Quick Reference Guide
3 Pages
Preview
Page 1
Nimbus Professional
Mattress Replacement System
• • • • • • • • • • •
Advanced dynamic flotation mattress replacement system Patient weight limit 250kg (39 stones) Auto-Matt® sensor pad for automatic adjustment Vent Valve Control for therapeutic patient management Head Section Deflate Anti-Sink torso section Heelguard® powered down heel section Triple action filtration including Bio-Filter Dynamic or static modes Transport mode supports patient for 12 hours Audible and visual alarm system
MRF-723 5-14
®
®
NIMBUS 4 & NIMBUS PROFESSIONAL Quick Reference Guide This Quick Reference Guide has been produced to assist Healthcare Professionals and other users with only the functional aspects of the system. The guide does not replace the general safety warnings, cautions and recommendations provided in the product instructions for use, part number 649933EN. Refer to the product instructions for use before using this product. Indications: •
The Nimbus 4 and Nimbus Professional mattresses are designed for patients weighing up to 250 kg (550 lb).
•
The Nimbus 4 and Nimbus Professional systems are indicated for the prevention and/or management of all categories of pressure ulcer, when combined with an individualised, comprehensive pressure ulcer protocol.
1
Contraindication: Do not use Nimbus 4 and Nimbus Professional systems for patients with unstable spinal fractures. Caution: If patients have other unstable fractures, or conditions which may be complicated by a soft or moving surface, advice should be sought from an appropriate clinician before use. Note: For more information on putting the patient in the prone position on the Nimbus Professional mattress refer to the Instructions for Use document, section 6 “Nimbus Professional Mattress: Patient Positioning Guide”.
Power Switch & Alarm Reset - Switches the mains power to the pump on and off. When the pump detects an alarm condition, it can be cancelled by switching the pump off and then back on. The Wait indicator is illuminated when the mattress is being inflated. The indicator will remain illuminated until the mattress has been fully inflated. This may take up to 15 minutes. Alarms - During an alarm condition the triangular alarm symbol will flash, an audible alarm will sound and one of the following indicators will be illuminated: Low Pressure, High Pressure, Pump Fault, Power (Fail). The Low Pressure indicator is illuminated whenever the pump detects low pressure within the mattress. An audible alarm will sound unless cancelled by the Mute button. The indicator will extinguish once normal pressure is reached. Make sure the CPR is closed, the tubeset is connected properly and the mattress is set to Normal. The High Pressure indicator is illuminated whenever the pump detects high pressure within the mattress. An audible alarm will sound unless cancelled by the Mute button. The indicator will extinguish once normal pressure is reached. Make sure the tubeset is not kinked. Pump Fault - Internal pump malfunction. Call Service Engineer. The Power (Fail) alarm indicator will illuminate when a mains power failure has been detected. An audible alarm will sound. The alarm will continue until the mains power is resumed or the pump is switched off using the Power switch on the pump. If the power failure is prolonged, switch to Transport mode and disconnect the tubing. Examples of the Power (Fail) alarm are mains lead removal and power failure. The Service indicator will illuminate and remain on after a set number of running hours indicating the pump is ready for a service. The pump will continue to function normally while the Service indicator is illuminated. During an alarm condition the audible alarm can be muted by pressing the Mute button. The indicator on the Mute button will illuminate to show the audible alarm has been silenced. After approximately 15 minutes the alarm mute will be cancelled, and the audible alarm will sound. Press the Static button to select Static mode. An audible tone will sound when the button is pressed. The indicator on the Static button will only illuminate when the pump is in Static mode. Press the button again to return to Dynamic mode. The mattress cell pressure can be manually adjusted for patient comfort using the rotary Comfort Control dial.
1. NPUAP/EPUAP International Pressure Ulcer Guideline, 2009.
Installation Nimbus 4 & Nimbus Professional Mattress Replacement (MR) Place the MR directly onto the bed base, with the tubeset and CPR located at the foot end of the bed. Use the fastener straps provided to attach the MR to the bed frame.
Transport Mode To transport a patient using the Nimbus Professional mattress, turn the Transport control at the foot end of the mattress clockwise to Transport. Switch off the pump and disconnect the tubeset. In this mode the mattress will support the patient for up to 12 hours.
If the bed has divided sections for independent elevation of a patient’s head and/or knees, attach the mattress to the movable parts of the bed frame only.
To resume operation, reconnect the pump, turn the transport control counterclockwise to Normal and switch on. Transport Mode
To complete mattress installation Zip the protective cover over the mattress. Make sure that the ArjoHuntleigh logo is uppermost and at the foot end of the mattress. Check that the transport control is set to Normal and the CPR control is closed and locked in position. Make sure that the mains power cable and tubeset are positioned to avoid causing a trip or other hazard, and are clear of moving bed mechanisms or other possible entrapment areas. Where cable management flaps are provided along the sides of the mattress, these should be used to cover the mains power cable. The CPR control must be visible and accessible at all times. If the pump is to be hung from the end of the bed, attach the bed bracket to the bed frame. Alternatively the pump can be placed underneath the bed, either upright or lying on its back. Make sure the tubeset is securely connected to the pump and the mattress.
To Connect Tubeset
To Disconnect Tubeset 1
2
Normal Mode Transport
Transport
Vent Valve Control The Nimbus 4 has 5 cells with Vent Valves at the foot end. The Nimbus Professional has 19 cells all with Vent Valves (the shoulder support cell has no Vent Valve and can not be deflated). Closed Open To deflate a cell, turn the Vent Valve 180° clockwise, to point downwards.
Note: Refer to “Guidelines for Selecting Mattress Vent Valves to Open” in the “Operation” section of the product instructions for use.
CPR Control IN THE EVENT OF CARDIAC ARREST
2
1
Normal
Normal
In the event of a patient suffering cardiac arrest and CPR needing to be administered: To Activate the CPR 1.
Insert the mains power connector into the socket on the side of the pump. Connect the mains power cable to a suitable mains power outlet.
Lift the red handle on the control panel at foot end of the mattress. FAST DEFLATE
NORMAL
CP
R
SAFETY SIDES
1
TRANSPORT
2
Regarding the use of safety sides, always refer to your local trust policy AND the product instructions for use.
3
2.
Turn the handle counterclockwise.
Quick Start When the Nimbus pump is switched on using the Power switch, the pump will run a self diagnostic check for approximately 3 seconds during which all indicators will be illuminated. The Wait and Low Pressure indicators will remain illuminated while the mattress inflates. Allow approximately 15 minutes for the mattress to inflate fully. Warning: Do not place the patient on the mattress until it is fully inflated and normal operating pressure has been reached. Once the mattress is fully inflated, the Wait and Low Pressure indicators will extinguish. Place a bed sheet over the mattress and tuck in loosely. Once the patient is on the mattress, the pump will automatically sense and adjust the pressure in the cells to support the patient. ArjoHuntleigh AB
FAST DEFLATE
NORMAL
CP
R 1
TRANSPORT
2
3
3.
Pull the handle away from the panel. FAST DEFLATE
NORMAL
1
TRANSPORT
R
CP
2
3
4.
The grey triangular seal will rotate and the air will rapidly exhaust from the mattress.
To Reset the CPR 1.
Turn the grey triangular seal clockwise and push onto the connectors.
2.
Turn the red handle clockwise.
3.
Fold the handle flat to lock in position.