KCI
Prevena Patient Guide Rev B June 2010
Patient Guide
2 Pages
Preview
Page 1
Prevena™ Incision Management System
Patient Guide Refer to the Prevena™ 125 Clinicians Guide for additional information.
Device Description This guide provides important information including a contraindication, cautions and warnings that you should read and review. You have been prescribed the Prevena™ Incision Management System. Do not remove the dressing or turn off the unit, unless directed by your treating physician or this guide. If an alert sounds, an action is required on your part; refer to “Indicators and Alerts”. For questions or concerns regarding this therapy, contact your treating physician.
Prevena™ 125 Carrying Case
Prevena™ 125 Therapy Unit !
!
Prevena™ Carrying Case can be worn on a belt, over the shoulder or diagonally across the chest (not around the neck) with adjustable strap (provided).
Instructions for Use Important:
Monitor the Pressure Indicator on the dressing regularly. Refer to the Indicators and Alerts section of this guide for more details.
Important:
Do not turn therapy unit off unless advised by the treating physician or there is bleeding, signs of infection or signs of an allergic response; see WARNINGS below.
CaUtIon:
Dressing should only be removed by or on the advice of the treating physician.
Bathing •Bathing in tub is not recommended. •Do not submerge therapy unit or dressing. Showering •Light showering is permissible. Protect therapy unit from direct spray. Avoid prolonged water contact with therapy unit and dressing. •When towel drying avoid disrupting or damaging the dressing. Cleaning The therapy unit and carrying case can be wiped with a damp cloth using a mild household soap solution that does not contain bleach. Strenuous Activity •Refer to treating physician for when and what level physical activities may be resumed. •Avoid strenuous activities when using the Prevena™ Incision Management System.
Indicators On/Off Button and Audio Pause (Push once to pause audible alerts for 60 minutes). �
!
Audio Pause Indicator
• Contraindication: Sensitivity to silver. •As with any prescription medical device, failure to carefully read and follow all instructions and safety information prior to use may lead to improper product performance.
Low Battery Indicator
•Bleeding: With or without using the Prevena™ Incision Management System, certain patients are at risk of bleeding complications due to the operative procedure or concomitant therapies and/or comorbidities. If bleeding develops suddenly or in large amounts during therapy, leave Prevena™ Incision Dressing in place, turn off Prevena™ Therapy Unit and seek immediate emergency medical assistance.
+
Push here to slide open battery cover.
+
_
Contraindication and Warnings
Caution Indicator �
Battery Replacement
+
Sleeping •Place therapy unit in a position where tubing will not become kinked or pinched. •Ensure therapy unit cannot be pulled off table, or dropped on the floor during sleep.
Replace with “AA” batteries. Lithium batteries recommended for optimal performance.
Prevena™ Incision Dressing with Pressure Indicator Use Prevena™ Patch Strips to seal leaks. Dressing Pressure Indicator in down position indicates therapy at acceptable pressure. Dressing Pressure Indicator in up position indicates therapy is not at the acceptable pressure. (Refer to Indicators and Alerts)
•If you experience any of the following, you should call your treating physician right away as your incision may have become infected: you become feverish and/or there is an increase in soreness, redness, swelling, itching, warmth, or if there is pus or a bad odor. Your physician will advise you as to whether Prevena™ Therapy should be discontinued. •Magnetic Resonance Imaging (MRI): Do not take the Prevena™125 Therapy Unit into the MR environment. The Prevena™ Incision Dressing can typically remain with minimal risk in an MR environment. •Hyperbaric Oxygen Therapy (HBO): Do not take the Prevena™ 125 Therapy Unit or Prevena™ Incision Dressing into the hyperbaric oxygen chamber, it is not designed for this environment and should be considered a fire hazard. •If defibrillation is required in the area of Prevena™ Incision Dressing placement, the dressing should be removed before defibrillation, follow clinician’s instructions. •If at any time while using Prevena™ Incision Management System, the canister becomes full of fluid other than blood, indicated by a “Maximum Capacity” alert or visual inspection, turn therapy unit off and contact the treating physician. •The Prevena™ Incision Dressing has an acrylic adhesive coating and a skin interface layer with silver, which may present a risk of an adverse reaction in patients who are allergic or hypersensitive to acrylic adhesives or silver. If a patient has a known allergy or hypersensitivity to these materials, do not use the Prevena™ Incision Management System. If any signs of allergic reaction, irritation or hypersensitivity develop, such as redness, swelling, rash, urticaria, or significant pruritus, patient should consult a physician immediately. If bronchospasm or more serious signs of allergic reaction appear, the patient should turn off the Prevena™ Therapy Unit and seek immediate emergency medical asistance.
Symbols Used
Indicators and Alerts Therapy On/Off
TURN OFF AUDIO PAUSE TURN ON Hold 2 seconds Hold 5 seconds Push once to mute Alerts for 60 minutes
!
!
Green LED Indicates Therapy is On
!
!
Device Life-cycle Expired
Call Treating Physician
!
Life Cycle 192 hours (8 Days)
CM
!
YES
Leak Alert Eliminated?
x1 Press firmly around dressing edge
A slow beep/flashing LED = Change batteries within six hours. A rapid beep/rapidly flashing LED = Change Batteries Immediately.
Maximum Capacity Alert
Consult Instructions for Use Sterile using Radiation
+
Hold 5 seconds to TURN OFF. NO
!
Authorized Representative in the European Community
!
!
!
Call Treating Physician.
NO
Use Prevena™ Patch Strips to seal leak.
3 “AA” Batteries Observe (+) and () for orientation.
+
Low Battery Alert
Ensure Canister is securely locked and side tabs are flush with unit
This product is designated for separate collection at an appropriate collection point. Do not dispose of as household waste.
Continue Therapy.
+
Check all connectors for leaks.
(Lithium Batteries Recommended)
Hold 2 seconds to TURN ON .
Date of Manufacture
Check for kinked or pinched tubing.
Single Use Only
Manufacturer
Do not Resterilize
Class II Device
Do not use if package is damaged or open
Refer to Clinician Guide
Type B Applied Part
Latex Free
Expiration Date
Keep Dry
Fragile
Full or near full?
x2
YES Perform visual inspection of canister.
Turn therapy unit off. Call Treating Physician immediately. See WARNINGS Section on opposite page.
System Error Alert
! x1
ETL Listed, Conforms to UL Std. 60601-1 certified to CAN/CSA C22.2 Std. No. 601.1.
3182664
Leak Alert
x1
Conforms with the Medical Device Directive (93/42/EEC) and has been subject to the conformity procedures laid down in the council directive
YES
!
!
Continue Therapy.
Therapy Running?
x1
Hold 5 seconds to TURN OFF.
Latex
NO
Hold 2 seconds to TURN ON .
Call Treating Physician.
Dressing Pressure Indicator YES UP Position
Check for kinked or pinched tubing.
Check for leaks. See Leak Alert section above.
Specifications Environmental Conditions: Storage Conditions: Temperature Range:
Operating Conditions: -4ºF (-20ºC) to 140ºF (60ºC) Temperature Range: 41ºF (5ºC) to 122ºF (50ºC)
Relative Humidity Range: 0-95%, non-condensing
Altitude Range for Optimum Performance: -50 to 8000 feet (-15.24m to 2438 m)
Continue Therapy. Non Sterile
Indicator Down? NO
Call Treating Physician.
Manufacturer Information KCI USA, Inc. San Antonio, TX 78219 USA
KCI Medical Products (UK) Ltd. Wimborne, Dorset, BH21 7SH United Kingdom www.kci-medical.com
Rx only Federal (USA) law restricts this device to sale by or on the order of a physician.
Contact Information . For additional information concerning the Prevena™ Incision Management System, contact your local KCI representative.
0473
Equipment not suitable for use in presence of flammable anesthetic mixture with air, oxygen or nitrous oxide. IPX4 - Ingress Protection Level
Electromagnetic Compatibility:
Treating Physician______________________
Electromagnetic Interference - Although this equipment conforms with the intent of the directive 89/336/EEC in relation to Electromagnetic Compatibility (EMC), all electrical equipment may produce interference. If interference is suspected, move equipment away from sensitive devices or contact the manufacturer. Prevena™ and KCI are trademarks of KCI Licensing, Inc. © 2010 KCI Licensing, Inc.
All rights reserved. 390152-WEB Rev B 06/10
Phone: ___________