AIR200G & AIR400G Air Supply
User Manual
TABLE OF CONTENTS Symbol References...2 Intended Use and Precautions...2 Part Identification...3 Product Specifications...3 Cleaning and Preventive Maintenance...4 Returns and Repairs...4
AIR200G & AIR400G User Manual Symbol Reference REFER TO MANUAL
THIS END UP
TYPE BF APPLIED PART
TEMPERATURE LIMITATION
ALTERNATING CURRENT
HUMIDITY LIMITATION
MANUFACTURER
DATE OF MANUFACTURE
GENERAL WARNING SIGN
KEEP DRY
FUNCTIONAL EARTH (GROUND)
CAUTION
CSA C22.2 No. 601.1, UL-60601-1
Intended Use and Precautions The Air200G and Air400G are single-speed air supplies used to inflate HoverTech’s HoverMatt® Air Transfer Mattresses and HoverJack® Air Patient Lifts.
• Use this product only for its intended purpose as described in this manual. Only use attachments and/or accessories that are authorized by HoverTech International.
INTENDED USE
• When using the air supply in the MRI environment, a 25 ft. specialty MRI hose is required (available for purchase).
• The caregiver/operator is the person handling the equipment.
• Avoid electric shock. Do not open air supply.
• The patient is not the intended operator.
• Reference product-specific user manuals for additional operating instructions.
CAUTIONS • Route the power cord in a manner to ensure freedom from hazard. • Avoid blocking the air intakes of the air supply. • Never leave patient unattended on an inflated device.
2 | HoverTech International
• Warning: To avoid the risk of electric shock, this equipment must only be connected to a supply main with protective earth.
INTENDED CARE SETTINGS Hospitals, long-term or extended care facilities and ambulance/transport.
Air200G/Air400G Manual Rev. E
AIR200G & AIR400G User Manual Part Identification ASWH-100 Webbed handle BC-100 Blower coupler
AS-HOOK Metal hook with webbing
ELC-100 Electrical cord BA-100 Blower adapter
ASPH100 Plastic handle HS-100 Handle screw
ASCB Cord clamp with screws
RG-100 Rubber grommet AS-Hose 5 foot hose (includes BC-100 and AS-Cuff)
AS-Cuff VacuFlex hose cuff
CAUTION: WITH THE EXCEPTION OF FILTERS AS DESCRIBED IN THE CLEANING AND PREVENTIVE MAINTENANCE SECTION, THERE ARE NO OTHER USER SERVICABLE PARTS.
Product Specifications Dimensions:
9 DIA x 16 H inches (22.8 cm x 40.6 cm)
CL ASSIFICATION
Weight:
11 lbs. (5 kg)
Not for use in the presence of flammable anesthetics or in a hyperbaric chamber or oxygen tent.
Material:
Lexan/Stainless Steel
Power Cord Length:
15 feet (457 cm)
Type:
SJT 16 AWG*3C
To remove supply mains, unplug equipment from wall.
Service Life:
5 years
Use Temperature:
50° to 104° F (10° to 40° C)
Power Input:
120 V~, 60 Hz, 10 A
Use Humidity:
10% to 70% Non-Condensing
Storage/Shipping Temperature:
-40° to 176° F (-40° to 80° C)
Storage/Shipping Humidity:
10% to 70% Non-Condensing
Model #Air200G – 800W
Type of protection against electric shock:
CLASS I EQUIPMENT
Degree of protection against electric shock:
TYPE BF APPLIED PART
Protection against ingress of water:
Ordinary (not protected).
Mode of operation:
CONTINUOUS OPERATION
Model #Air400G – 1100W LATEX FREE
Air200G/Air400G Manual Rev. E
www.HoverMatt.com | 3
AIR200G & AIR400G User Manual Cleaning Between patient use, clean and disinfect the surface of the air supply by wiping it down with EPA approved hospital-grade disinfectant wipes or disinfectant cleaner sprayed on a cleaning cloth. Follow disinfectant manufacturer’s directions for dwell time and other instructions for use. NOTE: DO NOT SPRAY CLEANERS/LIQUIDS DIRECTLY ON THE AIR SUPPLY.
PREVENTIVE MAINTENANCE
running water. Allow the air filter to dry prior to placing back in the air supply. The filter should be replaced when it is clogged with debris that does not break free when it is washed. The filter should also be replaced if it begins to lose its shape or deteriorate. NOTE: IF AIR SUPPLY NEEDS TO BE DISPOSED, CHECK YOUR LOCAL / STATE / FEDERAL / INTERNATIONAL GUIDELINES BEFORE DISPOSAL.
Prior to use, a visual inspection should be performed on the air supply to ensure the power cord is not frayed or nicked and that there is no visual damage that would render the air supply unusable.
INFECTION CONTROL
If any damage is found that would cause the air supply not to function as intended, the air supply should be removed from use and returned to HoverTech International for repair.
When a HoverTech air supply is used in a patient room where isolation protocols are being observed, the hospital should employ the same protocols/ procedures it utilizes for other equipment used in that patient room.
The air supply has an air filter over the motor housing. This filter can be accessed by removing the three small screws holding the canister to the motor. It is recommended that the air filter is assessed per your facility’s preventive maintenance schedule or annually. Filter should be cleaned if visibly dirty. Remove the filter from the air supply and hold it under warm
Between use with an airborne isolation patient, the air supply filter can be removed and disinfected or replaced if hospital protocol requires. If the air filter is disinfected, allow it to dry prior to placing back in the air supply. An air hose cover is available. These covers are disposable and come in a box of 25 (Model #ASHC).
Returns and Repairs All products being returned to HoverTech International (HTI) must have a Return Goods Authorization (RGA) number issued by the company. Please call 800-471-2776 and ask for a member of the RGA Team, who will issue you an RGA number. Any product returned without an RGA number will cause a delay in the repair time. Returned products should be sent to: HoverTech International Attn: RGA #____________ 4482 Innovation Way Allentown, PA 18109