English Thank you for purchasing the KANGAROO ePump™ Enteral Feeding Pump. With proper care, this device will provide you with years of precision service.
Table of Contents Page
Section I - General Information... 1 Section II - Safety and Warnings... 3 Section III - Icon Identification ... 4 Section IV - Initial Setup Attaching the A/C Adapter Power Cord... 5 Battery Setup... 6 Attaching Pole Clamp... 6
Section V - Instructions for Use
Quick Start... 7 General Startup... 8 Placement/Mounting... 8 A/C Power Operation... 8 Battery Power Operation ... 9 Power On/Off... 9 Language Selection, First Power Up... 9 Keep or Clear Prior Pump Settings... 9 Loading Pump Sets... 10 Prime Pump... 11 Auto Priming... 11 Hold-To-Prime... 12 Re-priming after Bag Empty... 12 Selecting Feed Mode... 12 Selecting EZMODE ... 13
e
KANGAROO Pump™
Enteral Feeding Pump
Table of Contents Selecting Continuous or Intermittent Mode... 13 Continuous Mode (Adjust Feed)... 14 Continuous Mode Feeding ... 14 Continuous Mode Flushing... 14 Intermittent Mode (Adjust Bolus)... 15 Intermittent Mode Feeding... 15 Intermittent Mode Feeding - Bolus Max... 16 Intermittent Mode Flushing... 16 Run... 17 KTO ... 18 Lock Screen Feature... 18 Hold... 19 Flush Now... 20 More Options ... 20 Buzzer Alarm Volume ... 20 History... 20 Language Selection... 21 Continuous/Intermittent Mode... 21 BioTech Menu... 21 Accessing the BioTech Screen... 21 BioTech Screen Data... 21 BioTech Options... 22 Lock Settings Feature (Program Locked)... 22 EZMode ... 22 Communication... 22
Section VI - Re-Certification of Performance System Performance Re-Certification... 23 Certifying the Accuracy of Pump Flow Rate ... 23
Section VII - Cleaning... 24 Section VIII - Battery Replacement... 25
e
KANGAROO Pump™
Enteral Feeding Pump
Table of Contents Section IX - Error/Warning/Informational Screens and Troubleshooting
System Error... 27 Hold Error... 28 Rotor Error... 28 Feed Error... 28 Flush Error... 29 Flow Error ... 29 Pump Set Dislodged Error... 29 Battery Low... 30 Feeding Complete... 30 Pump Set Use >24 Hours Warning... 30 LED Array... 31
Section X - Specifications and Symbols Specifications... 31 Symbols on Pump and Accessories... 34
Section XI - Customer Service... 35 Section XII - Maintenance Side Door Enclosing Pump Set Loading Area ... 35 Battery Pack ... 36 Battery Door ... 36 Power Cord ... 36 Pole Clamp ... 36
Section XIII - Service Part Numbers Accessory Items ... 37
Section XIV - Warranty... 38 Section XV - Electromagnetic Conformity Declaration... 39 Section XVI - AccuSystem√... 40 This product contains software solely owned by Tyco. Tyco grants the user a non-exclusive, limited license to use the software according to the operating instructions. A copy of the license can be obtained from Tyco.
e
KANGAROO Pump™
Enteral Feeding Pump
Section I – General Information The KANGAROO ePump enteral feeding pump is a simple-to-use, precision enteral feeding pump. It is easily programmed to provide patients with either continuous or intermittent feeding, and can provide automatic flushing capability when used with KANGAROO ePump Feed & Flush Sets.
User Interface: • Intuitive user interface • Large, backlit LCD display • Step-by-step prompts to guide programming and operation • “Stoplight” LED array visually indicates pump status in a bright or darkened room
Ergonomics: • Quiet operation • Compact, aesthetic design • Tabletop usage or IV pole mounting • Simple loading of Pump Sets • Transparent side door to protect Pump Set • Integrated handle for easy transport
Features: • MISTIC (Magnetic Intelligent Set Type Identification Connector) Feed-Only vs. Feed & Flush Set identification system to ensure match between the pump’s user interface and Pump Set type • Automatic Anti-Free-Flow (AFF) System • Audible alarm to indicate errors or Pump Set loading conditions • Attitude-independent design eliminates need for drip chamber on pump set • Sensor technology detects both upstream and downstream flow conditions • Continuous feed, intermittent feed, and flush capability • Auto-prime feature reduces the need for time-consuming manual priming • Hold feature with restart and Keep Tube Open (KTO) feature • Stores previous 72 hours of feeding and flushing history
Service: • Serial I/O port for data retrieval and software updates • Uses replaceable, rechargeable NiMH Battery Pack for 15 hours of backup power • Removable Power Cord
e
KANGAROO Pump™
Enteral Feeding Pump
Status Line
Menu Selections
Status LEDs
Input Settings Menu Selection Buttons
Output Totals Power Button
Graphic Display Battery Status
Front
Integrated Handle Feeding Set Cap Storage Area Pole Clamp Attachment Point
A/C Adapter Port
Serial I/O Port Door
Battery Compartment
Back Figure 1A. KANGAROO ePump features, front and rear views.
e
KANGAROO Pump™
Enteral Feeding Pump
'FFEJOH4FU$BQUVSF
1PMF$MBNQ
'FFEJOH4FU -PBEJOH"SFB
Figure 1B. KANGAROO ePump Pump Set loading area.
Section II - Safety and Warnings CAUTION: U.S. federal law restricts the sale of this device to physicians or to their direct representatives. 1. Read this booklet thoroughly before using the KANGAROO ePump enteral feeding pump. 2. Do not use this device near flammable anesthetics. 3. Disconnect power supply before cleaning or servicing. 4. Use only KANGAROO ePump enteral feeding pump Power Cord (with built-in A/C adapter) with pump. See Section XIII - Service Part Numbers for replacement of Power Cord. 5. Use only KANGAROO ePump enteral feeding pump Pump Sets with this device. Pump is not compatible with other Pump Sets. 6. For re-certification, see Section VI - Re-certification of Performance. For other integrity checks, consult with a qualified Biomedical technician or contact the manufacturer (Section XI - Customer Service). 7. For service or for technical information, please contact Customer Service (Section XI). 8. Do not open the main housing, as there are no user-serviceable parts inside. Opening may affect function of device and voids the warranty. To replace battery, see Section VIII. 9. Dispose of old battery-powered electronic equipment in a manner consistent with institutional policy for expired equipment disposal. 10. Cleaning frequency and practices must be consistent with institutional policy for cleaning of non-sterile devices. See Section VII - Cleaning, for instructions on cleaning the KANGAROO ePump enteral feeding pump. 11. See icon descriptions in Section III - Icon Identification and Section X - Specifications and Symbols for additional safety information.
e
KANGAROO Pump™
Enteral Feeding Pump
12. This device is designed to minimize the effects of uncontrolled electromagnetic interference and other types of interference from external sources. Avoid use of other equipment that may cause erratic operation or degradation in the performance of this device. 13. WARNING: The serial I/O port (see Figure 1A) is intended only for use by the manufacturer. Any accessory equipment that can be connected to the DB9 Serial I/O port, and which does not comply with the equivalent safety requirements of this equipment, may lead to a reduced level of safety of the pump. While use of the pump’s DB9 port is not recommended for customers, the following should be considered in the event that the customer decides to connect to this port: • The accessory is specifically certified as complying with the safety requirements for accessories used with the pump, whereas the safety certification of the accessory has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized standard. 14. CAUTION: This enteral feeding pump should only be used for patients who can tolerate the flow rates and accuracy levels delivered by the pump. Premature infants may require higher accuracy rates than specified for this enteral feeding pump.
Section III – Icon Identification Run Mode
Vertically-moving droplet in RUNNING screen indicates Run mode.
Battery Symbols • Full charge • 2/3 or less charge • 1/3 or less charge • Very low battery (15 minutes max.)
Auto Priming Indicator When auto priming, the bar will progress to show auto prime feature is active.
Volume Setting Indicates alarm volume setting.
Error Indicator Pump operations are stopped until the error is resolved.
Information Indicator This indicates information only, and does not require any immediate action.
e
KANGAROO Pump™
Enteral Feeding Pump
EZMODE Indicator Indicates that EZMODE feature is turned on.
Communication Indicator Indicates that communication feature is turned on.
Volume to be Delivered Indicator Indicates that VTBD feature is turned on.
Keep Tube Open Indicates that Keep Tube Open feature is turned on.
Section IV - Initial Setup Attaching the A/C Adapter Power Cord Plug the Power Cord into the A/C Adapter Port on the back of the pump. See Figure 1A to locate port on pump. Make note of the location of the pins and tab on the plug to properly orient the plug. "$"EBQUFS 1PSU
"$"EBQUFS 1MVH
Figure 2A. Alignment of A/C Adapter plug with pump’s adapter port.
6TFDBSFBMJHOJOHBEBQUFS XJUIQVNQDPOOFDUPS
The connector, which attaches to the back of the pump, has a spring-loaded locking sleeve that locks into the pump, thus helping to prevent the connector from being accidentally unplugged from the pump. To remove the connector, grasp the outer sliding sleeve member and pull. Do not detach the connector by pulling on the cord only. (A cord extension accessory is available that will allow the wall plug to be separated from the power transformer, see Fig. 2B. The cord extension order number is located on page 37.)
e
KANGAROO Pump™
Enteral Feeding Pump
Figure 2B.
Battery Setup The battery pack is fully charged before shipping, and is left unconnected to optimally preserve battery life. WARNING: The battery should be connected before using the pump. It is also recommended that the battery charge be “topped off” before battery-only operation is utilized. The battery pack will charge continuously whenever the pump is plugged into a wall outlet, including during normal use of the pump with A/C power. Eight hours of charging is required to fully recharge the battery pack. A new battery pack will provide approximately 15 hours of service before needing to be recharged.
Attaching Pole Clamp The KANGAROO ePump enteral feeding pump can be attached to a vertical pole using the Pole Clamp, included with the pump. The pole clamp provides the ability to retain the power cord to prevent accidental cord removal or cord loss. Simply place the power cord into the U-shaped groove located on the back of the pole clamp as shown in Fig. 3 and attach the pole clamp to the KANGAROO ePump enteral feeding pump. Be sure not to pinch the power cord or damage the cord insulation when installing. To attach the Pole Clamp to the KANGAROO ePump enteral feeding pump, align the hole on the Pole Clamp with the mounting hole on the back of the pump. Use a 3/8” - 16 bolt to fasten together, as shown in Figure 3.
e
KANGAROO Pump™
Enteral Feeding Pump
Figure 3. Attaching and using the Pole Clamp or Cord Clip.
If cord restraint is preferred but use of the pole clamp is not desired, a cord restraint clip is provided with the KANGAROO ePump enteral feeding pump. Using the pole clamp mounting bolt, follow the pole clamp assembly instruction to attach the cord clip (Refer to Fig. 3).
Section V - Instructions For Use Quick Start 1. Press
POWER on lower, right-hand corner of control panel.
2. Open blue door enclosing Pump Set loading area. 3. To load Pump Set (refer to Figure 4):
e
KANGAROO Pump™
Enteral Feeding Pump
• Grasp finger tab on valve and insert into left pocket (1). • Grasp black ring retainer and gently stretch tubing around rotor (2). Pull up on retainer and insert into right pocket (3). • Press down on finger tab to ensure that valve is fully seated. Finger tab should align with locating rib. • Close blue door. (Note: display status line should read SET LOADED). 4. To automatically prime the pump, press4PRIME PUMP and4AUTO PRIME. For Feed & Flush sets, auto priming will prime both lines. 5. To directly control pump priming, press4PRIME PUMP and then press and hold the HOLD TO PRIME button(s)4. If a Feed & Flush Set is used, remember to first press4HOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then press4HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the distal end connector. 6. Use buttons4on left side of LCD to set the feeding parameters. Set flushing parameters if a Feed & Flush Set is loaded. The continuous or intermittent feed mode selection is made under the MORE options menu. 7. Press4RUN when ready to start. The screen will display RUNNING. 8. To stop, press4HOLD or press and hold the
POWER button. Black Ring Retainer (MISTIC)
Finger Tab
1
3
Figure 4. Installation of Pump Set.
Valve
2 Rotor
General Startup Placement/Mounting There are two recommended placement methods for the KANGAROO ePump enteral feeding pump. • Attached to a vertical IV pole via the Pole Clamp peripheral device, included with the pump. • Placed feet down on any horizontal surface.
A/C Power Operation Plug the KANGAROO ePump enteral feeding pump into an A/C power outlet for normal operation. When the pump is not plugged in, or if A/C power is interrupted, a built-in rechargeable battery will run the pump.
e
KANGAROO Pump™
Enteral Feeding Pump
Battery Power Operation If A/C power is not available to the pump, or if A/C power is interrupted, the pump will automatically operate on backup battery power. The built-in battery recharges any time the pump is plugged into an outlet. A fully-charged, new battery will supply 15 hours of backup power at 125 mL/hr before requiring recharging. Contact Customer Service, Section XI, to replace the Battery Pack if battery life degrades.
Power On/Off To power up the pump, press the POWER button on the lower right of the front panel. To power down the pump, press and hold the POWER button until the pump display turns off.
Language Selection, First Power Up Immediately after powering up the pump for the first time, the language option screen will appear. Select the language using the ( È ) or ( Í ) buttons. When the language is selected, push the fifth button ( Á ) to enter the language selection. After the language is selected the first time the pump is powered up, the language selection screen will not appear during future power up operations. Should the language need to be changed after the initial power up, it can be changed through the “MORE OPTIONS” menu (see Figure 10A).
Keep or Clear Prior Pump Settings Immediately after powering up the pump, if the prior feed or flush settings were not cleared and the settings are not locked in the BioTech Screen, the pump will give the following two options:
KEEP SETTINGS - Select this option to start with the same settings that were most recently programmed into the pump. Any of the previously-programmed settings can be modified, if needed. The feed volume totals, “mL FED” (Continuous Feed Mode) or “BOLUSES FED” (Intermittent Feed Mode) and “mL FLUSH” are also maintained.
CLEAR SETTINGS - Select this option to reset all input feeding settings and the output feed totals to zero. It will then be necessary to program all settings before running the pump. The BioTech options settings will not change.
Note: The output feed totals (“mL FED,” “BOLUSES FED,” “mL FLUSH”) can be cleared at any time, and without clearing the input settings, by doing the following: • Start the pump RUNNING • Press4HOLD • Press4CLEAR VOLUME The status line of the pump will then show LOAD A SET (Figure 5), or, if the Pump Set is already loaded, it will show SET LOADED (Figure 7).
e
KANGAROO Pump™
Enteral Feeding Pump
Loading Pump Sets` The KANGAROO ePump enteral feeding pump will indicate LOAD A SET in the screen’s status line if a Pump Set is not installed. A blinking icon of a Pump Set will also appear in the upper left corner of the screen. Figure 5A shows the screen for continuous feeding mode and Figure 5B shows the screen for intermittent feeding mode (intermittent feeding mode not available in EZMODE).
Figure 5A. LOAD A SET opening menu screen for Continuous Feeding mode (EZMODE not activated).
Figure 5B. LOAD A SET opening menu screen for Intermittent Feeding mode (EZMODE not activated).
To load a Pump Set, follow the steps below: 1. Open the blue transparent door that encloses the rotor and Pump Set loading area. 2. Load the Pump Set by holding the finger tab on the valve, and inserting the valve down into the slot on the left. (Figure 6A) 3. Grasp the MISTIC retainer end (black ring retainer) and wrap the tubing counterclockwise around the rotor. (Figure 6B) 4. Carefully pull the MISTIC retainer end upwards to position it over the slot on the right, and then lower the retainer into the slot. (Figure 6B) 5. Press down on the valve’s finger tab to ensure proper seating. 6. Close blue transparent door. 1VNQ4FU
4FU-PBEJOH%PPS
'JOHFS5BC 7BMWF
.*45*$ 3FUBJOFS &OE 1PXFS #VUUPO
Figure 6A.
Figure 6B.
The display should read SET LOADED and will show one of the following opening menus, depending on the type of Pump Set detected, and whether continuous or intermittent feeding mode has been selected.
e
KANGAROO Pump™
10
Enteral Feeding Pump
Figure 7A. Opening menu for Continuous Feed mode with a Feed-Only set.
Figure 7B. Opening menu for Continuous Feed mode with a Feed & Flush set.
Figure 7C. Opening menu for Intermittent Feed mode with a Feed-Only set.
Figure 7D. Opening menu for Intermittent Feed mode with a Feed & Flush set.
Prime Pump The KANGAROO ePump enteral feeding pump may be primed automatically with a single button push, including the flushing line if a Feed & Flush Set is loaded. The pump can also be primed in a more interactive method using the hold-to-prime feature. After a Pump Set has been loaded and the status line shows SET LOADED, press4PRIME PUMP in the opening menu (Figures 7A7D) to get one of the PRIME PUMP menus shown in Figures 8A-8B.
Figure 8A. PRIME PUMP menu for Feed-Only set.
Figure 8B. PRIME PUMP menu for Feed & Flush set.
Auto Priming The AUTO PRIME option will be available if the pump senses that the currently installed Pump Set was not previously primed, and that there is no fluid in the line near the rotor. If auto-priming is desired, press and release4AUTO PRIME to automatically prime the line(s). For Feed & Flush Pump Sets, the auto prime feature will automatically prime both lines, starting with the flush line.
e
KANGAROO Pump™
11
Enteral Feeding Pump
Press4STOP to cancel AUTO PRIME. When auto-priming is completed, the status line will show AUTO PRIME COMPLETE and will no longer show the AUTO PRIME option. Be sure that the line(s) are fully primed. If not, use the hold-to-prime option(s) to complete the priming, as described below.
Hold-to-Prime The hold-to-prime options allow for precise interactive control of the priming process.
Feed-Only Pump Sets For Feed-Only Pump Sets, the screen will appear as in Figure 8A. Press and hold4HOLD TO PRIME FEED until the feed line has been primed down to the stepped end connector at the end of the Pump Set.
Feed & Flush Pump Sets For Feed & Flush Pump Sets, the display will appear as in Figure 8B. First press and hold4HOLD TO PRIME FLUSH to prime the flushing line, and then press and hold4HOLD TO PRIME FEED to prime the feed line down to the stepped end connector at the end of the Pump Set. If a Feed & Flush Pump Set is used, remember to first press4HOLD TO PRIME FLUSH until the flushing fluid has reached the valve, and then press4HOLD TO PRIME FEED until the feeding fluid has reached past the valve and down to the stepped end connector. If the feeding line is primed first, subsequent priming of the flushing line will force the air between the flush solution bag and valve into the main line, which already has been primed with enteral formula.
Re-priming after Bag Empty A Pump Set bag that has been emptied will trigger the FEED ERROR screen. In this condition the Pump Set bag can be refilled to continue the feeding, but only after the Pump Set has been re-primed. To re-prime the Pump Set, do the following: • Disconnect the feeding line from the patient • Refill the bag • Press4CONTINUE to begin the pump RUNNING • Press4HOLD • Press4ADJUST SETTINGS • Press4PRIME PUMP • Press4HOLD TO PRIME to prime the line(s) Pump Sets should not be reused after 24 hours of initial usage.
Selecting Feed Mode EZMODE Option The EZMode option is a limited, Continuous Mode only setting of ePump. It feeds non-stop (no VTBD option) at a rate
e
KANGAROO Pump™
12
Enteral Feeding Pump
0-400mL/hr as defined by the user until manually stopped. It has no Bolus, Flushing, Run Mode Screen Lock, or AutoResume capability and displays no History information. Turning on EZMode will automatically force the unit out of the Intermittent (Bolus Feed) mode if set. Although the Continuous Mode feed VTBD will be set to 0 while EZMode is enabled, the previous value of the Feed VTBD will be stored in flash memory and recalled when “EZMode” is deselected. The user can change the Feed Rate by normal means. All other user settings, including those not compatible with the EZMode (Bolus VTBD, Number of Boluses, Super Bolus, etc), that were set before the pump was put into the EZMode will be ignored and invisible while in EZMode but remain stored in the flash memory for when EZMode is deselected. EZMode essentially has no effect on Clear Settings; pressing Clear Settings while EZMode is enabled will clear all Feed and Flush settings and totalizers as normal, not just those that apply to EZMode. Selecting the Clear Settings option on power up will not clear Biotech Options such as EZMode or Lock Settings. If the user selects the Adjust Feed option from either the Load A Set or Set Loaded screens, control will pass directly to and from the Enter Feed Rate screen and bypass the Adjust Feed screen since without the ability to adjust VTBD in EZMode, the Adjust Feed screen is redundant.
Selecting EZMODE EZMode may be selected through the Biotech Options button which may be accessed on the Biotech screen (see page 21). The EZMode setting will be stored in the flash memory and saved when the unit is shut off. NOTE: If the user inserts a Feed/Flush set while the EZMode is enabled, EZMode will be cancelled and the unit will immediately revert back to the normal mode. If EZMode is desired, the user must then re-enable EZMode through the Biotech Options screen. When enabled and not in Certification mode, a small EZ icon will be visible on the bottom of the screen near the battery icon. However, it will not appear in the Power Down, System Error, or Low Battery screens.
Figure 9.
Selecting Continuous or Intermittent Feed Mode The KANGAROO ePump enteral feeding pump can be programmed to feed continuously or intermittently. The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has been delivered or until the supply has been exhausted. The Intermittent feeding mode delivers boluses of enteral nutrition at programmed time intervals. The bolus volume and feed rate are also programmed. To select continuous feed mode or intermittent feed mode, press4MORE on the opening menu, shown in Figures 5A-5B or Figures 7A-7D, to access the MORE OPTIONS menu, Figure 10A. The MORE option is available either before a Pump Set has been loaded (Figures 5A-5B) or after the Pump Set is loaded (Figures 7A7D). See subsection “More Options” for information on the other options in the MORE OPTIONS menu.
e
KANGAROO Pump™
13
Enteral Feeding Pump
Select4CONTINUOUS/INTERMITTENT under the MORE OPTIONS menu, Figure 10A. Then press 4CONTINUOUS MODE or4INTERMITTENT MODE from the SELECT MODE menu, Figure 10B. Press4DONE to exit.
Figure 10A. MORE OPTIONS menu.
Figure 10B. SELECT MODE menu for continuous or intermittent mode selection.
Continuous Mode - Adjust Feed Continuous Mode Feeding For CONTINUOUS feeding mode, the ADJUST FEED option is available in the opening menus (Figure 7A for Feed-Only sets or Figure 7B for Feed & Flush sets). Select this option to program the feeding parameters from the ADJUST FEED screen, Figure 11A. Press4FEED RATE in Figure 11A to define the rate of delivery. Press the4buttons in Figure 11B to program the rate from 1 to 400 in increments of 1 mL/hr. Press4ENTER (top button) to exit the menu. Press4FEED VTBD in Figure 11A to define the Volume To Be Delivered. Press the4buttons in Figure 11C to program the volume from 1 to 3000 mL in increments of 1 mL. Press4ENTER (top button) to exit the menu. If the FEED VTBD is set to zero, the pump will run until the supply is exhausted. NOTE: When a FEED VTBD is set, the remaining volume to be delivered (remaining VTBD) will be displayed next to the W icon on the lower right-hand portion of the screen (Figure 11A). If the FEED VTBD is reprogrammed during a feeding, from HOLDING mode, the pump will restart the feeding and deliver the entire volume of the new VTBD setting.
Figure 11A. ADJUST FEED menu.
Figure 11B. Setting the Feed Rate. Figure 11C. Setting the Feed Volume to be Delivered.
Continuous Mode Flushing (Not Available in EZMODE) For CONTINUOUS feeding mode with a Feed & Flush Set loaded, the opening menu will appear as shown in Figure 7B. Press4ADJUST FLUSH in Figure 7B to program the flushing parameters from the ADJUST FLUSH screen, shown in Figure 12A. Press4FLUSH VOLUME in Figure 12A to define the volume per flush occurrence. Press the4buttons in Figure 12B to program the flush volume from 10 to 500 in increments of 1 mL. Press4ENTER (top button) to exit the menu.
e
KANGAROO Pump™
14
Enteral Feeding Pump
Press4FLUSH INTERVAL in Figure 12A to define the time interval between starts of flushing. Press the4buttons in Figure 12C to program the time interval from 1 to 24 hours in increments of 1 hour. Press4ENTER (top button) to exit the menu. The pump will automatically limit flushing settings which exceed pump capabilities to deliver. CAUTION: The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when programming the flush volume so that it matches the patient’s need. NOTE: The pump will automatically flush an additional 25 mL of water after completing the programmed delivery of formula, as long as at least one flush had occurred during formula delivery.
Figure 12A. ADJUST FLUSH menu, continuous mode.
Figure 12B. Setting the Flush Volume.
Figure 12C. Setting the Flush Time Interval.
Intermittent Mode - Adjust Bolus (Not Available in EZMODE) Intermittent Mode Feeding For INTERMITTENT feeding mode, the ADJUST BOLUS option is available in the opening menus (Figure 7C for a Feed-Only set or Figure 7D for a Feed & Flush set). Select this option to program the feeding parameters from the ADJUST BOLUS screen, Figure 13A. Press4BOLUS RATE in Figure 13A to define the rate of delivery. Press the4buttons in Figure 13B to program the delivery rate from 1 to 400 in increments of 1 mL/hr. Press4ENTER (top button) to exit the menu. Press4NUMBER OF BOLUSES in Figure 13A to define the total number of boluses to be delivered. Press the4 buttons in Figure 13C to program the number of boluses, up to a maximum of 99. Press4ENTER (top button) to exit the menu. The number of boluses can be set as zero, in which case the pump will run until the supply is exhausted. Press4VOLUME PER BOLUS in Figure 13A to define the Volume to be delivered in each bolus. Press the4buttons in Figure 13D to program the volume per bolus from 1 to 3000 mL in increments of 1 mL. Press4ENTER (top button) to exit the menu. Press4BOLUS INTERVAL in Figure 13A to define the time interval between starts of bolus deliveries. Press the4buttons in Figure 13E to program the time interval between starts of bolus deliveries from 1 to 24 hours in increments of 1 hour. Press4ENTER (top button) to exit the menu. The pump will automatically limit bolus settings which exceed pump delivery capabilities. For example, a bolus delivery volume of 200 mL cannot be made once per hour if a rate of 150 mL/hr is already programmed. In this case, the maximum volume allowed would be 150 mL.
e
KANGAROO Pump™
15
Enteral Feeding Pump
NOTE: If the NUMBER OF BOLUSES is reprogrammed during a feeding (from HOLD mode), the pump will restart the bolus feeding and will deliver all the boluses as defined by the new setting.
Figure 13A. ADJUST BOLUS menu.
Figure 13B. Setting the Bolus Rate.
Figure 13C. Setting the Number of Boluses.
Figure 13D. Setting the volume per Bolus.
Figure 13E. Setting the time interval between starts of bolus.
Intermittent Mode Feeding – Bolus Max (Not Available in EZMODE) The BOLUS MAX option can be used to deliver a bolus at a high rate, similar to gravity feeding. Press4BOLUS MAX in the BOLUS RATE menu, Figure 13B, to program the BOLUS MAX mode. The feeding rate for BOLUS MAX is 999 mL/hr.
Intermittent Mode Flushing For INTERMITTENT feeding mode with a Feed & Flush Set, the opening menu, shown in Figure 7D, will show the ADJUST FLUSH option. Press4ADJUST FLUSH in Figure 7D to program the flushing parameters from the ADJUST FLUSH screen, Figure 14A.
e
KANGAROO Pump™
16
Enteral Feeding Pump
Figure 14A. ADJUST FLUSH menu, intermittent mode.
Figure 14B. Setting the Flush Volume.
Figure 14C. Setting the Flush Time Interval.
Note: The nature of two intermittent processes, periodic bolus feeding and periodic flushing, is such that occasional conflicts can arise where each process needs to be running at the same time. The flushing process is assigned precedence over the feeding process for Feed & Flush pump sets. “Flushing precedence” means that a flush cycle will run to its end even if a bolus delivery is scheduled to begin. It also means that a flush cycle will interrupt a bolus feeding cycle that is already in progress. The feeding cycle will pick up where it was interrupted after the flushing cycle is complete. It is therefore possible that the net feeding volume is reduced from what may be intended, over a given time period. Press4FLUSH VOLUME in Figure 14A to define the volume per flush cycle. Press the4buttons in Figure 14B to program the flush volume from 10 to 500 in increments of 1 mL. Press4ENTER (top button) to exit the menu. Press4FLUSH INTERVAL in Figure 14A to define the time interval between starts of flushing. Press the4buttons in Figure 14C to program the time interval from 1 to 24 hours in increments of 1 hour. Press4ENTER to exit the menu. The pump will automatically limit flushing settings which exceed pump capabilities to deliver. CAUTION: The KANGAROO ePump enteral feeding pump flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use care when programming the flush volume so that it matches the patient’s need. NOTE: The pump will automatically flush an additional 25 mL of water after completing the programmed delivery of formula, as long as at least one flush had occurred during formula delivery.
Run The KANGAROO ePump enteral feeding pump will only run when a Pump Set is properly loaded and the necessary feeding parameters have been programmed. For a Feed & Flush Set, the flushing parameters do not have to be set for the pump to begin operating. If flushing parameters are set as zeros, there will be no flushing. After the pump has been programmed, press4RUN to start operation. Figure 15A shows the RUNNING screen for continuous feeding with flushing. The status line will indicate RUNNING, and the droplet icon will move vertically on the screen. The green LED indicator will be illuminated to give a quick visual indication of positive status in a darkened room. The RUNNING screen shows the feed rate, amount fed, flushing data if a Feed & Flush Pump Set is used, and the remaining volume to be delivered (remaining VTBD). The remaining VTBD is shown next to the W icon. The RUN option can be selected from the following screens, if the proper parameters are programmed and the Pump Set is loaded: • SET LOADED screens (Figures 7A-7D) • ADJUST FEED screen, continuous mode (Figure 11A) • ADJUST FLUSH screens (Figure 12A, 14A) • HOLDING screens (Figure 18A)
e
KANGAROO Pump™
17
Enteral Feeding Pump