Golvo™ Instruction Guide English 7EN140105-02 2011-03-30 Applies to the following models: Golvo 8000 Golvo 8008 Golvo 8008 LowBase™
Prod. No. 2000014 Prod. No. 2000015 Prod. No. 2000019
Golvo 8008 LowBase™
Golvo 8000/8008
Product Description Golvo is a versatile mobile lift, intended for use in most lifting situations, for example, between bed and wheelchair, to/from toilet, bath and shower, to/from the floor, for weighing purposes and for lifting with a stretcher. Due to its foldable arm rests, Golvo can to advantage be used for gait training. The wireless hand control gives the caregiver the freedom to work closer to the patient, which provides extra security in the lifting situation. For the caregiver this also enables an ergonomically enhanced work posture. Golvo is equipped with the Liko Diagnostic System™, a monitoring system which continually registers operating data and automatically indicates when the lift is in need of service.
The battery of the lift has built-in charge indicator where the charging status of the battery can be read, even when the battery is not attached to the lift. The battery is of environmentally friendly NiMH type. Golvo LowBase™, with its extra low base, height clearance of 58 to 110 mm (2.28 to 4.33 inch.), is suitable for lifting to and from beds/gurneys. Individual fitting of the sling and other lifting accessories is of utmost importance for function and safety when using the lift.
is a warning triangle used for situations which require extra care and attention. In this document, the person being lifted is referred to as the patient, and the person helping is referred to as the caregiver.
IMPORTANT!
Read the instruction guides for both the patient lift and lifting accessories before use. Lifting and transferring a person always involves a certain level of risk. It is important to completely understand the contents of the instruction guides. The equipment should only be used by trained personnel. Please contact your Liko/HillRom representative in the event of any uncertainties or questions.
Table of Contents Safety Instructions... 2 Definitions... 3 Technical Data... 3 Dimensions... 4 Assembly... 5 Disassembly... 7 Operation... 8 HandControl IR... 10 Operation Panel... 10 Charging the Batteries... 11 Maximum Load... 12 Recommended Lifting Accessories... 12 Simple Troubleshooting... 15 Inspection and Maintenance ... 16 NOTE! This instruction guide contains important information for users of the product. All those who use the product should review and fully understand the contents of the instruction guide. Remember to keep the instruction guide in a place where it is always available to those using the product.
Safety Instructions Before using the first time make sure that: • the lift is assembled in accordance with the assembly instructions • the lifting accessory is properly attached to the lift • the batteries have been charged for at least 5 hours • you have read the instruction guides for the lift and lifting accessories • personnel using the lift are informed of the correct operation and use of the lift. Before lifting, always make sure that: • the lifting accessories is selected appropriately in terms of type, size, material and design with regard to the patient’s needs • the lifting accessories are not damaged • the lifting accessory is correctly attached to the lift • the lift strap is not twisted or worn and can move in and out of the lift • the lifting accessory hangs vertically and can move freely • the lifting accessory is correctly and safely applied to the patient in order to prevent injuries • the latches are intact; missing or damaged latches must always be replaced • the sling’s strap loops are correctly connected to the sling bar hooks when the sling strap is extended, but before the patient is lifted from the underlying surface. Never leave a patient unattended during a lifting situation!
Golvo 8000/8008/8008 LowBase™ have been tested by an accredited testing institute and fulfil the requirements specified in the Medical Device Directive for class 1 products (MDD 93/42/EEC). Golvo 8000/8008/8008 LowBase™ complies with the requirements in EN ISO 10535, IEC 60601-1, IEC 60601-1-2, ANSI/AAMI ES60601-1 and CAN/CSA C22.2 no. 60601-1 Under no circumstances must the lift be modified. If you have any questions, please contact Liko. Particular care must be observed when using strong sources of potential disturbance, such as diathermy, etc, so that diathermy cables are not positioned on or near the lift. If you have questions, please consult the responsible assistive-device technician or the supplier. The lift must not be used in areas where flammable mixtures may occur, for example in areas where flammable goods are stored. Golvo • 7EN140105-02
w w w . l i k o . com
Definitions Golvo 8000/8008
1. Lift Strap 2. Sling bar with latches
1
18
3. HandControl IR
17
4. Parking panel for sling bar
2
16
5. Holder for Quick Reference Guide and colour code for sling sizes
3
6. Product decal
15 14
4
13
5
7. Lift mast with inbuilt motor 8. Front wheels 9. Base 10. Rear wheels with brakes
6 12
7
11. Motor for base-width adjustment 12. C ontrol box with emergency stop built-in charger and operation panel (electrical emergency lowering/raising)
Forward direction
11
13. Control box IR receiver
10
14. Battery
8
9
15. Foldable arm rest 16. Handles
Golvo 8008 LowBase™
17. Emergency Lowering (mechanical) 18. Lift arm 19. Extra low base (Golvo LowBase)
19
Technical Data Maximum load:
200 kg (440 lbs)
Material:
Anodised aluminium
Weight:
8000: 42 kg (92 lbs) 8008: 44 kg (97 lbs) 8008 LowBase: 46 kg (101 lbs) Heaviest removable part: 8000: 18 kg (40 lbs) 8008: 19 kg (42 lbs) 8008 LowBase: 21 kg (46 lbs)
Wheels:
Turning diameter:
Front: 75 mm* (3 inch.) twin wheels. *Golvo LowBase: 46 mm (1.8 inch.) twin wheels. Rear: 75 mm (3 inch.) twin wheels with brakes Golvo 8000: 1240 mm (49 inch.) Golvo 8008: 1330 mm (52 inch.) Golvo 8008 LowBase: 1330 mm (52 inch.)
Operating forces, controls:
Buttons on hand control: 4 N Button on display: 4 N
HandControl IR:
Batteries: 2 AAA LR03 1.5 V, alkaline. Reach: 2 m (78.74 inch.)
Electrical data:
24 V
Intermittent operation:
Int. Op 10/90, active operation max 2 min. Only 10% of a given length of time may be active, yet no more than 2 min.
Battery:
NiMH cells, 2.2 Ah. New batteries provided by the supplier.
Battery charger:
CBL20001, built-in, 100-240 V AC, 50-60 Hz, max 600 mA.
Lift motor:
24 V, 6.5 A
Motor for basewidth adjustment:
24 V, 3.5 A
Emergency lowering device:
Mechanical and electrical
Lifting speed with no load:
Two speeds; 41 mm/s (1.61 inch/s) and 33 mm/s (1.3 inch/s)
The device is intended for use indoors.
Lifting interval:
1260 mm (50 inch.) (height-adjustable)
Type B, in accordance with the electrical shock
Sound level:
52,8 dB(A)
protection class.
Protection class:
IP X4
Class II equipment. Protected by Patent
Golvo • 7EN140105-02
w w w . l i k o . com
Dimensions
700 mm
Top view
M
(27.5 in)
Lateral view
D C
A D2 O
Lmax Lmin
EG
Golvo LowBase 600 mm
F1 F
(23.6 in)
E
B3 B2 B1 B
B4
Maximum reach
Table of measurements Model
A
C
D
B1
B2
B3*
B4
Golvo 8000
1995 1360 1115 800
435
435
-
938
654
826
Golvo 8008
2090 1455 1185 870
480
480
-
1020 735
Golvo LowBase
2090 1455 1185 870
480
480
600 1028 745
max. min.
B
(mm) D2*
E
F
F1
G
Lmax Lmin
M
O
543
0
975
105
25
840
1735
475
440
362
907
623
0
1100
105
25
940
1860
600
436
391
907
623
0
1100
60 / 105
22
940
1860
600
436
391
max. min. max. min.
(Inch.) Model
A
B3*
B4
B1
B2
Golvo 8000
78.5 53.5 43.9 31.5 17.1 17.1
-
Golvo 8008
82.3 57.3 46.6 34.2 18.9 18.9
-
Golvo LowBase
D
D2*
E
F
F1
G
Lmax Lmin
M
O
36.9 25.7 32.5 21.4
0
38.4
4.1
1.0
33.0
68.3
18.7
17.3
12.8
40.2 28.9 35.7
623
0
43.3
4.1
1.0
37
73.2
23.6
17.2
15.4
82.3 57.3 46.6 34.2 18.9 18.9 23.6 40.5 29.3 35.7
623
0
43.3
2.4 / 4.1
0.9
37
73.2
23.6
17.2
15.4
max. min.
B
C
max. min. max. min.
The lifting interval 1260 mm (50 inch.) is height-adjustable, see page 9. Lifting height dimensions are based on the lift being equipped with a standard sling bar and standard wheels. When changing to other lifting accessories, check that the lift still achieves the desired lifting height. * Reference measurement according to Standard EN ISO 10535:2006.
Golvo • 7EN140105-02
w w w . l i k o . com
Assembly Before assembly, make sure you have the following parts and tools: • lift mast with control box and sling bar with latches, 2 M6 screws, bolt and lock nut (M8) • armrest • base with width adjustment motor • battery • handcontrol IR
• bag containing: instruction guide, quick reference guide, handle reinforcement for emergency lowering, charger cable, charger connector cable, cable lid with screws (2), back cover with alternative suspension clip for the hand control, 2 adhesive buttons for hanging up the hand control on and a cord for HandControl IR. • tools: 13 mm spanner, 4, 5 and 6 mm Allen keys.
Lock the wheels on the base before starting to assemble the lift.
A B
1. Unscrew and remove the transportation guard (metal plate) at the bottom of the lift mast. Remove the transportation guard, its attachment screws and the red information decal. A) Position the lift mast between the two black plastic plugs on the base cross member. B) Then push the lift mast forwards as in the figure above so that it hooks onto the cross member.
2. Screw the two provided M6 screws into the upper holes on the lift mast. No screws in the lower holes!
B
A
3. A) Place the armrest part in the attachment on the lift mast, starting with the lower groove. B) Lower and load the armrest until it hooks onto the upper groove on the armrest part. Do not remove any of the pre-assembled M8 screws completely, but it may be necessary to loosen them.
Golvo • 7EN140105-02
4. Secure the armrest by tightening the two preassembled M8 screws.
w w w . l i k o . com
1 3
2
5. Assemble the sling bar (or other lifting accessory) with the provided bolt and lock nut (M8). Check that the bolt is properly tightened.
6. Connect the cables as follows: - Charger cable to socket 1 -C able from lift motor to socket 2 - Cable from motor for basewidth adjustment to socket 3.
7. Thread the cables through the opening in the cable lid. Push the lid up and secure with the provided Allen key and screws (2).
8. Place the battery in its bracket in the control box. Check to see that the battery is secured in its position.
9. H ang the hand control on the green “button” on the mast. Two extra self-adhesive “buttons” for positioning the hand control are provided.
10. P ress the handle reinforcement into place with the text “Emergency lowering” visible on the emergency lowering handle.
11. P lace the Quick Reference Guide in the holder on the lift mast.
12. Release the emergency stop by turning the button in the direction indicated by the arrows on the button.
13. Before initial use, charge the battery by plugging the charger cable into the connector cable. Then plug the charger cable into an electrical socket (100-240 V AC). The battery charges fully in about 5 hours.
Golvo • 7EN140105-02
w w w . l i k o . com
15. Activate the IR system by pressing any button on the operation panel.
14. Place the charger cable on the intended hook on the lift mast after completed charging.
After assembly, make sure that: • the motion of the lift arm corresponds to the buttons on the hand control/operating panel; • the emergency lowering functions are working (mechanical and electrical);
• the wheel brakes are working; • the base-width adjustment works; • the batteries are charged.
Disassembly 1. Begin by removing the sling bar or other accessory that is fitted to the lift. 2. Remove the armrest holder as described below:
A. Loosen the two M8 screws in the armrest holder on either side of the lift mast.
B. Remove the armrest holder using two screwdrivers.
3. Remove the cables from the controlbox, see page 6. 4. Remove the lift mast as described below: When the lift mast has been detached from the base, it must be supported to prevent it from falling.
B. T hen screw the safety screws into the lower holes on the lift mast. This releases the lift mast from the base, and it can now be removed.
A. Unscrew both the safety screws in the upper holes on the lift mast. Golvo • 7EN140105-02
w w w . l i k o . com
Operation Activate:
Reset:
Emergency stop Activate: Press the red button on the control box. Reset: Turn the button in the direction indicated by the arrows.
Electrical emergency lowering/raising The lift arm is lowered/raised electrically by pressing one of the markings for “down” and “up”, respectively, on the operation panel.
A
Replacing the Hand Control Batteries When the indicator A on the hand control flashes (yellow), the batteries in the hand control should be replaced. Use 2 batteries of type AAA LR03 1.5 V, alkaline. Remove the back cover using a cross-head screwdriver. Replace the batteries. Refit the back cover.
Installation of latches After installation, check that the latch locks and runs freely in the sling bar hook.
Lock
Unlock
Locking the Wheels The rear wheels can be locked to prevent rotating and turning. The locking/unlocking of the wheels is done with the foot. When lifting, the wheels should be unlocked so that the lift can be moved to the patient’s centre of gravity. The wheels should be locked however, if there is a risk of the lift rolling into the patient, for instance, when lifting from the floor.
Mechanical emergency lowering Emergency lower by moving the handle marked ”Emergency lowing” up and down. Repeat the motion until the patient in the lift is lowered onto a firm surface; draw down the sling bar manually and continue to pump the handle until the sling bar is low enough to enable the sling’s strap loops to be unhooked.
Locked wheels during lifting can increase the risk of tipping. Golvo • 7EN140105-02
w w w . l i k o . com
a
Adjustment of the lift interval level If it is necessary to reach a lower level with the sling bar, this can be arranged by lengthening the lift strap using the mechanical emergency lowering device. Do not lengthen the strap more than necessary as the highest obtainable lifting level is also affected. An example it would be advantageous to lower the lift interval is when lifting from the floor using a sling whose strap loops do not reach up to the sling bar hooks when the lift is in its lowest position. Do as follows: Push down the red emergency lowering handle at the same time as loading the sling bar (pull down the sling bar with your other hand). This extends the strap and the sling bar is lowered. Repeat until the required strap length is obtained.
After mechanical emergency lowering / Restoring the lifting level If the lift strap has been lengthened due to using the emergency lowing function, the lift interval height will be lower than previously. To restore maximum lifting height, the lift strap must be reset to its original length. Do as follows: 1. Unload the strap above the emergency lowering device. Do this by either placing the sling bar above the lift arm or by another person holding the sling bar up so that the strap hangs loose. 2. L ower and raise the handle marked ”Emergency lowering” with your left hand. At the same time tension the strap by turning the black knob (a) clockwise with your right hand. Repeat this procedure until the red mark on the strap is just above the emergency lowering device. WRONG!
WRONG!
Parking the sling bar Avoid unbalanced lifting When the lift is not in use or is being moved without a load, it Always ensure that the lift strap is vertical and parallel to can be beneficial to place the sling bar in the parking panel. the lift mast when lifting. To achieve this, think of Golvo’s position under the bed. Adjust the width and/or place Golvo The parking panel is intended for parking Universal SlingBars differently under the bed to achieve the correct position. 350, 450 and 600 (all models). Do not lift when the load is unbalanced on the lift strap When the sling bar is parked in the parking panel, the as this can cause instability as well as wear the strap. lift should not be raised since this could be dangerous Unbalanced lifting poses a tipping risk! and cause personnel injury or damage to the lift if the sling bar should release and swing out from the panel.
Position of the Lift when Lifting From/To:
Bed
Golvo • 7EN140105-02
Chair/Hygiene chair
Floor
NOTE: Place a pillow under the patient’s head for increased performance and comfort. Always have the wheels locked when lifting from the floor.
w w w . l i k o . com
Arm rests If needed, the armrests can be lowered from rest to support position. The armrests have two purposes: to help the patient feel more secure and to facilitate moving the lift on the part of the caregiver. When using the lift to transfer a patient between rooms, the armrests should be lowered to the support position!
HandControl IR
Before Use Always enable the HandControl IR with the lift in question, see Enable IR control. The hand control has 6 buttons: Up Down nable IR control: E Hold the HandControl IR approx. 20 cm (8 in) from the control box’s IR receiver and press the hand control’s large Up and Down buttons simultaneously for approx. 2 sec. When a sound signal is emitted, the function is enabled. The lift is now ready to be used with the HandControl IR.
Up (lower speed) Down (lower speed) Base widening Base narrowing
If the lift has not been used for more than 24 hours, the system must be enabled by pressing any button on the control box operation panel before the HandControl IR can be used. One HandControl IR can control different patient lifts from Liko, equipped with the same IR system. For more information, read the section ”Communication Mode” on page 11.
Operation Panel The lift can be operated directly from the operation panel on top of the control box. On the operation panel there are a number of indicator lights with corresponding symbols; see the description below. Operation Panel Indicator Lights:
Hand control enabled (green light) - illuminates when the control box is receiving signals from the hand control.
Overload (orange light) - illuminates if the lift’s recommended max. load is exceeded, and the lift stops. The indicator turns off after 3 sec. and the lift operates again. Read more about Maximum load on page 12.
The operation panel has 6 buttons: Up Down Up (lower speed) Down (lower speed)
Base widening Base narrowing
Golvo • 7EN140105-02
Service indication (orange light) - illuminates when service is needed. Contact the service technician authorized by Liko/Hill-Rom. ELT, Expected Life Time (red light) - when the light flashes during lifting, the Liko Diagnostic System™ indicates that the expected life time for the lift is approaching its end. Contact the service technician authorized by Liko/Hill-Rom.
If the ELT proceeds to shine continuously when lifting, the lift should be taken out of service. Contact the service technician authorized by Liko/Hill-Rom. 10
w w w . l i k o . com
Communication Mode (This lift is delivered in unique communication mode.) The lift has two different communication modes: Unique Communication Mode: The lift will only be controlled by the HandControl IR (one or more), which is enabled for IR control of the lift. Open Communication Mode: More than one lift can be controlled from one (or several) HandControl IR. Important! If more than one IR-controlled lift is used in the same room, each lift must be set to the unique communication mode in order to avoid several lifts from being affected simultaneously by one HandControl IR only. Switching Communication Mode. When necessary to switch or confirm the communication mode, the two large Up and Down buttons on the operation panel of the lift should be used. Press both buttons simultaneously for approx. 2 sec. Await an acoustic and a visual signal from the operation panel. Unique Communication Mode is indicated by: an acoustic and visual signal. Open Communication Mode is indicated by: two acoustic and visual signals. Finish by enabling the IR control with the HandControl IR, see page 10.
Charging the Batteries 2 3 4
1
Battery Capacity Check the battery capacity by pressing the i button (1). The following information is available: 2 - Green light: More than 50 % of the battery capacity remains. 3 - Orange light: 25–50 % of the battery capacity remains. 4 - Orange light: 0–25 % of the battery capacity remains. The indication lamps light up for about half a minute; they then go out. The battery capacity is also shown when the lift is in use.
To obtain max. service life, it is important to charge the batteries regularly. We recommend charging after use of the lift or every night. Maximum charge is obtained after about 5 hours. When the batteries are fully charged, the charger switches to trickle charging automatically. During charging the lights shine alternatively (orange, orange, green). For deeply discharged batteries, the indication for initiated charging can be delayed a few minutes. Completed charging is indicated by a flashing green light. When the operation time is noticeably shortened, the batteries probably need changing. Stop charging and switch batteries. Never charge batteries in a wet area. The lift cannot be charged with the emergency stop engaged. NOTE! If the lift is not to be used for a long period, the battery should be connected to charging.
Charging With built-in charger: heck that the charger cable is connected to the control box outlet; see p. 6. C Connect the charger cable to an electrical outlet (100-240 V AC). Completed charging is indicated by a flashing green light on the battery. If the charger cable is permanently stretched, it should be replaced in order to minimize the risk of the cable getting stuck and breaking.
NOTE! The lift cannot be used when the charger cable is plugged into a wall socket. Old batteries are to be deposited at the nearest recycling station or given to personnel authorized by Liko.
Golvo • 7EN140105-02
11
w w w . l i k o . com
Maximum Load
Different maximum loads may apply to different products on the assembled lift unit: lift, sling bar, sling and any other accessories used. For the assembled lift unit, the maximum load is always the lowest maximum load rating for any of the components. For example, a Golvo which is approved for 200 kg (440 lbs) can be equipped with an accessory which is approved for 300 kg (660 lbs). In this case, the maximum load of 200 kg (440 lbs) applies to the assembled lift unit. Study the markings on the lift and lifting accessory, contact your Liko/Hill-Rom representative if you have any questions.
Recommended Lifting Accessories
Using lifting accessories other than those approved can entail a risk. Generally recommended sling bars and accessories for Golvo mobile lifts are described below. When changing a sling bar or other lifting accessories, the highest possible lifting height of the lift is affected. Before changing lifting accessories you should always ensure that the lift, after change, can fulfill the desired lifting height in order to manage the lifting situations for which the lift is to be used. To select suitable slings and other lifting accessories, refer to the brochures “Golvo” and “Lifting accessories”. For additional guidance in selecting a sling, refer to the instruction guides for the respective sling models. Here, you will also find guidance for combining Liko’s sling bars with Liko’s slings. Contact your Liko/Hill-Rom representative or visit www.liko.com for advice and information on Liko’s product range. SlingBar Mini 220 Prod. No. 3156005 Max 205 kg (450 lbs)
Universal SlingBar 350* Prod. No. 3156074 Max 300 kg (660 lbs)
Universal SlingBar 450* Prod. No. 3156075 (Standard on Golvo) Max 300 kg (660 lbs)
Universal SlingBar 600* Prod. No. 3156076 Max 300 kg (660 lbs)
Universal TwinBar 670* Prod. No. 3156077 Max 300 kg (660 lbs)
Universal SideBars 450 Prod. No. 3156079 including bag Max 300 kg (660 lbs)
Sling Cross-bar 450* Prod. No. 3156021 Max 300 kg (660 lbs)
Sling Cross-bar 670* Prod. No. 3156018 Max 300 kg (660 lbs)
SlingBar Cover Paddy 30 Prod. No. 3607001 (fits Universal SlingBars 350, 450 and 600, as well as SlingBar Slim 350) * also available equipped with Quick-Release Hook Golvo • 7EN140105-02
12
w w w . l i k o . com
Bag for SlingBars
Prod. No. 2001025
Quick-Release Hook Liko’s Quick-Release Hooks are a system for quick change of lifting accessories on Liko’s mobile and stationary lifts. The Golvo must be equipped with Q-link in order to be used with the Quick-release Hook. The Quick-Release Hook Universal fits the Universal Bars 350, 450 and 600 (prod. no. 3156074 - 3156076). Quick-Release Hook TDM fits the SlingBar Mini 220 (prod. no. 3156005), Sling Cross-bar 450 and 670 (prod. no. 3156021 and 3156018) and Universal TwinBar 670 (prod. no. 3156077). See the “Guide to Liko’s Quick-Release Hook System”, which can be downloaded from our website www.liko.com, or contact your Liko/Hill-Rom representative for more information on the advantages of the Quick-Release Hook system.
Q-link Prod. No. 31590005
Quick-release Hook Quick-release Hook Universal TDM Prod. No. 3156508 Prod. No. 3156502
Standaid set for Golvo Prod. No. 2001020 The Standaid set is intended for shorter transfers of patients with sufficient physical stability, for example, between wheelchair and toilet or between bed and wheelchair. The patient is lifted to a half standing position using a Liko SafetyVest or Liko SupportVest. See the applicable instruction guide for usage of Liko SupportVest and Liko SafetyVest.
Stretchers Most stretchers in Liko’s product range can be used in combination with Golvo. Contact your local Liko/Hill-Rom representative for more information. Scale For weighing patients in combination with Golvo, we recommend using LikoScale 350 Max 350 kg (770 lbs). LikoScale 350 is certified according to the European Directive NAWI 90/384 (Non-Automatic Weighing Instruments). Contact your local Liko/Hill-Rom representative for more information. LikoScale 350
Support Springs for Jump/Gait Training Elastic springs are available as an accessory in order to produce a softer, springy motion, for example, during gait training. The springs are available in three different versions: Long, 28 cm (11 inch.), max 70 kg (154 lbs) Prod. No. 3156511 user weight/pair Short, 22 cm (8.6 inch.), max 70 kg (154 lbs) Prod. No. 3156512 weight/pair Short, 22 cm (8.6 inch.), max 100 kg (220 lbs) Prod. No. 3156513 weight/pair See the instruction guides for Liko MasterVest, models 60 and 64 or Liko LiftPants model 92, for more information.
Golvo • 7EN140105-02
13
Prod. No. 3156228
w w w . l i k o . com
Leg Protector set Golvo Grey, pair, fits: Golvo 7000, 8000 Grey, pair, fits: Golvo 7007, 8008
Prod. No. 2006011G Prod. No. 2006012G
Cord for HandControl IR
Prod. No. 2006030
Wall-mounted Battery Charger
Prod. No. 2004108
Extra battery
Prod. No. 2006107
Golvo • 7EN140105-02
14
w w w . l i k o . com
Simple Troubleshooting The lift cannot be controlled via HandControl IR.
1. Check that the emergency stop has not been activated. 2. M ake sure that the charging cable is not connected to an electric outlet. 3. Check if HandControl IR indicates battery change. 4. If the lift has been inactive for 24 hours, enable the system, press optional button on the operation panel. 5. Make sure that the HandControl IR control is enabled. 6. If the problem remains, see “The lift cannot be controlled via the operation panel”, below.
The lift cannot be controlled via the operation panel.
1. Make sure that the emergency stop has not been activated. 2. Check that the charger cable is not connected to an electric outlet. 3. Check the battery capacity. 4. M ake sure that the cables to the control box are connected correctly. 5. Check that the battery’s contact plates are not defective or broken. 6. If the problem remains, please contact Liko/Hill-Rom.
The charger doesn’t work.
1. Make sure that the emergency stop has not been activated. 2. Make sure that the battery is properly attached. 3. C heck that the battery pack’s contact plates are not defective or broken. 4. If the problem remains, please contact Liko/Hill-Rom.
The lift is stuck in the high position.
1. Make sure that the emergency stop has not been activated. 2. Check the battery capacity. 3. HandControl IR indicates battery change. 4. Use the indicated electrical emergency lowering to lower the patient onto a firm surface. 5. Use the indicated mechanical emergency lowering to lower the patient onto a firm surface. 6. If the problem remains, please contact Liko/Hill-Rom.
The lift does not reach maximum lifting height.
1. Check that the lifting interval level is correctly set. 2. If the problem remains, please contact Liko/Hill-Rom.
If you hear unusual sounds.
Contact Liko/Hill-Rom.
Golvo • 7EN140105-02
15
w w w . l i k o . com
Inspection and Maintenance Care and Maintenance For trouble-free use, certain details should be checked each day the lift is used: • Inspect the lift and check to make sure that there is no external damage. • Check the sling bar attachment. • Check the lift strap for wear, and ensure that the strap is not twisted. • Check the functionality of the latches. • Check the integrity of the lifting motion and the base-width adjustment. • Check that the lifting interval level is set correctly and that the emergency lowering is working as it should (both the electrical and the mechanical). • Charge the batteries each day the lift is used and make sure the charger works. When necessary, clean the lift with a moist cloth, using warm water or disinfectant and check that the wheels are free from dirt. NOTE! Do not use cleaning agents that contain phenol or chlorine, since these can damage aluminium and polyamide material. The lift should not be exposed to running water. Service A periodic inspection of the lift should be carried out at least once a year. Periodic inspection, repair and maintenance may be performed only in accordance with the Liko service manual by personnel authorized by Liko and using original Liko spare parts.
Expected Life Time The product has an expected life time of 10 years when correctly handled, serviced and periodically inspected in accordance with Liko’s instructions. Transport and Storage During transportation, or when the lift is not to be used for a long time, the emergency stop should be engaged. The environment where the lift is transported and stored should have a temperature of 10 to 40 °C (50–104 °F) and a relative humidity of 30 to 75 %. The air pressure should be 700–1060 hPa. Recycling For instructions on how to recycle your Liko product, please visit our website: www.liko.com. Product Changes Liko’s products undergo continuous development, which is why we reserve the right to make product changes without prior notice. Contact your Liko/Hill-Rom representative for advice and information about product upgrades. Design and Quality by Liko in Sweden Liko is quality certified in accordance with ISO 9001 and its equivalent for the medical device industry, ISO 13485. Liko is also certified in accordance with environmental standard ISO 14001.
w w w . l i k o . com Manufacturer: Liko AB SE-975 92 Luleå Sweden [email protected]
© Copyright Liko AB 2011-03
Service Agreement Liko offers the opportunity to enter into service contracts for the maintenance and regular inspection of your Liko product.