Stryker LIFEPAK CR2 Supplemental Operating Instructions Dec 2020

Supplemental Operating Instructions

1 Page

LIFEPAK® CR2 Defibrillator Supplemental Operating Instructions Refer to the LIFEPAK CR2 Defibrillator Operating Instructions for complete directions for use.  Warning Premature Battery Depletion. If the lid or lid magnet is missing, the LIFEPAK CR2 battery can become prematurely depleted.  Troubleshooting Tips This table explains problem conditions that you may encounter while using the defibrillator.  OBSERVATION  POSSIBLE CAUSE  CORRECTIVE ACTION  Defibrillator does not provide voice prompts or beeping tones after you open the lid (turn it on).  Lid magnet is missing  •  Use the ON/OFF button to turn the defibrillator on if required for emergency use.  •  Contact qualified service personnel.  Depleted battery  •  Replace the battery immediately. If a replacement is not available, order a new battery immediately.  Speaker system failure  •  Contact qualified service personnel.  Lid magnet is missing  •  Use the ON/OFF button to turn the defibrillator off.  •  Contact qualified service personnel.  Voice prompts continue after lid is closed.  Stryker Emergency Care 11811 Willows Road NE Redmond, WA 98052 USA Tel: 1 800 STRYKER Toll Free (USA only): 1 800 732 3081 Fax: 425 867 4121 strykeremergencycare.com Copyright © 2020 Stryker  12/2020  3345569-000