INO Therapeutics

INOmaxDS ir Pre-Use Checkout Rev 03 Dec 2012

Guide

2 Pages

Pre-Use Checkout A. Initial Connections (check all connections) 1. Check o-ring on regulator for damage 2. Connect high pressure regulator to INOMAX® cylinder and INOmax DSIR  B. Perform high pressure leak test Note: Make sure INOMAX cylinder icon is present and INOmax DSIR backup and INOblender are OFF 1. Open/Close INOMAX cylinder valve  Note: Once the calibration is completed (bars turn green) press the menu button twice to return to the main screen  D. Perform purge and alarm verification  4. Verify values OFF  NO = 14-26 ppm  8  1. Ensure oxygen flow of 10 L/min 2. Set INOMAX dose to 40 ppm 3. Verify values Acceptable NO Value 4. Turn the Acceptable NO2 Value INOMAX FiO2 dose to zero  3  4  5  6  1. O2 Flowmeter 2. O2 Tubing 3. 15 M x 4.5 mm Adapter 4. 22 M / 15 F x 22 M / 15 F Adapter 5. Injector Module 6. 300 mm of 22 mm hose 7. Gas Sample Tee 8. Patient Gas Sample Line with Nafion 9. NO/N2 Injector Tube 10. Injector Module Electrical Cable  35-45 ppm <1.5 ppm 95% ±3%  G. Perform INOblender Test 1. Remove the Pre-Use set-up oxygen tubing from the oxygen flowmeter and connect it to the front of the INOblender 2. Remove the injector module from the Pre-Use set up and reconnect the adapters  2 2  ON  3. Verify “Backup ON” alarm activates  5. Turn Backup delivery OFF  (second menu level)  9  2. Turn Backup delivery ON  F. Complete INOmax DSIR Performance Test  C. Perform low range calibration  10  1. Ensure oxygen flow of 10 L/min  NO2 = ≤1.0 ppm  2. Wait 30 seconds and ensure no pressure drop  1  E. Perform Backup delivery test  4  7  Assemble connectors and tubing as shown 1. Ensure INOMAX cylinder valve is closed 2. Connect to wall or cylinder oxygen 3. Set O2 Flowmeter to 10 L/min 4. Set INOmax DSIR dose to 40 ppm “Cylinder Valve Closed” alarm will occur 5. Purge is complete when  “Low NO/N2 Pressure” alarm activates 6. Open cylinder valve 7. Turn INOMAX dose to zero (note: The “Set Dose is Zero, Close Cylinder Valve” indicator will appear, at this point do not close the cylinder valve)  40 ppm 3  5  3 10 L/min 5 1  2  3. On the INOblender, set the INOMAX dose to 40 ppm and O2 flow to 10 L/min Acceptable NO Value  32 - 48 ppm  4. Verify values on the INOmax DSIR 5. Turn the dose and flow to zero and remove the Pre-Use set-up from the INOblender The INOmax DSIR is now ready to connect to the patient  Warning: If not starting therapy within 10 minutes, depressurize the INOMAX regulator  End of Pre-Use Checkout  WARNING: This reference card is not a substitute for the INOmax DSIR and INOblender Operation and Maintenance manuals. Refer to these manuals for all applicable cautions and warnings.
File Type: PDF
File Size: 545 KB
File Name: INO Therapeutics - 20144 - INOmaxDS ir Pre-Use Checkout Rev 03 Dec 2012 - 2012-12.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions