ICARE
ic100 Quick Guide Ver 1.2
Quick Guide
4 Pages
Preview
Page 1
QUICK GUIDE ENGLISH
QUICK GUIDE
TURNING THE TONOMETER ON AND LOADING THE PROBE 1. Press the Select or Measure button to turn the tonometer on. Illustrations of these alternative ways of starting: (Remember to use the wrist strap)
Pressing the Select button
Pressing the Measure button
2. Open the probe tube by removing the cap and insert 3. When the Play-symbol appears on the display, the the probe into the probe base as shown in the image.
MEASUREMENT Step 1. Ask the patient to relax and
look straight ahead at a specific point. Bring the tonometer near the patient’s eye.
tonometer will be ready for measurements.
Icare ic100 tonometer Quick Guide
ENGLISH
Step 2. Keep the probe horizontal and pointing perpendicularly to the center of the cornea. The distance
from the tip of the probe to the patient´s cornea should be 4-8 mm (5/32 - 5/16 inch). The probe base light must be green.
Green probe base light indication
Step 3. Press the Measure button to measure IOP. Probe will contact the eye during measurements: Press and hold the button to collect 6 measurements in a row OR Press the button briefly 6 times to collect 1 measurement at the time. ONE SHORT BEEP Indicates a correct single measurement. DOUBLE BEEP Indicates an incorrect single measurement. (See error messages from full manual). LONG BEEP Indicates that six succesfull measurements are done and IOP result is available at the display.
DISPLAY DURING MEASUREMENTS
Before
After the second measurement
After the sixth measurement
After the sixth After the sixth measurement but with measurement with too acceptable variation much variation
PARTS OF THE TONOMETER
Forehead support adjusting wheel Display Measure button Select button
Forehead support Probe base and probe base light Collar Navigation buttons
CHANGE/CLEAN THE PROBE BASE REGULARLY TO AVOID UNNECESSARY SERVICE. CLEANING THE PROBE BASE The probe base can be reused after careful cleaning. Make sure that the probe base is completely dry before using it. 1. Unscrew the probe base collar and carefully remove the probe base from the tonometer. 2. Invert the probe base over a container filled with pure alcohol solution, allowing the probe base to drop into the container and let soak for 5 - 30 minutes. 3. Dry the probe base by blowing some compressed air into it. 4. Carefully reinsert the completely dry probe base, screw the collar back into the tonometer.
www.icaretonometer.com
Please refer to the manual for additional information. (USB)
TA011-058 EN-1.2
Please refer to the manual for more detailed instructions of changing and cleaning the probe base.