Patient Warming System Instructions for Use
700-307 Revision: 8
2
Inspiration Healthcare
Contents 1.
Introduction General Description Description of Parts
3 3 3
2.
Symbols
6
3.
Safety Intended Use Indication for Use Contraindications Warnings Disclaimers User Profile
8 8 8 8 8 10 10
4.
System Description/Application AlphaCore5 Controller Mattresses and Blanket
10 10 11
5.
Connection and Disconnection Power to the AlphaCore5 Controller Power from the AlphaCore5 Controller ODU Connector Amphenol Connector Patient Probe
12 12 12 12 13 13
AlphaCore5 Controller Start-up AlphaCore5 Controller Start-up Cycle Bootloader Screen Bootloader HIGH PRIORITY ALARM Test Bootloader MEDIUM PRIORITY ALARM Test Bootloader LOW PRIORITY ALARM Test Commence Normal Operation
14 14 14 14 14 14 15
System Setup Control Descriptions Settings Screen (with active choices in green) Clock and Calendar Adjustment Screens Clock Adjustment Screens Calendar Adjustment Screens How to Use the AlphaCore5 Controller User Selectable Options Checking the Clock
16 16 16 17 17 17 18 21 22
6.
7.
Inspiration Healthcare Limited W inspiration-healthcare.com
8.
Normal Operation
22
9.
Checking of Thermal Cut-Out Monthly Thermal Cut-Out Test Notification Thermal Cut-Out Test Access Settings Menu Select Thermal Cut-Out Test Thermal Cut-Out Test Screen Thermal Cut-Out Detected Thermal Cut-Out Not Detected
23 23 23 23 24 24 25 25
10. Checking of Alarms Temperature Sensor Alarm Test Access Settings Menu Select Temperature Sensor Alarm Test Temperature Sensor Alarm Test (ALARM-3-TEMP-HIGH) Temperature Sensor Alarm Test (ALARM-2-ACCESSORY-THERM) Temperature Sensor Alarm Test (ALARM-2-ACCESSORY-CUT-OUT)
26 26 26 26 27
11. Storage
30
12. Infection Control and Cleaning
30
13. Service & Maintenance
30
14. Disposal
31
15. Problem Solving and Warning Messages
32
16. Power Status Indicators
35
17. Alarms, User Alerts and User Actions
36
18. Technical Description and Specification
42
© 2018 Inspiration Healthcare Limited
28 29
700-307 Revision: 8
Patient Warming System Instructions for Use
3
Patient Warming System 1. Introduction General Description The Inspiration Healthcare Limited Patient Warming System is designed for use in the operating room, recovery room, anaesthetic room, intensive care and emergency department. The AlphaCore5 Controller has only one mode of operation, providing safe and controlled warming to assist patients in maintaining normal body temperature. The applied parts are available as: • Mattresses for use under a patient • Blanket to be placed over a patient • Patient probe to indicate patient temperature (provided by third party manufacturer) Mattresses and blanket can be used in combination with each other simultaneously providing the total power draw of the combination does not exceed that required by two IHC-OTM1 mattresses (maximum total loading 170 watts). The patient’s core temperature should be regularly and independently monitored.
Different sizes and models of mattress and blanket are available from Inspiration Healthcare Limited for various procedures and situations. The AlphaCore5 Controller powers the mattresses and blanket at low voltage, ensuring safety for patients and operators. The temperature is controlled automatically to an operator selected level. An overtemperature safety cut-out is integrated into each mattress and blanket. The AlphaCore5 Controller is provided with a multi-purpose clamp and can be placed on a flat surface or mounted on an Intravenus pole (IV pole), anaesthetic machine, or Medi-rail. The system can control up to 5 accessory mattresses and/or blanket for one patient. It has a range of operating temperatures from 25-40°C and is designed to be operated continuously, maintaining a uniform heat under and/or over the patient. The serviceable life of the AlphaCore5 Controller is 5 years and serviceable life of the mattresses, blanket and connecting cables is 3 years under normal use conditions.
Description of Parts Figure 1 shown below is a picture of the AlphaCore5 Controller with mattresses and blanket.
Figure 1 – AlphaCore5 Controller with mattresses and blanket
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
4
Inspiration Healthcare
AlphaCore5 Controller Part Number
Part Description
IHC-1000
AlphaCore5 Controller
Note: Max Watts per channel = 85W, Sum of all channels Max Watts = 170W Table 1 - AlphaCore5 Controller Details
Operating Theatre Mattress Part Number
Part Description
Wattage
Warming Up Time
IHC-OTM1
O.T. Mattress, Full Length (1900mm x 585mm)
77W
45 minutes
IHC-OTM1-N
O.T. Mattress, Full Length, Narrow (1900mm x 535mm) 77W
45 minutes
IHC-OTM2
O.T. Mattress, ¾ Length (1200mm x 585mm)
65W
60 minutes
IHC-OTM2-N
O.T. Mattress, ¾ Length, Narrow (1200mm x 535mm)
65W
60 minutes
IHC-GTM1
O.T. Mattress, ½ Length (1070mm x 585mm)
55W
60 minutes
IHC-GTM1-N
O.T. Mattress, ½ Length, Narrow (1070mm x 535mm)
55W
60 minutes
Table 2 - Operating Theatre Mattress Details
Neonatal Mattress Part Number
Part Description
Wattage
Warming Up Time
IHC-NCM1
Neonatal Mattress, Small (610mm x 340mm)
55W
31 minutes
IHC-NCM3
Neonatal Mattress, Medium (610mm x 430mm)
40W
25 minutes
IHC-NCM5
Neonatal Mattress, Panda (660mm x 480mm)
35W
30 minutes
IHC-NCM9
Neonatal Mattress for LifeStart™
13W
37 minutes Table 3 - Neonatal Mattress Details
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
5
Recovery Room Blanket Part Number
Part Description
Wattage
Warming Up Time
IHC-RB1
Recovery Room Blanket (1660mm x 800mm)
58W
30 minutes
Table 4 - Recovery Room Blanket Details Note: Any Mattress can be used in combination with the Recovery Blanket Note: Warming Up Times shown are from 23°C to 37°C and are approximate
Mattress/Blanket Connection Cables Amphenol Connector (right-angled) Part Number
Part Description
IHC-1020-10
Connection Cable, 1m (right-angled)
IHC-1020-30
Connection Cable, 3m (right-angled) Table 5 - Mattress Blanket Amphenol Connecting Cable Details
Mattress/Blanket Connection Cables Amphenol Connector (straight) Part Number
Part Description
IHC-1021-10
Connection Cable, 1m (straight)
IHC-1021-30
Connection Cable, 3m (straight) Table 6 - Mattress Blanket Amphenol Connection Cable Details
Skin Probes Part Number
Part Description
TCFMT400/ AS-THT
Re-usable Skin Probe (supplied by Metko) (not for USA)
TCFMT400/ ASD Disposable Skin Probe (supplied by Metko) (not for USA) TCFMT400/ AECTHT
2 pin Redel Connector Table 7 - Skin Probe Details
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
6
Inspiration Healthcare
2. Symbols Warning: Read fully these operating instructions as they may affect safety or prevent damage The use of this equipment requires detailed knowledge and the observation of the Instructions for Use. This device is only intended for the described use Product complies with the provisions of the Medical Device Directive (93/42/EEC) Defibrillation Proof Type BF Equipment
1 2 3 4
1 The AlphaCore5 Controller is only compatible with the following patient warming accessories: Part Number
IHC – OTM1 IHC – OTM1-N IHC – OTM2 IHC – OTM2-N IHC – GTM1 IHC – GTM1-N IHC – NCM1 IHC – NCM3 IHC – NCM5 IHC – NCM9 IHC – RB1 2 Partial covering of the mattress/blanket device by pillows or other items having good insulating properties may impair the heating performance of the device 3 Mattresses and blanket should be used flat and unfolded 4 Mattresses and blanket can be used with direct patient contact or separated with a layer of fabric
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
Do not use Gel Pad(s)
Serial number
Patients lie on top of mattresses
Do not use if package is damaged
Patients lie underneath blanket
Keep dry
Printed side up
Keep away from sunlight
Do not fold it
Not for General Disposal
Roll it if necessary with print facing outwards
Humidity limitation (10% to 70% RH) 40
Patient temperature probe input
Manufacturer
Limit of temperature (-10°C to 40°C)
10
Caution: United States Federal law restricts this device for sale by or on the order of a physician
Mattresses and blanket may become damaged or perforated by sharp objects or instruments
Catalogue number
7
IP32
Mattresses / Blanket are rated to a degree of Ingress Protection (IP) 32
IP3X
AlphaCore5 Controller is rated to a degree of Ingress Protection (IP) 3X
Date of Manufacture
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
8
Inspiration Healthcare
3. Safety Intended Use The Inspiration Healthcare Limited Patient Warming System, consisting of a AlphaCore5 Controller, heating mattresses and blanket, is a system for use in supporting the maintenance of a patient’s core body temperature within the normothermic range between 36.5°C and 37.5°C. The patient applied parts of mattresses and blanket are for either neonatal or paediatric to adult patient ranges. An independent device should be used to regularly measure the patient’s core body temperature whilst the system is in use. Indication for Use To support the maintenance of normothermia in patients at risk or who may become at risk of inadvertent hypothermia within a hospital setting. The system should be used for all age ranges from Neonatal to Adult. Contraindications It is the responsibility of the user to determine whether warming is appropriate for each individual patient. The Patient Warming System should not be used on patients where clinical considerations indicate that warming of the patient is not advisable (e.g. aortic cross clamping). Caution should be used when using mattresses and blanket were patients have ischemic limbs. The Inspiration Healthcare Limited Patient Warming System and all its accessories are not MRI compatible. A potential increase in delivery of transdermal medications when in direct contact with either mattresses or blanket (for example, skin patches) may cause patient harm. Inspiration Healthcare Limited W inspiration-healthcare.com
Avoid direct contact with mattress or blanket in circumstances where patient’s skin is compromised. Warnings The user should carefully read the instructions for use before using the Patient Warming System AlphaCore5 Controller, mattresses and blanket to become familiar with the function of the equipment. On a monthly basis, the user should check the functionality of the thermal cut out by following the instructions in Section 9. This functional test must not be used when a patient is in contact with the warming accessory. Do not use the Patient Warming System if there is any sign of damage. Ensure the surface integrity of the mattresses and/or blanket and replace if necessary. Do not use the Patient Warming System in the presence of flammable anaesthetic gases or oxygen enriched atmospheres (zone of high risk) as these may pose an explosion hazard. Do not position the AlphaCore5 Controller in the path of a direct X-ray beam (mattresses and blanket are radiolucent and their performance will not be affected by X-ray activity). The time for the mattresses and blanket to reach a set temperature of 37°C from ambient (23°C ± 2°C) will vary depending on the size, with a maximum estimated time of 30 minutes for the smallest to 45 minutes for the largest (these may increase over the lifetime of the product). The AlphaCore5 Controller displays the heated element temperature, not the patient contact temperature. Whilst these may be very similar the actual contact surface may be between 1 and 3°C lower than the displayed temperature.
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
The AlphaCore5 Controller has an alarm system with an interruption of power supply mains alarm condition. Do not use the blanket/mattresses with any AlphaCore5 Controller or connection cables other than those supplied with the system, or approved by Inspiration Healthcare Limited.
Pooling of surgical preparation solutions under the patient is specifically contraindicated, due to risk of skin irritation. Do not allow surgical skin preparation solution to pool under the patient. The AlphaCore5 Controller is only compatible with the following patient warming accessories: IHC-OTM1
O.T. Mattress, Full Length (1900mm x 585mm)
IHC-OTM1-N
O.T. Mattress, Full Length, Narrow (1900mm x 535mm)
IHC-OTM2
O.T. Mattress, ¾ Length (1200mm x 585mm)
IHC-OTM2-N
O.T. Mattress, ¾ Length, Narrow (1200mm x 535mm)
IHC-GTM1
O.T. Mattress, ½ Length (1070mm x 585mm)
IHC-GTM1-N
O.T. Mattress, ½ Length, Narrow (1070mm x 535mm)
IHC-NCM1
Neonatal Mattress, Small (610mm x 340mm)
IHC-NCM3
Neonatal Mattress, Medium (610mm x 430mm)
IHC-NCM5
Neonatal Mattress, Panda (660mm x 480mm)
Do not position the AlphaCore5 Controller in such a way as to make it difficult to disconnect the mains lead.
IHC-NCM9
Neonatal Mattress for LifeStart™
IHC-RB1
Recovery Room Blanket (1660mm x 800mm)
If disconnection from the mains power should be necessary, the IEC mains lead should be detached from the AlphaCore5 Controller or from the mains socket.
Partial covering of the mattress/blanket device by pillows or other items having good insulating properties may impair the heating performance of the device.
In the event of a power failure, use a back-up Power Supply such as a UPS (Uninterruptable Power Supply) or Institutional back-up generator.
Mattresses and blanket should be used flat and unfolded.
The mattresses/blanket are not to be wrapped around the patient. Ensure the surface of the mattress or blanket has no folds or creases and check that the surface is not damaged prior to placing on the operating table or trolley. Do not use if there is any damage, perforation or wear. Do not partially cover the mattresses with insulating materials like pillows/blankets. It is important to ensure that the mattress is firmly secured and that the straps cannot slide off the operating table brackets’ in relation to the operating theatre mattress. No part or accessory to the Patient Warming System can be maintained or serviced while the system is in use. To avoid the risk of electric shock this equipment must only be connected to an electrical supply with a protected earth.
The Patient Warming System is not sterile. Where necessary the user should take appropriate precautions to protect the sterile field. Inspiration Healthcare Limited W inspiration-healthcare.com
9
Mattresses and blanket can be used with direct patient contact or separated with a layer of fabric. Channels which are not in use should be left covered.
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
10
Inspiration Healthcare
4. System Description/Application Disclaimers
AlphaCore5 Controller
Inspiration Healthcare Limited assumes no responsibility for the use or reliability of its Patient Warming System AlphaCore5 Controller, mattresses or blanket if they are not used correctly as a system in conjunction with appropriate accessory mattresses, blanket and AlphaCore5 Controllers authorised for use by Inspiration Healthcare Limited.
The AlphaCore5 Controller is a precision temperature Controller to be used only in conjunction with the low voltage, electrically heated Inspiration Healthcare Limited branded blanket/mattresses.
Inspiration Healthcare Limited makes no warranty of any kind with regard to other manufacturer’s Controller , mattress or blanket applications that are created to be used in conjunction with Inspiration Healthcare Limited devices by the user. This document contains confidential and proprietary information that is protected by copyright. All rights are reserved. Any unauthorised copying, storage, reproduction or translation of this document in any form is strictly prohibited. User Profile Designed for use by, or under the supervision of, a registered medical professional (Doctor, Midwife Nurse or Practitioner) in the operating room, recovery room, anaesthetic room, intensive care, emergency department, and/or medical, obstetric, maternity and surgical wards.
Mains voltage is supplied to the AlphaCore5 Controller via an IEC socket and IEC mains lead. There is a device on/off button at the top of the unit and a touchscreen interface along with the active/alarm bar which illuminates when power is connected, and the on/off button is pressed. At the top is a carrying handle and to the rear is a clamp to allow the unit to be mounted on an IV pole, anaesthetic trolley or Medi-rail. At the front of the unit is a touchscreen GUI (Graphical User Interface) (see Figure 2) control panel from which all the functions can be accessed. There are five yellow ODU-style connection points, three at the front and one on either side to allow for the connection of up to five accessories: mattresses in combination with blanket. The connection cables are provided for mattresses/blanket in a number of lengths and configurations. There is a blue ODU-style connector on the right-hand side of the device for a patient temperature probe. There is an IEC socket on the left-hand side of the device for mains power input. The AlphaCore5 Controller input is protected by two 3Amp fuses which are accessible on the side of the unit underneath the IEC socket (for qualified service personnel only). The AlphaCore5 Controller provides a 26Vdc supply to the mattresses/blanket and receives temperature information back. The temperature setting is controlled and indicated on the GUI. When the mattresses/blanket are warming up the channel indicator will be blue, if passively cooling from a
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
higher set temperature to a lower one the channel indicator will be amber and green when at the set temperature.
37
A pressure relief pad is integrated into the mattress underneath the flexible warming surface to work in conjunction with the operating table mattress. This may help reduce the incidence of pressure sores, without any attenuation of the warming performance. Integral straps are used to retain the mattress in-situ.
C
o
The blanket is designed to operate over the patient and constructed in the same manner as the mattresses, but without the pressure relief pad which is replaced by a soft insulating material.
37.9oC
Figure 2 - AlphaCore5 Controller GUI
The maximum working power output is equivalent to controlling two full size mattresses (IHC-OTM1) irrespective of configuration (170 watts). There is no active cooling with this device, reduction in accessory temperature is passive and will not decrease below ambient. Mattresses and Blanket Mattresses must be fitted with the marked side uppermost, and the plain surface facing downwards. The blanket is to be placed over the patient with the writing uppermost, away from the patient. Mattresses are available in various sizes and are only designed to operate under the patient. They provide an even temperature over the whole surface and are extremely flexible. The basic construction of mattresses across the range is the same. The mattresses are water and solvent resistant and have been tested for bio-compatibility to allow skin contact during use. All seams are fully sealed by RF Welding.
Inspiration Healthcare Limited W inspiration-healthcare.com
11
On initial start-up, if the ambient temperature is low, the mattress/blanket will take longer to achieve the desired temperature. If the desired temperature is not achieved within the defined warming up time, there will be a visual and audible indication. Should this happen, the AlphaCore5 Controller will continue to power up the mattress/blanket until the desired temperature is achieved. The mattresses/blanket contain a special conductive heating material and will not cause any environmental heating. The heat emitted by this material will only be felt by the user/patient when applying pressure or weight to the mattress. It is normal that the mattresses/blanket do not feel particularly warm to the touch when left uncovered. The AlphaCore5 Controller monitors temperature and when the mattress/blanket reaches the desired temperature the AlphaCore5 Controller will maintain this automatically. If the maximum allowed temperature is exceeded, a visual warning indicator will be displayed, and the over-temperature alarm will sound. An internal safety cut-out system will operate if any fault condition causes the mattress/blanket to exceed a temperature of 41°C.
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
12
Inspiration Healthcare
5. Connection and Disconnection Refer to Section 1: Introduction for the list of connection cable options and accessories. Power to the AlphaCore5 Controller Power can be applied from two sources: a mains cable and the internal battery. The internal battery will only support the independent use of the NCM-9 mattress for a minimum of 30 minutes (from fully charged). The mains power input is 100-240Vac, 50-60Hz. The battery is on continuous charge whilst mains power is connected, and the AlphaCore5 Controller turned on. To connect the mains cable, hold the connector body and push into the mains socket. To remove the cable, hold the connector body and pull, do not twist or pull by the cable. The AlphaCore5 Controller can be powered via other medical devices, provided that they have an isolated earth.
Inspiration Healthcare Limited W inspiration-healthcare.com
Power from the AlphaCore5 Controller Connection cables are provided to take power and control signals between the mattresses/blanket and AlphaCore5 Controller . Warning: Do not force the connector into the socket. Ensure it is correctly orientated by aligning the directional arrow on the connector body with the marker on top of the terminal housing. ODU Connector For connection to the AlphaCore5 Controller, there is a choice of right angled and straight ODU-style connectors which are colour coded yellow. The right-angled connector is designed to point the cable downwards, away from the display. The connector pushes into the yellow terminals on the AlphaCore5 Controller. The connectors are keyed – align keying before pushing home – never force connectors. To disconnect, hold the main body of the connector and pull. Do not twist or force the connection.
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
Amphenol Connector
Patient Probe
The Amphenol connector at the opposite end of the connection cable plugs into the connector fitted to the mattress/blanket. Align the plug and socket then push together and rotate the black collars clockwise to lock the connection as shown below.
The patient probe is connected with an ODU-style connector to the blue connection on the righthand side of the AlphaCore5 Controller This can be connected and disconnected to the AlphaCore5 Controller as described in ODU Connector section above.
13
Warning: Clinical decisions should not be made based on the patient temperature probe reading of the AlphaCore5 Controller – it is for indication only. Do not attempt to force the connectors and pay attention to the orientation of the plugs for proper connection.
The ODU-style connectors are all keyed (and colour coded) so that only a mattress cable can be plugged in to the mattress output; only the patient probe can be connected to the patient probe input. To disconnect, hold the connector locking rings, rotate counter-clockwise and then pull them apart.
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
14
Inspiration Healthcare
6. AlphaCore5 Controller Start-Up AlphaCore5 Controller Start-Up Cycle
Bootloader MEDIUM PRIORITY ALARM Test
After pressing the power button and the unit powers up, the bootloader screen is displayed. At the final stage of the bootloader process the HIGH, MEDIUM, and LOW alarm audible signals should each be initiated for a signal alarm burst. At the same time, the visual alarm bar at the top of the unit should be illuminated RED or AMBER according to the priority of the audible alert being sounded.
The visual alarm bar should be illuminated in AMBER. The audible alarm should complete at least one full sequence of the MEDIUM PRIORITY ALARM tone.
Bootloader Screen In the first stage of the bootloader, the visual alarm bar should be GREEN and there should be no audible alarms sounding.
Bootloader LOW PRIORITY ALARM Test The visual alarm bar should be illuminated in AMBER. The audible alarm should complete at least one full sequence of the LOW PRIORITY ALARM tone.
Bootloader HIGH PRIORITY ALARM Test The visual alarm bar should be illuminated in RED. The audible alarm should complete at least one full sequence of the HIGH PRIORITY ALARM tone.
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
15
Commence Normal Operation After completing the alarm system test, the bootloader should complete and the unit commence normal operation. At this point all audible alarms should be turned OFF and the visual alarm bar illuminated GREEN (unless real alarms are active).
37
C
o
Inspiration Healthcare Limited W inspiration-healthcare.com
37.9oC
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
16
Inspiration Healthcare
7. System Setup Connect and utilise the AlphaCore5 Controller , mattresses/blanket with consideration to safety as outlined in Section 3. Control Descriptions Below is a list of meanings of all the buttons on the display and indicator lights:
Power on and off: 3 second hold
8 2
1 5 6
37
3
C
o
9
1. Settings menu 2. Power source and battery status 3. Mains connected 4. Accessory connection status 5. Increase desired temperature 6. Decrease desired temperature
37.9oC
7
10
7. Initial accessory connection indication (with associated active alarm bar amber)
4
8. Device status and alarm bar green okay and amber active 9. Set temperature target of accessories 10. Patient probe connected and temperature measured
Settings Screen (with active choices in green) V1.0.0 (PMU v1.0)
o
C
.
F
o
,
2
1. Clock and Calendar adjustment option screens
3
2. Degrees Celsius or Fahrenheit (for patient probe temperature display) 3. Decimal point or comma
4 5
4. Screen locked or unlocked 5. Tick press to confirm
1
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
17
Clock and Calendar Adjustment Screens Calendar Adjustment Screen
Clock Adjustment Screen
1
00:00
6
2
00-00-00
5
DD - MM - YY
5
3 4
1. Adjustment up arrow 2. Adjustment down arrow 3. Hours 4. Minutes 5. Tick for confirmation of settings 6. Day – Month – Year
1. Settings icon faded whilst screen lock is active
1
37
2. Screen Locked icon
C
o
2
Inspiration Healthcare Limited W inspiration-healthcare.com
37.9oC
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
18
Inspiration Healthcare
How to Use the AlphaCore5 Controller Securely mount the AlphaCore5 Controller to either a pole or Medi-rail using the clamp provided. The device can be used free standing; however, it should be positioned with due consideration to the local environment and safety. Rail Mount
Pole Mount
Connected to the mains supply, press the on/off switch to the top right of the display for 3 seconds. The device will power up with the touchscreen displaying the target temperature. To the left of this temperature value are up/down arrows which, when pressed, adjust the temperature in 1°C steps to enable the desired temperature to be set. Note: pressing and holding the arrows will not increase or decrease the value by more than 1°C. The temperature range is 25-40°C in 1°C increments. The touchscreen will display icons as indicated below.
37
C
o
37.9oC
On the top right of the display a battery icon, mains power connection icon, charging icon and a status bar will be shown. If a patient probe is connected, a smaller blue icon will be visible displaying the probe temperature (below and to the right of the main temperature setting display). If a temperature probe is not attached, this icon will not be displayed.
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
Patient Warming System Instructions for Use
Along the bottom of the screen, there are five accessory boxes related to the potential number of accessories that can be active at any one time (five in all). If none are connected, they will all remain grey.
37
19
C
o
37.9oC
Once an accessory mattress or blanket is connected the related grey box will flash amber, a repeating beep will be heard and the status bar above the general display will be amber.
37
C
o
The flashing amber box should be pressed to confirm connection of the accessory. The status bar will turn green, the beeping will stop, and the associated accessory box will be blue if under temperature and warming.
37.9oC
37
C
o
37.9oC
AlphaC
Inspiration Healthcare Limited W inspiration-healthcare.com
© 2018 Inspiration Healthcare Limited
700-307 Revision: 8
20
Inspiration Healthcare
If the attached accessory is above the set temperature, the associated accessory box will turn amber.
37
C
o
37.9oC
51.4784
When the set temperature is reached, the accessory box will turn green. Note that there is no active cooling.
37
C
37
C
o
If the accessory monitored temperature is 1oC over the set target temperature for 1 minute, the device will alarm with a repeating beep. The related accessory box will flash amber and the status bar will also turn amber. To silence this and confirm observation, press the accessory box flashing amber. The box will remain amber until the temperature of the accessory normalises. The alert condition will return if the temperatures do not become normalised. This applies to any number of accessories connected individually.
37.9oC
o
37.9oC
51.4784
If an accessory is disconnected, the device will alarm with a repeating beep, the status bar will turn amber, and the related accessory box will flash amber with a white ‘disconnected plug’ icon. To accept the disconnection, press the related icon and the alarming box will return to grey. Should the disconnection be accidental then reconnect to the AlphaCore5 Controller and reset the alarm condition by pressing the associated box (the ‘unplugged’ icon will disappear). Accept connection by pressing the accessory box again and the colour of the box will change from amber to the appropriate status as previously described.
Inspiration Healthcare Limited W inspiration-healthcare.com