Articulated Stand, L - Shaped
IFU Articulated stand with quick-release coupling, base and clampingjaws Instructions For Use
20 Pages
Preview
Page 1
en
Instructions for use Articulated stand with quick-release coupling, base and clamping jaws
03-2021 Copyright © All product illustrations, product descriptions, and texts are the intellectual property of KARL STORZ SE & Co. KG. Their use and reproduction by third parties require the express approval of KARL STORZ SE & Co. KG. All rights reserved.
Table of contents
Table of contents 1 General information... 4 1.1 Read the instructions for use ... 4 1.2 Read the instructions for use of combinable products... 4 1.3 Scope... 4 1.4 Description of warning messages... 4 2 Normal use... 5 2.1 Intended use ... 5 2.2 Indications... 5 2.3 Contraindications... 5 2.4 Target user populations ... 6 2.5 Patient population... 6 3 Safety ... 7 3.1 Serious incidents ... 7 3.2 Unsterile product ... 7 3.3 Correct handling ... 7 3.4 Damaged products ... 7 4 Product description ... 8 4.1 Product overview ... 8 4.2 Possible combinations... 9 4.3 Symbols on the packaging ... 10 4.4 Ambient conditions ... 10 5 Preparation... 11 5.1 Unpacking the product ... 11 5.2 Inspecting the product... 11 5.3 Assembling the product... 11 5.4 Inserting the instrument or telescope into the clamping jaw ... 13 5.5 Assembling product with KSLOCK... 14 6 Disassembly... 15 6.1 Disassembling the product ... 15 7 Maintenance, servicing, repairs, and disposal... 16 7.1 Repairing the product ... 16 7.2 Disposing of the product ... 16 8 Accessories and spare parts ... 17 8.1 Recommended accessories ... 17 8.2 Spare parts list ... 17 9 Subsidiaries... 18
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
3
General information
1 General information 1.1 Read the instructions for use If the instructions for use are not followed, patients, users, and third parties may be injured or the product may be damaged. Read the instructions for use carefully and follow all the safety notes and warnings. Keep the instructions for use clearly visible next to the product.
1.2 Read the instructions for use of combinable products If the instructions for use of combinable products are not followed, patients, users, and third parties may be injured or the product may be damaged. Read the instructions for use of the combinable products carefully and follow all the safety notes and warnings.
1.3 Scope These instructions for use are valid for the following products: Mechanical holding system with KSLOCK, consisting of base, articulated stand and clamping jaw Base
Articulated stand
Clamping jaw
Accessories
28172HK
28272HA
28272UGN
28272CN
28172HR
28272HB
28272UKN
28172HM
28272HC
28272UGK
28272HD
28272UKK 28272UL 28272UFN
1.4 Description of warning messages To prevent any injury to persons or damage to property, the warnings and safety notes in the instructions for use must be observed. The warning messages describe the following levels of danger. WARNING WARNING Designates a possible imminent risk. If this is not avoided, it could lead to death or serious injuries. CAUTION CAUTION Designates a possible imminent risk. If this is not avoided, it could lead to minor injuries. ATTENTION ATTENTION Designates a possibly harmful situation. If this is not avoided, the products could be damaged.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
4
Normal use
2 Normal use 2.1 Intended use Intended use of holding system Depending on the technical design, the holding system is used for the short-term flexible fastening of KARL STORZ exoscopes, endoscopes and instruments during open, microsurgical or minimally invasive procedures. The holding system does not come into contact with the patient and is therefore non-invasive. Intended use of base The base is used for fastening holding systems to operating tables. Intended use of clamping jaw Depending on the technical design, the clamping jaw is used for the short-term fastening of rigid endoscopes, flexible endoscopes, camera heads, or instruments to holding systems during minimally invasive or open procedures. The clamping jaw does not come into contact with the patient and is therefore non-invasive. The clamping jaw is a supplementary part of a holding system.
2.2 Indications –
The holding system is indicated for holding retractor blades, hooks, nasal specula, uterus manipulators or similar instruments to retract tissue in order to create working space for the surgeon during open, microsurgical and endoscopic procedures.
–
The holding system is indicated for holding sheaths or working-sheaths (e.g. Lotta System) during endoscopic interventions to ensure a stable work environment.
–
The holding system is indicated for fixing laryngoscopes during surgery to ensure a stable work environment.
–
The holding system is indicated for holding exoscopes or overview cameras during microsurgical or open procedures to ensure a stable camera image and view to the site.
–
The holding system is indicated for holding mediastinoscopes to ensure a stable work environment.
–
The holding system is indicated for holding medical equipment (e.g. input devices, manmachine interfaces, electromagnetic generators for navigation systems) in place.
–
The holding system is indicated for holding patient or surgeon suspension accessory (e.g. armrests).
–
The holding system is indicated for holding working trocars or sheaths (e.g. EasyGo, TEO) during minimally invasive or microsurgical interventions to ensure a stable work environment.
–
The holding system is indicated for holding endoscopes during minimally invasive surgery or endoscopic inspection/diagnosis to ensure a stable work environment and to relieve the camera assistant.
2.3 Contraindications No contraindications relating directly to the product are currently known. Use of the instrument is contraindicated if, in the opinion of the responsible physician, the patient's health would be endangered through its use.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
5
Normal use
The holding systems are not used in direct contact with the patient, ensuring instead the positioning stability of the corresponding applied parts. Use with the applied parts is contraindicated if use of these applied parts is contraindicated. 28272UFN only The clamping jaw is not suitable for use with endoscopic camera heads or VITOM® systems. The clamping jaw is also not suitable for use if, in the opinion of the responsible physician, the patient’s health would be endangered through its use.
2.4 Target user populations The medical device may only be used by doctors and medical assistants with a relevant specialist qualification.
2.5 Patient population There are no restrictions in terms of patient groups for this product.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
6
Safety
3 Safety 3.1 Serious incidents According to the Medical Device Regulation (MDR), a “serious incident” includes incidents that directly or indirectly had, could have had, or could have any of the following consequences (MDR, Art. 2, No. 65 [1]): –
Death of a patient, user, or another person
–
Temporary or permanent serious deterioration in the medical condition of a patient, user, or another person
–
A serious threat to public health The manufacturer and appropriate authority must be notified of all serious incidents.
3.2 Unsterile product The product is not sterile when delivered. The use of non-sterile products poses a risk of infection for patients, users, and third parties. Reprocess the product in line with the reprocessing instructions before initial use and every subsequent use.
3.3 Correct handling If the product is not handled correctly, patients, users, and third parties may be injured. Only persons with the necessary medical qualification and who are acquainted with the application of the product may work with it. Check that the product is suitable for the procedure prior to use. Check the product for the following points before and after every use: –
Completeness
–
Good working order
–
Rough surfaces left inadvertently
–
Sharp corners
–
Burred edges
–
Correct assembly of the components
–
Functionality Do not leave broken-off components inside the patient. Do not overload the product with mechanical stress. Do not bend bent products back to their original position.
3.4 Damaged products Damaged products can result in injury to patients, users, or third parties. Before each use, check all components of the product for damage. Do not use damaged products.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
7
Product description
4 Product description 4.1 Product overview Example illustrations of base - articulated stand - clamping jaw
6
4
5
3 9
1
7
8
2
1
Rail clamp
2
Stand clamp
3
Stop screw
4
Quick release coupling with KSLOCK interface, female
5
Thumb screw for quick release coupling
6
Central clamp
7
Quick release coupling with KSLOCK interface, male
8
Instrument clamp
9
Fixation screw for instrument clamp
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
8
Product description
4.2 Possible combinations Possible combinations between the individual components of the articulated stand Base
Articulated stand
Clamping jaw
Accessories
28172HR
28272HA
28272UGN
28272CN*
28172HK
28272HB
28272UKN
28172HM
28272HC
28272UGK
28272HD
28272UKK 28272UL 28272UFN
* The clamping cylinder (accessory) 28272CN can be combined only with clamping jaws 28272UGN and 28272UGK. Possible combinations between articulated stand and operating/examination tables Bases 28172HR and 28172HK of the articulated stand allow operating and examination tables to be fastened to rails according to European and US standards. Possible combinations between articulated stand and telescopes/instruments KARL STORZ instruments with KSLOCK interface (male) can be adapted to the KSLOCK interface (female) of the articulated stand directly. KARL STORZ telescopes and instruments without KSLOCK interface can be adapted to the articulated stand using a clamping jaw. The size of the instrument/telescope is limited by the clamping range of the clamping jaw. Clamping jaw
Clamping range
Suitable for
28272UGN
16.5 - 23 mm
Telescopes with square head
4.8 - 12.5 mm
Telescope/instrument sheaths
0 - 18 mm
Universal
28272UGK, with ball joint 28272UKN 28272UKK, with ball joint 28272UL 28272UFN
KARL STORZ videoscope and fiberscopes
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
9
Product description
4.3 Symbols on the packaging Symbol
Meaning Manufacturer
Date of manufacture
Article no.
Number of products in the product packaging
Batch code
Serial number
Medical device
Unique Device Identifier
Unsterile
Consult instructions for use
In accordance with US federal law (21 CFR 801.109), this product may only be sold to or on prescription from a licensed physician. CE conformity mark With this mark, the manufacturer declares the compliance of the products with the applicable regulation (EU) 2017/745. A code number after the CE mark indicates the responsible notified body.
4.4 Ambient conditions There are no special transport and storage conditions for this product.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
10
Preparation
5 Preparation 5.1 Unpacking the product 1.
Carefully remove the product and accessories from the packaging.
2.
Check the delivery for missing items and evidence of shipping damage.
3.
In the case of damage, hidden defects, and short deliveries, document their nature and extent and contact the manufacturer or supplier immediately.
5.2 Inspecting the product 1.
Inspect products for visible contamination. Do not use contaminated products.
2.
Inspect the products for the following characteristics:
–
Good working order
–
Functionality
–
Correct assembly of the components
–
Completeness Inspect the product for damage, e.g.:
–
Rough surfaces
–
Sharp corners
–
Burred edges
–
Protruding parts
5.3 Assembling the product Inserting the articulated stand into the base 1.
Place the base on the rail of the operating table and fix using the rail clamp.
2.
Remove the stop screw from the articulated stand. Insert the articulated stand into the opening in the base.
3.
Screw the stop screw into the articulated stand and tighten.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
11
Preparation
4.
Set the articulated stand to the desired height and, if necessary, direction (for rotation socket 28172 HR) and fix in place using the stand clamp.
Inserting the clamping jaw into the quick release coupling 5.
Select a clamping jaw suited to the instrument.
6.
Fit the clamping jaw on the quick-release coupling. The clamping jaw must engage (audible click).
7.
Secure by turning the thumb screw in the "CLOSE" direction as far as it will go.
When assembled correctly, the pin of the quick-release coupling is no longer visible.
If necessary, the height of the articulated stand can be extended using the column extension 28172HM.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
12
Preparation
5.4 Inserting the instrument or telescope into the clamping jaw 1.
Open the instrument clamp of the clamping jaw.
2.
Insert the instrument into the clamping jaw.
3.
Secure the instrument in position using the clamping jaw fixation screw. Ensure a secure hold.
Inserting a telescope into the clamping jaw 28272UGN When using the clamping jaw 28272UGN, it is absolutely essential to ensure that the telescope is correctly clamped.
With large telescope heads (10 mm telescopes), the clamping jaw must be mounted flush with the upper edge of the telescope head
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
13
Preparation
When using rigid 10 mm telescopes, the clamping cylinder 28272CN can also be mounted on the sheath of the telescope. The telescope and clamping cylinder are fixed using the clamping jaws 28272UGK or 28272UGN.
5.5 Assembling product with KSLOCK 1.
Attach the product with KSLOCK interface (male) directly to the quick release coupling of the articulated stand. The coupling must engage (audible click).
2.
Secure by turning the thumb screw in the "CLOSE" direction as far as it will go. When assembled correctly, the pin of the quick-release coupling is no longer visible.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
14
Disassembly
6 Disassembly 6.1 Disassembling the product 1.
Products adapted via a clamping jaw: Release instrument clamp and pull the product out of the clamping jaw.
2.
Products with KSLOCK coupling: Undo the thumb screw of the quick release coupling by turning in the "open" direction and pull the clamping jaw from the KSLOCK quick release coupling.
3.
Remove the stop screw from the articulated stand.
4.
Release the stand clamp and remove the articulated stand from the base.
5.
Release the rail clamp and remove the base from the rail of the operating table.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
15
Maintenance, servicing, repairs, and disposal
7 Maintenance, servicing, repairs, and disposal 7.1 Repairing the product Repair work may only be performed by KARL STORZ or by a company authorized by KARL STORZ. The interventions described in this instruction manual are exempt from this rule. Please contact your local KARL STORZ subsidiary or authorized dealer (see the list of subsidiaries). Contaminated devices may not be shipped. To prevent contact infections and airborne infections, products must first be decontaminated. KARL STORZ reserves the right to send back contaminated products.
7.2 Disposing of the product For disposal, the following measures are necessary: 1.
Decontaminate the products prior to disposal.
2.
Country-specific national laws and regulations must be observed.
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
16
Accessories and spare parts
8 Accessories and spare parts 8.1 Recommended accessories Item no.
Designation
28272CN
Clamping cylinder, folding, for flexible mounting of 10 mm telescopes to telescope sheath, autoclavable. The clamping cylinder allows vertical movement and rotation of the telescope.
28172HM
Column extension, 50 cm, with lateral clamp for height adjustment of the articulated stand, for use with articulated stands 28272 HA/HB/HC and mounting bases 28172HK/HR
041150-20
Drape, with elastic tip, package of 20
8.2 Spare parts list Item no.
Designation
ET35-91-090
Thumb screw, for fixing the retaining rod, for base 28172HK
28172HRS
Thumb screw, for fastening the rotation socket 28172HR to the operating table
28172HZ
Stop screw, for articulated stands 28272HA/HB/HC/HD
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
17
Subsidiaries
9 Subsidiaries KARL STORZ SE & Co. KG Dr.-Karl-Storz-Straße 34, 78532 Tuttlingen/Germany Postfach 230, 78503 Tuttlingen/Germany Phone: +49 7461 708-0 , Fax: +49 7461 708-105 Email: [email protected]
KARL STORZ SE & Co. KG Representation Office Kęstučio st. 59 / Lenktoji st. 27, 08124 Vilnius, Lithuania Phone: +370 5 272 0448, Mobile: +370 685 67 000 Email: [email protected]
KARL STORZ Endoskope Berlin GmbH Scharnhorststr. 3, 10115 Berlin/Germany Phone: +49 30 3069090, Fax: +49 30 30 19452
KARL STORZ Endoskopi Danmark A/S Skovlytoften 33, 2840 Holte, Denmark Phone: +45 45162600, Fax: +45 45162609 Email: [email protected]
KARL STORZ Endoscopy Canada Ltd. 7171 Millcreek Drive, Mississauga, Ontario L5N 3R3 Canada Phone: +1 905 816-4500, Fax: +1 905 816-4599 Toll free (Canada only) Phone: 1-800-268-4880, Fax: 1-800-482-4198 (Canada only) Email: [email protected] KARL STORZ Endoscopy-America, Inc. 2151 East Grand Avenue, El Segundo, CA 90245-5017, USA Phone: +1 424 218-8100, Fax: +1 424 218-8525 Toll free (USA only) Phone: 800 421-0837, Fax: 800 321-1304 (USA only) Email: [email protected] KARL STORZ Veterinary Endoscopy-America, Inc. 1 South Los Carneros Road, Goleta, CA 93117, USA Phone: +1 805 968-7776, Fax: +1 805 685-2588 Email: [email protected] KARL STORZ Endoscopia Latino-America, Inc. 815 N. W. 57th Avenue, Suite 480, Miami, FL 33126-2042, USA Phone: +1 305 262-8980, Fax: +1 305 262-8986 Email: [email protected] KARL STORZ Endoscopia México S.A. de C.V. Edificio Atlantic, Oficina 3G, Calle D e/ 1ra y 3ra, 10400 Vedado, Havanna, Cuba Phone: +537 836 95 06, Fax: +537 836 97 76 Email: [email protected] KARL STORZ Endoscopia México S.A. de C.V. Av. Ejercito Nacional No. 453 Piso 2, Colonia Granada, Alcaldia Miguel Hidalgo, C.P. 11520 Ciudad de México Phone: +52 (55) 1101 1520 Email: [email protected] KARL STORZ Marketing América Do Sul Ltda. Rua Joaquim Floriano, nº. 413, 20º andar – Itaim Bibi, CEP-04534-011 São Paulo, Brasil Phone: +55 11 3526-4600, Fax: +55 11 3526-4680 Email: [email protected] KARL STORZ Endoscopia Argentina S.A. Zufriategui 627 6° Piso, B1638 CAA - Vicente Lopez, Provincia de Buenos Aires, Argentina Phone: +54 11 4718 0919, Fax: +54 11 4718 2773 Email: [email protected] KARL STORZ Endoskopi Norge AS Stamveien1, 1483 Hagan, Norway Phone: +47 6380 5600, Fax: +47 6380 5601 Email: [email protected] KARL STORZ Endoskop Sverige AB Storsätragränd 14, 127 39 Skärholmen, Sweden Phone: +46 8 505 648 00 Email: [email protected] KARL STORZ Endoscopy Suomi OY Taivaltie 5, 01610 Vantaa, Finland Phone: +358 (0)96824774, Fax: +358 (0)968247755 Email: [email protected]
KARL STORZ Endoscopy (UK) Ltd. 415 Perth Avenue, Slough, Berkshire, SL1 4TQ, United Kingdom Phone: +44 1753 503500, Fax: +44 1753 578124 Email: [email protected] KARL STORZ Endoscopie Nederland B. V. Displayweg 2, 3821 BT Amersfoort, Netherlands Phone: +31 (0)33 4545890 Email: [email protected] KARL STORZ Endoscopy Belgium N. V. Phone: +31 (0)33 4545890 Email: [email protected] KARL STORZ Endoscopie France S. A. S. 12, rue Georges Guynemer, Quartier de l’Europe, 78280 Guyancourt, France Phone: +33 1 30484200, Fax: +33 1 30484201 Email: [email protected] KARL STORZ Endoskop Austria GmbH Landstraßer Hauptstr. 148/1/G1, 1030 Wien, Austria Phone: +43 1 71 56 0470, Fax: +43 1 71 56 0479 Email: [email protected] KARL STORZ Endoscopia Ibérica S. A. Parque Empresarial San Fernando, Edificio Munich – Planta Baja, 28830 Madrid, Spain Phone: +34 91 6771051, Fax: +34 91 6772981 Email: [email protected] KARL STORZ Endoscopia Italia S. r. l. Via dell’Artigianato, 3, 37135 Verona, Italy Phone: +39 045 8222000, Fax: +39 045 8222001 Email: [email protected] KARL STORZ Croatia d.o.o. Capraška 6, 10000 Zagreb, Croatia Phone: +385 1 6406 070, Fax: +385 1 6406 077 Email: [email protected] KARL STORZ Endoskopija d.o.o. Cesta v Gorice 34b, 1000 Ljubljana, Slovenia Phone: +386 1 620 5880, Fax: + 386 1 620 5882 Email: [email protected] KARL STORZ Polska Sp. z o.o. ul. Bojkowska 47, 44-100 Gliwice, Poland Phone: +48 32 706 13 00, Fax: +48 32 706 13 07 Email: [email protected] KARL STORZ Endoszkóp Magyarország Kft. Toberek utca 2. fsz. 17/b, HU-1112 Budapest, Hungary Phone: +36 195 096 31, Fax: +36 195 096 31 Email: [email protected] KARL STORZ Endoscopia Romania srl Str. Prof. Dr. Anton Colorian, nr. 74, Sector 4, 041393 Bukarest, Romania Phone: +40 (0)31 4250800, Fax: +40 (0)31 4250801 Email: [email protected] KARL STORZ Endoskope Greece M.E.P.E.* Patriarhou Grigoriou E’ 34, 54248 Thessaloniki, Greece Phone: +30 2310 304868, Fax: +30 2310 304862 Email: [email protected] *Repair & Service Subsidiary
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
18
Subsidiaries
KARL STORZ Industrial** Gedik Is Merkezi B Blok, Kat 5, D 38-39, Bagdat Cad. No: 162, Maltepe Istanbul, Turkey Phone: +90 216 442 9500, Fax: +90 216 442 9030 **Sales for Industrial Endoscopy
KARL STORZ Endoscopy (Shanghai) Ltd., Chengdu Branch Room 803-805, 8F Jin Jiang International Building, No. 1 West Linjiang Road, Wuhou District, 6100414, Chengdu, People’s Republic of China Phone: +86 28 86587977, Fax: +86 28 86587975 Email: [email protected]
OOO KARL STORZ Endoskopy – WOSTOK Derbenyevskaya nab. 7, building 4, 115114 Moscow, Russia Phone: +7 495 983 02 40, Fax: +7 495 983 02 41 Email: [email protected]
KARL STORZ Endoscopy (Shanghai) Ltd., Shenyang Branch Room 2001-2005, 20F N-MEDIA International Center, No. 167 Youth Avenue, Shenhe District, 110014, Shenyang, People‘s Republic of China Phone: +86 24 23181118, Fax: +86 24 23181119 Email: [email protected]
TOV LLC KARL STORZ Ukraine Avenue Geroyiv Stalingrada Str. 2D, office 717 Kyiv, 04210/Ukraine Phone: +38 095 000-895-0, +38-097-000-895-0, +38 073 000-895-0 Email: [email protected] KARL STORZ SE & Co. KG Representation Office Sabit Orudschow 1184, apt. 23, 1025 Baku, Azerbaijan Phone: +99 450 613 30 60 Email: [email protected] KARL STORZ ENDOSKOPE – East Mediterranean and Gulf (Offshore) S.A.L. Spark Tower 1st floor Charles Helou St., Horch Tabet – Sin El Fil, Beirut, Lebanon Phone: +961 1 501105, Fax: +961 1 501950 Email: [email protected] KARL STORZ Endoscopy (South Africa) (Pty) Ltd. P.O. 6061, Roggebaai, 8012 Cape Town, South Africa Phone: +27 21 417 2600, Fax: +27 21 421 5103 Email: [email protected] TOO KARL STORZ Endoskopy Kasachstan Saryarka, 6, BC “Arman“, off. 910, 010000 Astana, Republic of Kazakhstan Phone: +7 7172 552-549, 552-788, Fax: -444 Email: [email protected] KARL STORZ ENDOSKOPE East Mediterranean & Gulf (branch) Building West Side 7A – Unit 7WA – 3008, Dubai Airport Free Zone, P.O. Box 54983, Dubai - United Arab Emirates Phone: +971 (0)4 2958887, Fax: +971 (0)4 3205282 Service Hotline: +971 (0)4 3415882 Email: [email protected] KARL STORZ Endoscopy India Private Limited 11th Floor, Dr. Gopal Das Bhawan, 28, Barakhamba Road, New Delhi 110001, India Phone: +91 11 4374 3000, Fax: +91 11 4374 3010 Email: [email protected] KARL STORZ SE & CO. KG Interchange 21 Tower, Level 33, 399 Sukhumvit Road, North Klongtoey, Wattana, 10110 Bangkok, Thailand Phone: +84 28 3823 8000 Fax: +84 28 3823 8039 Email: [email protected] KARL STORZ SE & Co. KG Resident Representative Office 14th Floor, MPlaza Saigon, 39 Le Duan, District 1, Ho Chi Minh City, Vietnam Phone: +84 28 3823 8000, Fax: +84 28 3823 8039 Email: [email protected] KARL STORZ Endoscopy China Ltd. Room 2503-05, 25F AXA Tower, Landmark East, No. 100 How Ming Street, Kwun Tong, Kowloon, Hong Kong, People’s Republic of China Phone: +852 28 65 2411, Fax: +852 28 65 4114 Email: [email protected]
KARL STORZ Endoscopy (Shanghai) Ltd., Guangzhou Branch Room 02B & 03 & 04A, 35F Teem Tower, No. 208 Tianhe Road, Tianhe District, 510620, Guangzhou, People’s Republic of China Phone: +86 20 87321281, Fax: +86 20 87321286 Email: [email protected] KARL STORZ Endoscopy Asia Marketing Pte Ltd. No. 8 Commonwealth Lane #03-02, Singapore 149555, Singapore Phone: +65 69229150, Fax: +65 69229155 Email: [email protected] KARL STORZ Endoscopy Singapore Sales Pte Ltd No. 8 Commonwealth Lane #03-02, Singapore 149555, Singapore Phone: +65 69229150, Fax: +65 69229155 Email: [email protected] KARL STORZ SE & Co. KG Representative Office Indonesia Sinarmas MSIG Tower Level 37, Jl. Jend. Surdirman No. Kav. 21, Jakarta Selatan DKI Jakarta 12920 Email: [email protected] KARL STORZ Endoscopy Korea Co. Ltd. 9F Hyowon-Building, 97, Jungdae-ro, Songpa-gu, 05719 Seoul, Korea Phone: +82-70-4350-7474, Fax: +82-70-8277-3299 Email: [email protected] KARL STORZ Endoscopy Taiwan Ltd. 12F, No. 192, Sec. 2, Chung Hsin Rd., Sindian District, New Taipei City, Taiwan Phone: +886 933 014 160, Fax: +886 2 8672 6399 Email: [email protected] KARL STORZ SE & Co. KG Representative Office Philippines 1901 Picadilly Star Bldg., 4th Avenue, BGC, Taguig City 1636, Philippines Phone: +63 2 317 45 00, Fax: +63 2 317 45 11 Email: [email protected] KARL STORZ Endoscopy Japan K. K. Stage Bldg. 8F, 2-7-2 Fujimi, Chiyoda-ku, Tokyo 102-0071, Japan Phone: +81 3 6380-8622, Fax: +81 3 6380-8633 Email: [email protected] KARL STORZ Endoscopy Australia Pty. Ltd . 68 Waterloo Road, Macquarie Park NSW 2113, P O Box 50 Lane Cove NSW 1595, Australia Phone: +61 (0)2 9490 6700, Fax: +61 (0)2 9420 0695 Toll free: 1800 996 562 (Australia only) Email: [email protected] www.karlstorz.com
KARL STORZ Endoscopy (Shanghai) Ltd., Beijing Branch Room 1805-1807, Building B, 18F Beijing IFC, No. 8, Jianguomenwai Street, Chaoyang District, 100022, Beijing, People’s Republic of China Phone: +86 10 5638188, Fax: +86 10 5638199 Email: [email protected] KARL STORZ Endoscopy (Shanghai) Ltd., Shanghai Branch Room 701A Building 5 & Room 501 Building 7, No. 3000 Longdong Avenue, Pilot Free Trade Zone, 201203, Shanghai, People’s Republic of China Phone: +86 21 60339888, Fax: +86 21 60339808 Email: [email protected]
Instructions for use • Articulated stand with quick-release coupling, base and clamping jaws • OUK808_EN_V1.0_03-2021_IFU_CE-MDR
19