KARL STORZ ENDOSKOPE

HOPKINS Telescope

10970BA, DCI HOPKINS Telescope 30°

12 Pages

Processing Instruction 10970BA, DCI® HOPKINS® Telescope 30°  Table of contents 1  RI/10970BA/E Version: 2.0015  1 Safety section ... 3 2 Accessories... 5 3 Preparation for cleaning and disinfection ... 5 4 Manual precleaning... 5 4.1 Disassembly... 5 4.2 Brushing the surfaces ... 5 4.3 Auxiliary pretreatment with ultrasound ... 5 5 Manual cleaning ... 5 6 Manual disinfection ... 5 7 Machine cleaning and disinfection ... 6 7.1 Automated cleaning/thermal disinfection ... 6 7.2 Slide-in tray and instrument holder... 6 8 Assembly, inspection and care... 6 9 Packaging Systems ... 6 10 Sterilization... 7 10.1 Validated sterilization ... 7 Steam sterilization using the fractionated prevacuum procedure ... 7 Hydrogen peroxide (H₂O₂) sterilization – ASP STERRAD®... 7 Hydrogen peroxide (H₂O₂) sterilization – STERIS® AMSCO®... 8 Ethylene oxide sterilization (EO) ... 8 Chemical low-temperature sterilization with STERIS® System 1®... 9 Chemical low-temperature sterilization with peracetic acid – STERIS® System 1E® ... 9 10.2 Optional sterilization ... 9 Low-temperature steam and formaldehyde process (LTSF) ... 9 11 Limits of reprocessing... 10 12 Contact data ... 11  NOT APPLICABLE FOR THE US Page 2 of 12
File Type: PDF
File Size: 770 KB
File Name: KARL STORZ ENDOSKOPE - RI-10970BA-E - 10970BA, DCI HOPKINS Telescope 30° - Ver2.0015.pdf