KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System
User Guide
Notice Congratulations on your purchase of the KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System. The KODAK 8000C unit is the KODAK 8000 equipped with the cephalostat unit. Thank you for your confidence in our products and we will do all in our power to ensure your complete satisfaction. The User Guide for the KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System includes information on the cephalometric features. For the panoramic features, see the KODAK 8000 Extraoral Imaging System (SM722) User Guide. We recommend that you thoroughly familiarize yourself with this Guide in order to make the most effective use of your system. WARNING: We recommend that you consult the “Safety, Regulatory and the Technical Specification User Guide” before using the KODAK 8000C Extraoral Imaging Systems.
The information contained in this Guide may be subject to modification without notice, justification or notification to the persons concerned. No part of this Guide may be reproduced without the express permission of Carestream Health, Inc. The US Federal law restricts this device to sale by or on the order of a physician. This document is originally written in English. Manual Name: KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide Part Number: SM735 Revision Number: 02 Print Date: 03/2010 The brand names and logos reproduced in this Guide are copyright. KODAK is a trademark of KODAK used under License. KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System, complies with Directive 93/42/CEE relating to medical equipment. .
0086
Manufacturer Care stream Hea lth, Inc. 150 Verona Street Roche ster NY 14 608
Authorized Representative in the European Community
EC REP TROPHY 4, Rue F. Pelloutier, Croissy-Beaubourg 77435 Marne la Vallée Cedex 2, France
Contents 1-About This Guide Conventions in this Guide... 1–1 2-KODAK 8000C UNIT OVERVIEW General Overview... 2–1 Mobile Components... 2–2 General Functional Components... 2–3 Digital Sensor Locations... 2–4 Laser Locations... 2–5 Control Panel... 2–6 X-Ray Remote Control Overview... 2–7 Positioning Accessories and Replacement Parts... 2–8 3-IMAGING SOFTWARE OVERVIEW Computer System Requirements... 3–1 General Software Overview... 3–2 KODAK Dental Imaging Software... 3–2 Cephalometric Acquisition Interface Module... 3–2 Cephalometric Acquisition Interface Module... 3–3 Cephalometric Acquisition Window Overview... 3–3 Cephalometric Program Pane... 3–5 Cephalometric Patient Pane... 3–6 Cephalometric Parameter Pane... 3–7 4-GETTING STARTED Switching on the Unit... 4–1 Starting the Imaging Software... 4–2 Creating a Patient Record... 4–2 Accessing the Cephalometric Acquisition Window... 4–3 5-ACQUIRING CEPHALOMETRIC IMAGES Acquiring a Lateral Image... 5–1 Preparing the Unit and Setting the Acquisition Parameters... 5–1 Acquiring a Frontal AP or PA Image... 5–5 Preparing the Unit and Setting the Acquisition Parameters... 5–5 Preparing and Positioning the Patient... 5–6 Acquiring an Oblique Image... 5–9 Preparing the Unit and Setting the Acquisition Parameters... 5–9 Preparing and Positioning the Patient... 5–10 Acquiring a Submento-Vertex Image... 5–12 Preparing the Unit and Setting the Acquisition Parameters... 5–12 Preparing and Positioning the Patient... 5–13
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
iii
Contents
Acquiring a Carpus Image... 5–16 Preparing the Unit and Setting the Acquisition Parameters... 5–16 Preparing and Positioning the Patient... 5–17 X-Ray Dose Emission Information... 5–19 6-MAINTENANCE Daily... 6–1 Monthly... 6–1 Annually... 6–1 7-TROUBLESHOOTING Quick Troubleshooting... 7–1
iv
Chapter 1
About This Guide Conventions in this Guide The following special messages emphasize information or indicate potential risk to personnel or equipment:
WARNING Warns you to avoid injury to yourself or others by following the safety instructions precisely.
CAUTION Alerts you to a condition that might cause serious damage.
IMPORTANT Alerts you to a condition that might cause problems.
NOTE Emphasizes important information.
TIP Provides extra information and hints.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
1–1
Conventions in this Guide
1–2
About This Guide
Chapter 2
KODAK 8000C UNIT OVERVIEW The KODAK 8000C digital panoramic and cephalometric unit is designed to carry out the following radiological examinations: • • • • • • • •
Panoramic Maxillary Sinus Temporomandibular Joints (TMJ) Lateral cephalometric Frontal (PA or AP) cephalometric Oblique cephalometric Submento-vertex cephalometric Carpus cephalometric
General Overview The KODAK 8000C digital panoramic and cephalometric unit is composed of the following functional components: • • • • • • • • • • • • • • • •
The unit head that contains all the electronic control The rotative arm The fixed arm with a control panel The panoramic digital sensor The x-ray source assembly The x-ray remote control The chin rest and bite block The chin rest base The panoramic chin rest and bite block The temple supports The hand grips The cephalostat arm The cephalostat head The head clamps and ear cones The nasion support The acquisition software (see “Imaging Software Overview”)
The following figures illustrate the general overview of the KODAK 8000C digital panoramic and cephalometric units.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
2–1
General Overview
IMPORTANT The Cephalostat can be positioned either on the right or the left side of the KODAK 8000 unit.
Mobile Components Figure 2-1 illustrates the up and down movement of the KODAK 8000C digital panoramic and cephalometric units mobile component and the rotation of the rotative arm. Figure 2–1
2–2
KODAK 8000C UNIT OVERVIEW
KODAK 8000 and KODAK 8000C Units Mobile Components
General Overview
General Functional Components Figure 2-2 illustrates the general functional components of the KODAK 8000C digital panoramic and cephalometric units. Figure 2–2
KODAK 8000 and KODAK 8000C Units Functional Components
14
16 1
15 17 16
11 8 3
9
12 7
10 6 4 2
13
5
1
ON/OFF button
10
Collimator selector
2
Chin rest and bite block
11
Unit rotative arm
3
Temple supports
12
X-Ray remote control
4
Temple supports control knob
13
PC hosting the imaging and the acquisition software
5
Hand Grips
14
Cephalostat arm
6
Control panel
15
Cephalostat head
7
Height adjustment buttons
16
Head clamps and ear cones
8
Panoramic digital sensor
17
Nasion support
9
X-Ray source assembly
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
2–3
General Overview
Digital Sensor Locations Figure 2-3 illustrates the locations of the digital panoramic and digital cephalometric sensors of the KODAK 8000C digital panoramic and cephalometric units.
Figure 2–3
KODAK 8000 and KODAK 8000C Units Digital Sensor Locations
system
2–4
KODAK 8000C UNIT OVERVIEW
General Overview
Laser Locations Figure 2-4 illustrates the location of the lasers of the KODAK 8000C digital panoramic and cephalometric units. Figure 2–4
KODAK 8000 and KODAK 8000C Units Laser Beam Locations
4
3
2
1
1
Mid-sagittal plane positioning laser beam
2
Frankfort plane positioning laser beam
3
Canine plane positioning laser beam
4
Cephalometric Frankfort plane positioning laser beam
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
2–5
Control Panel
Control Panel The control panel is an alphanumeric, digital soft touch console. It allows the operator to control certain unit functions. It also displays the operating parameters and error messages.
Figure 2–5
Unit Control Panel
8
Digital Panoramic and Cephalometric System
kV
1
2–6
mA
S
2
3
4
1
X-Ray emission LED: Yellow, indicates the x-rays are being emitted.
2
Display Screen: Displays the current acquisition parameters and the error messages.
3
Reset button: Resets the unit arm to the initial position to enable the patient to enter and exit the unit.
4
Laser beam button: Activates the laser positioning beams to correctly position the patient.
KODAK 8000C UNIT OVERVIEW
X-Ray Remote Control Overview
X-Ray Remote Control Overview The x- ray remote control enables you to launch a radiological image acquisition via the exposure button from outside the x-ray room. You must press and hold the exposure button until the end of acquisition. Premature release of the exposure button interrupts the acquisition.
Figure 2–6
X-Ray Remote Control
1
1
Exposure button: launches image acquisition.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
2–7
Positioning Accessories and Replacement Parts
Positioning Accessories and Replacement Parts The following accessories are used when positioning a patient. They are delivered with the KODAK 8000 digital panoramic and KODAK 8000C digital panoramic and cephalometric unit. Table 2-1 and Table 2-2 list the panoramic and cephalometric positioning accessories.
Table 2–1
Panoramic Positioning Accessories and Replacement Parts Accessory
Description
Panoramic chin rest and TMJ x2
Maxillary sinus chin rest
TMJ x2 and TMJ x4 nose rest
Standard bite block
Bite block for edentulous patients
A set of right and left temple supports x2
Single use sheaths for bite blocks (500 pcs box)
2–8
KODAK 8000C UNIT OVERVIEW
Positioning Accessories and Replacement Parts
Table 2–2
Cephalometric Positioning Accessories and Replacement Parts Accessory
Description Head clamps with ear cones x2
Nasion support
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
2–9
Positioning Accessories and Replacement Parts
2–10
KODAK 8000C UNIT OVERVIEW
Chapter 3
IMAGING SOFTWARE OVERVIEW Computer System Requirements This section specifies the minimum computer system requirements for KODAK 8000C digital panoramic and cephalometric extraoral imaging system software. IMPORTANT It is MANDATORY to check that the computer system configuration is compatible with the computer system requirements for the KODAK 8000C software. If necessary you MUST update your computer system configuration. KODAK 8000C MUST be connected to the computer via a point-to-point Ethernet link and not via a LAN. DO NOT place the computer and the peripheral equipment connected to it in the immediate vicinity of the patient in the unit. Leave at least 1.5 m distance from the unit. The computer and the peripheral equipment must conform to the IEC 60950 standard. Table 3–1
Minimum Computer System Requirements Item
Viewing
Acquisition
CPU
2 GHz Intel Duo Core
2 GHz Intel Duo Core
RAM
2 GB
2GB
•
Hard disk drive
•
1.2 GB for software installation
Comments
RAM has a major impact on system performance.
1.2 GB for software installation
80 GB free space to use the software
Graphic board
Nvidia / ATI based board supporting Nvidia / ATI based board The video RAM has major Open GL 1.2 with 256 MB of video supporting Open Glide 1.2 with impact on system performance. RAM on AGP x8 video bus (for 256 MB of video RAM on AGP x8 example: Nvidia GeForce 6800 GT) video bus (for example: Nvidia GeForce 6800 GT)
Monitor
• • •
Operating system (32 bits)
• • •
1 monitor 17" or larger 1024 x 768 minimum screen resolution - 32 bits color mode
• • •
Windows 2000 SP4
•
Windows XP Home / Pro edition SP2
•
1 monitor 17" 1280 x 880 minimum screen resolution
Your monitor is a vital component in displaying quality images. Low-quality screens will prevent you from proper diagnoses and treatment.
Windows XP Home / Pro edition SP2 Windows Vista
Windows Vista
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
3–1
General Software Overview
Table 3–1
Minimum Computer System Requirements (Continued) Item
Viewing
Acquisition
Comments
1 Ethernet interface
1 Ethernet interfaces (100Mbits)
If there is a LAN connection a second Ethernet interface is required.
CD/DVD drive
A DVD-BURNER drive is required.
A DVD-BURNER drive is required.
Backup Media
Removable/portable, external hard disk drive
Removable/portable, external hard We strongly recommend a daily disk drive. backup of x-ray images and patient records.
Ethernet interface
General Software Overview The KODAK 8000C digital panoramic and cephalometric extraoral imaging system operates with the following software: • •
KODAK dental imaging software Cephalometric acquisition interface module
KODAK Dental Imaging Software The KODAK dental imaging software is a user-friendly working interface that was designed and developed specifically for radiological diagnosis. It is the common imaging platform for all our digital systems for dentistry. The KODAK dental imaging software has the following features: • •
Patient record management using Patient Window features Extraoral and intraoral image management using Imaging Window features
NOTE For a complete information on the KODAK Dental Imaging Software, see the “KODAK Dental Imaging Software, Quick Start Guide”.
Cephalometric Acquisition Interface Module The cephalometric acquisition interface module is a user-friendly working interface that was designed and developed specifically for KODAK 8000C digital panoramic and cepahlometric extraoral imaging system.
3–2
IMAGING SOFTWARE OVERVIEW
Cephalometric Acquisition Interface Module
Cephalometric Acquisition Interface Module Cephalometric Acquisition Window Overview The Cephalometric Acquisition Window is the main cephalometric interface with the KODAK 8000C cephalometric extraoral imaging system that provides you with imaging acquisition functions. Figure 3–1
1
Cephalometric Acquisition Window
Information button: • About: Identifies the Software and the Firmware versions
• •
Reset of the Values: Resets to the manufacturing parameter settings Memorize settings: Memorizes the user preference settings for each patient type (kV, mA and seconds)
2
Preview Screen: Displays the acquired image in real time.
3
Selected Parameter Display: Displays the current acquisition parameter settings.
4
System Status Screen: Displays various alert or warning messages originating from the unit.
5
X-Rays ON / OFF button: Turns off the x-ray emissions to demonstrate the acquisition process for the patient.
6
Generator Cooling indicator: Indicates the automatic cooling time (mm:ss) required for the generator to reach 0 for a new acquisition.
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System User Guide (SM735)_Ed 02
3–3
Cephalometric Acquisition Interface Module
7
Positioning laser button: Activates the laser positionning beams to correctly position the patient.
8
Reset button: Resets the rotative arm in the start position.
9
Ready Indicator LED • Green indicates the unit is ready to start acquisition.
•
Black indicates the unit is not ready to start acquisition.
10
Exit button: Closes the Acquisition Window.
11
X-Ray Emission indicator: Yellow, indicates the x-ray emission status.
12
Selector Button: Selects different acquisition setting options. • Click Program to select examination type options.
• •
Click Patient to select patient type parameters. Click Parameters to select exposure parameter options.
The Selector button enables you to access the following 3 panes: • • •
3–4
Program pane: Examination type options Patient pane: Patient type parameter options Parameters pane: Exposure parameter options
IMAGING SOFTWARE OVERVIEW