TruDi™ Navigation System Information for Use
Software Version 1.4 UG-2000-001 (02A) November 2018
Please read this document carefully before using the TruDi™ Navigation System
© Acclarent, Inc. 2016 - 2018. All rights reserved.
CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. Third-party trademarks used herein are trademarks of their respective owners. The TruDi™ Navigation System is covered by one or more of the following US patents owned by Biosense Webster: 6,226,542; 6,484,118; 6,650,927; 6,690,963. For Acclarent patent information, see www.acclarent.com/patentmarking. Some pictures of the workstation are provided courtesy of Dell Computer Corporation, ©Dell Computer Corporation 2007.
P.N.: UG-2000-001 (02A) 15 November 2018
Table of Contents iii
Table of Contents About the TruDi ™ Navigation System ... 1 Typical Setup ... 2 Notes, Cautions, and Warnings ... 3 Safety Information ... 4 Indications for Use ... 5 Qualified Users and Training ... 5 Warnings and Precautions... 5 Clinical Safety... 5 Physical Setup ... 5 Preventing Fire and Electrical Shock ... 6 Interference with Other Equipment... 6 Location Accuracy ...7 Security ...7 Electromagnetic Compatibility (EMC) ... 8 Symbols... 13 External Product Labels ... 15 Standards Compliance and Classifications ... 16 Hardware ... 17 System Specifications... 18 Environmental Conditions ... 19 System Components ... 20 Console Connections ... 23 Instrument Hub Connections ... 24 Supported Acclarent Devices...25 Workstation Connections ...25 Chair Adapter ... 26 Periodic Maintenance ... 30 Cleaning the System Components ... 30 Overview of User Interface ... 32 Starting Up and Shutting Down the TruDi™ Navigation System ... 33 Opening Screen... 34 Main Screen ... 35 Image Brightness and Contrast Adjustment... 36 Main Toolbar ... 37 Status Indication ... 42 Adjusting the View ... 43 Advanced Navigation Options... 45 Registration Screen ... 47
UG-2000-001 (02A)
TruDi™ Navigation System IFU
iv
Table of Contents Performing a Procedure ... 50 Setup... 51 Loading a CT Image...52 Preview Mode ... 54 Marking TruDi™ Planning Points ... 55 Setting Up a TruDi™ TruView... 56 Measuring Distance ... 58 Deleting Navigation Elements ... 59 Performing Registration ...60 Starting a Procedure... 65 FAM - Fast Anatomical Mapping ...66 Starting FAM... 67 Troubleshooting ... 68 System Messages ...69 System Report ... 74 Index ... 75
TruDi™ Navigation System IFU
UG-2000-001 (02A)
About the TruDi™ Navigation System 1
Chapter 1
About the TruDi™ Navigation System In This Chapter: Typical Setup ... 2 Notes, Cautions, and Warnings ... 3
The TruDi™ Navigation System is an image-guided navigation system designed for use during intranasal and paranasal surgical procedures. When using the System, ENT physicians can track and display the real-time location of the tip of navigated instruments relative to pre-acquired CT images. The TruDi™ Navigation System is composed of several components. •
The TruDi™ Emitter Pad emits a known electromagnetic field throughout the procedure.
•
The Patient Tracker provides a location reference relative to the . The Patient Tracker is a patch with an embedded magnetic sensor, applied to the patient's forehead.
•
The Console drives the TruDi™ Emitter Pad signals and communication between the System components.
•
The Workstation, with the information from the Console and displays it on the monitor.
•
The Registration Probe is used to trace the pre-acquired image to the patient's head at the beginning of the procedure and mark corresponding points.
•
The Patient Tracker, Registration Probe, and Navigated surgical devices connect to the Instrument Hub.
installed, processes
A Technical Service representative is responsible for installing the TruDi™ Navigation System. NOTE Images in this document are for illustrative purposes only. See also: •
System Components
UG-2000-001 (02A)
TruDi™ Navigation System IFU
2
About the TruDi™ Navigation System
Typical Setup The TruDi™ Navigation System is designed for use in either a hospital Operating room or in an ENT Treatment Room setting. In the chair configuration, the TruDi™ Emitter Pad is attached to its dedicated Chair Adapter and is placed on an existing patient chair. In the patient bed configuration, the TruDi™ Emitter Pad is placed on the same surface as the patient's head and the Chair Adapter is not required. CAUTION The Chair Adapter is intended to support patients whose weight is less than 440 pounds (200 kg). Because the patient's head is positioned in the middle of the TruDi™ Emitter Pad during a procedure, ensure that the TruDi™ Emitter Pad is fixed securely in place before beginning the procedure. When not in use, all System components can be stored on the cart. NOTE Always ensure that all relevant System components are connected only to the Isolation Transformer that is provided with the System.
Figure 1: TruDi™ Navigation System TruDi™ Navigation System IFU
UG-2000-001 (02A)
About the TruDi™ Navigation System 3
Notes, Cautions, and Warnings This document uses various types of notes. An explanation of each type, in the appropriate format, is given below. NOTE A note provides important information, emphasizing or supplementing the main text. The information does not relate directly to issues that might cause injury to patients or users, or damage to the System.
CAUTION A caution provides information relating to issues that might cause injury to patients or users, or damage to the System.
WARNING A warning provides information relating to issues that might cause serious, lifethreatening injury to patients or users, or serious, irreparable damage to the System.
UG-2000-001 (02A)
TruDi™ Navigation System IFU
4
Safety Information
Chapter 2
Safety Information In This Chapter: Indications for Use ... 5 Qualified Users and Training ... 5 Warnings and Precautions... 5 Clinical Safety... 5 Physical Setup ... 5 Preventing Fire and Electrical Shock ... 6 Interference with Other Equipment... 6 Location Accuracy ...7 Security ...7 Electromagnetic Compatibility (EMC) ... 8 Symbols... 13 External Product Labels ... 15 Standards Compliance and Classifications ... 16
This chapter provides important information regarding the safe functioning of the TruDi™ Navigation System. Follow the instructions in this manual carefully. In addition, consider current medical practices regarding patient care and safety. WARNINGS •
Failure to observe the guidelines and instructions provided in this document could result in patient injury.
•
Failure to observe the guidelines and instructions provided in this document could result in faulty functioning of the TruDi™ Navigation System.
TruDi™ Navigation System IFU
UG-2000-001 (02A)
Safety Information 5
Indications for Use The TruDi™ Navigation System is intended for use during intranasal and paranasal image-guided navigation procedures for patients who are eligible for sinus procedures. See also: •
Location Accuracy
•
Interference with Other Equipment
Qualified Users and Training Training is available for physicians and medical staff designated to be TruDi™ Navigation System users. Please contact your Acclarent representative to arrange training.
Warnings and Precautions The order of the following warning statements does not indicate their relative importance.
Clinical Safety •
Distortion to the System's magnetic field may degrade the accuracy of the Navigated surgical device indicator position in relation to the registered CT image.
•
Never advance or retract the ACCLARENT® NAVWIRE™ Sinus Navigation Guidewire, balloon catheter, or other Navigated surgical devices against unknown resistance. This can cause tissue trauma or device damage.
•
Read and follow all device Instructions for Use for navigation devices which are used with the System. Use only Acclarent-approved devices with the TruDi™ Navigation System.
Physical Setup CAUTION The TruDi™ Navigation System is ready for first clinical use only after it has been installed by an Acclarent representative and fully tested and approved as part of the initial installation process. If the procedure chair is changed or moved to a different procedure environment, the device operation should be verified by Acclarent prior to use. •
Do not cover the vent openings of the Console or the Workstation while the System is functioning. Doing so may stop the study and damage the equipment.
•
Do not modify the components of the TruDi™ Navigation System in any way.
•
Use only parts and accessories manufactured or recommended by Biosense Webster. Parts and accessories used must meet the requirements for the applicable IEC 60601 series safety standards. The TruDi™ Navigation System does not interface with any external medical device. UG-2000-001 (02A)
TruDi™ Navigation System IFU
6
Safety Information •
Connect the System components only to the power supply that is provided with the TruDi™ Navigation System. Do not use a different power supply.
•
It is recommended that the System components be powered from an uninterruptible power supply (UPS). This will enable continuous operation during power mains interruptions.
•
Take precautions against electrostatic discharge when working with the System.
•
Before turning on the equipment, ensure that all System cables are securely connected.
Preventing Fire and Electrical Shock •
Ensure that all TruDi™ Navigation System components are connected to the common protective ground in the procedure room, or to a separation device. Do not connect System electrical components to a multiple socket power outlet. Using a multiple socket outlet might compromise safety.
•
Do not connect or disconnect power cables and TruDi™ Emitter Pad cables while the System is turned on. Doing so could result in minor shock to the user and/or damage to the System.
•
Do not connect or disconnect the System power supply while it is turned on. Doing so could result in damage to the System.
•
To reduce the risk of fire, do not use the TruDi™ Navigation System in the presence of flammable anesthetic mixtures or other flammable gases.
•
For continued protection against the risk of fire, only use replacement fuses that have the same type and rating as the original fuses (115 V~, T6.3 / 230 V~, T3.15).
Interference with Other Equipment •
System function might interfere with minute ventilation rate-adaptive pacemakers. When performing a procedure on a patient who has such a device, consider changing the operating mode of the pacemaker to a minute ventilation insensitive mode before performing the procedure. If necessary, consult the pacemaker manufacturer's instructions.
•
Metallic and conductive objects in or near the electromagnetic navigation field can degrade navigational accuracy and the TruDi™ Zone. If metallic distortion causes excessive error, navigation may be disabled. To restore navigation, move metallic and conductive objects away from the TruDi™ Zone. Electrical noise in or near the TruDi™ Zone can degrade navigational accuracy. If electrical noise introduces excessive error, the system may automatically disable navigation. To restore navigation, move devices that produce electrical noise (such as cellular phones) away from the TruDi™ Zone.
•
TruDi™ Navigation System IFU
UG-2000-001 (02A)
Safety Information 7
Location Accuracy The navigation capability of the TruDi™ Navigation System is based on the interactions between a known electromagnetic field (emitted by the TruDi™ Emitter Pad) and magnetic sensors (embedded in navigated devices). System accuracy is better than 2 mm. The TruDi™ Emitter Pad's working volume is called the TruDi™ Zone. The TruDi™ Zone is 7.9 in x 7.9 in x 6.3 in (20 cm x 20 cm x 16 cm), with an additional reduced accuracy zone of 0.8 in (2 cm) in each direction. CAUTION The TruDi™ Navigation System can display only one device at a time. Ensure that the TruDi™ Emitter Pad is fixed securely in place before beginning registration. Do not allow the TruDi™ Emitter Pad to move during the procedure. Movement of the TruDi™ Emitter Pad after the beginning of the registration process might affect location accuracy. Ensure that device shafts and connectors do not come close to one of the transmitters on the TruDi™ Emitter Pad. If the shaft or connector is in proximity to a transmitter, location accuracy might be affected. Metal interference caused by a ferromagnetic material placed within the TruDi™ Zone might affect location accuracy, which can cause difficulty in tracking navigated devices. Keep ferromagnetic material outside the TruDi™ Zone. If metal interference is detected: •
The status indicator for the affected device displays the error status. (See Status Indication.)
•
When there is a low level of interference (outside the TruDi™ Zone), the crosshairs are hidden and an orange indicator is displayed. When there is metal interference inside the TruDi™ Zone, the affected device is not displayed in the 3D rendered view or in the Multi-Planar Reconstruction (MPR) view.
If accuracy seems to be affected, check for ferromagnetic objects in the TruDi™ Zone. If removing objects from the TruDi™ Zone does not help, exit the study. Continue to use the endoscope without navigation support and gently remove the navigation device.
Security The Workstation (computer) supplied with the TruDi™ Navigation System is a dedicated computer for the System. Do not install unauthorized software on the System. To protect against computer viruses, do not use the Workstation for any other purpose. See also: •
TruDi™ Navigation System Security User Guide, UG-2000-091
UG-2000-001 (02A)
TruDi™ Navigation System IFU
8
Safety Information
Electromagnetic Compatibility (EMC) The following tables present data regarding TruDi™ compliance with IEC 60601-1-2, 4th edition. Table 1: Guidance and Manufacturer’s Declaration Electromagnetic Emissions Medical Electrical Equipment and Systems
For All
The TruDi™ Navigation System is intended for use in the electromagnetic environment specified below. The customer or user of the TruDi™ Navigation System should assure that it is used in such an environment. Emissions Test
Compliance
Electromagnetic Environment Guidance
Group 2
The TruDi™ Navigation System uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class A
Harmonic emissions IEC 61000-3-2
Class A
The TruDi™ Navigation System is suitable for use in all establishments other than domestic.
Voltage fluctuations or flicker emissions IEC 61000-3-3
Compliance
RF emissions CISPR 11
WARNING The TruDi™ Navigation System (including its peripheral components) is intended for use by healthcare professionals only. This System (including its peripheral components) might cause radio interference or might disrupt the operation of nearby equipment.
TruDi™ Navigation System IFU
UG-2000-001 (02A)
Safety Information 9 Table 2: Guidance and Manufacturer’s Declaration Electromagnetic Immunity (Specifications for all Medical Electrical Equipment and Systems) The TruDi™ Navigation System is intended for use in the electromagnetic environment specified below. The customer or user of the TruDi™ Navigation System should assure that it is used in such an environment. Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment Guidance
Electrostatic discharge (ESD) IEC 61000-4-2
±8 kV contact ±15 kV air
±8 kV contact ±15 kV air
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.
Electrical fast transient or burst IEC 61000-4-4
±2 kV for power supply lines
±2 kV for power supply lines
±1 kV for input/output lines
N/A
Mains power quality should be that of a typical commercial or hospital environment.
Surge IEC 61000-4-5
±1 kV line(s) to line(s)
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±2 kV line(s) to earth
0% UT; 0.5 cycle
0% UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
At 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315°
0% UT; 1 cycle
0% UT; 1 cycle
and
and
70% UT; 25/30 cycles
70% UT; 25 cycles
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Single phase: at 0°
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
Single phase: at 0°
0% UT; 250 /300 cycle
0% UT; 250 cycle
30 A/m
30 A/m @ 50 Hz
Mains power quality should be that of a typical commercial or hospital environment. Mains power quality should be that of a typical commercial or hospital environment. If the user of the TruDi™ Navigation System requires continued operation during power mains interruptions, it is recommended that the TruDi™ Navigation System be powered from an uninterruptible power supply (UPS).
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment
NOTE UT is the AC mains voltage prior to application of the test level.
UG-2000-001 (02A)
TruDi™ Navigation System IFU
10
Safety Information Electromagnetic Immunity (Specifications for Medical Electrical Equipment and Systems That Are Not Life Supporting) The TruDi™ Navigation System is intended for use in the electromagnetic environment specified below. The customer or user of the TruDi™ Navigation System should assure that it is used in such an environment. Immunity Test Conducted RF
IEC 60601 Test Level
Compliance Level
3 Vrms
3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz
150 kHz to 80 MHz (V1)
Radiated RF
3 V/m
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
80 MHz to 2.5 GHz (E1)
Electromagnetic Environment Guidance Portable and mobile RF communication equipment should be used no closer to any part of the TruDi™ Navigation System, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance:
80 MHz to 800 MHz 800 MHz to 2.5 GHz Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strength from fixed RF transmitters as determined by an electromagnetic site survey,1 should be less than the compliance level in each frequency range.2 Interference may occur in the vicinity of equipment marked with the following symbol:
1
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the System is used exceeds the applicable RF compliance level, the System should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the System. 2
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.
TruDi™ Navigation System IFU
UG-2000-001 (02A)
Safety Information 11 The TruDi™ Navigation System is intended for use in the electromagnetic environment specified below. The customer or user of the TruDi™ Navigation System should assure that it is used in such an environment. Immunity Test
IEC 60601 Test Level
Immunity from Proximity field from wireless communicatio ns equipment
IEC 610004-3
Compliance Level
Electromagnetic Environment Guidance
List of frequencies, from 9 V/m up to 28 V/m, PM (18 Hz or 217 Hz), FM 1 kHz
NOTES •
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
•
These guidelines might not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
UG-2000-001 (02A)
TruDi™ Navigation System IFU
12
Safety Information Table 4: Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and TruDi™ (Specifications for Medical Electrical Equipment and Systems that are Not Life-Supporting) Recommended separation distances between portable and mobile RF communications equipment and the TruDi™ Navigation System The TruDi™ Navigation System is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The TruDi™ Navigation System customer or user can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the TruDi™ Navigation System as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter [W]
Separation distance according to frequency of transmitter [m] 150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2
d = 1.2
d = 2.33
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTES •
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
•
These guidelines might not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
TruDi™ Navigation System IFU
UG-2000-001 (02A)
Safety Information 13
Symbols The following table displays the symbols used on the TruDi™ Navigation System hardware and in the accompanying documentation (including this manual). Table 5: Legend to Symbols Appearing on the System Hardware and Documentation Symbol
Description Accompanying Documents The device is shipped with accompanying documents. Please read all documents before operating the System. Instructions for Use Refer to accompanying Instructions for Use. Instructions for Use Consult accompanying Instructions for Use. (Packaging label.)
Equipotential Ground Pin The device comes equipped with an equipotential earth pin. This pin should be connected as per standard hospital procedure. Date of Manufacture Hardware supplied with the device is labeled with the date of manufacture, as indicated by this symbol. Manufactured By The manufacturer for the device is marked by this symbol. Distributed By The Distributer for the device is marked by this symbol. Canadian Standards Association Certification Mark The device has been CSA certified to comply with the Safety Standard for Medical Devices IEC 60601-1. Customer Support/Call Center The contact information for the authorized Service and Support Department representatives is marked with this symbol.
UG-2000-001 (02A)
TruDi™ Navigation System IFU
14
Safety Information Symbol
Description Separate Collection for Electrical and Electronic Equipment All electrical and electronic equipment provided with Systems released after 13 August 2005 is marked with a Separate Collection for Electrical and Electronic Equipment symbol, indicating that this equipment must undergo separate collection for disposal, in countries where EU directive 2002/96/EC is in effect. Please contact Customer Support regarding disposal of all labeled equipment with this label. Heavy Object To prevent injury, precautions should be taken when lifting or moving objects marked with this symbol. Non-ionizing Electromagnetic Radiation Fragile: Handle with Care This medical device can be broken or damaged if not handled carefully. Keep Dry This medical device must be protected from moisture. Humidity Limitation This label indicates the range of humidity to which the medical device can be safely exposed. Atmospheric Pressure Limitation This label indicates the range of atmospheric pressure to which the medical device can be safely exposed. Type BF applied part y.
Alternating Current
TruDi™ Navigation System IFU
UG-2000-001 (02A)
Safety Information 15
External Product Labels
Figure 2: System Label
Figure 3: Console to Power Supply Label
UG-2000-001 (02A)
TruDi™ Navigation System IFU
16
Safety Information
Standards Compliance and Classifications The optical communication of the Workstation is based on a Class I LED component. The TruDi™ Navigation System complies with the current edition of the following standards and regulations. Table 6: TruDi™ System Compliance Source
Standard Number
CAN/CSA-C
22.2 No.60601-1
MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE
EN
1041
INFORMATION SUPPLIED BY THE MANUFACTURER OF MEDICAL DEVICES
EN
ISO15223-1
MEDICAL DEVICES - SYMBOLS TO BE USED WITH MEDICAL DEVICE LABELS, LABELLING AND INFORMATION TO BE SUPPLIED - PART 1: GENERAL REQUIREMENTS
IEC
60601-1
MEDICAL ELECTRICAL EQUIPMENT - PART 1: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE
IEC
60601-1-2
MEDICAL ELECTRICAL EQUIPMENT - PART 1-2: GENERAL REQUIREMENTS FOR BASIC SAFETY AND ESSENTIAL PERFORMANCE - COLLATERAL STANDARD: ELECTROMAGNETIC COMPATIBILITY - REQUIREMENTS AND TESTS
Title
The Biosense Webster Quality System complies with the current edition of the following standards and regulations. Table 7: Quality System Standards Source
Standard Number
ISO
13485
MEDICAL DEVICES - QUALITY MANAGEMENT SYSTEMS REQUIREMENTS FOR REGULATORY PURPOSES
FDA
21 CFR Part 820
QUALITY SYSTEM REGULATION
ISO
14971
MEDICAL DEVICES - APPLICATION OF RISK MANAGEMENT TO MEDICAL DEVICES
IEC
62304
MEDICAL DEVICE SOFTWARE - SOFTWARE LIFE CYCLE PROCESSES
ISO
14001
ENVIRONMENTAL MANAGEMENT SYSTEMS REQUIREMENTS WITH GUIDANCE FOR USE
TruDi™ Navigation System IFU
Title
UG-2000-001 (02A)