Huntleigh Healthcare
Huntleigh Dopplers other models
dopplexAssist Vascular Assist Short Form PVR Leg Arteries Guide Ver 5 sw
Guide
2 Pages
Preview
Page 1
Vascular Assist Short Form Guide Version 5 Software Application
PVR Leg Arteries
1.
Connect the cuff and inflating bulb to the pneumatic connections on the module.
2.
Press to enter patient details and follow step 3, or press follow step 4.
to collect waveforms and
Patient Details 3.
Using the QWERTY touch screen, enter ‘assist’ for the password, then enter the patients Surname or Ref. no. and press
.
If patient details already exist, use the
Alternatively, press
keys to select the name from the list.
to create a new patient.
Complete the Forename and I.D. No. fields. After completing the I.D. no., press
to exit the registration process. Check the details on the screen before pressing save the information. 4.
Use the touch screen to select PVR – Leg Arteries.
5.
Wrap the cuff around the left Ankle and select Left Ankle using the touch screen.
6.
Gently inflate the cuff to 60mmHg.
7.
Press
to freeze the waveform.
8.
Press
once to accept the waveform.
9.
Wrap the cuff around the Right Ankle.
710394-4
and
to
10. Select Right Ankle using the touch screen to start the waveform capture. 11. Gently inflate the cuff to 60mmHg.
12. Press
to freeze the waveform.
13. Press
to accept the waveform.
14. Press
to save the trace (only possible if patient details have been entered first).
15. Press
to print the report and press
16. Press
to return to the test selection screen.
710394-4
to print.