MAYFIELD Imaging Overview and IFU Refresher Guide
Guide
22 Pages
Preview
Page 1
MAYFIELD® Imaging Overview and IFU Refresher
MAYFIELD is a registered trademark of SM USA Inc. and is used by Integra under license.
0688771-1
MAYFIELD® Imaging Products
•
Head Clamp, Base Unit and Accessories • • •
• • •
Radiolucent Horseshoe Headrest Spine Table Adapter (Jackson Table) Radiolucent Skull Pins • •
•
Metal Free Conversion Accessory Link Arm Extensions Tri-Star Swivel Adapter
Titanium Sapphire
Radiolucent Budde® Halo System (this system is outside the scope of this training)
MAYFIELD is a registered trademark of SM USA Inc. and is used by Integra under license. Budde is a registered trademark of Integra LifeSciences Corporation in the United States and/or other countries.
0688771-1
MAYFIELD Imaging: Head Clamp
•
Clamp: • • • •
Made of carbon-filled PEEK Little or no artifact Only metal is the spring in the torque knob Metal-free conversion accessory offers a completely metal free set-up (different torque knob)
Head Clamp (Standard Configuration)
Metal Free Conversion Accessory
0688771-1
MAYFIELD Imaging: Head Clamp 2
1
•
Standard Clamp 1. XR2 Standard Torque Screw (439A1095) 2. XR2 Adult Rocker (439A1094) 3. Swivel Lock Knob 4. Skull Clamp Base or Body 5. Plunger 6. Starburst Attachment 7. Skull Clamp Ratchet Extension
•
3
Accessories a. XR2 Force Applicator, 439A1093 (included with metal free conversion kit, 439A1092) b. Stop Wheel c. Metal free ratchet extension d. XR2 Child’s Rocker (439A1091)
Head Clamp (Standard Configuration)
Metal Free Conversion Accessory a
7
6
5
4
b c
d
0688771-1
MAYFIELD Imaging: Base Unit
•
Base Unit
Base Unit: • •
Metal free zone around the head clamp attachment / scanning area Width adjustable – 2 different sizes
Item No.
OR Table Width(s) (Side rail to side rail)
Cross Bar Width
A2079
19.5in to 24in (49.6 cm to 61 cm)
Approx. 28”
A2079E
22in to 28in (55.9 cm to 71.1 cm)
Approx. 32”
Metal free zone
0688771-1
MAYFIELD Imaging: Base Unit
•
Identification:
Base Unit
1. Skull Clamp/Headrest Locking Knob 2. Swivel Adaptor 3. Link Arm Locking Knobs 4. Yoke 5. Crossbar 6. Base Handle 7. Side Rail Bracket 8. Locking Lever 9. Base Handle Adjustment Knob 10.Auxiliary Side Rail 11.Side Rail Bracket Locking
0688771-1
MAYFIELD Imaging: Base Unit Accessories
•
Tri-Star Swivel Adapter, A2111
•
Extension Arms 437A2224, 5 inch
437A2225, 7 inch
0688771-1
MAYFIELD Imaging: Horseshoe Headrests Pediatric Horseshoe Headrest, A2011
Adult Horseshoe Headrest, A2010
Starburst to attach to standard swivel or tri-star swivel adapter
0688771-1
MAYFIELD Imaging: Horseshoe Headrests
Spine Table Adapter, A2600R
Using the Spine Table Adapter with the Jackson Table
Metal free zone
0688771-1
MAYFIELD Imaging Radiolucent Products (XR2) Instructions for Use Refresher
10 0688771-1
MAYFIELD Imaging Base Unit Review of Components
Rail end bracket
Std. Swivel Adaptor (Blue ring) Side Rail
Locking Handle with Locking Lever
Link Arm (Yellow ring)
Base Handle Adjustment Knob
Tri-Star Swivel Adaptor (Blue ring) (Available as an Accessory)
0688771-1
Refresher on Instructions for Use Step 1. With the rail end bracket toggle locks loosened and the locking handles unlocked, hold the base unit as shown for attachment to the OR Table; grasp the side rails on the end brackets with each hand.
Second Locking handles are unlocked & opening “up”
First Rail end bracket toggle locks are loosened to allow easy attachment to the table side rails
Safety Latch
0688771-1
Refresher on Instructions for Use Step 2. Slide the base unit brackets onto the side rails of the operating room table.
Note that locking handles are opening “up” and NOT DOWN
0688771-1
Refresher on Instructions for Use Step 3. Secure the brackets to the side rails.
Turning rail end bracket toggle locks Tighten securely to side rails Cautions: Keep finger clear of hinge points when closing the base unit locking levers. It is recommended that the levers be closed using the palm of the hand. Always be sure the locking mechanisms are secure after completing table adjustments. Verify that the locking lever is secure by confirming that the safety latch is engaged.
0688771-1
Refresher on Instructions for Use Step 4. Attaching the Link and Adaptor to the Base Unit Assembly.
Note the color-coded sleeves – Line them up, yellow-to-yellow & blue-to-blue
0688771-1
Refresher on Instructions for Use Step 5. Positioned to await attachment of the skull clamp.
Locking handles are unlocked Link and swivel adaptor are connected but not tightened for easy movement and attachment to the skull clamp Note: The head section of the OR table can be attached prior to the patient entering the OR.
0688771-1
Refresher on Instructions for Use Step 6. With the skull clamp attached to the patient and in position, the Adaptor is brought up for attachment to the skull clamp
Attach in this order 1. Attach swivel adaptor to skull clamp
Locking Handle
2. Lock both locking handles (Use an “Open-Palm” to close handles. Make certain that the locking levers are engaged)
Swivel/Clamp Torque Knob
3. Tighten Swivel Adaptor knob (Blue) 4. Tighten Link attachment (Yellow) 5. Double-check all connections for being tight and secure (double-check does not mean double-tighten)
Link/Swivel Torque Knob
Link torque knob 0688771-1
Refresher on Instructions for Use Step 7. With the system now locked and ready, confirm that all knobs and levers are locked and secure.
0688771-1
Refresher on Instructions for Use Step 8. Knobs should not be overtightened. Over-tightening the mechanisms’ adjustment screws may result in damage to the unit. Please remember that overextending or overloading the base unit may result in unintended movement, shortened product life and/or damage to the unit
0688771-1
Always Remember The Indications for Use, Warnings and Cautions