HS CLEMENT CLARKE INTERNATIONAL

Perkins Mk3 Quick Guide Issue 1 March 2015

Quick Guide

4 Pages

Cleaning methods Manual cleaning methods appropriate for non-immersable, opto-electronic devices should be used. External components that come into contact with intact skin and do not come into contact with the patient, should be wiped with a clean, non absorbent cloth and detergent solution or isopropyl alcohol 70%.  Technical Specification Specifications  Environmental conditions  Regulatory  Part numbers  Measurement range  0-50 mmHg (0 – 66.5 kPa)  Standard divergence  0.49 mN  Reverse span  =/< 0.49 mN  Weight  1.0 kg (Packed)  Light source  LED 450 – 490nm  Power supply  3.6V NiMh battery  Temperature  -10 to +35 C  Humidity  30% to 90%  Atmospheric pressure  800 to 1060 hPa  Classification  CE Class IIa  Standards  MDD 93/42/EEC ISO 8612:2009 ISO 15004-1:2006 EN60601-1 EN60825-1  Perkins tonometer (UK) Perkins tonometer with Tonosafe Perkins tonometer (EU) Perkins tonometer (AUS)  5806000A 5806001A 5806001B 5806004  PERKINS TONOMETER Instructions For Use (IFU) Quick Guide  HAAG-STREIT UK Edinburgh Way Harlow Essex CM20 2TT Phone (01279) 881974 perkins@haag-streit-uk.com www.haagstreituk.com/ifu  0120  ©HAAG-STREIT UK 2015 E & OE. 1902547/ISSUE 1/March 2015
File Type: PDF
File Size: 834 KB
File Name: HS CLEMENT CLARKE INTERNATIONAL - Perkins Mk3 Quick Guide Issue 1 March 2015 - 2015-03.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions