Bipolar Cord 30-1536
Integra LifeSciences Production Corporation 11 Cabot Boulevard Mansfield, MA 02048 USA
LCN 300-1536-85 Rev M 10/19 1393193-1 ©2019 Integra LifeSciences Corporation. All rights reserved.
ENGLISH IMPORTANT INFORMATION Please Read Before Use
Bipolar Cord 30-1536 Failure to follow these instructions will shorten the useful life of this device. This device is provided nonsterile and must be cleaned and sterilized before use, including initial use.
PRECAUTIONS
Leave cord loose. Do not secure cord. Do not rest anything on the cord in the pack. Visually inspect cord before each use. Replace cord if there is any sign of deterioration such as broken insulation or frayed wires. This bipolar cord is designed for use with bipolar electrosurgical units only. Do not use this bipolar cord at generator power settings that exceed peak-to-peak RF voltage of 400 volts. The maximum allowable power settings for the CMC®-III are as follows: • Coagulation: 150 Malis Units (50 watts) = 100 volts • Cutting: 80 Malis Units (60 watts) = 90 volts This cord is not recommended for use with the CODMAN® MALIS® CMC-V Bipolar Electrosurgical Generator. Codman recommends use of bipolar cord 80-1186 with this Generator. For other generators, refer to the instructions for use to determine the maximum allowable power settings.
Cleaning After each use, wipe the bipolar cord with a damp, soft cloth. 2