ON-Q® PUMP INSERT (MODEL SPECIFIC INFORMATION)
with
Fixed Flow Rate
MANUFACTURED BY:
I-Flow Corporation 20202 Windrow Drive Lake Forest, CA 92630 U.S.A.
Kimberly-Clark N.V. Belgicastraat 13 1930 Zaventem, Belgium
ON-Q® with Fixed Flow Rate
IMPORTANT INFORMATION This insert is supplemental to the ON-Q Pump Directions for Use (DFU), which contains additional information on usage of the ON-Q Pump, including Indications/ Contraindications as well as Cautions and Warnings regarding patient/user’s safety. Please read the entire insert before operating the ON-Q with Fixed Flow Rate device. Follow all instructions carefully to ensure the safety of patient and/or user.
CAUTIONS
1. Temperature: The ON-Q Flow Controller should be in direct contact with the skin (88°F, 31°C). Temperature will affect solution viscosity, resulting in faster or slower flow rate. Flow rate will increase approximately 1.4% per 1°F/0.6°C increase in temperature and will decrease approximately 1.4% per 1°F/0.6°C decrease in temperature.
INDICATIONS FOR USE Refer to the ON-Q Pump Directions For Use for complete information.
CONTRAINDICATIONS Refer to the ON-Q Pump Directions For Use for complete information.
3
ON-Q® with Fixed Flow Rate
ON-Q® with Fixed Flow Rate
DESCRIPTION OF DEVICE: See Figure 1
The ON-Q Fixed Flow Rate delivers medication at a continuous infusion rate. 1. Fill Port 2. ON-Q Pump 3. Clamp 4. Air Eliminating Filter 5. Flow Controller (tape to skin) 6. Distal Luer
Fill Port
➊ ON-Q Pump➋
Fig. 1
➌
NOTE: Refer to the ON-Q Pump Directions For Use for information on filling the pump and other usage instructions. This document includes specific instructions for ON-Q® pump models with Fixed Flow Rate only.
IMPORTANT: Start by reading the ON-Q Pump Directions For Use, then continue with this document.
Clamp
➍
Use Aseptic Technique
PRIMING THE ADMINISTRATION SET
Caution: It is important to completely prime the pump tubing. Failure to do so may prevent the pump from infusing. • Open clamp. • Remove tubing cap to start priming (up to 15 minutes). • When all air has been removed from the entire tubing and fluid flow is observed at end of the distal luer, the administration set is primed. • Close clamp and replace tubing cap until ready for use.
➏
➎
If administration set does not prime, follow these steps: 1. Attach a luer adapter or stopcock to the distal luer. 2. Attach a small syringe (10 ml preferred) to the other side of the adapter and pull back on the syringe to create suction. 3. Continue to create suction until all air is removed from the tubing and fluid flow is observed from the distal luer. Repeat as necessary. 4. Disconnect syringe and luer adaptor or stopcock, and observe pump for complete priming. 5. If this does not work, check to see if something else is impeding flow, such as medication precipitate, clamp is closed or kinked tubing.
STARTING THE INFUSION
• Connect catheter to pump tubing. Open clamp to begin infusion. • Tape Flow Controller to skin to ensure flow rate accuracy.
4
5
ON-Q® with Fixed Flow Rate
TECHNICAL SPECIFICATIONS
CAUTION: Filling the pump less than labeled fill volume increases flow rate. Filling the pump greater than labeled fill volume decreases flow rate. TABLE 1: DELIVERY TIME INFORMATION PRODUCT INFORMATION
0.5
1
2
2
4
5
4
5
6
10
Labeled Fill Volume (ml)
65
100
100
270
270
270
400
400
400
400
Maximum Fill Volume (ml)
65
125
125
335
335
335
550
550
550
550
Retained Volume (ml)
≤3
≤4
≤4
≤9
≤9
≤9
≤ 15 ≤ 15 ≤ 15 ≤ 15
MODEL
Labeled Flow Rate (ml/hr)
APPROXIMATE DELIVERY TIME
FILL VOLUME (ml)
24 hrs
1 day
65
36 hrs
1.5 days
80
48 hrs
2 days
60 hrs
2.5 days
72 hrs
3 days
45
84 hrs
3.5 days
96 hrs
4 days
120 hrs
5 days
60
370
100 125
215
255
460
250
290
80
175
285
330
50
90
195
315
340
420
55
100
215
380
475
65
120
250
450
430
540
DELIVERY ACCURACY: When filled to the labeled volume, flow accuracy is ± 15% of the labeled infusion rates when infusion is started 0-8 hours after fill and delivering nominal saline as the diluent at 88°F (31ºC) against a back pressure of 40 cm of water. CAUTION: Temperature will affect solution viscosity, resulting in faster or slower flow rate. Flow rate will increase approximately 1.4% per 1°F/0.6°C increase in temperature and will decrease approximately 1.4% per 1°F/0.6°C decrease in temperature.