAGEMENT BOLUS RATE 100 mL/hr VARIABLE TIME BOLUS RATE 400 mL/hr MINIMUM LOCKOUT 5 MINUTES PAIN MANAGEMENT PROG. TYPES C B C+B PAIN MANAGEMENT UNITS mL mg mcg PAIN MANAGEMENT MAX RATE 25 mL/hr PAIN MANAGEMENT MAX LD 25 mL PAIN MANAGEMENT LD RATE 125 mL/hr PAIN MANAGEMENT DOSE LIMIT ALL LIMITS PAIN MANAGEMENT MAX PRIME 8.3 mL PIGGYBACKING YES WEIGHT DOSED UNITS CUSTOM ENABLE PRECEDEX YES WEIGHT DOSED LD MAX TIME 60 MINUTES WEIGHT DOSED LD MAX RATE 1000 mL/hr
***************************************** * VERIFIED BY: * ***************************************** ***************************************** * END OF DEFAULT MODE CONFIGURATION * *****************************************
***************************************** * VERIFIED BY: * ***************************************** ***************************************** * END OF CUSTOM MODE CONFIGURATION * *****************************************
Figure 1. Sample printout of factory DEFAULT settings for a 7-Therapy pump.
Figure 2. Sample printout of CUSTOM settings for a 7-Therapy pump.
GEMSTAR SYSTEM CONFIGURATION GUIDE
7
430-10991-002 (Rev. 06/06)
Enhanced Configuration Selection Sheet
Facility Name ______________________________________________
Select this menu Item To configure the following functions: Default Settings Reset Defaults NA YES GENERAL PUMP 01 THERAPIES T (TPN) P (Pain Management) I (Intermittent) C (Continuous) W (Weight Dosed) mL (mL/hr Only) V (Variable) 02 FORCED REVIEW YES 03 SHIFT TOTALS YES 04 DISPLAYED TOTALS CONTAINER
KEYPAD LOCK
BOLUS
05 CONTACT PHONE NUMBER 06 AIR SENSITIVITY
blank ALL SELECTIONS
07 DISP OT WARN (Display Operation Test Warning) 08 OT FREQUENCY (Operation Test Frequency) 09 OTCALL BACK (Operation Test Call Back) 01 LOCK CODE 02 LOCK REMINDER 03 UNLOCK PROMPT 04 AUTO LOCK
NO
01 BOLUS PLACEBO 02 PM MAX BOLUS (Pain Management Maximum Bolus) 03 PM BOLUS RATE (Pain Management Bolus Rate) 04 VT BOLUS RATE (Variable Time Bolus Rate) 05 MINIMUM LOCKOUT
GEMSTAR SYSTEM CONFIGURATION GUIDE
365 Days DISABLED (If enabled, 7 days) 13000 NO NO 1 NO AUTO LOCK
YES 25 mL 125 mL / hr 400 mL / hr 5 minutes (Domestic) 3 minutes (International)
8
Facility Setting Circle one: YES NO Circle desired therapies: T (TPN) P (Pain Management) I (Intermittent) C (Continuous) W (Weight Dosed) mL (mL/hr Only) V (Variable) Circle one: YES NO Circle one: YES NO Circle one: SHOW CONTAINER SHOW SHIFT Any combination up to 16 ____________________ Circle one: ON (0.5mL) 2 mL ALL SELECTIONS (ON, 2mL, OFF) Circle one: YES NO 1-365 Days (Enter desired number of days) ____________ 1-364 Days (Enter desired number of days) ____________ Any combination of 3 - 5 Digits ____________ Circle one: YES NO Circle one: YES NO Circle one: NO AUTO LOCK RATE CHANGE CONTAINER FULL Circle one: YES NO 0.1 - 25 mL (Enter desired amount) ____________ 1 - 125 mL / hr (Enter desired rate) _________ 1 - 1000 mL / hr (Enter desired rate) ___________ 1 - 5 minutes ____________ (Enter minimum lockout interval ___________)
430-10991-002 (Rev. 06/06)
Select this menu Item To configure the following functions: Default Settings 01 PAIN MGMT (PM) PM (PAIN MANAGEMENT (Pain Management) CONT (Continuous) SPECIFIC) BOLUS (Bolus) C & B (Continuous & Bolus) 02 PM UNITS (Pain Management Units) mg, mcg, mL 03 PM MAX RATE (Pain Management Maximum Rate) 04 PM MAX LD (Pain Management Maximum Loading Dose) 05 PM LD RATE (Pain Management Loading Dose Rate) 06 PM DOSE LIMIT (Pain Management Dose Limit)
OTHER
25 mL / hr 25 mL 125 mL / hr
8.3 mL
03) WD PRECEDEX 04) WD LD MAX TIME
NO 60 minutes
05) WD LD MAX RATE
1000 mL/hr
GEMSTAR SYSTEM CONFIGURATION GUIDE
Circle desired combination of the three units of delivery: mg mcg mL 0.1 - 25 mL / hr (Enter desired rate) _________ 0.1 - 25 mL (Enter desired amount) _________ 1 - 125 mL / hr (Enter desired rate) _________ Select only one dose limit if all limits are not desired: 1 HOUR 4 HOUR # BOLUSES / HOUR ALL LIMITS 4.6 - 8.3 mL (Enter desired amount) _________
ALL LIMITS
07 PM MAX PRIME (Pain Management Maximum Prime Volume) 01 PIGGYBACKING 02 WD UNITS (Weight Dosed)
Facility Setting Circle desired combination of delivery types: CONT (Continuous) BOLUS (Bolus) C & B (Continuous & Bolus)
YES
Circle one: YES NO Indicate Numerical Preference: mcg / kg /min _______ mcg / kg / hr ______ mcg / min ______ mcg / hr ______ mg / kg / min ______ mg / kg /hr ______ mg / min ______ mg / hr _______ grams / hr ______ ng / kg / min _______ U / kg / hr _______ U / min _______ U / hr _______ mUn / min ______ mEq / hr _______ mL / hr ______ NOT USED ______ Circle one: YES NO 1 - 60 minutes (Enter maximum time) _________ 1 - 1000 mL/hr
mcg / kg /min mcg / kg / hr mcg / min mg / kg / min mg / kg /hr mg / min mL / hr NOT USED NOT USED NOT USED NOT USED NOT USED NOT USED NOT USED NOT USED NOT USED
9
430-10991-002 (Rev. 06/06)
NOTES
GEMSTAR SYSTEM CONFIGURATION GUIDE