Q2ROLLER Lateral Turning Device User Manual Rev F
User Manual
4 Pages
Preview
Page 1
Lateral Turning Device
User Manual Visit www.HoverMatt.com for other languages
TABLE OF CONTENTS
Intended Use and Precautions...2 Part Identification - Q2Roller...2 Part Identification - Air Supply...2 Air Supply Keypad Functions...3 Instructions for Use...3 Product Specifications/Required Accessories...3 Cleaning & Maintenance...4 Returns and Repairs...4
User Manual Intended Use and Precautions • Only for use with the HT-Air air supply adjustable button.
INDICATIONS FOR USE
Patients who are unable to turn or laterally position themselves
• Make sure patient is centered on the Q2Roller before inflating. Reference Center Line Indicator on the Q2Roller.
CONTRAINDICATIONS
Patients who are experiencing thoracic, cervical or lumbar fractures that are deemed unstable
• Caregivers must ensure patient is attended and stabilized during Q2Roller inflation, deflation or when inflated at angles greater than 30 degrees.
INTENDED CARE SETTINGS
• Do not launder. • Reference product-specific user manuals for additional operating instructions.
Hospitals, long-term or extended care facilities
PRECAUTIONS
• Ensure the Q2Roller meets the patient’s needs before using. • Always use a minimum of two caregivers when operating the Q2Roller. • Ensure valves are not in contact with the patient while on the Q2Roller.
PRECAUTIONS – HOVERTECH INTERNATIONAL AIR SUPPLY • Route the power cord in a manner to ensure freedom from hazard.
• Product should only be used by trained personnel.
• Avoid blocking the air intakes of the air supply.
• Only use attachments and/or accessories that are authorized by HoverTech International.
• Use this product only for its intended purpose as described in this manual.
• Use the Q2Roller according to instructions.
• CAUTION: Avoid electric shock. Do not open air supply.
Part Identification - Q2Roller
Part Identification - Air Supply
Center Line Indicator
Black Inflation Valve (2) Red Deflation Valve (2)
Power Cord Strap
Carrying Handle/ Attachment Hook
Hose Release Button
Control Panel Label Power Cord
Turning Chambers
Hose Nozzle
Hose
Air Filter Cover
Securing Apron
WARNING: The HT-Air is not compatible with DC power supplies. The HT-Air is not intended for use with the HoverJack Battery Cart. CAUTION: NO USER SERVICEABLE PARTS. Only qualified service personnel shall perform repairs on the HoverTech International Air Supply.
2 | HoverTech International
Q2RollerManual Rev. F
User Manual Air Supply Keypad Functions Adjustable Speed Indication (Green Flashing LED)
The ADJUSTABLE keypad function has four different settings. Each press of the button increases the air pressure and rate of inflation. The Green Flashing LED will indicate the inflation speed by the number of flashes (i.e. two flashes equals the second inflation speed).
Power Indication Green LEDs
STANDBY: Used to stop inflation/air flow (Amber LED indicates STANDBY mode).
Adjustable/ Pressure Setting
Transfer Speed High
Standby (Amber LED)
Transfer Speed Low
Instructions for Use 1. Ensure the red deflation valves are capped tightly before beginning inflation. 2. Place the Q2Roller on the bed mattress, and center using the Center Line Indictor. 3. Tuck the securing apron of the Q2Roller under the bed mattress to keep it in place during use. 4. Remove the backing from the adhesive strips on the Q2R Pad, and center it on top of the Q2Roller (if the patient is already on the mattress, place the device and pad under the patient using a log-rolling technique).
6. Locate the inflation valve on the OPPOSITE side of the turn direction, and place the air supply hose over the valve. Press the ADJUSTABLE button on the HT-Air keypad to initiate air flow. If necessary, press this button up to 3 more times to increase air flow and rate of inflation. 7. When the turn is complete, remove the hose from the valve and press the STANDBY button on the air supply keypad to stop air flow. Chamber may not need to be fully inflated to achieve desired position. 8. To quickly deflate the Q2Roller, remove the red cap from the deflation valve. To slowly deflate or adjust the chamber, depress the center of the black inflation valve.
5. With the patient centered on the device, ensure all bed/stretcher rails are up, or follow your facility’s protocol.
Product Specifications/Required Accessories Q2ROLLER ® LATERAL TURNING DEVICE
Q2R ® PAD
Material:
Polyurethane
Material:
Non-Woven, Breathable PE
Construction:
RF Welded
Construction:
Thermal Bonding
Width:
Chambers: 46.5" (118 cm) (combined chamber width) Apron: 72" (183cm)
Width:
39” (99 cm)
Length:
39” (99 cm)
Length:
Chambers: 44.5" (113 cm) Apron: 44" (112 cm)
LATEX FREE Model #: HTR-200 Patient Weight Limit: 600 lbs. (272 kg) Single-Patient Use
Model #: HTR-CHX-5PACK (5 ea.) Model #: HTR-CHX-CASE (50 ea.)
REQUIRED ACCESSORY:
Model #: HTAIR1200 (North American Version) – 120V~, 60 Hz, 10A Model #: HTAIR2300 (European Version) – 230V~, 50 Hz, 6A Model #: HTAIR1000 (Japanese Version) – 100V~, 50/60 Hz, 12.5A Model #: HTAIR2356 (Korean Version) – 230V~, 50/60 Hz, 6A
Q2RollerManual Rev. F
www.HoverMatt.com | 3
User Manual Cleaning & Maintenance CLEANING
The Q2Roller is a single-patient use device. If the device becomes soiled, completely wipe it down using a germicidal cleaner (phenolic disinfectant, quaternary solution or other intermediate level disinfectant according to facility procedure) to disinfect. For best results, follow the cleaner manufacturer’s recommended dwell time and instructions for use. Apply germicidal cleaner directly into hard-to-reach areas. Let air dry. Do not launder.
INFECTION CONTROL
The Q2R Pad is recommended to cover the Q2Roller to keep the device from getting soiled (also available for separate purchase). Other materials, such as pads or linens, may also be placed on top of the Q2Roller to keep it clean. If the Q2Roller is used on an isolation patient, the hospital should employ the same protocols/procedures it utilizes for the bed mattress and/or for the linen in that patient room.
AIR SUPPLY CLEANING AND MAINTENANCE
See air supply manual for reference. NOTE: CHECK YOUR LOCAL/STATE/FEDERAL/INTERNATIONAL GUIDELINES BEFORE DISPOSAL.
PREVENTIVE MAINTENANCE
Prior to use, a visual inspection should be performed on the Q2Roller to ensure that there is no visible damage that would render the Q2Roller unusable. The Q2Roller should be periodically inspected to ensure the following: • All inflation valves are self-sealing with no evidence of leakage • There are no punctures or tears in the Q2Roller If any damage is found that would cause the Q2Roller not to function as intended, the Q2Roller should be discarded.
Returns and Repairs All products being returned to HoverTech International (HTI) must have a Return Goods Authorization (RGA) number issued by the company. Please call 800-471-2776 and ask for a member of the RGA Team, who will issue you an RGA number. Any product returned without an RGA number will cause a delay in the repair time. Returned products should be sent to: HoverTech International 4482 Innovation Way Allentown, PA 18109 Attn: RGA #______________________ Phone: 800-471-2776 Fax: 610-694-9601
4482 Innovation Way Allentown, PA 18109 800.471.2776 Fax 610.694.9601 www.HoverMatt.com [email protected]
4 | HoverTech International