Liko
Viking XS Instruction Guide Oct 2007
Instruction Guide
12 Pages
Preview
Page 1
Viking XS Instruction Guide English 7EN139102-02 2007-10-19 Applies to the following models: Viking XS Prod. No. 2040007
Product Description Viking XS is an easily manoeuvred mobile lift, especially developed for use in environments with narrow doorways or otherwise restricted space. Viking XS has electrical raising and lowering of the lift arm as well as mechanical base-width adjustment. Viking XS is appropriate for all common lifting situations; e.g., transfers between bed and wheelchair,
to and from the toilet and bathtub or for lifting to and from the floor. For optimal function and safety, it is essential that the lift and any accessories you intend to use be individually tested.
In this document, the person being lifted is referred to as the “patient” and the person helping them is referred to as the “caregiver”. “Caution!” triangles are used to warn of situations that demand extra care and attention.
CAUTION! Carefully read instructions for the lift and lifting accessories before use. Lifting and transferring a person always presents a potential risk. A thorough understanding of the contents of this instruction guide is essential. Only trained personnel should use the equipment. If you have questions, please contact Liko or your local Liko representative.
Contents Safety Instructions...3 Definitions...3 Technical Data...4 Measurements...4 Assembly... 5-6 Operation...7 Charging the Batteries...8 Maximum Load...9 Recommended Lifting Accessories... 9-10 Simple Troubleshooting...11 Care and Maintenance...12
NOTE! This instruction guide contains important information about the use of the product. All personnel who use this product must be thoroughly familiar with the contents of this instruction guide. Remember to keep the instructions where they are easily accessible for users of the product.
Viking XS • 7EN139102-02
www.lik o.com
Safety Instructions
Before using the lift, make certain that: • the lift is assembled according to the instructions • lifting accessories are correctly applied to the lift • the batteries have been charged for at least 10 hours • you have read and understood the instruction guides for the lift and lifting accessories • personnel using the equipment have received appropriate instructions and training • you have selected the correct type, size, material, and design of slings and accessories to safely meet the patient’s needs. Before lifting, always make certain that: • the lifting accessories are not damaged • the lifting accessory is correctly and securely applied to the patient, so that no personal injury can occur • the lifting accessory is correctly applied to the lifting equipment • the sling’s strap loops are correctly fastened to the slingbar hooks when the sling strap is extended, but before the patient is lifted from the underlying surface. Never leave a patient unattended in a lifting situation!
Viking XS has been tested by an accredited testing institute and complies with the requirements of the directive for Class 1 medical-technical products (MDD 93/42/EEC). Viking XS complies with the requirements of IEC 60601-1, IEC 60601-1-2, EN ISO 10535, UL 60601-1 and CAN/CSA C22.2 No. 601.1. Particular caution must be observed in connection with the use of strong sources of electromagnetic interference such as diathermy equipment and similar devices. Ensure that, e.g., diathermy cables are not placed near the lifting equipment. In the event of uncertainty, consult the responsible technician or supplier. Equipment not suitable for use in the presence of flammable mixtures. Maximum load: 160 kg (350 lbs.)
Definitions 15
1
14 2 13
12
3 11 4
XS
5
S
MS
M
L
XL
XX L
10 9 8
7
16
1. Lift arm 2. Flexlink 3. Slingbar with safety latches 4. Motor/actuator for lift mast 5. Product decal 6. Front wheels 7. Base 8. Rear wheels with brakes 9. Pedal for base-width adjustment 10. Locking handles 11. Control unit with battery, emergency stop and electrical emergency lowering 12. Handcontrol 13. Handles 14. Lift mast 15. Emergency lowering (mechanical) 16. Holder for Quick Reference Guide with colour codes for sling sizes
6
Viking XS • 7EN139102-02
www.liko.com
Technical Data Lifting speed: 3.0 cm/s without load Batteries: 2 pcs. 12V 2.9 Ah valve-regulated lead-acid gel-type batteries. New batteries available from supplier. Battery External charger, 100-230 VAC, charger: 50-60 Hz, max. 220 mA Motor (liftarm): 24V, permanent magnetic motor with mechanical safety mechanism and safety nut. Material: Aluminium
Emergency lowering: Wheels: Intermittent operation: IP class:
Mechanical (patented) and electrical Front: 75 mm twin castor wheels. Rear: 75 mm brake-equipped twin castor wheels. Int. Op. 10/90, active operation max 2 min. Out of a time of 100, active must be less than 10, though not more than 2 min. IP X4
Built for indoor operation. Type B, according to the degree of protection against electric shock.
Measurements G
A L
E
B1
F F1
D C
B
A: Maximum load and weight in kg. Measurement in mm. B: Maximum load and weight in Ibs. Measurement in inch. Viking XS
Max. load
A
160
B
350
Max
L
Min
Max
1550
410
1780
61
16
70
Viking XS • 7EN139102-02
A
B
B¹
C
D
E
F
F¹
G
Weight total part
1190
1100
760
540-950
410-820
960
110
15
750
25
14
1170
47
43
30
21-37
16-32
38
4.3
0.6
29.5
55
31
46
Min
Turningdiameter
www.lik o.com
Assembly Before assembly, make sure you have the following: • L ift mast and lift arm, motor for lift arm, slingbar and control unit • Base • Handcontrol with cable
• Locking handles (1 pair) • Bag containing Instruction Guide and Quick Reference Guide • Battery charger
1. Lock both rear wheels. Place the lift mast in the socket on the base.
2. Using the accompanying locking handles, secure the lift mast in the base. Adjust the handles so that they point downward.
XS
MS
M
L
XL
XX L
4. Place the Quick Reference Guide in the holder on the control unit.
3. Connect cables as follows: - Cable handcontrol to outlet 1 - Cable motor for lift arm to outlet 2
Viking XS • 7EN139102-02
S
www.liko.com
5. Reset the emergency stop by turning the button in the direction indicated by the arrows.
After assembly, check to ensure that: • wheel brakes work properly • the indicator lamp on the front of the control unit lights during charging.
• lift mast motions correspond with the buttons on the handcontrol • emergency lowering works properly (mechanical and electrical)
Viking XS • 7EN139102-02
www.lik o.com
Operation Operation Viking XS is operated by pressing the buttons on the handcontrol. For raising and lowering the lift arm: press and , respectively. Lifting motion stops as soon as the push-button is released. If the battery needs charging, a lamp (A) is lit on the handcontrol. There is, however, enough power left in the battery for a few more lifts.
Base-width adjustment The base width can be increased and decreased mechanically by pressing the foot pedal. Press the right side to widen the base and press the left side to make the base narrower.
Electrical emergency lowering The lift arm is lowered electrically by pressing “down” ( control unit.
) on the
Mechanical emergency lowering Mechanical emergency lowering is performed by turning the red emergency-lowering cylinder in the direction of the arrows.
Emergency stop For emergency stop: Press the red Emergency stop button. To release/reset: Turn the button in the direction of the arrows.
Locking the wheels Press down the locking lever on the wheel. The wheel locks for rotation and turning. In most lifting situations, the wheels should not be locked. We recommend that wheels be locked only when there is a risk for personal injury if the lift should start to roll, e.g., when lifting a person from a horizontal position on the floor. Release the lock by pressing the lever up. . Locked wheels during lifting increases the risk of the lift tilting over. Never pull the lift by the actuator! This may result in damage to the lift and possible injury.
Viking XS • 7EN139102-02
www.liko.com
Charging the Batteries A display on the control unit shows current battery capacity. When the handcontrol is activated, a diode/lamp indicating battery capacity lights up for 10 seconds. A Green: min. 50% B Yellow: 25-50% C Yellow: 0-25% D The lamp indicates that charging/maintenance charging is in progress. This lamp is always illuminated when the charger is connected.
A B C D
When the yellow lamp (C) lights and the power of the battery has become too low, the lift will automatically shut off, this in order to protect the battery from being completely discharged. The lift then has to be charged before it can be used again. To ensure maximum battery life, batteries must be charged regularly. We recommend charging after the lift has been used, or each night. If the lift is not in daily use, we recommend that the Emergency Stop button be pushed in after use to break the current and save battery power. NOTE! Do not charge batteries in a wet area!
Maximum charge is reached after about 12 hours. When the batteries are fully charged, the charger switches automatically to maintenance charging. To check battery condition, activate the handcontrol after maximum charging and read the control unit display. The green lamp indicates good battery condition. If either of the yellow lamps is illuminated, batteries should be replaced.
Charging Procedure
With external charger: 1. Connect charger cable to outlet for charger, under the control unit. 2. Connect charger to power supply (100-230 VAC).
1
2
NOTE! The lift must not be used while connected to the battery charger!
Old batteries are to be left at the nearest recycling station or given to personnel authorized by Liko.
Viking XS • 7EN139102-02
www.lik o.com
Maximum Load
Different maximum allowable loads may apply to different products on the assembled lift system: lift, slingbar, sling and other accessories. For the total lift system, the lowest max. allowable load indicated for the respective products on the system always applies. For example: A Viking XS that is approved for 160 kg (350 lbs.) may be equipped with a slingbar that is approved for 300 kg (660 lbs.). In this case, the applicable max. load is 160 kg (350 lbs.) for the total lift system. Study markings on the lift and lifting accessories or contact your Liko representative if you have any questions.
Recommended Lifting Accessories
Slingbars and accessories recommended for use with the Viking series of mobile lifts are described below. Changing slingbars and adding extra accessories affects the maximum lifting height of the lift. Before changing slingbars and accessories, it is important to ensure that it will still be possible to achieve the desired lifting height for situations in which the lift will be used. For choice of appropriate slings and other lifting accessories, see e.g., the brochure ”Lifting accessories”. For further guidance on choosing a sling, consult the instruction guide for the respective sling model. Here, you will find advice on suitable combinations of Liko slingbars and Liko slings. Contact your Liko representative or visit www.liko.com for advice and information on Liko’s product range.
Slingbar Mini 220 Max 205 kg / 450 lbs.
Prod. No. 3156005
Universal Slingbar 350 Max 300 kg / 660 lbs.
Prod. No. 3156074
Universal Slingbar 450 (Standard for Viking XS) Max 300 kg / 660 lbs.
Prod. No. 3156075
Universal Slingbar 600 Max 300 kg / 660 lbs.
Prod. No. 3156076
Universal Twinbar 670 Max 300 kg / 660 lbs.
Prod. No. 3156077
Universal Sling Sidebars 450 incl. storage bag Max 300 kg / 660 lbs.
Prod. No. 3156079
The following slingbars can only be used to a limited extent (applies to Viking XS) Sling Cross-bar 450 Max 300 kg / 660 lbs.
Prod. No. 3156021
Sling Cross-bar 670 Max 300 kg / 660 lbs.
Prod. No. 3156018
Viking XS • 7EN139102-02
www.liko.com
Quick-release Hook Liko’s Quick-release Hook system enables quick and easy exchange of lifting accessories on Liko’s mobile and stationary lifts. Viking can be equipped with Q-link 13 for use with Quick-release Hook, which is mounted on the lifting accessory. Quick-release Hook Universal fits Universal Slingbar 350, 450 and 600 (Prod. No. 3156074-3156076). Quick-release Hook TDM fits Slingbar Mini 220 (Prod. No. 3156005), Sling Cross-bars 450 and 670 (Prod. No. 3156021 and 3156018) and Universal Twinbar 670 (Prod. No. 3156077). When a slingbar mounted with the Quick-release Hook system is used, the lifting height will be 33 mm shorter than with a permanently mounted slingbar. See ”Guide to Liko´s Quick-release Hook System”, which can be downloaded from our website, www.liko.com, or contact Liko for more information on the advantages and use of the Quick-release Hook system.
Storage bag for slingbars
Quick-release Hook Universal Prod. No. 3156508
Quick-release Hook TDM Prod. No. 3156502
Q-link 13 Prod. No. 3156509
Prod. No. 2001025
Slingbar cover Paddy 30 Prod. No. 3607001 (fits Universal Slingbar 350, 450 and 600 and Slingbar Slim 350)
Lever for base width adjustment
Prod. No. 2037010
Lever for base-width adjustment. Mounted on existing base-width pedals.
Cable reel EU, UK, AU/NZ Prod. No. 2107104-6 Accessory that enable assembly of external charger and charger cable on the lift.
Scale for Viking When mounting LikoScale on Viking XS the lifting height is decreased with 132 mm, therefore we do not recommend Viking XS to be used in combination with a scale.
0
LikoScale 350 Prod. No. 3156228
Viking XS • 7EN139102-02
10
www.lik o.com
Simple Troubleshooting The lift does not work up/down
1. Check that the emergency stop button is not pushed in (p. 7). 2. Check that the cables are connected correctly (p. 5). 3. Check that the charging cable is not connected to the wall socket. 4. Check that the battery is charged (p. 8). 5. If the lift still does not work satisfactorily, contact Liko.
Battery charging does not work
1. Check that the emergency stop button is not pushed in (p. 7). 2. Check to ensure that the wall socket is power-supplied. 3. Check to ensure that the charging cable is correctly connected (p. 8). 4. If the lift still does not work satisfactorily, contact Liko.
The lift stops in the elevated position
1. Check that the emergency stop button is not pushed in (p. 7). 2. U se the electrical emergency lowering device to safely lower the patient (p. 6). 3. Use the mechanical emergency lowering device (marked) to lower the patient (p. 7). 4. Check that the battery is charged (p. 8). 5. If the lift still does not work satisfactorily, contact Liko.
If you hear unusual sounds
Contact Liko
Viking XS • 7EN139102-02
11
www.liko.com
Care and Maintenance For safe and trouble-free operation, a few routine procedures should be performed every day the lift is used: • Visually inspect the patient lift and check for external damage or wear. • Check that all screws and lock nuts on the slingbar are tight. • Check that the safety latches on the slingbar work properly. • Check the handcontrol and the lift and width movements. • Check the emergency lowering function (both electrical and mechanical). • Charge the battery every day the lift is used, and check charger function. When necessary, clean the lift with a moist cloth, using common surface cleaners or disinfectants and check to ensure that the wheels are free of dirt and hair. NOTE! Do not use cleaning agents that contain phenol or chlorine, since these can damage aluminium and polyamide material. Transportation and storage During transportation, or when the patient lift is not to be used for some time, the emergency stop button should be pushed in. The environment where the patient lift is transported and stored should have a temperature between 10˚C and 40˚C and a humidity between 30% and 75%. The air pressure should be between 700 and 1060 hPa. Inspection Viking XS must be inspected at least once per year. Pay particular attention to parts that are subject to wear. Inspection must be performed according to the Liko service manual and by authorized personnel.
Service agreements Liko invites you to sign a service agreement for regular maintenance and testing of your Liko product. Contact your local Liko representative for more information. Liko's products are constantly being updated and refined. Liko reserves the right to change aspects of the products without prior notice. Contact your local Liko representative for updated information and advice. Recycling For instructions on how your Liko product should be recycled, please visit our website www.liko.com. Made in Sweden Liko is quality certified according to ISO 9001 and its equivalence for the medical device industry, ISO 13485. Liko is also certified according to environmental standard ISO 14001.
www.liko.com Manufacturer
National Distributor in USA and Canada:
Liko Distributor in UK:
Liko AB Nedre Vägen 100 SE-975 92 Luleå Sweden [email protected]
Liko North America 122 Grove Street Franklin, Ma 02038, USA Phone: 508-553-3993, 1-888-545-6671 Fax: 508-528-6642 info @ l i k o . c o m
Liko (UK) Ltd. Brunel Centre, Stroudwater Business Park Stonehouse, Gloucester GL10 3RU, United Kingdom Phone: +44 (0) 1453 827272 Fax: +44 (0) 1453 828844 [email protected]
© Copyright Liko AB 2007-10
IMPORTANT! Repairs and maintenance may only be carried out by personnel authorized by Liko using original Liko spare parts.