KERN
Scales General and Multi Purpose
MPS_M, MTS_M, MXS_M and MWS_M Scales Operating Instructions Ver 1.5 March 2012
Operating Instructions
46 Pages
Preview
Page 1
KERN & Sohn GmbH Ziegelei 1 D-72336 Balingen E-mail: [email protected]
Tel.: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Website: www.kern-sohn.com
Operating Instructions Personal weighing scale, Handrail scale, Bariatric scale, Wheelchair scale
KERN MPS_M / MTS_M / MXS_M / MWS_M Version 1.5 03/2012 GB
MPS / MTS / MXS / MWS-BA-e-1215
GB
KERN MPS 200K100M / PM KERN MTS 300K100M KERN MXS 300K100M KERN MWS 300K100M KERN MWS 400K100DM Version 1.5 03/2012
Operating Instructions Personal weighing scale without column / with column, Handrail scale, Bariatric scale, Wheelchair scale Contents 1
Technical data ... 4
2
Declaration of conformity ... 6
2.1
Explanation of graphical symbols ... 6
3
Basic directions (general information) ... 8
3.1
Use ... 8
3.1.1 3.1.2
Indication ... 8 Contraindication ... 8
3.2 3.3 3.4 3.5
Intended use ... 8 Inappropriate use ... 9 Guarantee ... 9 Monitoring the test substances ...10
4
Basic safety directions ... 11
4.1 4.2 4.3
Observing the directions included in the Operating Instructions ...11 Personnel training ...11 Avoidance of contamination ...11
5
Transport and storage ... 11
5.1 5.2
Check upon delivery ...11 Packaging / return transport...11
6
Unpacking, installation and starting ... 12
6.1 6.2 6.3
Place of installation, place of use ...12 Unpacking ...12 Installation and setting of scale ...13
6.3.1 6.3.2
Scope of delivery ... 23 Installation direction for a model with wall bracket ... 23
6.4 6.5
Mains socket ...24 Battery operation/ Rechargable battery operation (optional) ...24
6.5.1 6.5.2
Battery operation ... 25 Rechargeable battery operation (optional) ... 27
6.6 6.7
Initial start-up ...29 Menu overview of verified scales ...29
2
MPS / MTS / MXS / MWS-BA-e-1215
7
Operation ... 30
7.1
Operation elements - 20 keys of the terminal ...30
7.1.1 7.1.2 7.1.3
Display... 30 Display view... 30 Overview of keyboard ... 31
8
Using scale ... 32
8.1 8.2 8.3 8.4
Weighing ...32 Taring ...32 HOLD function ...33 Determination of Body Mass Index ...33
8.4.1
Classification of BMI values ... 34
8.5
PRE-TARE function ...34
8.5.1
PRE-TARE function with 5 memories ... 35
8.6
Printing function ...36
8.6.1
Parameters of RS232 interface ... 36
9
Error messages ... 37
10
Service, maintenance, disposal ... 37
10.1 10.2 10.3
Cleaning ...37 Service, maintenance ...38 Disposal ...38
11
Troubleshooting ... 39
12
Verification ... 40
12.1 12.2 12.3
Adjusting ...40 Adjustment key and seals ...42 Checking the scale settings concerning scale verification ...44
12.3.1
Menu overview in the service mode (adjustment switch in the adjustment position) ... 44
12.4
Validity period of verification (present status in Germany) ...46
MPS / MTS / MXS / MWS-BA-e-1215
3
1 Technical data MPS 200K100M/PM
MTS 300K100M 6-position one
MXS 300K100M
Weighing range (max.)
200 kg
300 kg
300kg
Minimal load (min.)
2 kg
2 kg
2 kg
Verification value (e)
100 g
100 g
100 g
KERN Display
Display Recommended calibration weight, (class)
LCD with 25 mm high digits 200 kg (M1)
300 kg (M1)
Signal rise time (typical)
2–3 s
Warm-up time
10 min
Operating temperature
+5°C …. +35°C
Storage temperature
-20°C … +60°C
Air humidity
300 kg (M1)
max. 80% (non-condensing) mains adapter 15 V / 300 mA (EN60601-1) operation with 6 x 1.5 V battery supply, AA type batteries
Power supply operation time 38 h Auto-Off function
Scale plate mm Total weight kg (net)
operation time 47 h
after 3 min without load change (possibility of setting)
Terminal (S x G x W) mm Scale ready for operation (W x D x H) mm
operation time 47 h
210 x 110 x 48 275 x 295 x 60 with column: 275 x 460 x 1010
550x550x1100
550x550x80
275 x 295 x 60
550x550
550x550
4.8
20.0
14.0
Verification according to 90/384/EEC
medical, class III
Medical device according to 93/42/EEC
class I with measuring function
Rechargeable battery operation (optional)
4
Loading time: 23 h; Loading time: 23 h; Loading time: 23 h; operating time: 35 h;
operating time: 45h;
operating time: 45h;
7,2 V / 2000 mA
7,2 V / 2000 mA
7,2 V / 2000 mA
MPS / MTS / MXS / MWS-BA-e-1215
KERN
MWS 300K100M
Display
MWS 400K100DM 6-position one
Weighing range (max.)
300 kg
300kg; 400kg
Minimal load (min.)
2 kg
2 kg
Verification value (e)
100 g
100 g; 200g
Display Recommended calibration weight, (class)
LCD with 25 mm high digits 300 kg (M1)
400 kg (M1)
Signal rise time (typical)
2 – 3 sec.
Warm-up time
10 min; 10 min
Operating temperature
+ 5° C …. + 35° C
Storage temperature
- 20°C … + 60°C
Air humidity
max. 80 % (non-condensing) mains adapter 15V / 300 mA ( EN60601-1) operation with 6 x 1.5 V battery supply, AA type batteries
Power supply
operation time 47 h Auto-Off function
operation time 47 h
after 3 min without load change (possibility of setting)
Terminal (B x T x H) mm
210 x 110 x 48
210 x 110 x 48
Scale ready for operation (W x D x H) mm
1155x800x60
1250x1060x68
900x740
1000x1000
26,5
39
Scale plate mm Total weight kg (net) Verification according to 90/384/EEC
medical, class III
Medical device according to 93/42/EEC Rechargeable battery operation (optional)
class I with measuring function Loading time: 23 h; operating time: 45 h;
MPS / MTS / MXS / MWS-BA-e-1215
7,2 V / 2000 mA
5
2 Declaration of conformity Declaration of conformity: refer to the separate document with serial number of the device
CE marking:
93/42/EEC
0297 year
M
0103
2.1
90/384/EEC The directive relating to non-automatic weighing instruments
Explanation of graphical symbols
M SN WY 070563
This EC verification sign means that this scale is compliant with the EC Directive 90/ 384/EEC concerning non-automatic weighing instruments. The scales marked with this sign are permitted for medical applications in European Community. Serial number marking of each device (located on the device and packaging) (example number here)
Production date marking of medical devices (year and month here as an example)
2008-03 „ Note: follow the directions included in the attached document or „Follow the Operating Instructions”
6
MPS / MTS / MXS / MWS-BA-e-1215
Marking of the medical device manufacturer with its address Kern & Sohn GmbH D – 72336 Balingen Ziegelei 1
„Electro-medical equipment” with usable part of B type
Protection class II device
Old devices do not belong to municipal waste. They may only be delivered to collection points of municipal waste.
+70°C
Temperature limitation with lower and upper limit (storage temperature on the packaging) (example temperature data)
-30°C
9 V DC / 500 mA
Data concerning the scale supply voltage with indication of polarity (polarity and example values)
MPS / MTS / MXS / MWS-BA-e-1215
7
3 Basic directions (general information) According to the Directive 90/384/EEC scales must be verified to the following application purposes. Article 1, paragraph 4 „Determination of weight in the course of medical practice i.e. weighing of patients for the purpose of health monitoring, diagnosis and medical treatment.” 3.1
Use
3.1.1 Indication - Determination of body weight in the course of medical practice. - Used as a „non-automatic scale” i.e. a person is to be placed carefully in the centre of the scale plate or in the suitable supporting device in the case of wheelchair or chair scales. In the case of baby scales, a baby is to be always laid down or seated on the scale pan. For wheel chair scales, a wheel chair together with the person sitting in it is pushed over the ramp into the center of the weighing plate and/or the electric wheel chair is moved electrically driven onto the weighing plate. The weighing value can be read off after a stable weighing value has been obtained. 3.1.2 Contraindication No contraindications. 3.2 Intended use Depending on the model, the scale is used to determine the weight of standing persons, sitting persons and/or persons sitting in the wheel chair at a wheel chair scale and the weight of lying babies in medical treatment rooms The scales are intended to diagnose, prevent and monitor diseases. The scales equipped with serial interface can only be connected to the equipment compliant with EN60601-1 standard. In the case of personal weighing scales, a person to be weighed is to be placed carefully in the centre of the scale plate and left at rest or in the case of chair scales, a person is to be sat in the centre of seat and left at rest. The whole wheel chair is to be pushed onto the weighing plate and/or the electric wheel chair is moved onto the weighing plate and the wheels have to be fixed for weighing. The weighing value can be read off after a stable weighing value has been obtained. The scale is designed for continuous operation.
The scale platform can only be walked on by people that can stand on it securely with both feet, or sit calmly (in the case of chair scale and wheel chair scale).
8
MPS / MTS / MXS / MWS-BA-e-1215
Scale platforms or footrests are equipped with anti-slip material which cannot be removed or covered when people are weighed. When scales equipped with height measure are used, pay attention that the top flap is always folded down after their use to avoid danger of injury. Before any use, the scale must be checked for correct condition by an authorised person. 3.3 Inappropriate use Do not use the scales for dynamic weighing. Do not leave a permanent load on the weighing plate. This can damage the measuring equipment. Be sure to avoid impact shock and overloading the scale in excess of the prescribed maximum load rating (max.), minus any possible tare weight that is already present. This could result in damage of the scale. Never operate the scale in hazardous locations. The series design is not explosionproof. Attention should be paid that flammable mixture may also be formed from anaesthesiological means that contain oxygen or laughing gas (nitrous oxide). Construction alterations may not be made to the scale. This can lead to incorrect weighing results, faults concerning safety regulations as well as to destruction of the scale. The scale may only be used in compliance with the described guidelines. Other areas of application/planned use must be approved by KERN in writing. 3.4
Guarantee
The guarantee shall become void in the event of the following: non-observation of our guidelines in the Operating Instructions, use outside the described applications, alteration to or opening the device, mechanical damage or damage caused by media, liquids, usual wear and tear, inappropriate erection or electric installation, overloading of the measuring equipment, scale falling down.
MPS / MTS / MXS / MWS-BA-e-1215
9
3.5 Monitoring the test substances The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the responsible user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN‟s home page (www.kernsohn.com) with regard to the monitoring of balance test substances and the test weights required for this. Test weights and balances can be adjusted quickly and at a reasonable price at KERN‟s accredited DKD (Deutsche Kalibrierdienst) calibration laboratory (return to national standard). In the case of the scales for weighing people provided with the scale to determine a body size, it is recommended to carry out the check of its measuring accuracy because determination of the human body size is always connected with a very large inaccuracy.
10
MPS / MTS / MXS / MWS-BA-e-1215
4 Basic safety directions 4.1
Observing the directions included in the Operating Instructions Carefully read this operation manual before setup and commissioning, even if you are already familiar with KERN balances.
4.2
Personnel training
The medical staff must apply and follow the operating instructions for proper use and care of the product. 4.3 Avoidance of contamination To avoid cross contamination (mycosis, ...), the scale plate is to be cleaned regularly. Recommendation: after each weighing which could result in potential contamination (e.g. when there is a direct skin contact during weighing).
5 Transport and storage 5.1 Check upon delivery Please check the packaging immediately upon delivery and the device during unpacking for any visible signs of external damage. 5.2
Packaging / return transport Keep all parts of the original packaging for a possibly required return. Only use original packaging for returning. Prior to dispatch disconnect all cables and remove loose/mobile parts. Reattach possibly supplied transport securing devices. Secure all parts such as the weighing platform, power unit etc. against shifting and damage.
MPS / MTS / MXS / MWS-BA-e-1215
11
6 Unpacking, installation and starting 6.1 Place of installation, place of use The scale is designed in such a way that reliable weighing results can be achieved under normal application conditions. By selecting the correct location for your scale, you will be able to work quickly and precisely. Therefore, please observe the following when choosing a place of installation: -
Place the scale on a firm, level surface;
-
Avoid extreme heat as well as temperature fluctuation caused by installing the scale next to a radiator or in the direct sunlight;
-
Protect the scale against direct draughts due to open windows and doors;
-
Avoid shaking during weighing;
-
Protect the scale against high humidity, vapours and dust;
-
Do not expose the device to extreme dampness for longer periods of time. Inadmissible bedewing (condensation of air moisture on the device) can occur if a cold device is taken into a significantly warmer environment. In this case, please keep the device for approx. 2 hours at room temperature after it has been disconnected from mains supply;
-
Avoid static charge build-up on the scale and people to be weighed;
-
Avoid contact with water.
Major display deviations (incorrect weighing results) are possible if electromagnetic fields occur (e.g. coming from mobile phones or radio equipment) as well as due to static charging and instable power supply. It is necessary then to change the scale location or remove disturbance source. 6.2 Unpacking Carefully remove individual scale parts or the whole scale from its packaging and position the scale in its intended working location. When the mains adapter is used, be careful not to cause the danger of falling over the power cable.
12
MPS / MTS / MXS / MWS-BA-e-1215
6.3
Installation and setting of scale
Personal weighing scale MPS with wall bracket:
Scope of delivery:
MPS / MTS / MXS / MWS-BA-e-1215
13
Personal weighing scale MPS-PM with column:
Scope of delivery:
Balance with display unit and tripod Mains adapter 4 screws
14
MPS / MTS / MXS / MWS-BA-e-1215
Assembly: Remove cap (1) Unscrew the screw (2)
Pull cable with Plug-in connection (3) through the supporting foot (4) and pull it out at the end (5)
Place supporting foot next to the balance
MPS / MTS / MXS / MWS-BA-e-1215
15
Introduce cable completely into the tripod tube (6)
Reinsert cap (1) Screw-in screw (2)
When screwing the screw ensure that the plug-in connector inside the tripod foot is not squeezed. Use the 4 screws to attach the tripod to the bottom of the balance
Level balance with foot screws until the air bubble of the water balance is in the prescribed circle. Check levelling regularly.
16
MPS / MTS / MXS / MWS-BA-e-1215
Adjust foot screw of the tripod in a way that the tripod has a safe base and is not loose.
MPS / MTS / MXS / MWS-BA-e-1215
17
Scale MTS with handrail:
Scope of delivery:
18
MPS / MTS / MXS / MWS-BA-e-1215
Assembly: Fasten 3 corner elements to the platform, using 4 screws each time.
Place the handrail on 3 corner elements and screw it.
Fasten the terminal holder to the handrail with 3 screws.
Remove the side rubber plugs at both sides of the display. Fasten the display to the holder with both handwheels. Adjust the display position with handwheels. MPS / MTS / MXS / MWS-BA-e-1215
19
Bariatric scale MXS:
Scope of delivery:
20
MPS / MTS / MXS / MWS-BA-e-1215