INOmax Connection Guide to Sensormedics 3100 series HFOV with a Filtered Circuit

Guide

1 Page

INOmax® DSIR Connection to a Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Filtered Circuit Warning:  Omission of the one-way valve may result in high NO delivery.  Caution:  Refer to page 42 of the Operation and Maintenance manual for a list of general precautions.  1  2  1  3  2  4  3  4 5  6  15 14 9  7  15 14  8  9  9 10  5  6  7 8 9 10 a  a c  b  13  12  13  11 12  b  11  6.  NO/N2 Injector Tube  12. Bias Flow Tube  a.  2 . Ventilator Outlet  7.  One-Way Valve  b.  22F Inlet  3.  Injector Module  8.  Paw Limit Valve Control  13. Patient Gas Sample Line Connection  c.  22M / 15F Outlet  4.  Injector Module Electrical Cable Connection  9.  Filter  14. Dump Valve Control  5.  INOmax DSIR  11. Humidifier Outlet  1.  Sensormedics 3100A/B Ventilator  10. Humidifier Inlet  Injector Module  15. Paw Control Valve  Figure 1 Example: High Frequency Oscillatory Ventilator Diagram  Note:  Connections to various ventilators are unique to each manufacturer as well as their corresponding disposable circuits. For technical assistance, please contact Customer Service on 1300 198 565.  c