KODAK 9000C AND KODAK 9000C 3D Extraoral Imaging System
User Guide
Notice Congratulations on your purchase of the KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The User Guide for the KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System includes information on the cephalometric features. For the panoramic or the 3D features, see the KODAK 9000 and KODAK 9000 3D Extraoral Imaging System (SM710) User Guide. We recommend that you thoroughly familiarize yourself with this Guide in order to make the most effective use of your system. The information contained in this Guide may be subject to modification without notice, justification or notification to the persons concerned. No part of this Guide may be reproduced without the express permission of Carestream Health, Inc. The US Federal law restricts this device to sale by or on the order of a physician. This document is originally written in English. Manual Name: KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide Part Number: SM713 Revision Number: 01 Print Date: 04/2008 The brand names and logos reproduced in this Guide are copyright. KODAK is a trademark of KODAK used under License. KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System, complies with Directive 93/42/CEE relating to medical equipment.
Contents 1-About This Guide Conventions in this Guide... 1–1 Note to the User... 1–2 Warning and Safety Instructions... 1–2 Marking and Labeling Symbols... 1–4 2-KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW General Overview... 2–1 Mobile Components... 2–2 General Functional Components... 2–3 Digital Sensor Locations... 2–4 Laser Locations... 2–5 Label Locations... 2–6 Control Panel Overview... 2–7 X-Ray Remote Control Overview... 2–8 Positioning Accessories and Replacement Parts... 2–9 3-IMAGING SOFTWARE OVERVIEW PC System Requirements... 3–1 General Software Overview... 3–2 KODAK Dental Imaging Software... 3–2 Cephalometric Acquisition Interface Module... 3–2 Cephalometric Acquisition Interface Module... 3–3 Cephalometric Acquisition Window Overview... 3–3 Cephalometric Program Pane... 3–5 Cephalometric Patient Pane... 3–6 Cephalometric Parameter Pane... 3–7 4-GETTING STARTED Switching On the Unit... 4–1 Starting the Imaging Software... 4–2 Creating a Patient Record... 4–2 Accessing the Cephalometric Acquisition Window... 4–3 5-ACQUIRING CEPHALOMETRIC IMAGES Acquiring a Lateral Image... 5–1 Preparing the Unit and Setting the Acquisition Parameters... 5–1 Preparing and Positioning the Patient... 5–2 Acquiring a Frontal AP or PA Image... 5–5 Preparing the Unit and Setting the Acquisition Parameters... 5–5 Preparing and Positioning the Patient... 5–6 Acquiring an Oblique Image... 5–9
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
iii
Contents
Preparing the Unit and Setting the Acquisition Parameters...5–9 Preparing and Positioning the Patient...5–10 Acquiring a Submento-Vertex Image...5–12 Preparing the Unit and Setting the Acquisition Parameters...5–12 Preparing and Positioning the Patient...5–13 Acquiring a Carpus Image...5–16 Preparing the Unit and Setting the Acquisition Parameters...5–16 Preparing and Positioning the Patient...5–17 X-Ray Dose Emission Information...5–19 6-MAINTENANCE Daily...6–1 Monthly...6–1 Annually...6–1 7-TROUBLESHOOTING Quick Troubleshooting...7–1 8-TECHNICAL SPECIFICATIONS Compliance with International Standards...8–1 Manufactured by:...8–1 Compliance with International Regulations...8–1 Compliance with European and International Standards...8–1 Classification in Accordance with EN/IEC 60601-1...8–2 Classification in Accordance with EN/IEC 60601-1-2...8–2 Ambient Operating Conditions...8–2 Storage Conditions...8–2 Transport Conditions...8–2 Cephalometric Unit Technical Specifications...8–3 Unit Electronic Specifications...8–4 X-Ray Generator Technical Specifications...8–5
iv
Chapter 1
About This Guide Conventions in this Guide The following special messages emphasize information or indicate potential risk to personnel or equipment:
WARNING Warns you to avoid injury to yourself or others by following the safety instructions precisely.
CAUTION Alerts you to a condition that might cause serious damage.
IMPORTANT Alerts you to a condition that might cause problems.
NOTE Emphasizes important information.
TIP Provides extra information and hints.
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
1–1
Note to the User
Note to the User WARNING X-rays can be harmful and dangerous if not used properly. The instructions and warnings contained in this guide must be therefore carefully followed. As a manufacturer of radiology units that conform to stringent radiological protection standards in force throughout the world, we guarantee a maximum degree of protection against radiation hazards. Nonetheless, you are handling a radiology unit specially designed to emit x-ray doses in order to carry out a medical diagnosis. The room in which your radiology unit is to be installed must comply with all official regulations applicable to protection against radiation. You must install your radiology unit in a room protected against x-ray emission. This room must reduce to at least 12db the frequency interferences of the 30MHz to 1GHz band. Your local representative will assist you in the initial use of your radiology unit and will supply any relevant information you may require. To use and operate your panoramic unit and your digital imaging software you must follow the instructions contained in this guide.
Warning and Safety Instructions When operating the KODAK 9000C and KODAK 9000C 3D unit, observe the following warning and safety instructions:
DANGER OF ELECTRIC SHOCK This is an electrical unit. DO NOT expose it to water spray. Such action may cause an electric shock or a malfunction of the unit.
LASER WARNING
LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT Maximum laser output:1mW,650nm IEC60825-1:1993+A1:1997+A2:2001
1–2
About This Guide
For maximum safety, advise the patient not to look at the beam. Before turning on the beams, lower the Frankfurt plane beam to the lowest level. While making adjustments, ensure that the beam is not directed into the patient's eyes.
Warning and Safety Instructions
WARNINGS:
!
!
!
!
! !
! ! ! ! ! ! ! ! ! ! ! !
You are responsible for the operation and maintenance of this unit. Only legally qualified persons can operate this unit. They MUST have training to use the radiological equipment. DO NOT open the cover of the unit.When necessary, have a trained authorized service technician carry out inspection and maintenance operations. Install this unit in an x-ray room that complies with current installation standards. From this location, you must be able to maintain visual or audio communication with the patient and be able to access the Acquisition interface module during exposure. This unit must be permanently connected to the ground with a fixed power supply cable. DO NOT place the PC and the peripheral equipment connected to it in the immediate vicinity of the patient in the unit. Leave at least 1.5 m distance between the patient and the unit. The PC and the peripheral equipment must conform to the IEC60950 standard. See your computer installation guide for details of the data processing system, PC and screen. Leave a sufficient amount of clear space around the CPU to ensure that it is properly ventilated. To obtain maximum image quality and visual comfort, position the screen to avoid direct light reflections from internal or external lighting. DO NOT operate the unit if there is the threat of an earthquake. Following an earthquake, ensure that the unit is operating satisfactorily before using it again. Failure to observe this precaution may expose patients to hazards. X-ray equipment is hazardous to patients and the operator if you do not observe the exposure safety factors and operating instructions. DO NOT place objects within the field of operation of the unit. The patient should wear a protective lead-lined shoulder apron, unless other Radiation Protection Protocols apply locally. Disinfect any parts of the unit that come into contact with the patient and the operator after each patient has been exposed to x-rays. While adjusting the height of the unit, ensure that the patient is kept clear of the mechanism. When the unit is not in use, ensure that the ON/OFF switch is set to OFF (O). If the unit develops a fault, switch it to off (O), display an “Unserviceable” notice and contact a service technician. To dispose of the unit or its components, contact a service technician. Ask the patient to refrain from moving during the entire period of exposure. Ask the patient to remain still until the unit arm has stopped moving and the RESET movement has completed. DO NOT use this unit in conjunction with oxygen-rich environments. This unit is not intended for use with flammable anesthetics or flammable agents. DO NOT hang from the cephalostat.
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
1–3
Marking and Labeling Symbols
Marking and Labeling Symbols
Type B device symbol complying with the IEC 60601-1 standard
In the EEC, this symbol indicates: DO NOT discard this product in a trash receptacle; use an appropriate recovery and recycling facility. Contact your local sales representative for additional information on the collection and recovery programs available for this product WARNING and IONIZING RADIATION symbols warn you about radiation dangers.
LASER WARNING
LASER RADIATION DO NOT STARE INTO BEAM CLASS 2 LASER PRODUCT Maximum laser output:1mW,650nm IEC60825-1:1993+A1:1997+A2:2001
1–4
About This Guide
Laser radiation. DO NOT stare into the beam. Class 2 laser product. Maximum output power: 1mW, 650 nm (IEC 60825-1 standard) This unit emits laser radiation.
Chapter 2
KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW The KODAK 9000C and KODAK 9000C 3D digital imaging unit is compliant with the requirements of the EEC and international medical standards. The KODAK 9000C and KODAK 9000C 3D unit has been designed to carry out the following radiological examinations: ! ! ! ! !
Lateral Frontal (PA or AP) Oblique Submento-vertex Carpus
General Overview The KODAK 9000C unit is composed of the following functional components: ! ! ! ! ! ! ! ! ! ! ! ! ! !
The rotative arm The fixed arm with a control panel The panoramic digital sensor The x-ray source assembly The x-ray remote control The chin rest base The panoramic chin rest and bite block The temple supports The hand grips The cephalostat arm The cephalostat head The head clamps and ear cones The nasion support The acquisition software (see “Imaging Software Overview”)
The KODAK 9000C 3D unit is composed of the following functional components: ! ! ! !
The rotative arm The fixed arm with a control panel The panoramic digital sensor The 3D digital sensor KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
2–1
General Overview
! ! ! ! ! ! ! ! ! ! ! !
The x-ray source assembly The x-ray remote control The chin rest base The panoramic chin rest and bite block The 3D chin rest and bite block The temple supports The hand grips The cephalostat arm The cephalostat head The head clamps and ear cones The nasion support The acquisition software (see “Imaging Software Overview”)
The following figures illustrate the general overview of the KODAK 9000C and KODAK 9000C 3D unit.
Mobile Components Figure 2-1 illustrates the up and down movement of the KODAK 9000C and KODAK 9000C 3D mobile component and the 180° rotation of the rotative arm.
Figure 2–1
2–2
KODAK 9000C and KODAK 9000C 3D Unit Mobile Components
KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW
General Overview
General Functional Components Figure 2-2 illustrates the general functional components of the KODAK 9000C and KODAK 9000C 3D. Figure 2–2
KODAK 9000C and KODAK 9000C 3D Unit Functional Components
5b 11 1 5
12
9 7 6
8
14 5a 13
2
5 4
15
3
10
1
ON/OFF button
7
Sensor
2
Unit fixed arm
8
X-ray source assembly
3
Control panel
9
Unit rotative arm
4
Hand Grips
10
X-Ray remote control
5
Chin rest base
11
Chephalostat arm
5a
Panoramic chin rest
12
Cephalostat head
5b
3D chin rest and bite block
13
Head clamps and ear cones
6
Temple supports
14
Nasion support
15
PC hosting the imaging and the acquisition software
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
2–3
General Overview
Digital Sensor Locations Figure 2-3 illustrates the locations of the panoramic, 3D and cephalometric digital sensors of the KODAK 9000C and KODAK 9000C 3D.
IMPORTANT * When the unit is ONLY panoramic, the 3D option is not available. Therefore, the location of the digital sensor indicates that it is ready for the 3D option.
Figure 2–3
KODAK 9000C and KODAK 9000C 3D Unit Panoramic, 3D and Cephalometric Digital Sensor Locations
Digi ta l 3 D
*
2–4
KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW
General Overview
Laser Locations Figure 2-4 illustrates the location of the lasers of the KODAK 9000C and KODAK 9000C 3D.
Figure 2–4
KODAK 9000C and KODAK 9000C 3D Unit Laser Positions 2
3 4 B
5 A 1
B
A
3 5
4
1
2
1
2
1
3D central positioning laser beam
2
Mid-sagittal positioning laser beam
3
Horizontal positioning laser beam
4
3D Field of View (FoV) positioning laser beam
5
Cephalometric Frankfurt positioning laser beam
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
2–5
General Overview
Label Locations Figure 2-5 illustrates the location of all the labels of the KODAK 9000C and KODAK 9000C 3D. The label locations of the KODAK 9000C are represented with *.
IMPORTANT The labels presented are ONLY for the KODAK 9000C 3D option. The “*Panoramic” presents the labels and their locations for the KODAK 9000C.
Figure 2–5
KODAK 9000C and KODAK 9000C 3D Unit Label Locations
* 2008/04/23 K9000 C 3D XPNK004 0.90A
2008/04/23
*
X-Ray Source Assembly :CU 227
330 2,5 mm AI 2,5 mmAI
CEI OPX 105 6448
* Panoramic
2008/04/23 K9000 C XPNK004 0.90A
2–6
KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW
Type : CG932
Control Panel Overview
Control Panel Overview The control panel is an alphanumeric, digital soft touch console. It allows the operator to control certain unit functions. It also displays the operating parameters and error messages.
Figure 2–6
Unit Control Panel 8
kV
mA
1
2
3
S
mem
9
10
11
4
5
6
1
Height Adjustment button: Adjusts the height of the unit to the height of the patient.
2
3D Head Adjustment button: Adjusts the patient head to the x-ray beams.
3
3D Adjustment button: Adjusts the unit rotative arm movements to correctly position the patient for 3D acquisition.
4
Laser Beam button: Activates the beams to correctly position the patient.
5
3D Position Verification button: Positions the unit rotative arm at the selected or memorized 3D position.
6
Reset button: Resets the unit arm to the initial position to enable the patient to enter and exit the Unit.
7
3D Memorization button: Memorizes the 3D current positioning parameter settings that override the default parameters.
8
Display Screen: Displays the current acquisition parameters and the error messages.
9
Ready Indicator LED: Green, indicates the unit is ready for acquisition.
10
X-Ray Emission LED: Yellow, indicates the x-rays are being emitted.
11
System Status LED: Red, indicates the error alerts.
7
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
2–7
X-Ray Remote Control Overview
X-Ray Remote Control Overview The x- ray remote control enables you to launch a radiological image acquisition via the exposure button from outside the x-ray room. You must press and hold the exposure button until the end of acquisition. Premature release of the exposure button interrupts the acquisition.
Figure 2–7
X-Ray Remote Control
1
1
2–8
Exposure button: launches image acquisition.
KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW
Positioning Accessories and Replacement Parts
Positioning Accessories and Replacement Parts The following accessories are used when positioning a patient. They are delivered with the KODAK 9000C and KODAK 9000C 3D digital imaging unit.
Table 2–1
Cephalometric Positioning Accessories and Replacement Parts Accessory
Description Head clamps with ear cones x2
Nasion support
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
2–9
Positioning Accessories and Replacement Parts
2–10
KODAK 9000C AND KODAK 9000C 3D UNIT OVERVIEW
Chapter 3
IMAGING SOFTWARE OVERVIEW PC System Requirements This section specifies the minimum PC system requirements for KODAK 9000C and KODAK 9000C 3D extraoral imaging system software. IMPORTANT It is MANDATORY to check that the PC system configuration is compatible with the PC system requirements for the KODAK 9000C and KODAK 9000C 3D software. If necessary you MUST update your PC system configuration. KODAK 9000C and KODAK 9000C 3D MUST be connected to the PC via a point-to-point Ethernet link and not via a LAN. DO NOT place the PC and the peripheral equipment connected to it in the immediate vicinity of the patient in the unit. Leave at least 1.5 m distance from the unit. The PC and the peripheral equipment must conform to the IEC 60950 standard. IMPORTANT For KODAK 9000C, you can use the “Viewing” graphic board configuration for acquisition. Table 3–1
Minimum PC System Requirements Viewing
Acquisition *
CPU
2 GHz Intel Duo Core
3 GHz Intel Pentium 4
RAM
2 GB
2GB
Item
Hard disk drive
! !
1.2 GB for software installation
Comments
RAM has a major impact on system performance.
1.2 GB for software installation
80 GB free space to use the software
Graphic board
Nvidia / ATI based board supporting Nvidia board on PCI Express video The video RAM has major impact Open Glide 1.2 with 256 MB of video bus, minimum GeForce 8800 GT on system performance. RAM on AGP x8 video bus (example: 512 MB of video RAM, or Quadro Nvidia GeForce 6800 GT) FX 4600 768 MB of video RAM
Monitor
! ! !
1 monitor
! !
Windows 2000 SP4
!
Windows Vista 32 bits
Operating system
17" or larger 1024 x 768 minimum screen resolution - 32 bits color mode
Windows XP Home / Pro edition SP2
! ! !
1 monitor 17" 1280 x 880 minimum screen resolution
Your monitor is a vital component in displaying quality images. Low-quality screens will prevent you from proper diagnoses and treatment.
Windows XP Home / Pro edition SP2
KODAK 9000C and KODAK 9000C 3D Extraoral Imaging System User Guide (SM713)_Ed 01
3–1
General Software Overview
Table 3–1
Minimum PC System Requirements (Continued) Item
Acquisition *
Viewing
Ethernet interface
1 Ethernet interface
Comments
2 Ethernet interfaces (100Mbits)
CD/DVD drive
A CD-ROM drive is required to install A CD-ROM drive is required to the product. install the product.
Backup Media
Removable/portable, external hard disk drive
Removable/portable, external hard disk drive.
We strongly recommend a daily backup of x-ray images and patient records.
* For the KODAK 9000C, panoramic and cephalometric acquisition, you need only the “Viewing” graphic board configuration.
General Software Overview The KODAK 9000C and KODAK 9000C 3D extraoral imaging system operates with the following software: ! !
KODAK dental imaging software Acquisition interface module
KODAK Dental Imaging Software The KODAK dental imaging software is a user-friendly working interface that was designed and developed specifically for radiological diagnosis. It is the common imaging platform for all our digital systems for dentistry. The KODAK dental imaging software has the following features: ! ! !
Patient record management using Patient Window features Extraoral and intraoral image management using Imaging Window features 3D image management using 3D Imaging Window features
NOTE For a complete information on the KODAK Dental Imaging Software, see the “KODAK Dental Imaging Software, Quick Start Guide”.
Cephalometric Acquisition Interface Module The cephalometric acquisition interface module is a user-friendly working interface that was designed and developed specifically for KODAK 9000C and KODAK 9000C 3D extraoral imaging system.
3–2
IMAGING SOFTWARE OVERVIEW