Setup Pre-Use Procedures Checklist Calibration Connection Diagrams Monthly Checkout Checklist Circuit Connection Diagrams 1
Contents Setup Setup Checklist INOvent delivery system, front viexxw Injector Module INOvent delivery system, rear view User controls and indicators on the front Pre-Use Procedures Pre-Use Procedures Checklist Circuit connections for purge and performance test Sample tee connection to the Manual NO delivery system Calibration Connection Diagrams Connection diagram for NO and NO2 high range calibration Connection diagram for O2 high range calibration Monthly Checkout Monthly System Checkout Checklist Low pressure system leak test connection diagram Circuit connections for purge and performance test Circuit Connection Diagrams Neonatal circuit Pediatric circuit Adult circuit Anesthesia system with ventilator BIO-MED MVP-10 or Infrasonics Infant Star 100 ventilator Self-filling resuscitator bag Hyper-inflation bag Spontaneous breathing circuit Connectors and Adapters List Changing Cylinder & Purging Regulator
3 4 4 5 6 7 7 8 9 9 10 11 12 13 14 15 16 17 18 19 20 21 22
w Important: This guide is provided as a convenience and for general information only. Never use this product without clearly and thoroughly understanding the most recent revision of the Operation and Maintenance manual. The manual is the source for specific, updated information regarding checklists, 2 diagrams, and/or instructions contained in this Guide.
Setup Setup Checklist ❑
1. Connect supply hoses to INOvent delivery system and therapy gas.
❑
2. Check cylinders for correct product identity labels and NO concentration.
❑
3. Ensure both cylinders have adequate supply pressure: not less than 200 psig / 14 kPa x 100 (bar).
❑
4. Empty and clean the fluid trap if necessary.
❑
5. Connect sample line to the INOvent delivery system.
❑
6. Connect the injector module tubing and electrical cable to both the INOvent delivery system and injector module.
❑
7. Connect the manual NO delivery system.
❑
8. Connect the electrical power cord to an emergency-powerbacked, hospital-grade outlet.
3
Setup INOvent delivery system, front view Fluid Trap Filter Cartridge
Display and operating panel
Manual NO Delivery System Activity Indicator
Patient Gas sample line connector
Fluid Trap Bottle Injector Module Electrical Cable connector
Hose clip
NO/N2 Injector Tube connector
Injector Module Electrical cable connection to the INOvent Delivery System
Inspiratory limb breathing circuit; from ventilator 4
NO/N2 injector tube connection to the INOvent Delivery System Inspiratory limb breathing circuit; to humidifier
Setup INOvent delivery system, rear view
NO2 NO Sensor Sensor
Nurse Call connector
Serial Port connector Power Entry Module Power Cord connector
Connector to the O2 Flowmeter
External Ground connection point O2
NO/N2 Input Hose Connectors (2)
Sample Sensor Cooling Fan and Filter gas outlet
ON/STANDBY switch
Regulator Low Pressure Hose with Quick Connectors (2)
5
Setup User controls and indicators on the front
Alarm Silenced Countdown Timer
Alarm message area
Inspired NO2 Measured Value
Inspired NO Measured Value
Alarm History indication area Alarm History button
Alarm Silence button
NO Bargraph
Mains Power light Inspired O2 Measured Value Alarms button
6
Setup Calibration Pause Set NO Flow button button button button
Control Wheel
Pre-Use Procedures Pre-Use Procedures Checklist ❑
1. Do the “Initial connections and leak test.”
❑
2. Do the “System purge and performance tests.”
❑
3. Do the “Manual NO delivery system purge and performance test.”
Circuit connections for purge and performance test O2 Flowmeter Injector Module Electrical Cable
NO/N2 Injector Tube
Patient Gas sample line
O2 Tubing
15M x 4.5 mm Adapter
22M/15F x 22M/15F Adapter
Injector Module
300 mm of 22 mm hose
Sample Tee
7
Pre-Use Procedures Sample tee connection to the Manual NO delivery system NO/O2 outlet
O2 inlet
O2 supply hose from the O2 flowmeter
Oxygen tubing (user supplied)
Patient Gas sample line Sample Tee 15M x 4.5 mm Adapter
8
NO/N2 input hoses from the regulators
Pressure Compensated O2 flowmeter (user supplied), shown set to OFF (no float activity)
Calibration Connection Diagrams Connection diagram for NO and NO2 high range calibration Patient Gas sample line
Calibration gas cylinder pressure gauge
One-way valve
Cylinder ON/OFF control
Vent to the atmosphere or scavenge - Do not occlude
NO or N2 calibration gas cylinder
Connection diagram for O2 high range calibration O2 to the Sample System
INOvent delivery system Sample Line
100% O2
Flowmeter
Sample Tee
To the Atmosphere
9
Monthly Checkout Monthly System Checkout Checklist The monthly system checkout consists of:
10
❑
1. Initial connections and leak tests.
❑
2. Purge and system alarms tests.
❑
3. Calibration and monitoring alarms.
❑
4. INOvent delivery system performance.
❑
5. Manual NO delivery system purge and performance.
Monthly Checkout Low pressure system leak test connection diagram
NO/N2 Injector Tube
Injector Module Electrical Cable
NO/O2 outlet
O2 inlet
O2 tubing
15M x 4.5 mm Adapter
Injector Module
22M/15F x 22M/15F Adapter
NO/N2 input hoses from the regulators
Squeeze Bulb
Pressure Compensated O2 flowmeter
O2 supply hose from the O2 flowmeter 11
Monthly Checkout Circuit connections for purge and performance test
O2 Flowmeter Injector Module Electrical Cable
NO/N2 Injector Tube
Patient Gas sample line
O2 Tubing
15M x 4.5 mm Adapter
12
22M/15F x 22M/15F Adapter
Injector Module
300 mm of 22 mm hose
Sample Tee
Patient Gas sample line Ventilator Inspiratory port 22M/15F x 22M/15F Adapter To INOvent delivery system
9 mm I.D.
15 mm I.D. to 3/8 in. Tubing Adapter
22M /15F x 15M Adapter with gas sample tee
15 mm O.D. to 3/8 in. Tubing Adapter
22F x 15M Adapter Humidifier
13
Circuit Connection Diagrams Pediatric circuit
Patient Gas sample line Ventilator Inspiratory port 22M/15F x 22M/15F Adapter To INOvent delivery system
15 mm, 6 inch Inspiratory Extention
14
22M/15F x 15M Adapter with gas sample tee
22F x 15M Adapter Humidifier
Circuit Connection Diagrams Adult circuit
Patient Gas sample line Ventilator Inspiratory port 22M/15F x 22M/15F Adapter To INOvent delivery system 22M/15F x 15M Adapter with gas sample tee
22M/15F x 22M/15F Adapter
22F x 15M Adapter Humidifier
15
Circuit Connection Diagrams Anesthesia system with ventilator
Patient Gas sample line connection (“Sample Inlet” port)
Patient Gas sample line
INOvent delivery system
Injector Module Connections on absorber Inspiratory port Expiratory port
Patient wye
16
Sample 22M/15F x 22M/15F Tee Adapter
Hose
Circuit Connection Diagrams BIO-MED MVP-10 or Infrasonics Infant Star 100 ventilator Patient Gas sample line
Ventilator
Ventilator Inspiratory port
INOvent delivery system
Patient wye
22M/15F x 22M/15F Adapter Sample Tee
Expiratory Breathing Circuit Hose Inspiratory Ventilator Expiratory Breathing Circuit Hose Valve
Injector Module
NO/N2 Injector Tube
Injector Module Electrical Cable
17
Circuit Connection Diagrams Self-filling resuscitator bag O2 Flowmeter
Resuscitator bag with O2 reservoir Injector Module NO/N2 Electrical Cable Injector Tube
O2 Tubing Sample Tee
O2 Tubing
Patient Gas sample line 15M x 4.5 mm Adapter
22M/15F x 22M/15F Adapter
Injector Module
15M x 4.5 mm Adapter
O2 Tubing
w Caution: Tubing length downstream of the O2 tubing sample tee must be minimized. The INOvent monitoring system operates at pressures up to 70 cm H 2O above ambient at the sample tee. Excessive tubing length can result in “Monitoring Failure” alarm from over-pressurization. To help ensure proper INOvent delivery system operation, test the circuit configuration before using it. 18
Circuit Connection Diagrams Hyper-inflation bag
O2 Flowmeter
Patient Gas sample line
Injector Module Electrical Cable
O2 Tubing Sample Tee
NO/N2 Injector Tube O2 Tubing
O2 Tubing
15M x 4.5 mm Adapter
22M/15F x 22M/15F Adapter
Injector 15M x Module 4.5 mm Adapter
Pressure Gauge
Hyperinflation Bag
19
Circuit Connection Diagrams Spontaneous breathing circuit
O2 Flowmeter Injector Module Electrical Cable
O2 Tubing
15M x 4.5 mm Adapter
20
NO/N2 Injector Tube
Sample Tee
Patient Gas sample line
One-way Valves
Injector Breathing 22M/15F x Face Module Circuit 22M/15F Mask Hose Adapter 22M/15F x Breathing 22M/15F Circuit Adapter Tee Breathing Circuit Bag