Directions for Use
258 Pages
Preview
Page 1
Contents 1 LCSU 4 Overview –– Important Information 4 –– LCSU 4 - 800 ml (Cat. No. 880051/880052)
4
–– LCSU 4 - 300 ml (Cat. No. 880061/880062)
4
2 Description and Intended Use –– Intended Use 5 –– Cautions and Warnings 5 –– Limited Warranty 5
3 LCSU 4 Setup –– Prepare the LCSU 4 for Operation
6
–– Assemble 800 ml Version
6
–– Operation with High Efficiency Filter Kit
6
–– Assemble 300 ml Version
7
4 User Instructions –– Check Before Each Use
7
–– Power Source Options 8 –– Control Panel and Indicator Symbols
8
–– Suction Level Setting 8 –– How to Operate and Adjust Suction level
9
5 Battery Information –– Battery Charging 9 –– Battery Test 9 –– Charging Operations 10 –– External Battery Charger
10
6 Cleaning and Maintenance –– Cleaning 11 –– Device Test 11
7 Troubleshooting 12 8 Main Units, Accessories and Parts
13
9 Glossary of Symbols
14
10 Specifications 15 11 Regulatory Information
16
3
English
1 LCSU 4 Overview Important Information These Directions for Use cover two main versions of LCSU 4; one configured with 800 ml Canister, and one with 300 ml Canister. Both versions can be purchased with or without RTCA compliance. Unless otherwise specified, the information in these Directions for Use applies to all versions. Important Inspect all parts when unpacking. If there are signs of damage or parts are missing - immediately notify the seller. Do not attempt to use the LCSU 4 if parts are damaged or missing, or if the Battery has not been sufficiently charged.
LCSU 4 - 800 ml (Cat. No. 880051/880052)
LCSU 4 - 300 ml (Cat. No. 880061/880062)
Items Included:
Items Included:
•
LCSU 4 Main Unit
•
LCSU 4 Main Unit
•
800 ml Disposable Canister
•
300 ml Disposable Canister
•
Patient Tube 1.8 m (6')
•
Patient Tube 0.9 m (3')
•
AC/DC Adapter charger, with AC Power-cord
•
AC/DC Adapter charger, with AC Power-cord
•
Battery
•
Battery
•
Directions for Use
•
Directions for Use
•
Carry Bag (for 800 ml version)
•
Carry Bag (for 300 ml version)
•
Wire Stand
•
Vacuum Tube
Disposable Canister (800 ml) with Internal Filter
Disposable Canister (300 ml)
Control Panel
Internal Filter Vacuum Inlet
Wire Stand
Vacuum Tube
Vacuum Port Connector (Canister - white)
Patient Tube 1.8 m (6')
Patient Port
Vacuum Regulator 12V DC Power Input
Battery Cover
Patient Tube 0.9 m (3')
Vacuum Inlet Connector (LCSU 4 )
•
The main suction unit is the same for both versions.
•
Each version can easily be converted to the other Canister option by ordering additional parts.
•
For a complete parts overview (see Chapter 8).
4
Intended Use The LCSU 4 is a portable, electrically powered, medical suction device intended for field and transport use. It is intended for intermittent operation to remove secretions, blood or vomit from a patient’s airway to allow ventilation. Higher vacuum levels are generally selected for oropharyngeal suction, and lower vacuum levels are usually selected for tracheal suctioning and the suctioning of children and infants. Important •
Do not use this unit until you have carefully read and fully understood these Directions for Use. Contact Laerdal Medical or its authorized distributor for additional information if required.
•
Federal law (USA) restricts the LCSU 4 to sale by or on the order of a physician or other licensed medical authority.
•
Use only Laerdal accessories supplied directly by Laerdal Medical or one of its authorized distributors to ensure that the LCSU 4 operates satisfactorily.
Cautions and Warnings Cautions •
The LCSU 4 is not suitable for use in the presence of flammable liquids or gases; danger of explosion or fire.
•
Do not use the LCSU 4 under environmental conditions that are outside ranges specified. This can endanger safety and adversely affect device operation.
•
Ingress of suctioned material into the pump can damage and/or disable the device. If suctioning of liquid from the Canister or patient into the pump is suspected, do not use the LCSU 4. Contact Laerdal Medical or your authorized distributor for advice.
Warnings •
The LCSU 4 should only be used by persons trained in the use of medical suction equipment, and according to local protocol.
•
Unauthorized service attempts, opening or tampering with the LCSU 4 or its electrical components can damage or disable the device, and will void the Limited Warranty.
•
Cat. No. 880052/880062 are approved according to RTCA/DO-160G Section 21 Category M but limited to battery operation use only. Use of the AC/DC Adapter charger (Cat. No. 886111) or DC Power cord (Cat. No. 884500) for charging or operation inside an aircraft must be avoided.
•
Cat. No. 880051/880061 are not approved for use in aircrafts.
Limited Warranty The LCSU 4 comes with a two (2) year limited warranty, excluding the Canisters, Tubing systems and Battery. See “Laerdal Global Warranty” for terms and conditions, available on www.laerdal.com. The Battery is warranted for 90 days. Laerdal does not provide Service Parts for this product. Excluding the internal Battery, there are no user-serviceable/user-replaceable parts inside the LCSU 4 Main Unit.
5
English
2 Description and Intended Use
English
3 LCSU 4 Setup Prepare the LCSU 4 for Operation Important All models are shipped with the Battery inside the unit, but not connected. Connect the Battery and charge it fully before using the unit. See Charging Instructions (Chapter 5).
Assemble 800 ml Version 1
2
3
Canister Ports A - Vacuum Port B - Patient Port
1 Secure the Lid to the Canister.
2 Place the Canister into Wire Stand.
3 Ensure that the Patient Port is accessible.
4
5
6
7
4 Attach the Connector to the Vacuum Port Connector on the Suction Unit.
5 Attach the WHITE Connector to the Vacuum Port Connector on the Canister.
6 Check that all Vacuum Tube connections are firmly attached.
7 Attach the Patient Tube to Patient Port on the Canister.
A B
Important The 800 ml Canister has an internal Filter in the Lid. The Canister is disposable and cannot be cleaned. The Filter automatically stops suction/flow when the Canister is full, or the Filter becomes saturated if the unit tips onto its side during use. Caution Always use the 800 ml Canister supplied by Laerdal, which has an internal Filter. Never connect any type of Patient Tubing directly to the LCSU 4 Vacuum Inlet Port Connection. Overflow of suctioned material into the LCSU 4 pump will result in loss of suction and permanent damage to the unit. In the event of overflow, do not use the LCSU 4. Contact Laerdal Medical or your authorized distributor.
Operation with High Efficiency Filter Kit To increase filtration efficiency the Vacuum Tube may be replaced with a High Efficiency Filter Kit (Cat. No. 886116). Ensure Filter IN points towards the WHITE connector, and OUT towards the BLUE. For Cleaning and Maintenance, see Chapter 6.
6
IN OUT
Assemble 300 ml Version
Check Before Each Use
1 Push the upper port connection into the Vacuum Inlet and check that the lower part of the Canister clicks in place.
1 The Unit should not be damaged.
2 Connect the Patient Tube to the Patient Port. Ensure that all connections are secure to prevent leakage. 1
2
2 The Unit should be clean. 3 All parts should be properly assembled (Canister, Tubes etc.). 4 Perform Device Test after each reassembly (see Chapter 6). 5 Check Battery level: While performing the Device Test the Battery level should not illuminate RED. If the Battery level illuminates RED, the Battery should be charged. See Charging Instructions (Chapter 5).
300m 300mll
250m 250mll 200m 200mll
150m 150mll
100m 100mll 50m 50mll
Important •
The 300 ml Canister is sealed, with an internal Filter. The Canister is disposable and cannot be cleaned. The Filter automatically stops suction/flow when the Canister is full, or the Filter becomes saturated if the unit tips onto its side during use.
•
The 300 ml Canister (Cat. No. 886100) can also be used with the previous model LCSU 3. However, the LCSU 4 cannot be used with LCSU 3 Canisters.
•
To improve the upright stability, a Wire Stand (Cat. No 886115) is offered as an Optional Accessory. This adds both a handle and a foot to the unit.
Important For both models - always have an extra Canister available in case the first Canister is filled completely, or the unit becomes tipped on to its side and the Filter becomes saturated and stops the suction/flow. Warning If the Canister on either model fills and the shut-off mechanism activates, and you have no spare Canister ready for immediate replacement, shut off LCSU 4 and utilize alternative methods according to local protocol to clear patient’s airway. Continued efforts to suction with LCSU 4 Canister(s) filled may cause overflow that will prevent suctioning, damage the pump, void the unit warranty, and lead to prolonged downtime.
Caution Do not attempt to install or use LCSU 3 Canisters on LCSU 4.
Cat. No. 886115
7
English
3 LCSU 4 Setup
English
4 User Instructions Power Source Options
Control Panel and Indicator Symbols ON / OFF switch
Internal Battery Operation LCSU 4 is equipped with an internal Battery, NiMH 12 volt 1.6 Ah. If the unit is to be run on Battery power, any external power source must be unplugged. Unplugging the external power source whilst running will stop the unit from operating. To restart, press the on/off switch.
Suction Level Setting
External 12V DC Operation DC Power-cord for connection to vehicle 12V DC is required. Plug the smaller power connector into the LCSU 4 12V DC Power Input Connection. Plug the larger connector into the vehicle 12V DC power receptacle. External AC Operation AC/DC Adapter Charger required. Plug the smaller DC output cord power connector into the LCSU 4 12V DC Power Input Connection. Plug the AC input line cord power connector into a grounded AC mains supply receptacle. When in use, it is normal that the AC/DC Adapter Charger becomes warm.
Green LED Yellow LED Red LED
Suction Level Setting •
The Scale illuminates ‘Green’ to indicate the level of vacuum/suction strength
•
The ‘Light Blue’ area indicates reduced suction levels for infants and small children
The LEDs have two brightness levels. Half illuminated indicates a halfway vacuum level, e.g., 175 is indicated by a fully illuminated 150 and a half illuminated 200 LED. LED Indicator
Status
Green
External Power is connected
Yellow
Battery is charging (Will go off when Battery is fully charged)
Red
Battery level is low
Warning If the Low Battery symbol illuminates, immediately switch to an external power source to avoid interrupted operation. If the LCSU 4 does not receive external power, the Low Battery indicator will remain on and the unit performance will drop rapidly leading to complete LCSU 4 shutdown.
8
How to Operate and Adjust Suction level
Battery Charging
1 Unwind the Patient Tube (check that no kinks will obstruct flow).
Important
2 Turn the unit “ON” by pressing the button (
Use only Laerdal Battery Cat. No. 886113.
).
3 Block the Patient Tube. 3
Caution 4
Use of other than Laerdal brand Battery may result in errors related to the Battery status indicator, reduction of the Battery operation time, failure to effectively operate LCSU 4, and/or give rise to hazards to operator and/or patient. When in operational use 1 An empty Battery must charge up to 5 hours to reach full capacity.
4 Set desired suction vacuum level by turning the Vacuum Regulator. • •
Turn clockwise (+) to increase vacuum Turn counter-clockwise to decrease vacuum
5 The vacuum level will display on the 50 – 550+ mmHg scale.
2 Battery run time: Approximately 45 minutes of continuous operation at zero vacuum level (free flow), fully recharged. 3 Always fully charge Battery - before and after each use. To prolong Battery lifetime it is recommended to place the Battery on continuous charge. It will not harm the unit. If continuous charging is not possible, charge Battery for minimum 24 hours once a month. Completely discharging the Battery will shorten its useful life. When stored unused for > 3 months
6 When the desired vacuum level is reached; unblock and then block the Patient Tube. Result: Unit should return to the desired level. 7 Apply necessary suction therapy. If desired, a suction tip or catheter could be attached to the Patient Tube.
•
Fully charge Battery before storing
•
Recharge every 3-6 months
Battery Test When in operational use, the Battery should be tested every 6-12 months. 1 Start the test with a fully charged Battery. 2 Set maximum vacuum level.
Important
3 Let the unit run for 20 minutes (free flow).
If LCSU 4 does not maintain the desired suction performance level, refer to Troubleshooting instructions (see Chapter 7).
4 Block the Patient Tube. 5 If vacuum level fails to reach 550+ mmHg the Battery should be replaced.
After Each Use 1 When finished with suctioning, let LCSU 4 run for a moment to allow all suctioned material to flow from the Patient Tube into the Canister. 2 Disconnect and dispose of Canister and Patient Tube. 3 Clean the exterior and any reusable parts of the LCSU 4 assembly according to instructions (see Chapter 6). 4 Perform Device Test (see Chapter 6). 5 Place Battery on charge (see Chapter 5). 9
English
5 Battery Information
English
5 Battery Information Charging Operations
External Battery Charger (Cat. No. 886112) The Battery can be charged externally by removing it from the LCSU 4 unit and using the External Battery Charger.
Caution Do not operate the unit for more than a few minutes if the RED Low Battery indicator is illuminated. Recharge the Battery as soon as possible.
External Battery Charger Power cord
LED Indicator
AC/DC Adapter Charger (Cat. No. 886111)
Battery Connector
Primarily the Battery is charged while installed within the LCSU 4 unit by using the standard AC/DC Adapter Charger. Battery
AC/DC Adapter Charger
An empty Battery must charge up to 5 hours to reach full capacity. Follow the LED Indicator and charge the Battery as necessary.
Power Cord
Follow the LED status on the Control Panel on the LCSU 4 and charge the Battery as necessary.
Green LED Yellow LED Red LED
LED Indicator
Status
Red LED illuminated
Battery level low
Yellow LED illuminated
Charging
Yellow LED flashes
Battery almost charged
Yellow LED switches off
Battery is fully charged
Green LED illuminated
Connected to AC/DC
10
LED Indicator
Status
LED not illuminated
Plugs not connected
Yellow LED flashes
Charging is pending
Yellow LED illuminated
Charging
Green LED illuminated
Battery is fully charged (*)
Red LED flashes
Charge failure
* The Battery can be left on continuous charge, even if the Green LED is illuminated. It will not harm the units. Caution Do not cover the Charger. When in use, it is normal that the Charger and the Battery becomes warm.
Cleaning Cautions
Carry Bags Wipe Bags according to instructions provided above for Main Cabinet. Do not launder.
•
Disconnect the LCSU 4 from external power prior to cleaning.
•
To prevent any electrical shock hazard, use minimum liquids when cleaning. Do not immerse the LCSU 4, or allow it to stand in water or other liquids. This can damage the device and/or cause electrical shocks resulting in injury to persons.
After each reassembly, and before returning a LCSU 4 to operational use, the Device Test should be performed:
Warning
3 Block the Patient Tube
Do not pump any cleaning solution or other liquids through the vacuum pump, i.e. through the Vacuum Connector. This can damage the LCSU 4.
4 Set Suction Vacuum level to 550+ mmHg
Main Cabinet 1 Disconnect from external power supply. 2 Clean Cabinet surfaces by carefully wiping with a soft cloth or sponge with mild detergent. Use hand dishwashing liquid or similar that is compatible with the Material Chart (see Chapter 10). 3 Dry all surfaces using a clean cloth or paper towel. Canisters and Patient Tubes Dispose after use. Important The Canisters and Patient Tubes are all disposable items. Do not attempt to clean or reuse these. Due to the risk of cross contamination, all disposable items must be replaced after each use. They are for single-patient use only. Vacuum Tube (for 800 ml version) and Wire Stand Wash by immersing and scrubbing in hand dishwashing liquid or similar. 1 Rinse thoroughly in pure water. 2 Allow to dry. Disinfect if desired. High Efficiency Filter Kit (for 800 ml Version) •
Filter cannot be cleaned or disinfected.
•
Replace Filter immediately if contamination or discoloration is observed, or if it gets wet.
* If the unit is used on patients in areas where cross contamination is an issue, it is recommended that the Filter is replaced after each use.
Device Test
1 Start the test with a fully charged Battery 2 Turn the unit “ON”
5 Unblock and then Block the Patient Tube again. 6 Result: Unit should return to 550+ mmHg setting. 7 Repeat procedure for 300 mmHg and 50 mmHg settings.
Cautions Do not attempt to use any LCSU 4 that has not passed the above test. If the LCSU 4 unit does not test satisfactorily, recheck all parts of the assembly and test once again. If necessary, contact Laerdal Medical or one of its authorized distributor.
Disposal When discarding the LCSU 4, we recommend it be discarded according to local protocol. This appliance is marked according to the European directive 2012/19/EU on Waste Electrical and Electronic Equipment (WEEE). By ensuring this product is disposed of correctly, you will help prevent potential negative consequences for the environment and human health, which could otherwise be caused by inappropriate waste handling of this product. The symbol on the product, or on the documents accompanying the product, indicates that this appliance may not be treated as household waste. Instead it shall be handed over to the applicable collection point for the recycling of electrical and electronic equipment. Disposal must be carried out in accordance with local environment regulations for waste disposal. For more detailed information about treatment, recovery and recycling of this product, please contact your local city office, your household waste disposal service or the Laerdal representative where you purchased the product.
11
English
6 Cleaning and Maintenance
English
7 Troubleshooting Troubleshooting Warning Potential for Electrical shock. Do not attempt to open or disassemble pump or electrical accessories. LCSU 4 will not power “ON” 1 Verify that the Battery is properly installed and charged. 2 Check alternative 12V DC or AC sources. LCSU 4 runs, no vacuum, or insufficient vacuum 1 Verify proper Canister and Tubing connections. 2 Check system for leaks in Canister and Tubing connections. 3 Check Canister overflow protection float (800 ml). 4 Check if the Filter has been shut-off (300 ml). Vacuum too high or too low Turn Vacuum Regulator to increase or reduce vacuum level. Battery will not charge 1 Verify that the Battery is installed and connected. 2 Re-connect source power and observe charge (all connections). Battery seems insufficient 1 Charge the Battery for 5 hours. 2 Perform the Battery Test. Important If the LCSU 4 condition is not resolved, contact Laerdal Medical or one of its authorized distributors for advice.
12
Main Units, Accessories and Parts LCSU 4 (Main Unit versions) Cat. No. 880051 880052 880061 880062
Item LCSU 4, 800 ml LCSU 4, 800 ml (RTCA) LCSU 4, 300 ml LCSU 4, 300 ml (RTCA)
Complete unit Complete unit Complete unit Complete unit
Consumables Cat. No. 886100 886102 886104 886105 886106
Item 300 ml Disposable Canister w/tubing 800 ml Disposable Canister w/tubing 800 ml Disposable Canister w/o tubing Patient Tube 1.8 m (6’) Disposable Vacuum Tube
Quantity Qty. 1 Qty. 1 Qty. 6 Qty. 1 Qty. 1
Accessories Cat. No. 886108 886115 886109 886110 886111 886112 884500 886107 886116
LCSU 4 L A E R DA L
LCSU 4 L A E R DA L
LC
SU
LA
ER
DA
L
LC
SU
LA
4
ER
DA
L
Item Wire Stand (for 800 ml) Wire Stand (for 300 ml) Carry Bag (for 300 ml) Carry Bag (for 800 ml) AC/DC Adapter Charger w/o Power-cord External Battery Charger DC Power-cord Replacement Filter High Efficiency Filter Kit
886115
LCSU 4 Spare Parts
LC
L A E R DA L
L A E R DA L
L A E R DA L
LC
SU
LA
ER
DA
L
4
4
886108
LCSU 4
Quantity Qty. 1 Qty. 1 Qty. 1 Qty. 1 Qty. 1 Qty. 1 Qty. 1 LCSU 4 Qty. 10 Qty. 1
LC
SU
LA
ER
DA
L
4
Cat. No. 886113 886123 886125 886126 886127 886128
LCSU 4 L A E R DA L
Item
SU
LA
ER
DA
L
886111
4
LCSU 4 L A E R DA L
LC
SU
Quantity
Battery, 12 V DC NiMH, Rechargeable LCSU 4 Battery Cover (door) Shoulder Strap for bags AC Power-cord US AC Power-cord EU AC Power-cord UK LC
L A E R DA L
SU
LA
LC
SU
LA
ER
DA
L
ER
DA
L
LA
ER
DA
L
Qty. 1 Qty. 1 Qty. 1 Qty. 1 Qty. 1 Qty. 1
4
4
4
886113
AC Power-cord
886126
886127
886128
13
English
8 Main Unit, Accessories and Parts
English
9 Glossary of Symbols Product Symbols
Definition
Definition
This product is in compliance with the essential requirements of Council Directive 93/42/EEC Medical Device Directive, as amended by Council Directive 2007/47/EC, class IIa The product is in compliance with Council Directive 2011/65/EU on restriction of the use of certain hazardous substances (RoHS 2).
Do not cut
D
Single Use
Fragile. Handle with care
IP12
The degree of protection provided by the chassis according to IP12
h
Unique product type identification
f
Serial number
Humidity
UL Classified
Atmospheric pressure
0434
4NG8
Center positive polarity indicator Direct Current
N
Date of production Warning / Caution Important
Recycle Type BF applied part Disposal must be carried out in accordance with local environment regulations for waste disposal.
Not made with natural rubber latex
14
Packaging Symbols
p 70 °C 158 °F
l
-40 °C -40 °F
Keep dry Transport storage temperature range
Power Source Options Battery Rechargable, NiMH 12 volt 1.6 Ah DC Power-cord (12 V) Dry Location Use Only AC/DC Adapter Charger Brand: Bridgepower / Model: MENB1040A1240F01 Input: 100-240 V, 50-60 hz, 1.2 A Output: +12 V, 3.4 A Dry Location Use Only
Vacuum - Max.: 550+ mmHg Vacuum - Range: 50 - 550+ mmHg Vacuum Indicator accuracy: +/- 5% of full scale
Filter Specifications Internal Filter of the 300 ml Disposable Canister Porous plastic filter, PE material. Internal Filter of the 800 ml Disposable Canister Aerostate HEPA rated bacterial filter. White, unfinished surface, cellulose with wet strength resin. High Efficiency Filter Kit
External Battery Charger Input: 110-240 V, 50-60 hz, 250 mA Output: +18.5 V, 0.6 A Dry Location Use Only
With the High Efficiency Filter Kit installed the unit is in accordance with ISO 10079-1. The Kit reduces the Air flow and Battery run time. The filter is HEPA rated with an efficiency of 99.97% down to a particle size of 0.3 μm.
Environmental Conditions
Material Chart
Operating Temperature Range: 0 °C (+32 °F) – +40 °C (+104 °F)
Cabinet: Battery Cover: Vacuum Regulator: Bottom Cover: Vacuum Inlet Connector: Control Panel: 800 ml Disposable Canister: - Lid: - Internal Filter: Vacuum Tube: Vacuum Port Connector: Vacuum Inlet Connector: High Efficiency Filter: Filter housing: 300 ml Disposable Canister: - Internal Filter: Patient Port: Patient Tube: Wire Stands:
Operating Relative Humidity: 0 - 95% (non condensing) Operating Atmospheric Pressure: 10.2 Psi (70 kPA) – 15.4 Psi (106 kPA) Storage & Transport Temperature: -40 °C (-40 °F ) – +70 °C (+158 °F) Storage & Transport Relative Humidity: 0 - 95% (non condensing) Storage & Transport Atmospheric Pressure: 7.3 Psi (50 kPA) – 15.4 Psi (106 kPA)
Physical Characteristics
PC PC PC PC TPR PVC GPPS HDPE Aerostate Silicone, K-Resin TPR PC PP K-Resin PS PE PP PVC Steel, PVC
Dimensions •
880051/880052 (LCSU 4, 800 ml): 23.6 cm x 19 cm x 23.6 cm (9.3" x 7.5" x 9.3")
•
880061/880062 (LCSU 4, 300 ml): 18.5 cm x 26.2 cm x 8.12 cm (7.3" x 10.3" x 3.2")
Weight •
880051/880052 (LCSU 4, 800 ml): 1.97 kg (4.35 lbs)
•
880061/880062 (LCSU 4, 300 ml): 1.53 kg (3.375lbs)
Canister Capacity •
300 ml
•
800 ml
Performance Air flow at Vacuum Inlet (without canister attached) All configurations: 30 LPM (free flow) typical (May be less when running from internal battery) 15
English
10 Specifications
English
11 Regulatory Information International Travel This suction unit is equipped with an AC/DC Adapter Charger allowing operation on any AC voltage (100-240 VAC, 50/60Hz). However the correct power cord must be used to connect adapter to wall power.
Regulatory Information Classification This product is in compliance with the essential requirements of Council Directive 93/42/EEC Medical Device Directive, as amended by Council Directive 2007/47/EC, class IIa The product is in compliance with Council Directive 2011/65/EU on restriction of the use of certain hazardous substances (RoHS 2). •
Electrically powered medical suction equipment for field and transport use, according to ISO10079-1:1999
•
High flow/High vacuum, 50 – 550+ mmHg
•
Not suitable for the use in the presence of flammable liquids or gases.
•
Internally powered/class I equipment type BF, according to IEC 60601-1
•
Protection class IPI2 and standard power supply
•
Intermittent Operation: 30 minutes on, 30 minutes off
Certifications IEC 60601-1 : 1988 (2nd edition); IEC 60068-2-6/IEC 60068-2-64/IEC 60068-2-27/IEC 60068-2-31; CAN/CSA-C22.2 No. 601.1-M90, 2005; UL 60601-1, EN 60601-1-2 : 2007, EN ISO 10079-1 : 2009 (Except requirement of 500 ml volume in claim 59.11.1, due to the 300 ml Canister option) Standards Cat. No. 880052/880062: Meets RTCA/DO-160G - Section 21 Category M (for battery operation only; commercial aircraft, airborne equipment).
16
20-05448 Rev J
Manufactured for: Laerdal Medical AS, Tanke Svilandsgate 30 P.O. Box 377, 4002 Stavanger, Norway Tel. +4751511700 Fax +4751523557 Distributed in USA by: LAERDAL MEDICAL CORPORATION 167 Myers Corners Road P.O. Box 1840, Wappingers Falls New York 12590-8840 Tel. +1 (800) 431-1055, +1 (845) 297-7770 Fax +1 (800) 227-1134, +1 (845) 298-4545