MAICO
MA 50 Operating Instructions Oct 2004
Operating Instructions
28 Pages
Preview
Page 1
Operating Instructions MA 50
MAICO Diagnostic GmbH, Salzufer 13/14, D 10587 Berlin, Tel.: ++4930/70 71 46 50, Fax: ++4930/70 71 46 99 MAICO Diagnostics, 7625 Golden Triangle Drive, Eden Prairie, MN 55344, Toll Free ++1/888.941.4201, Fax ++1/952.903.4200
MAICO
Operating Instructions MA 50
Table of Contents
Page
1. Introduction
2
2. Description
3
3. Getting started 3.1 Unpacking your instrument 3.2 Preparing the MA 50 for use 3.3 Getting familiar with your MA 50 3.31 Using the keyboard of the MA 50 3.32 The display of the MA 50
4 4 5 6 7 7
4. Pure tone audiometry 4.1 Air Conduction (AC) Testing 4.2 Uncomfortable Hearing Level (UCL) Testing 4.3 Bone Conduction (BC) Testing 4.4 Masking of the opposite ear
8 8 9 10 10
5. Quick reference guide
13
6. Recommended literature
14
7. Computer interface
15
8. Care and maintenance of the instrument
16
9. Trouble shooting
17
10. Technical data
18
11. Warranty, maintenance and after-sales service
22
12. Safety regulations
23
13. List for subjective instrument check
24
14. Certificate of conformity
25
15. Front panel layout and rear panel connections
27
GEBA50_e.04a.wpd
1
851406/5 10/04
MAICO
Operating Instructions MA 50
1 Introduction Thank you very much for purchasing a quality product from the MAICO family. This MA 50 audiometer is manufactured to meet all quality and safety requirements, and has been certified with the CE-symbol according to Medical Directive 93/42/EEC. In designing the MA 50 we placed particular importance in making it a userfriendly device, meaning its operation is simple and easy to understand. And because MA 50 functions are software controlled, upgrading later to new, extended measurement functions will be simple and inexpensive. That means that you have invested in a device that will adjust to your future needs. This user manual should make it as easy as possible for you to become familiar with the functions of the MA 50. Please open out the flap of illustrations on the last page. The description of the position (e.g. ! ) of controls, displays and connections, found again in the text, will make it easier for you to learn how to operate the MA 50. If you have problems or have ideas for further improvements, please get in touch with us. Simply call. Your MAICO-team
GEBA50_e.04a.wpd
2
851406/5 10/04
MAICO
Operating Instructions MA 50
2 Description The MA 50 is an audiometer for pure tone audiometric tests. It can be used for ENT diagnostics in the office, and for mobile audiometry in clinics, homes and abroad. Tests can be performed using the TDH 39 headphones (AC), B 71 bone conduction receiver (BC) or optional insert phones. Built-in test signals are pure tone, pulse tone, warble tone and narrow band noise. The MA 50 audiometer delivers 11 air conduction (AC) test frequencies from 125 Hz to 8 kHz, with levels from -10 dBHL to 120 dBHL. Bone conduction (BC) can be tested with 11 test frequencies from 125 Hz to 8 kHz with levels from -10 dBHL to 70 dB HL (with the standard bone conductor B 71 the frequency range is limited from 250 Hz to 6 kHz). The optional insert phones EAR 3A submit levels from -10 dBHL to 120 dBHL with 11 test frequencies from 125 Hz to 8 kHz. The hearing level is controlled independently for the signal and the masking channel by two detented knobs on the left and right of the instrument which can be operated from the side or the top. The level steps are 5 dB. The signal STIM bar and the frequency up/down keys are just beside the level control knobs for easy one handed control of level, frequency and signal presentation. The bright LED-display shows level and frequency for each channel. As a result of modern microprocessor technology, the MA 50 is easy to use and is extremely reliable. The audiometer is designed to be extremely service friendly. Automatic test programs make trouble shooting and the yearly calibration as easy as possible. PC-Interface: A serial RS 232C interface for data transfer to a connected computer is built in. The optional MAICO-audiometry module for NOAH enables the automatic data transfer of all test results of the MA 50, like tone audiogram. For more information see chapter 7.
GEBA50_e.04a.wpd
3
851406/5 10/04
MAICO
Operating Instructions MA 50
3 Getting started 3.1 Unpacking your instrument Your MA 50 was carefully inspected and packed for shipping. However, it is a good practice to thoroughly inspect the outside of the shipping box for signs of damage. If any damage is noted, please notify the carrier immediately. Carefully remove the instrument from the shipping box. Remove the plastic bag from the instrument and inspect the case for any damage. Notify the carrier immediately if any mechanical damage is noted. This will assure that a proper claim is made. Save all packing material so the claim adjuster can inspect it as well. Notify your dealer or MAICO when the adjuster has completed the inspection. SAVE ALL THE ORIGINAL PACKING MATERIAL AND THE SHIPPING CONTAINER SO THE INSTRUMENT CAN BE PROPERLY PACKED IF IT NEEDS TO BE RETURNED FOR SERVICE OR CALIBRATION. All accessories are already packaged in the compartment connected with the MA 50. Please check that all accessories listed below are received in good condition. If any accessories are missing or damaged, immediately notify your dealer or MAICO. Standard accessories: 1 Headphone TDH 39 1 Bone conduction receiver B 71 with headband 1 Patient response switch 1 Power cable 1 Audiogram form (50 sheets) Calibration of the device: The instrument, headphones, bone conduction receiver as well as the optional insert phone match one another and have the same serial number (e.g. 6631520). Because they have been calibrated with this particular instrument, use of other transducers is not allowed. If one of the acoustic transducers is replaced, the instrument must be recalibrated. The use of non-calibrated audiometers leads to incorrect measurements!
GEBA50_e.04a.wpd
4
851406/5 10/04
MAICO
Operating Instructions MA 50
3.2 Preparing the MA 50 for use The MA 50 with its integrated compartment for the accessories is perfect for portable use. The rugged housing, light weight and the comfortable handle make it easy and convenient to transport the instrument. Carry it, as shown, with the bottom away from the leg. That helps protect the front cover from damage, and due to the asymmetrical handle position it provides more space for your leg. To get started first move the handle under the housing. Unlatch the cover by pressing in the two black locks located on the left and right sides near the front of the instrument. Open the front cover and rest Picture 1 Transport of MA 50 it on the back cover of the accessory compartment. To open the compartment, press the two black locks in the upper middle of the instrument. Open the compartment cover as seen in picture 2. Take the headphone, the bone conduction receiver, the patient response switch and the microphone out of the compartment and connect the power cable to power. The MA 50 operates with voltages from 96 to 240 V~ AC, 44 Hz to 440 Hz.To avoid pinching the cables when closing the cover, lay the cables in the slots Picture 2 Portable with provided. Close the back cover and latch it with slight open cover pressure. The MA 50 should be operated in a quiet room, so that the audiometric examinations are not influenced by outside noises. For use in noisier environments headphones with optional sound insulation muffs are available. Electro-medical instruments, which emit strong electromagnetic fields (e.g. microwaves - radiotherapy devices), can influence the function of the audiometer. Therefore the use of these instruments is not allowed in close proximity to the audiometer. Picture 3 MA 50 in operation The test room must be at normal temperature, usually 15/ C / 59/ F to 35/ C / 95/ F, and the instrument should be switched on about 10 minutes before the first measurement to guarantee precise measuring results. If the device has been cooled down (e.g. during transport), please wait until it has warmed up to room temperature.
GEBA50_e.04a.wpd
5
851406/5 10/04
MAICO
Operating Instructions MA 50
3.3 Getting familiar with the MA 50 Turn on the instrument with the power switch " which is located at the right side of the MA 50. The - 10 1. 0 0 0 30 device performs its initial test and autocalibration. If an error is detected the test is stopped and the code of the error is shown on the LED-display #. In this case please contact your local dealer or service. If the test is passed, the instrument is MAICO setup to air conduction and pure tone for the right ear and narrow band 16 18 26 10 masking noise for the left ear. The corresponding LED ' s above the Picture 4 MA 50 power switch, display and receiver $ and the test channel selection buttons for receiver and test channel selector % are lighted. The frequency is set to 1 kHz and the level to 30 dBHL for the right test channel and to -10 dB 1. 0 0 0 30 for the left masking channel. All these - 1 0 settings are also shown on the display # (see picture 5). Picture 5 MA 50 display with initial settings The hearing level can be easily adjusted with the right knob & and the masking level with the knob ' on the left side of the instrument. For optimal convenience these level control knobs - 10 1. 0 0 0 30 are usable from the top or the side of the instrument. They are detented in 5 dB steps. The STIM bars ( and ) are located beside the level control knobs ', &. The STIM LED * light up when you press the STIM bar ) to present the test signal. The STIM LED MAICO + lights continuous because the the masking noise is continuous presented. 1 3 5 23 25 24 14 The frequency is changed with up , 2 4 and down ! buttons. The design of the MA 50 makes it easy to control Picture 6 MA 50 level and frequency control level, signal presentation and frequency with one hand. STIM
STIM
PAT
+
FM
Hz
STIM MODE
PULSE
INSERT BC AC
TRACK
L R
MA 50
STIM
STIM
PAT
+
Hz
FM
INSERT BC AC
PULSE
STIM MODE
TRACK
L R
MA 50
GEBA50_e.04a.wpd
6
851406/5 10/04
MAICO
Operating Instructions MA 50
3.3.1 Using the keyboard of the MA 50 All main functions of the MA 50 are directly accessable by pressing a single button. Following is the description of the functions of each button: , frequency up: change to next higher frequency ! frequency down: change to next lower frequency - FM - modulation (warble tone): changes test signal from pure tone - 10 1. 0 0 0 30 to warble tone. $ receiver selector: changes the receiver from AC to BC or INSERT (if option assembled). The lighted LED above shows the current selection. MA 50 MAICO . pulse tone: enables pulsing of pure tone or 5 9 11 15 warble tone. 4 10 14 16 / stim mode: Picture 7 MA 50 keyboard changes the signal presentation from presenter to interrupter mode. 0 tracking function: enables the automatic tracking (see chapter 4.4 masking). % channel selector key: selects the right or left test ear. STIM
STIM
PAT
+
Hz
FM
INSERT BC AC
PULSE
STIM MODE
TRACK
L R
3.3.2 The display of the MA 50 The bright LED-Display # of the MA 50 shows levels and frequency. The display shows the individual levels for the left channel on the left and the right - 1 0 1. 0 0 0 30 channel on the right. The frequency is displayed in the middle. Also the Picture 8 MA 50 display for pure tone HL-test position of the dot in the middle indicates the kind of measurement. If the dot position is left from the center HL test is indicated. The dot in the center stands for UCL measurement and a dot right from the middle is MCL measurement. The example picture 8 shows the initial setting for AC threshold test. The right test channel is set to 30 dBHL , the left masking channel is set to -10 dBHL narrow band noise for masking. The test frequency is 1000 Hz and the test is a HL measurement (dot left from the middle).
GEBA50_e.04a.wpd
7
851406/5 10/04
MAICO
Operating Instructions MA 50
4 Pure tone audiometry 4.1 Air Conduction (AC) Testing In the hearing threshold test, the hearing threshold of the patient is measured in comparison with the normal hearing threshold for air conduction. The test is started on the ear with better hearing. The patient should sit at a distance of at least 1 m from the device. Eliminate any obstructions which will interfere with the placement of the earphone cushions on the ear (i.e. hair, eyeglasses). Ensure the headphones are put on correctly. Red side on the right, blue side on the left. Adjust the headband of the headphones so that the receivers are at the correct height (the sound output grid 1 exactly facing the ear canal). Start with the “better” indicated ear at 1 kHz. (After switching on, the frequency is automatically set to 1 kHz.) In the following example we assume that the right Picture 10 Headphone is the “better” ear. Set the receiver selector $ to AC. Set the left level control knob & to a value just below expected hearing loss. - 10 1. 0 0 0 30 Explain to the patient that he should press the button of the patient response switch if he just hears the test tone. Press the STIM bar ) for a certain time to present the test tone. The STIM LED * should MAICO light on. 23 25 If the patient hears the test 1 3 5 2 4 10 12 16 18 24 tone, the patient response Picture 11 MA 50 controls and display for display 2 lights red. In this case airconduction threshold test decrease the level with the level control knob &. Proceed with presenting the test signal as described before. If the patient doesn't hear the test tone, increase the level with the level control knob &. Proceed with presenting the test signal as described before. When you find a stable threshold value, note the level and frequency. If the MA 50 is connected to a PC, the value was stored with your last STIM presentation. Test through the frequencies. Starting at 1 kHz, set the higher frequencies STIM
STIM
PAT
+
Hz
FM
INSERT BC AC
PULSE
STIM MODE
TRACK
L R
MA 50
GEBA50_e.04a.wpd
8
851406/5 10/04
MAICO
-
-
Operating Instructions MA 50
first, then the lower frequencies. Use the frequency up key , to select the higher frequencies and use the frequency down key ! to select the lower frequencies. Select the next frequency, increase the level again and proceed with presenting the test signal as described before. Once all frequencies are tested choose the poorer ear and repeat the hearing threshold test. You can do this just by pressing the test ear button %. The corresponding LED L lights and you test the left ear using the controls on the right side. The masking is controlled with the level knob ' on the left side of the instrument. (For masking see chapter 4.4). The correct marks in an audiogram are: O (red) = right and X (blue) = left Pulse tone: If required, the test can also be performed with a pulsed tone. Press the PULSE button . and the pure tone will be switched 0.25 s on and 0.25 s off. Warble tone: If required, the test can also be performed with a warble tone. Press the FM button - and the pure tone will frequency modulate. The warble tone can also be pulsed as described before. For hygienic reasons it is important to clean the ear cushions of the headphone (see chapter 8).
4.2 Uncomfortable Hearing Level (UCL) Testing Testing of UCL can be measured using pure tone stimuli. The purpose is to determine the dB level at which the stimuli becomes uncomfortable to the patient. This information is valuable for determining the patient 's upper dynamic range limit for proper hearing aid fittings and for determining symptoms of recruitment. Warning! Because the UCL test uses high sound pressure levels, it is extremly important to perform this test using the utmost caution to avoid damaging the ear. To prevent the possibility of extreme discomfort by the patient, it is important to start the test with levels near the patient's MCL (Most Comfortable Level). The UCL level is described as the level between very loud and loud perception of the test signal. Press the test channel selector button % longer than 2 seconds. The dot in the middle of the LED-display 1 0. 0 0 60 # moves to the center of the - 1 0 display. (see picture 12). Picture 12 MA 50 display for UCL-test left Start as described in chapter 4.1 with a test level of 60 dBHL. GEBA50_e.04a.wpd
9
851406/5 10/04
MAICO
-
Operating Instructions MA 50
Present the tone briefly (max. 1s) If the signal was recognized by the patient as not uncomfortable increase the level and proceed as described before. If the signal was uncomfortable for the patient note the values. Proceed accordingly with other test frequencies. For hygienic reasons it is important to clean the ear cushions of the headphone (see chapter 8).
4.3 Bone Conduction (BC) Testing Bone conduction, i.e. the transmission of sound waves through the skull directly to the inner ear conways information about the function of the inner ear. For a neural hearing loss the values of air conduction (chapter 4.1) and bone conduction are the same. In this case a hearing loss of the middle ear can be eliminated.
Picture 13 Bone conductor
-
Place the bone conduction receiver so that the flat, circular side of the transducer Y is placed on the mastoid, at the noticeable ledge of the cranial bone behind the auricle (see picture 14). The other side of the headband is placed in front of the opposite ear.
-
Set the receiver selector $ to BC.
-
Perform the test the same way as for air conduction (see 4.1). Enter the measurements for all frequencies on the form, connect all points with dotted lines marked on the audiogram form as follows: > = right and < = left.
Picture 14 correct seating of the bone conductor
For hygienic reasons it is important to clean the bone conductor (see chapter 8). 4.4 Masking of the opposite ear The basics of masking are explained below. To begin testing immediately, please go directly to 4.4.3 Manual masking.
GEBA50_e.04a.wpd
10
851406/5 10/04
MAICO
Operating Instructions MA 50
4.4.1 Crossover: When measuring a pure tone audiogram you assume that the measured hearing threshold is correct. But if you recognize that sound is also transmitted through bone conduction over the entire skull it is probable that the opposite ear has also received sound. This is called “crossover”. Crossover can also occur when measuring air conduction because a small amount of air conducted sound is received by the skull and transmitted by bone. Whether the crossover signal can heard by the opposite ear depends on its inner ear function. Relevant to crossover is the sound level which is received by the opposite ear. The difference between the original test signal at the test ear and the received signal at the opposite ear is called “interaural attenuation”. For bone conduction measurement the interaural attenuation is 0 to 15 dB. Bone conduction crossover is therefore possible even with a slight difference in hearing loss between ears. Important: Please advise the patient to tell you in which ear he hears the test signal. It is thereby easier to detect crossover. 4.4.2 Masking theory: To ensure that the patient will not experience crossover you must mask the opposite ear. Masking increases the hearing threshold of the opposite ear. The masking is done with a noise signal which is transmitted by the headphone. For pure tone audiometry a narrowband noise is used. This noise changes its center frequency according to the frequency of the test signal. 4.4.3 Manual masking: If you have to mask use the common masking rules. See chapter 7 for recommend literature. Masking is an important part of practical audiometry. It is necessary to be familiar with this topic to avoid errors which would lead to a wrong diagnosis. Select the ear you like to test with the channel selector key %. For the opposite ear narrow band masking noise is automatically set. Adjust the required masking level with the levelcontrol knob '. Notice that the masking sound is continuously presented for effective masking. You can interrupt the masking signal by pressing the corresponding Picture 15 Headphone interrupter key (. GEBA50_e.04a.wpd
11
851406/5 10/04
MAICO
Operating Instructions MA 50
To mask when performing bone conduction testing, place the headphone on the opposite ear so that the receiver is at the correct height (the sound output grid 1 exactly faces the ear canal). Adjust the headband of the headphones. The receiver on the side where the bone conductor is placed should sit directly on the cheek bone. The marking for air conduction with masking should be done with the symbols 3 = the right side and 4 = Picture 16 correct seating the left side on the audiogram form. The marking for of the bone conductor bone conduction with masking should be done with the symbols [ = the right side and ] = the left side on the audiogram form. 4.4.4 Automatic masking: With the manual masking, as described before, you have to readjust the masking level every time you change the test signal level. The MA 50 has a tracking feature for easy masking. Adjust for the test signal using the level control knob &, then with the masking level control knob ' adjust for the corresponding masking level. Now press the TRACK button 0. The masking level is automatically changed if you adjust the test signal level. (e.g. if the test level was 30 dB and the masking level 50 dB and you change the test level to 45 dB the masking level is automatically adjusted to 65 dB.
GEBA50_e.04a.wpd
12
851406/5 10/04
MAICO
Operating Instructions MA 50
5 Quick reference guide 6.1 Startup settings Air conduction on both channels Pure tone with a level of 30dBHL on the right ear Narrow band noise with a level of -10dBHL on the left (masking) ear Presenter mode 6.2 Receiver selection Press the RECEIVER key . to select the receiver (AC, BC, INSERT). The corresponding LED above the key lights. Not available receivers will be skipped. After a new selection the level will be reset to 30dBHL for the signal and 10dBHL for masking. 6.3 Frequency selection Use the frequency keys , and ! to select the frequency. The maximum and minimum frequency depends on the receiver you have selected. 6.4 Warble tone (FM) Press the FM key - to activate the warble tone. Press the FM key - again to switch back to pure tone. 6.5 Pulse tone Press the PULSE key . to activate the pulse tone. Press the PULSE key . again to deactivate pulsing. 6.6 Test channel selection Press the test channel selector key % to select the left or right ear for testing. 6.7 Tracking function Press the TRACK key 0 to activate tracking. The selected test channel tracks now the masking level at the opposite ear. Press the TRACK key 0 again to deactivate tracking. 6.8 STIM Mode selection Press STIM MODE / key to switch between continuous presentation and presentation by pressing the STIM bar ). When in continuous presentation mode the STIM bars ( and ) function as interrupters. The light above the STIM MODE key / is on if continuous presentation is selected. Masking noise is always in continuous presentation. Press the STIM MODE key 5 again to deactivate continuous presentation.
GEBA50_e.04a.wpd
13
851406/5 10/04
MAICO
Operating Instructions MA 50
6 Recommended literature Audiometric Interpretation: A Manual of Basic Audiometry Lloyd, Lyle L., and Harriet Kaplan Baltimore: University Park Press, 1980
Auditory Disorders: A Manual for Clinical Evaluation Jerger, Susan, and James Jerger Boston: College Hill Press, 1981
Handbook of Clinical Audiology Katz, Jack Baltimore: William & Wilkins, 1994
Roeser's Audiology Desk Reference Roeser, Ross J. New York / Stuttgart: Thieme, 1996
Auditory Diagnosis Silam, Shlomo and Carol A. Silvermann San Diego / London: Singular Publishing Group, 1997
GEBA50_e.04a.wpd
14
851406/5 10/04
MAICO
Operating Instructions MA 50
7 Computer interface The MA 50 has a built-in serial RS 232 PC-interface for data transfer to a connected PC. The serial cable to connect the MA 50 with the PC should be a normal, non twisted 9-pin extension cable. It has a male and a female connector and all wires are direct 1:1 connected. It is also sold as monitor extension cable. Caution: The computer you are using must meet electrical safety requirements, such as IEC 950 or UL. This is to avoid electrical shock of either the patient or you. All test results are online transferred to the PC and stored in the database. S In pure tone audiometry initially 1. 0 0 0 3 0 all measured data is transferred - 1 0 as hearing threshold data. This is Picture 17 MA 50 display with indicated by the dot left from the pure tone threshold setting center of the LED-display # (see picture 17). To test the uncomfortable level 1 0. 0 0 8 0 (UCL) press the test channel - 1 0 selector key % longer than 2 Picture 18 MA 50 display with seconds and the dot moves to UCL setting the center of the LED-display #. All measured data is now transferred as UCL data. To test the most comfortable level (MCL) press the test channel - 1 0 1 0 0. 0 8 0 selector button % longer than 2 Picture 19 MA 50 display with seconds and the dot moves to a MCL setting position right of the center of the LED-display #. All measured data is now transferred as MCL data.
GEBA50_e.04a.wpd
15
851406/5 10/04
MAICO
8
Operating Instructions MA 50
Care and maintenance of the instrument Disconnect the power plug before cleaning!
-
-
To clean the instrument, headphones, bone conduction receiver, loudspeaker and other accessories use a soft cloth dampened with a little warm soapy water or washing-up liquid; no alcohol or spirits should be used. The ear cushions Z of the headphone can be detached for cleaning. To remove the ear cushion Z pull it gently from the earphone. To assemble the cleaned or changed ear cushion press it on the earphone and make sure that it sits properly. The sound outlet hole must be exactly in the middle of the earphone. Picture 20 changing of the ear cushions
-
During cleaning, please ensure that no liquid runs into the switches, level control, headphone capsules or loudspeaker openings.
GEBA50_e.04a.wpd
16
851406/5 10/04
MAICO
Operating Instructions MA 50
9 Trouble shooting If you should find that your instrument is no longer working properly during a test run, please check the following points: Diodes in the buttons do not light up and there is nothing to see on the LED-display #: Is the power cord plugged in correctly? -
Check the power fuses 6.
100 - 240 V / 50 - 60 Hz
1 AT
Disconnect the power plug, unscrew the fuses 6 alongside the power cord socket and check the fuses. If the wire in the glass vial is broken, please insert new fuses (1 A slow blow).
A
RS 232
B
Picture 21 power socket and fuses
Diodes light up, but test tone is absent: Green STIM LED + or * does not light: Press STIM bar ( or ) (When the diode lights up and the tone is available, the device works in presenter mode). -
Green STIM LED + or * does light: Is level control ' or & set to an audible level position? Are the acoustic receivers connected to the correct socket? Is the required receiver with button 7 or 8 selected - is the correct LED lit? Is the lead connecting the headphones or the bone conductor loose or defective?
If your instrument still does not work properly after this short check, please consult your dealer or service center.
GEBA50_e.04a.wpd
17
851406/5 10/04
Operating Instructions MA 50
MAICO
10 Technical Data The MA 50 audiometer is an active, diagnostic medical 0124 product according to the class IIa of the EU medical directive 93/42/EEC. Standards: Test-Frequencies: Level step:
IEC 645-1/ EN 60 645-1 : Type 3 ANSI S3.6-1996 : Type 3 125 Hz, 250 Hz, 500 Hz, 750 Hz, 1 kHz, 1.5 kHz, 2 kHz, 3 kHz, 4 kHz, 6 kHz, 8 kHz 5 dB level steps
Sound Pressure Level: AC with earphone Telephonics TDH 39: - 10 dBHL ... 120 dBHL ( 500 Hz, 4 kHz, 6 kHz ( 250 Hz, 8 kHz ( 125 Hz
-10 dBHL ... 115 dBHL) -10 dBHL ... 100 dBHL) -10 dBHL ... 80 dBHL)
BC with bone conduction receiver Radioear B 71: - 10 dBHL ... 70 dBHL
( 500 Hz, 750 Hz, 3 kHz, 4 kHz, 6 kHz -10 dBHL ... 60 dBHL) ( 250 Hz -10 dBHL ... 40 dBHL) ( 125 Hz, 8 kHz no output ) ( speech -10 dBHL ... 85 dBHL)
Insert earphone with optional EAR 3A: - 10 dBHL ... 120 dBHL
( 750 Hz, 1.5 kHz -10 dBHL ... 115 dBHL) ( 500 Hz, 4 kHz -10 dBHL ... 110 dBHL) ( 250 Hz -10 dBHL ... 105 dBHL) ( 6 kHz -10 dBHL ... 95 dBHL) ( 125 Hz -10 dBHL ... 90 dBHL) ( 8 kHz -10 dBHL ... 85 dBHL)
Test Signal: Modulation:
Pure tone, pulse tone, warble tone, narrowband noise. Pulse tone: 0.5 s on/off Warble tone: triangular, freq. devitation 5%, repetition rate 5 Hz
Functions:
Tone presenter / interrupter Tracking (fixed level difference between both channels)
Warm up time:
less than 10 min after power on
Environment conditions:
GEBA50_e.04a.wpd
+ 15 ... + 35 /C / + 59 ... + 95 /F (operation) + 5 ... + 50 /C / + 41 ... + 122 /F (storage) Maximum humidity 90 % (storage and operation) 18
851406/5 10/04
Operating Instructions MA 50
MAICO
Dimensions:
W x D x H: 36 x 46 x 15 cm / 14.2” x 18.1” x 5.9”
Weight:
4.6 kg / 10.1 lbs (with accessories 5.8 kg / 12.8 lbs)
Power Supply:
100 - 240 V~ 50/60 Hz ±10 %
Power Consumption: app. 25 VA Device Fuses:
2x 1A slow blow
Connection plugs:
Connection
Specification
9 6 :
power socket
left/right=power, middle=prot.GND
(100 ... 240 V~ 50/60 Hz)
; < = >
patient response switch BC (bone conduction receiver) AC (headphone) right channel AC (headphone) left channel
power fuses serial PC-interface
96 - 240 V
1=GND, 2=RX, 3=TX, 5=GND, 6=DTR sleeve=GND, tip=in sleeve=GND, tip=out sleeve=GND, tip=out sleeve=GND, tip=out
1 A slow blow RS 232 C RI= 500 S ZA= 4 S, UA= 8 Veff ZA=10 S, UA=1 Veff ZA=10 S, UA=1 Veff
/ 50 - 60 Hz
1 AT
A
B
PATIENT RS 232 INTERFACE RESPONSE
C
D
BC
R
AC
N
T V
L
Picture 22 rear of MA 50 with connection plugs
GEBA50_e.04a.wpd
19
851406/5 10/04