Laerdal Medical

Silicone Resuscitator Function Test Guide

Guide

2 Pages

Laerdal Silicone Resuscitator Function Test 1. Intake Valve a) Compress the ventilation bag with one hand and close its neck opening with your other hand. Release the grip on the bag. Rapid bag re-expansion confirms efficient air intake.  1. a  b) Close the neck opening and try to compress the bag. If the bag cannot be compressed with reasonable force, or if bag compression forces the air out between your hand and neck of the bag, the valve efficiently prevents backward leakage of air.  1.b  2.1 Patient valve a) Assure that a (single) Lip Valve has been installed in the Patient Valve. Attach the Patient Valve to the bag. Hold a reservoir bag over the patient port connector pressing with your thumb on the reservoir bag connector. Ensure tight seal between the patient port and Reservoir Bag. Compress the bag with your other hand several times. Inspect that the lip valve opens during compression  2.1. a  Filling of the Reservoir Bag in this set-up confirms that the Patient Valve efficiently directs air to the patient b) With the filled Reservoir Bag held firmly to the valve connector, compress the Reservoir Bag while watching the external Disk Membrane. Lifting of the Disk Membrane from its seat confirms that air is correctly directed to atmosphere instead of being returned to the ventilation bag.  2.1. b
File Type: PDF
File Size: 57.9 KB
File Name: Laerdal Medical - Silicone Resuscitator Function Test Guide.pdf

By clicking Download you are confirming this is the correct document for your purposes
and that you agree to our Terms & Conditions