INOmax Connection Guide to Sensormedics 3100 series HFOV with a Rigid or Flexible Circuit

Guide

1 Page

INOmax® DSIR Connection to a Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Rigid or Flexible Circuit Warning:  Omission of the one-way valve may result in high NO delivery.  Caution:  Refer to page 42 of the Operation and Maintenance manual for a list of general precautions. 1  2  1  2  3  3 4  5  5  6  7 11  8  11  6  7 8 a  9  10  1.  Sensormedics 3100A/B Ventilator  2.  Ventilator Outlet  3.  Injector Module  4.  INOmax® DSIR  5.  NO/N2 Injector Tube Connection  4  a c  9  b  b  10  Injector Module Electrical Cable Connection  10. Patient Gas Sample Line Connection  a. b.  22F Inlet  7.  One-Way Valve  11. Bias Flow Tube  c.  22M / 15F Outlet  8.  Humidifier Inlet  9.  Humidifier Outlet  6.  Injector Module  Figure 1 Example: High Frequency Oscillatory Ventilator Diagram  Note:  Connections to various ventilators are unique to each manufacturer as well as their corresponding disposable circuits. For technical assistance, please contact Customer Service on 1300 198 565.  c