Laerdal Medical
Moyo Fetal Heart Rate Monitor User Guide Rev E
User Guide
56 Pages
Preview
Page 1
Intended Use The "Moyo - Fetal Heart Rate Monitor" by Laerdal Global Health is intended for healthcare professionals to monitor fetal heart rate during labor. Moyo uses a Doppler ultrasound sensor to measure and analyze the fetal heart rate. The calculated fetal heart rate is presented on the display together with an audible (doppler) sound. Moyo provides an alarm in the case of prolonged abnormal fetal heart rate. Intended medical indication is to screen for or monitor fetal distress on pregnant women - particularly in the setting of labor. There are no known contraindications to the use of Moyo. –––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
Illustrations in this User Guide may vary from actual product.
2
Contents 1 Overview 2 Cautions & Warnings 3 Get Started 4 Alarms 5 Cleaning & Charging 6 Features in Detail 7 Troubleshooting 8 Technical Description
3
4 5 7 15 21 29 43 47
Overview Items included: 1 Moyo main unit with ultrasound sensor, 1 neck strap, 3 reusable abdominal transducer belts, 1 charger, 1 charger cable and this User Guide.
Speaker Status Light
Display Maternal heart rate pads (electrodes) * Power on/off button Speaker on/off button History graph button
Ultrasound transducer (sensor) *
Cable
* Patient applied part (intended for patient contact)
4
Cautions & Warnings Please ensure that following precautions are taken to ensure that Moyo functions properly during use.
This product is intended for use by authorized healthcare professionals only. Read this User Guide and become familiar with the operation of the device prior to use. The device is intended for monitoring only one fetus at a time. In case of monitoring of multiple fetuses, alternative devices/methods should be used. In case the device displays erratic or unreliable FHR, try to reposition the ultrasound transducer or use alternative monitoring devices/methods. Excessive handling and movement of the product may cause lost or erroneous heart rate readings. Do not connect the charger to Moyo while in use with a patient. The device shall be turned off and cleaned/disinfected between patient uses. Additional equipment connected to medical electrical equipment must comply with the respective IEC or ISO standards (e.g. IEC 60950 for data processing equipment). Furthermore all configurations shall comply with the requirements for medical electrical systems (see IEC 60601-1-1 or clause 16 of the 3rd ed. of IEC 60601-1, respectively). Anybody connecting additional equipment to medical electrical equipment configures a medical system and is therefore responsible that the system complies with the requirements for medical electrical systems. Attention is drawn to the fact that local laws take priority over the above mentioned requirements. If in doubt, consult your local representative or the technical service department. Do not modify this equipment without authorization of the manufacturer. If this equipment is modified, appropriate inspection and testing must be conducted to ensure continued safe operation. Take special precautions regarding electromagnetic compatibility (EMC) when using medical electrical equipment.You must operate your monitoring equipment according to the EMC information provided: • Measurements can be affected by ESD, portable and mobile RF communications equipment. • Use of high frequency electrosurgical equipment in the vicinity of this equipment may produce interference and cause incorrect measurements. Do not immerse this product in liquid. Use of Moyo does not replace clinical supervision. 5
3 Get Started To start operating Moyo, you need to learn the basic functions first.
7
Get Started To get started, this manual will take you through a few simple scenarios where you will learn the main functions of Moyo.
What you need: MOYO
Ultrasound gel
Abdominal transducer belt
Only water-based ultrasound gel should be used. If ultrasound gel is not available, water can be used for short time monitoring. You should charge Moyo's battery regularly to ensure that it is always ready for use. To learn more about how to charge Moyo, refer to page 28.
8
Get Started
Use sequence
?
?
?
87 134 83
87 85
135 15:36
95
87
To start using Moyo, press and hold the power button
195
PRACTICE
OBSERVE
Turn on Moyo.
Do you see 97 the display turning on? Do you hear the startup sound? If you encountered a problem, go to
?
Troubleshooting on page 43. 9
97
Use sequence
Get Started
?
?
?
87 134 83
87 85
135 15:36
95
?
87
means that no heart rate (HR) is found by the sensor.
195
SIMULATION NOTE
97
You can tap on the sensor to manually simulate a HR when there is no dedicated FHR simulator available.
? 97 135 10
134
134
134
Get Started
To measure fetal heart rate (FHR), you first need to apply gel on the ultrasound sensor and place it on the mother’s abdomen (if possible over the fetal back). If no HR sound is heard, reposition the ultrasound sensor until you hear the rhythmic HR sound.
Palpate to identify fetal position. Best results will be achieved with the ultrasound sensor placed over the fetus’s back.
PRACTICE Apply gel on the ultrasound sensor. Place the ultrasound sensor on the mother’s abdomen. (Reposition the ultrasound sensor if you do not hear a rhythmic HR sound) 11
Get Started
Moyo shows a HR number. Green color means that HR is in the normal range for a fetus. You can hear rhythmic HR sound from the speaker. Notice when FHR is shown, the status light blinks at the same rate as the FHR together with the rhythmic HR sound.
OBSERVE Do you hear the rhythmic HR sound? Do you read the numbers on the display?
12
Get Started
?
?
?
87 134
87
83
85
135 15:36
95
87 195
If you are not sure that the found heart rate is the fetal, you can instruct the mother to hold the pads to measure mother’s HR. If the two numbers are similar, then the HR found by the ultrasound sensor might be the maternal HR.
97
The ultrasound sensor might measure or be affected by mother's HR, a twin (multiples) or excessive patient movement/handling. In case the device displays unreliable or erratic FHR, try to reposition the ultrasound transducer or use alternative monitoring devices/ methods. Refer to page 32 for details about the maternal HR display.
?
PRACTICE 97 135
Instruct the mother to hold the pads. Compare the HR found by the ultrasound sensor with the maternal HR.
134 13
134
134
Get Started
When you are sure that you have found the FHR, keep the ultrasound sensor in place using the abdominal transducer belt.
PRACTICE Keep the sensor in place using the abdominal transducer belt.
14
4 Alarms Moyo provides alarms if the FHR is lost or if prolonged abnormal FHR is detected. In this chapter, you will learn about the alarms.
15
Alarms
Use sequence Ox2O
Lost FHR alarm
?
?
?
87 134
87
83
85
135 15:36
95
87 195
Moyo will give an alarm if the ultrasound sensor 97has not picked up any HR for 60 seconds. On the display, you will see the blinking.
?
If you hear the alarm, press the speaker button ? alarm.
PRACTICE Remove the ultrasound sensor from the mother, or stop simulating FHR, and wait more than 60 seconds.
to silence (acknowledge) the
97
OBSERVE
135
Did you hear the alarm sound? Did you silence the alarm?
134
134
16
134
Alarms
When
?
is shown, reposition the ultrasound sensor to find a HR.
If HR is not found after repositioning the sensor, continue with standard procedures according to your medical facility, e.g. using alternative devices to obtain FHR. If you are not sure that the detected heart rate is the fetal, you can instruct the mother to hold the pads to obtain the mother’s HR.
PRACTICE Search and find the FHR.
17
Alarms Abnormal FHR alarm Moyo is designed to monitor FHR, and provide visual and auditory alarm when abnormal FHR is present over a prolonged period. Moyo shows the number and status light in 3 different colors as shown below:
More than 160 bpm
Abnormally high FHR (yellow and red)
Between 110 - 160 bpm
Normal FHR (green)
Less than 110 bpm
Abnormally low FHR (yellow and red)
Yellow color turns red when abnormal FHR is present for a period of time.
To learn more about abnormal FHR alarm rules, refer to page 36.
PRACTICE Simulate / tap on the sensor for 3 minutes at the rate of e.g. 80 bpm (beats per minute) to trigger an abnormal FHR alarm. After you have triggered the alarm, continue to the next page. 18
87 134 83
Alarms
87 85
135 15:36
95
87 195 97
? 97 135 Moyo will give an alarm when the FHR has been abnormal for a prolonged time. You can 134sound. 134 134 see the number in red color, and hear the alarm If you hear the alarm, press the speaker button alarm.
to silence (acknowledge) the
OBSERVE Did you hear the alarm sound? Did you silence the alarm? Continue standard procedures - respond in a timely and appropriate manner. 19
5 Reprocessing In this chapter, you will learn how to clean, disinfect and inspect Moyo, and how to charge its battery so that it is ready for the next use.
21
Reprocessing Moyo and its accessories must be cleaned and disinfected between patient uses to minimize the risk of cross-contamination. Moyo must also be inspected for damage between patient uses. Turn off Moyo before starting the cleaning process.
Main unit, cable and sensor Cleaning
Thoroughly clean all surfaces using a cloth dampened with soapy water. A toothbrush dipped in soapy water can be used to remove difficult soil from the sensor. The main unit must not be scrubbed with a brush to avoid getting water into the device.
Wipe Moyo with a clean cloth dampened with clean water, then leave to dry.
22
Reprocessing
Inspect Moyo to confirm that it is clean and dry. Repeat cleaning process until all visual soil has been removed.
Disinfecting Effective disinfection is not possible without first performing a thorough cleaning. Care should be taken while handling the product between cleaning and disinfection.
Spray 70% ethanol on the surface of Moyo, or wet Moyo with a clean cloth dampened with 70% ethanol. 23