Integra LifeSciences Corporation
Non-sterile Integra Devices Inspection , Cleaning and Disinfection Instructions
2 Pages
Preview
Page 1
• The cases/trays used for cleaning must always be loaded correctly to ensure proper cleaning. • After cleaning, check instruments for cleanliness (visible dirt). This especially applies to cannulated instruments or those with blind holes and crevices.
Cleaning, Disinfection and Inspection of Non-sterile Integra® Devices
• To ensure proper instrument functioning, verify that all movable parts have been thoroughly cleaned. • Pay special attention to slots, ratchets, joints and box locks, narrow lumens, blind holes, and other areas that are hard to access. • Deionized or distilled water should be used for the final rinse.
Description and Intended Use Integra LifeSciences Corporation reusable surgical device sets consist of various configurations of orthopedic devices in cases or trays. The sets are constructed from durable metal and/or plastic materials. The reusable surgical devices are intended for use in orthopedic surgical procedures according to the Instructions for Use and Surgical techniques that accompany Integra® implants. Reusable devices are to be cleaned, inspected, disinfected and sterilized between uses. Inspection Before Use Reusable Devices can be used indefinitely if not damaged or worn. Device systems should be cleaned and then inspected between uses. DO NOT use broken or damaged devices. Contact Integra for repair or replacement of damaged items. If damage or malfunction is detected, the device should not be used. Disposable Devices should be disposed of according to hospital procedure and any applicable laws. Preparation/General Guidance for Cleaning and Disinfection Verify that all instruments required for use are present in the case. In addition, it is not recommended to use chloride containing cleaning solutions since its use has been linked to corrosion of metallic instruments, especially stainless steel. Please also note the following: • Disinfect and clean devices immediately after use in order to avoid device encrustations. • Solutions used for cleaning must always be prepared in accordance with the manufacturer’s instructions. • Never use metal brushes or metal sponges for manual cleaning. • Use a suitably sized non-metallic bristle brush for cleaning lumens, cannulations, blind holes, and cavities, making sure that every part of the inner surface can be properly accessed. • Clean jointed instruments in closed as well as open positions. • Disassemble instruments as far as possible before cleaning. • Be sure to arrange the items so that the water can easily flow out of cannulations, blind holes, and cavities. • For instruments with long or narrow lumens, standard processing should be used only if the disinfectant can flow easily through the lumens and safe rinsing is guaranteed.
Automated Cleaning and Disinfection Procedure The following steps should be completed in sequence. Please note that all instructions provided are as validated by Integra LifeSciences. The disinfection validation protocols were performed in accordance with ISO 15883-1: 2009 Washer-disinfectors, Part 1; General requirements, terms and definitions and tests. a) Prepare an enzymatic detergent using lukewarm deionized water as per the manufacturer’s recommendation. b) Fully immerse the devices and allow to soak for a minimum of two (2) minutes. c) Following the soak time, flush any lumens of the device using a syringe. d) Rinse the devices under lukewarm running deionized water for a minimum of one (1) minute, while agitating the devices. Agitation includes actuating all movable parts, such as opening and closing hinges and moving the devices around under the running water. e) Use a clean soft bristled brush and/or pipe cleaner to brush and aid in the rinse for the exterior and interior of device components. Use a syringe to flush any lumens. f) Place the devices back into the designated locations of the case/tray, and load the case/tray set into an automated washer disinfector (Steris 444 or equivalent). g) The washer disinfector cycle parameters are as follows:
Phase
Recirculation Time (Min.)
Water Temperature (Min.)
Detergent
Pre-Wash
02:00
Cold Water
NA
Enzyme Wash
01:00
Hot Water
Enzymatic Cleaner
Wash
02:00
60°C
Neutral Detergent
Thermal Rinse
10:00
82°C*
NA
* Note: The highest grade of water available should be used during the final rinse cycle, distilled or deionized is recommended.
h) After washing and disinfecting, dry the devices using a clean lint free cloth and visually examine to determine if all adherent visible soil has been removed. i) Repeat the cleaning and disinfection procedure if visible debris is detected. Inspection After Cleaning Following cleaning and disinfection, the instruments must be macroscopically clean, i.e. free from visible dirt or deposits. All movable parts, working tips and blades (scissors) should be inspected with particular care. In addition, the instruments must be sterilized prior to use as per the noted reprocessing instructions referenced below. Non Sterile Device Sets Validated for Use
Associated Reprocessing Instructions for Use (IFU)
Cadence® Total Ankle Trials, Insertion, and Removal Instruments Cadence® Total Ankle Resection Instruments Cadence® Flat Cut Instruments
LC-04-1020-0006
Cadence® Flat Cut Instruments, Alternate Ti6® Internal Fixation System Low Profile Screw Set Ti6® Internal Fixation System Solid Screw Tray1 Ti6® Internal Fixation System Base Instrument Set
LC-04-0000-0004
Ti6® Internal Fixation System Headless/Quicksnap Screws Ti6® Internal Fixation System Digital Fusion Screws1 Total Foot System 2 Titan™ Modular Shoulder System, 2.5 Glenoid Tray Titan™ Modular Shoulder System, 2.5 Humeral Tray I Titan™ Modular Shoulder System, 2.5 Humeral Tray II
LC-04-0000-0013
Titan™ Reverse Shoulder System Titan™ Reverse Shoulder System-S 1 Titan™ Modular Shoulder System, 2.5 Monoblock Sounder Instrumentation 1
This system does not have a registered CE mark. Manufacturer: Ascension Orthopedics, Inc. 11101 Metric Blvd Austin, TX 78758 USA Phone: 1 (800) 654-2873 Fax: 1 (888) 980-7742 integralife.com n
Authorized Representative: Integra LifeSciences Services (France) Immeuble Sequoia 2 97 allée Alexandre Borodine Parc Technologique de la Porte des Alpes 69800 Saint Priest–France Telephone: +33 (0) 4 37 47 59 50 Fax: +33 (0) 4 37 47 59 25
2797
Cadence, Ti6, Integra, and the Integra logo are registered trademarks of Integra LifeSciences Corporation or its subsidiaries in the United States and/or other countries. Titan is a trademark of Integra LifeSciences Corporation or its subsidiaries. ©2020 Integra LifeSciences Corporation. All rights reserved. LC-04-0000-0016 Rev E 02/20 0555798-6