Integra LifeSciences Corporation
The LICOX System Product Presentation Sept 2007
Product Presentation
22 Pages
Preview
Page 1
® The LICOX
System
Brain Tissue Oxygen Monitoring System Product Presentation NS1581-09/07
What is the LICOX® system? • Measures interstitial brain tissue oxygenation (PbtO2) in mmHg and brain temperature (°C) • Probe inserted approximately 35mm below the dura into the white matter of the brain • PbtO2 used in conjunction with current ICP/CPP monitoring methods
11/2/2007
2
Who needs the LICOX® system? Patients at risk for developing cerebral hypoxia or ischemia Head trauma patients Aneurysm patients Subarachnoid hemorrhage patients Stroke patients © 2007 Dr. Marcos Tatagiba. Reprinted with permission.
11/2/2007
3
When is the LICOX® system placed? • Within the first 24-48 hours of injury – The sooner cerebral hypoxia is detected, the better secondary injury can be prevented
• Generally, when an ICP catheter is required, a LICOX® probe should be considered
11/2/2007
4
How is the LICOX® system placed? • Can be either bolted or tunneled • Can be placed independently or with an ICP catheter • Requires minimal additional effort • Does not need to be zeroed prior to placement, SMART card is included with each O2 probe with all calibration data • Monitor need not be present when placing in the OR 11/2/2007
5
Where is the LICOX® system placed? • Placement is up to clinical discretion • The idea is to prevent SECONDARY injury by ensuring living tissue is receiving adequate oxygen
11/2/2007
• LICOX® probe can be placed in either the injured side or non-injured side of brain • Should not be placed directly into a lesion
6
What are the probe options? • BOLTED Double Lumen – Camino® ICP channel and LICOX® PMO catheter channel (IP2P)
• TUNNELED Licox® PMO combined oxygen and temperature catheter (IT2)
• Probe kits come with required drill bits and other accessories
11/2/2007
7
How accurate are the probes? Oxygen Accuracy: PbtO2 0-20 mmHg accuracy is ± 2 mmHg PbtO2 21-50 mmHg accuracy is ± 10% PbtO2 51-150 mmHg accuracy is ± 13%
Temperature Accuracy: ± 0.2 °C
11/2/2007
8
What about the nursing staff? • Simple monitor set-up and use • The monitor displays a digital oxygen and temperature reading • Alarms are managed through a connection to the bedside monitor
11/2/2007
9
What is a “normal” reading? • Normal: 25-35 mmHg • Risk of death increases – < 15 mmHg for 30 minutes – < 10 mmHg for 10 minutes • PbtO2 < 5 mmHg – high mortality • PbtO2 < 2mmHg - neuronal death1 1- Bardt T, Unterberg A, et al. Monitoring of brain tissue PO2 in traumatic brain injury: effect of cerebral hypoxia on outcome. Acta Neurochirurgica.1998;71(Suppl):153-156.
11/2/2007
10
How is patient outcome affected? • Its been found: – Head injured patients who undergo aggressive therapy to maintain ICP/CPP at normal levels still experience periods of severe brain hypoxia1 – Interventions previously thought to improve tissue oxygenation may improve ICP and CPP but actually decrease PbtO22 1. Bardt T, Unterberg A, et al. Monitoring of brain tissue PO2 in traumatic brain injury: effect of cerebral hypoxia on outcome. Acta Neurochirurgica.1998;71(Suppl):153-156. 2. Zauner A, Doppenberg E, et al. Extended neuromonitoring: new therapeutic opportunities? Neurological Research. 1998;20(Suppl 1):85-90.
11/2/2007
11
How is patient outcome affected? • The PbtO2 number can provide: – Notification of hypoxic episodes – Independent predictors of unfavorable outcome and death1 • Treatments to maintain PbtO2 correspond to more favorable patient outcomes2 1. Zauner A, Doppenberg E, et al. Extended neuromonitoring: new therapeutic opportunities? Neurological Research. 1998;20(Suppl 1):85-90. 2. Valadka A, Gopinath S, et al. Relationship of brain tissue PO2 to outcome after severe head injury. Critical Care Medicine. 1998;26(9):1576-1581.
11/2/2007
12
Is the LICOX® system cost-efficient? • Added costs the Licox® system are justified by significantly improved outcomes
• Implementation of new protocols using PbtO2 – Improve patient care1 – Better utilize resources1 – Probable reduction of • Ventilator days1,2 • ICU days1,2 • Overall hospital days1,2
1. Spain D, McIlvoy L, Fix S, et al. Effect of clinical pathway for severe traumatic brain injury on resource utilization. The Journal of Trauma. 1998;45(1):101-105. 2. Simons R, Eliopoulos V, Laflamme D, Brown D. Impact on process of trauma care delivery 1 year after introduction of trauma program in a provincial trauma center. The Journal of Trauma. 1999;46(5):811-816. 11/2/2007
13
How is the LICOX® system different than a Jugular Venous Bulb? • Jugular bulb oximetry, SjvO2, measures oxygen saturation of venous blood – Measures global oxygen reduction – Cannot identify regional cerebral ischemia – may lead to secondary injury1,2 1. Clay H. Validity and reliability of the SjO2 catheter in neurologically impaired patients: A critical review of the literature. Journal of Neuroscience Nursing. 2000;32(4):194-203. 2. Mayberg T, Lam A. Jugular bulb oximetry for the monitoring of cerebral blood flow and metabolism. Neurosurgery Clinics of North America. 1996;7(4):755-765.
11/2/2007
14
How is LICOX® different than a Jugular Venous Bulb? • SjvO2 measurements are shown to be unreliable – Good quality data are only obtained about 50% of the placement time1,2,3 – SjvO2 is difficult to use in children due to small vein size4 1. Clay H. Validity and reliability of the SjO2 catheter in neurologically impaired patients: A critical review of the literature. Journal of Neuroscience Nursing. 2000;32(4):194-203. 2. Kiening K, et al. Monitoring of cerebral oxygenation in patients with severe head injuries: Brain tissue PO2 versus jugular vein oxygen saturation. Journal of Neurosurgery. 1996;85:751-757 3. Meixensberger J. et al. Multimodality hemodynamic neuromonitoring - Quality and consequences for therapy of severely head injurted patients. Acta Neurochirugica. 1998;71(Suppl):260-262. 4. Palmer S, et al. The impact on outcomes in a community hospital setting of using the AANS traumatic brain injury guidelines. The Journal of Trauma. 2001;50(4):657-664. 11/2/2007
15
A few institutions that have published clinical studies with the LICOX® systems: • Mission Hospital – Mission Viejo, CA
• Harborview – Seattle, WA
• Creighton University – Omaha, NE
• University of Pennsylvania Hospital – Philadelphia, PA
11/2/2007
16
Thank you for your time! Please contact us anytime should you have any further questions.
References Bardt T, Unterberg A, et al. Monitoring of brain tissue PO2 in traumatic brain injury: effect of cerebral hypoxia on outcome. Acta Neurochirurgica.1998;71(Suppl):153-156. Clay H. Validity and reliability of the SjO2 catheter in neurologically impaired patients: A critical review of the literature. Journal of Neuroscience Nursing. 2000;32(4):194-203. Mayberg T, Lam A. Jugular bulb oximetry for the monitoring of cerebral blood flow and metabolism. Neurosurgery Clinics of North America. 1996;7(4):755-765.
11/2/2007
18
References (cont.) Kiening K, et al. Monitoring of cerebral oxygenation in patients with severe head injuries: Brain tissue PO2 versus jugular vein oxygen saturation. Journal of Neurosurgery. 1996;85:751-757 Meixensberger J. et al. Multimodality hemodynamic neuromonitoringQuality and consequences for therapy of severely head injurted patients. Acta Neurochirugica. 1998;71(Suppl):260-262. Palmer S, et al. The impact on outcomes in a community hospital setting of using the AANS traumatic brain injury guidelines. The Journal of Trauma. 2001;50(4):657-664.
11/2/2007
19
References (cont.) Prasad S. et al. Cerebral oxygenation in major pediatric trauma: Its relevance to trauma severity scores and outcomes. Paper presented at: 35th Annual Meeting of the American Pediatric Surgical Associtation; May 27-30, 2004; Ponte Vedra Beach, Fla. Simons R, Eliopoulos V, Laflamme D, Brown D. Impact on process of trauma care delivery 1 year after introduction of trauma program in a provincial trauma center. The Journal of Trauma. 1999;46(5):811-816.