MAQUET
SERVO-i ADULT Data Sheet rev 07
Data Sheet
20 Pages
Preview
Page 1
DATA SHEET VENTILATION SERVO-i UNIVERSAL
HIGHLIGHTS For all patient categories - available in two editions
NAVA - improved synchrony and unique monitoring capabilities
Intuitive user interface
The Open Lung Tool®
Designed for cost-efficiency
Continuity of ventilatory care: inter- and intrahospital transport and MR use*
Flexible placement Modular – interchangeable plug-in units Supports invasive as well as non invasive ventilation
Enables Heliox delivery
| Critical Care | SERVO-i UNIVERSAL | 3 |
SERVO-i VENTILATOR FAMILY SERVO-i UNIVERSAL
UBE
SERVO-i Ventilatory Configurations The SERVO-i Ventilator family consists of four configurations:
Heliox NAVA
Infant Adult UBE UEE
Universal Basic Edition Universal Extended Edition
Nasal CPAP Non Invasive Ventilation (NIV) Y Sensor Measuring
Standard Configuration Options
CO2 Analyzer
Not applicable
Nebulizer Bi-Vent
The SERVO-i Ventilator family addresses the very different requirements of neonatal, pediatric and adult needs from a single ventilation platform. All four configurations are the same ventilator, equipped with different functions. They can be customized and upgraded with different options for future needs. (SERVO-i Adult
Alarm output connector All patient category software ®
Open Lung Tool (OLT)
and SERVO-i Infant can also be upgraded to SERVO-i Universal.) ®
The same ventilator can be used at the bedside, during transport
Automode
and in the MR-room facilitating training, operation and
SIMV (PRVC) + PS
maintenance, increasing efficiency and flexibility.
PRVC SERVO-i Universal is an intensive care ventilator for all patient
VS (Volume Support)
categories and is available in two editions - Basic and Extended, the latter featuring a wide range of modes and functions as standard.
SIMV (VC) + PS VC (Volume Control)
*An agreement with MAQUET must be signed, see conditions in
SIMV (PC) + PS
the SERVO-i MR declaration (Order no. 66 71 670) and the
PC (Pressure Control)
Interhospital Transport Declaration (Order no. 66 64 721) respectively.
PS/CPAP
UEE
| 4 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
The system - General
Non-operating conditions The device complies with requirements of Medical Device Directive 93/42/EEC.
Impact:
Peak acceleration: 15 g. Pulse duration: 6 ms. Number of impacts: 1000.
Classification:
Class I equipment. According to IEC/EN 60 601-1.
Storage temperature:
–25 to +60º C (–13 to 140º F)
Storage Relative Humidity:
< 95%
Standards:
IEC/EN 60 601-1 (Type B). IEC/EN 60 601-2-12. EN 794-1.
Storage Atmospheric Pressure: 470 to 1060 hPa
Electromagnetic compatibility (EMC): Emission:
According to IEC/EN 60601-1-2 (Edition 2:2001).
Immunity- Extended test to 30 v/m:
According to IEC/EN 60601-1-2.
Power supply Power supply, automatic range 100 – 120 V AC ±10%, 50 – selection: 60 Hz, or 220 – 240 V AC ±10%, 50 – 60 Hz. Plug-in battery module:
The 'EMC Declaration, Information to the Responsible
Battery backup:
Two battery modules are delivered with the ventilator. Up to six battery modules can be included.
Battery capacity:
Rechargable, 12 V, 3.5 Ah each.
Recharge time:
Approximately 3 h/battery.
Battery backup time:
At least 3 h, when using six batteries.
Organization' is available from MAQUET. Patient range: Adult, invasive and NIV:
Weight 10 – 250 kg.
Infant, invasive ventilation:
Weight 0.5 – 30 kg.
Infant, NIV:
Weight 3 – 30 kg.
Infant, Nasal CPAP:
Weight 0.5 – 10 kg.
Operating conditions Operating temperature:
+10 to +40ºC
Relative humidity:
15 to 95% non-condensing
Atmospheric pressure:
660 to 1060 hPa
Lowest pressure in breathing system:
–400 cmH2O
External 12 V DC:
12.0 V – 15.0 V DC, 10 A
Max power consumption:
At 100 – 120 V: 2 A, 190 VA, 140 W. At 220 – 240 V: 1 A, 190 VA, 140 W.
| Critical Care | SERVO-i UNIVERSAL | 5 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
The ventilator - General Dimensions:
User interface (See dimensional drawings page 16-17)
User Interface:
W 355 x D 53 x H 295 mm
Patient Unit:
W 300 x D 205 x H 415 mm
Weight:
Approximately 20 kg (Patient Unit 15 kg, User Interface 5 kg)
Method of triggering:
Flow and pressure
Max. operating pressure:
Approximately 115 cmH2O
Bias flow: Adult:
2 l/min
Infant:
0.5 l/min
Weight:
Approximately 5 kg
Attachment:
Can be attached to the mobile cart, a table, rail or pole (15 – 30 mm diameter).
Screen Type:
TFT-LCD module
Size:
31 cm (12.1”) diagonal
Viewing area:
246.0 x 184.5 mm
Inspiratory channel Pressure drop:
Gas supply Inlet gas pressure:
200 – 650 kPa / 2.0 – 6.5 bar / 29 – 94 PSI
Connection standards available: AGA, DISS, NIST, or French standard. Unavailable gas/loss of gas pressure:
The flow from an unavailable gas (air or O2) is automatically compensated for so that the patient gets the preset volume and pressure.
Patient system gas connectors Conical fittings: Gas exhaust port:
Male 22 mm / female 15 mm. In accordance with ISO 5356-1. Male 30 mm cone
Max. 6 cmH2O at a flow of 1 l/s
Internal compressible factor:
Max. 0.1 ml/cmH2O
Gas delivery system:
Microprocessor controlled valve
Inspiratory flow range: Adult:
0 to 3.3 l/s
Infant:
0 to 0.55 l/s
Expiratory channel Pressure drop:
Max. 3 cmH2O at a flow of 1 l/s
Internal compressible factor:
Max. 0.1 ml/cmH2O
PEEP regulation:
Microprocessor controlled valve
Rise time, expiratory flow measurement:
<12 ms for 10 – 90 % response at flow of 0.05 – 3.2 l/s
Expiratory flow range:
0 to 3.2 l/s
| 6 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Alarms
Alarms
Airway pressure (upper):
Battery:
Adult, invasive ventilation:
16 - 120 cmH2O
Adult, Non Invasive Ventilation:
16 - 60 cmH2O
Infant, Invasive ventilation:
16 - 90 cmH2O
Infant, Non Invasive Ventilation
16 - 60 cmH2O
Adult:
0.5 – 60 l/min
Infant:
0.01 – 30 l/min
Expired minute volume (Lower alarm limit): Adult:
0.5 – 40 l/min
Infant:
0.01 – 20 l/min
Apnea: Adult:
15 – 45 s
Infant:
5 – 45 s
Respiratory frequency:
1 – 160 breaths/min
High end expiratory pressure:
0 – 55 cmH2O
Low end expiratory pressure:
0 – 47 cmH2O. Note. Setting the alarm to 0 (zero) is equal to alarm off.
High continuous pressure:
End-tidal CO2 (upper and lower 0.5 – 20%. 4 – 100 mm Hg. limit): 0.5 – 14 kPa. Leakage out of range in NIV:
Expired minute volume (Upper alarm limit):
Set PEEP level + 15 cmH2O exceeded for more than 15 seconds.
O2 concentration:
Set value ±5 vol% or 18 vol%
Gas supply:
Below 200 kPa / 2.0 bar / 29 PSI and above 650 kPa / 6.5 bar / 94 PSI
Limited battery capacity: 10 min. No battery capacity: less than 3 min. Low battery voltage.
Yes. Described in the User’s manual.
No patient effort in NIV: Adult:
45 s
Infant:
15 s
Technical:
Yes. Described in the User’s manual.
Autoset (alarm limits) specification:
Invasive ventilation, controlled modes only.
High airway pressure:
Mean peak pressure +10 cmH2O or at least 35 cmH2O.
Upper minute volume:
Expiratory minute volume + 50%.
Lower minute volume:
Expiratory minute volume – 50%.
Upper respiratory frequency: Breathing frequency + 40%. Lower respiratory frequency: Breathing frequency – 40%. High end expiratory pressure: Mean end expiratory pressure + 5 cmH2O. Low end expiratory pressure: Mean end expiratory pressure – 3 cmH2O. Upper end tidal carbon End tidal carbon dioxide dioxide concentration (etCO2): concentration + 25%. Lower end tidal carbon End tidal carbon dioxide dioxide concentration (etCO2): concentration – 25%.
| Critical Care | SERVO-i UNIVERSAL | 7 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Ventilation Modes – Invasive ventilation Controlled ventilation:
Ventilation Modes - Invasive ventilation Automode
Pressure Control (PC):
Pressure controlled ventilation.
Volume Control (VC):
Volume controlled ventilation.
PRVC (optional with Universal Pressure regulated volume Basic Edition): controlled ventilation.
Control mode: VC <-> Support mode: VS Control mode: PC <-> Support mode: PS Control mode: PRVC <-> Support mode: VS
Neurally controlled ventilation (optional) NAVA
Supported ventilation:
Assisted ventilation based on diaphragm activity
Volume Support (VS) (optional Volume supported ventilation. with Universal Basic Edition): Pressure Support (PS)/CPAP: Pressure supported ventilation / Continuous Positive Airway Pressure ventilation.
Ventilation modes – Non Invasive Ventilation (optional) NIV Pressure Control:
Non-invasive pressure controlled ventilation.
Combined ventilation:
Comes with the corresponding controlled ventilation mode.
NIV Pressure Support:
Non-invasive pressure supported ventilation.
SIMV (VC) + PS:
Synchronized Intermittent Mandatory Ventilation based on volume controlled ventilation with pressure support.
Nasal CPAP:
Nasal Continuous Positive Airway Pressure ventilation.
SIMV (PC) + PS:
Synchronized Intermittent Mandatory Ventilation based on pressure controlled ventilation with pressure support.
SIMV (PRVC) + PS (optional Synchronized Intermittent with Universal Basic Edition): Mandatory Ventilation based on pressure regulated volume controlled ventilation with pressure support. Bi-Vent (optional):
Pressure controlled ventilation on two independently adjustable levels, allowing unrestricted spontaneous breathing on both levels.
Waveform and loop presentations Real time waveforms: Pressure waveform. Flow waveform. Volume waveform. CO2 waveform*. Edi waveform** * Requires SERVO-i CO2 Analyzer option.
** Requires SERVO-i Neurally Controlled Ventilation option.
Loops:
Volume / Pressure***. Flow / Volume***. A reference loop and three overlaying loops can be displayed. ***Displayed simultaneously with Open Lung Tool graphical trends, if requested.
| 8 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Displayed
Trended
Displayed
Trended
Monitoring
value
value*
Breathing frequency:
Yes
Yes
Monitoring
value
value*
Work of Breathing ventilator:
Yes
Yes
Spontaneous breaths per minute (RRsp):
Yes
Yes
Shallow Breathing Index (SBI): Yes
Yes
Total PEEP:
Yes
Peak Airway Pressure:
Yes
No
Yes
Static Compliance:
Yes
Mean Airway Pressure:
Yes
Yes
Yes
Inspiratory Resistance:
Yes
Yes
Pause Airway Pressure:
Yes
Yes
Expiratory Resistance:
Yes
Yes
End Expiratory Pressure:
Yes
Yes
I:E ratio:
Yes
No
CPAP Pressure:
Yes
Yes
Edi peak:
Yes
Yes
End Expiratory Flow:
Yes
Yes
Edi min:
Yes
Yes
Inspired Tidal Volume:
Yes
Yes
Edi Catheter position:
Yes
No
Expired Tidal Volume:
Yes
Yes
Elastance:
Yes
Yes
Inspired Minute Volume:
Yes
Yes
Time Constant:
Yes
No
Expired Minute Volume:
Yes
Yes
Measured O2 Concentration:
Yes
Yes
Dynamic Characteristics:
Yes
Yes
Yes
Yes
Ti/Ttot:
Yes
No
CO2 End tidal concentration (etCO2):
Work of Breathing patient:
Yes
Yes
*Stored trend values for up to 24 hours
CO2 Minute elimination ( CO2): Yes
Yes
Tidal CO2 elimination (VTCO2): Yes
Yes
MVe sp / MVe:
Yes
No
Spontaneous Exp. Minute Volume (MVe sp):
Yes
Yes
Leakage fraction in NIV (%):
Yes
Yes
P0.1 measurement:
Yes
Yes
Supply pressure (O2 and air):
Yes
No
Battery remaining time:
Yes
No
Barometric pressure:
Yes
No
| Critical Care | SERVO-i UNIVERSAL | 9 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Log function Event log:
Service log:
Alarms. Ventilator settings. Apnea periods. Immediate functions. Technical alarms. Test results. Preventive maintenance. Service history. Configuration log.
Parameter settings
Setting range:
Parameter:
Infant
Adult
Back-up pressure above PEEP 5 – (80 - PEEP) 5 – (120 (cmH2O): PEEP) NIV Back-up rate (breaths/min): 4 – 40
4 – 20
O2 concentration (%):
21 – 100
21 – 100
I:E ratio:
1:10 – 4:1
1:10 – 4:1
TInsp (s):
0.1 – 5
0.1 – 5
NAVA level (cmH2O/µV):
0 – 30
0 – 30
Edi trigger sensitivity (µV):
0.1 – 2.0
0.1 – 2.0
NIV Back-up TInsp (s):
0.3 – 1
0.5 – 2
Parameter settings
Setting range:
Parameter:
Infant
Adult
Inspiratory tidal volume (ml):
2 – 350
100 – 4000
TPause (s):
0 – 1.5
0 – 1.5
Inspiratory minute volume (l/min):
0.3 – 20
0.5 – 60
TPause (% of breath cycle time): 0 – 30
0 – 30
5 – 45
15 – 45
Flow trigger sensitivity level (fraction of bias flow):
0 – 100%
Apnea, time to alarm (s): Automode Trigger timeout (s):
3 – 15
7 – 12
Press. trigg sensitivity (cmH2O): -20 – 0
-20 – 0
PC/PS above PEEP (cmH2O)
5 – (80 -PEEP) 0 – (120 PEEP)
Insp. rise time (% of breath cycle time):
0 – 20
0 – 20
PC/PS above PEEP in NIV (cmH2O)
0 – (32 - PEEP) 0 – (32 - PEEP)
Insp. rise time (s):
PEEP (cmH2O)
0 – 50
0 – 50
PEEP in NIV (cmH2O)
2 – 20
CPAP pressure (cmH2O): CMV frequency (breaths/min):
0 – 100%
0 – 0.2
0 – 0.4
Insp. cycle off (% of peak flow): 1 – 70
1 – 70
Insp. cycle off in NIV (% of peak 10 – 70 flow):
10 – 70
2 – 20
2 – 20
–
Nebulizer time (min):
5 – 30
4 – 150
4 – 150
SIMV frequency (breaths/min):
1 – 60
1 – 60
Breath cycle time, SIMV (s):
0.5 – 15
1 – 15
PHigh (cmH2O)
(PEEP +1) - 50 (PEEP +1) - 50
THigh (s):
0.2 – 10
0.2 – 10
TPEEP (s):
0.2 – 10
0.2 – 10
PS above PHigh (cmH2O):
0 – (80 - PHigh) 0 – (120 PHigh)
5 – 30
| 10 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Parameter:
Setting range
Non Invasive Ventilation (optional)
Oxygen breaths:
100% for 1 minute
Ventilation modes:
Start breath:
Initiation of 1 breath (In SIMV mode initiation of 1 mandatory breath)
Pause hold:
Insp. or exp (0 – 30 seconds)
Alarm silence/reset:
2 minute silence and reset of latched alarms
Compliance compensation:
On/Off
Automode (optional with Universal Basic Edition):
Automode On/Off
Servo Ultra Nebulizer (optional): Nebulizer On/Off
NIV Pressure Support. NIV Pressure Control. Nasal CPAP (The CPAP pressure is kept constant by providing an automatic variable flow).
Max. leakage compensation level: Adult:
40 l/min
Infant, NIV:
15 l/min
Infant, Nasal CPAP:
10 l/min
Leakage overrange detection:
Automatic
Suction Support
Disconnect detection:
Automatic
Pre oxygenation time:
Max. 2 min
Connect detection:
Post oxygenation time:
Max. 1 min
Manual, or automatic via bias flow
Suction phase time:
No maximum level
Adjustable oxygen level:
21 – 100 %
Backup rate (in NIV Pressure Support):
Manually adjustable breathing rate. The breathing rate is controlled by the ventilator in the event of apnea.
Saving of data Recording of current waveform 20 seconds of data will be and parameter values: recorded (10 seconds before and 10 seconds after activation).
Communication/Interface Serial port:
RS-232C - isolated. For data communication via the Communication Interface Emulator (CIE).
Network connection (optional): MIB (Medical Information Bus) monitor connection Data transfer (optional):
Via Ventilation Record Card
Screen picture transfer (optional)
Via Ventilation Record Card
| Critical Care | SERVO-i UNIVERSAL | 11 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Open Lung Tool (OLT) (optional with Universal Basic Edition)
Alarm output connector
Three simultaneous graphical trends, presented breath-by-breath:
(optional)
1.
EIP and PEEP (End Inspiratory Pressure and Positive End Expiratory pressure).
Connector:
4-pole Modular connector
Ratings:
Max 40 V DC, Max 500 mA, Max 20 W
2.
VTi and VTe (Inspiratory and Expiratory Tidal volume).
Service
3.
C dyn i and VTCO2*
Regular maintenance:
(Dynamic inspiratory Compliance [= VTi/(EIP – PEEP)] and Tidal CO2 elimination*). * Requires SERVO-i CO2 Analyzer option. Values stored breath by breath: Up to 21.600 breaths can be stored. Cursor function:
An activated cursor can be moved with the main rotary dial or via the touch screen. When moving it along the graph, the numeric parameter values valid for that actual moment will be shown at the right hand side of the graphs.
Zoom function:
Time resolution of the x-axis can be selected in five different steps.
Time marking:
Hours and minutes (when values are measured).
Once every 12 months or at least after 5000 operating hours.
Note For inaccuracies and more detailed technical specifications please refer to the User’s manual.
| 12 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
ACCESSORIES
®
Servo Ultra Nebulizer (optional)
Aeroneb Pro Nebulizer System (optional)
Weight:
Approx. 125 g
See separate data sheet.
Dimensions:
H 105 mm x L 108 mm x W 60 mm
Connection cable length:
2.0 m
Nebulizer T-piece connections: Inlet/outlet: 22/15 mm outer/inner diameter and 22 mm inner diameter, ISO standard. Pediatric patient tubes: Adapters 22/10 mm outer diameter and 15/10 mm outer diameter. Internal volume:
60 ml
Ultrasonic generator frequency: 2.4 MHz Particle size (water):
Mass Median Diameter (MMD) = approximately 4.0 µm, measured distally in endotracheal tube 8 mm inner diameter.
Output from nebulizer (water) – Min. 0.1 ml water/min at 0.1 l minimum water flux: gas flow/s. Min. 0.3 ml water/min at 0.5 l gas flow/s. Min. 0.5 ml water/min at 1.0 l gas flow/s (Adult only). Buffer liquid:
Sterile water
Max. medication temperature:
55º C (131º F)
Volume, medication cup:
Max. 10 ml
Noise level:
Max. 50 dBA, measured at 0.3 m distance.
Y Sensor measuring (optional) Size: Y Sensor Module:
154 x 90 x 43 mm
Y sensor adult:
Length 84 mm
Y sensor infant:
Length 51 mm
Weight: Y Sensor Module:
0.4 kg
Y sensor adult:
10.5 g
Y sensor infant:
7.5 g
Sensor material:
Makrolon polycarbonate.
Tubing:
2.0 m. Medical grade PVC.
Power source – Y Sensor Module supply voltage:
Powered from the SERVO-i. <5 W at 12 V (normal operation).
Y sensor measuring – Performance Measuring method:
Fixed orifice, differential pressure.
Parameters:
Airway pressure. Airway flow. Inspiratory and expiratory volumes.
Measuring range: Adult:
2 to 180 l/min
Infant:
0.125 to 40 l/min
Airway adapter dead space: Adult:
< 9.0 ml
Infant:
< 0.45 ml
| Critical Care | SERVO-i UNIVERSAL | 13 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
SERVO-i CO2 Analyzer (optional)
SERVO-i CO2 Analyzer – Performance
Standard:
Measuring method:
Mainstream, dual-wavelength, non-dispersive infrared.
Parameters:
Capnogram. CO2 End tidal concentration (etCO2).
EN 864, ISO 9918. IEC/EN 60601-1 (Type BF, defibrillator proof)
Size: CO2 Analyzer Module:
154 x 90 x 43 mm
Sensor:
32.0 x 42.4 x 21.6 mm
CO2 Minute elimination* ( CO2). Tidal CO2 elimination* (VTCO2). * Note: Elimination of CO2 from the body takes place via expired gas from the lung (Production of CO2 takes place in cellular compartments and body tissues).
Weight: CO2 Analyzer Module:
0.45 kg
Sensor:
18 g
Airway adapter:
10 g Measuring range:
0 to 100 mm Hg CO2 partial pressure. 0 to 13.3 kPa CO2 partial pressure. 0 to 13.2% CO2 volume (at a barometric pressure of 1013 hPa).
CO2 Analyzer Module supply Powered from the SERVO-i. voltage:
Step response time:
<25 ms (10 to 90% step response).
Power consumption:
Warm-up time:
30 s to initial CO2 indication, max. 5 min to full specification.
Oxygen concentration compensation:
Automatic. Values supplied from the SERVO-i Ventilator System.
Barometric pressure compensation:
Automatic. Values supplied from the SERVO-i Ventilator System.
Digitizing rate:
87 Hz
Connectors and cables: CO2 Analyzer Module:
15-pole D-sub female connector.
Sensor:
20-pole, 2.4 m cable.
Power source:
Sensor:
8 W at 12 V, during warm up. 6.5 W at 12 V, during normal operation. Powered from the CO2 Analyzer Module.
Airway adapter dead space: Adult:
<5 ml
Infant:
<0.5 ml
| 14 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Neurally controlled ventilation (optional) Standards: IEC/EN 60601-1 (Type CF, defibrillator proof)
SERVO-i Mobile Cart (optional) Weight:
20 kg
Dimensions:
H 1015 mm x L 640 mm x W 560 mm (see dimensional drawing below)
Size: Edi module:
154 x 90 x 21 mm
Edi catheter cable:
Length: 2 m SERVO-i Drawer kit (optional) 0.25 kg
Weight:
4.5 kg
Power source - Edi module supply voltage:
Powered from SERVO-i. <3 W at 12 V (normal operation)
Dimensions:
H 240 mm x L 210 mm x W 300 mm
Parameters:
Edi waveform Edi leads waveforms NAVA estimated pressure waveform (Pest)
Weight - Edi module:
Edi catheters:
6 Fr, length 50 cm 8 Fr, length 100 cm 8 Fr, length 125 cm 12 Fr, length 125 cm 16 Fr, length 125 cm
SERVO-i Holder (optional) Weight:
3.5 kg
Dimensions:
H 352 mm x L 247 mm x W 159 mm (see dimensional drawing below)
SERVO-i Shelf Base (optional) Weight:
1.2 kg
Dimensions:
H 29 mm x L 205 mm x W 159 mm (see dimensional drawing below)
| Critical Care | SERVO-i UNIVERSAL | 15 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
Gas cylinder restrainer (optional) Max load:
2 x 5-liter bottles
SERVO-i IV Pole (optional) Max load (total):
6 kg
Gas trolley (optional) Max load:
2x10 kg bottles
Docking:
Dockable to SERVO-i Mobile Cart. Dockable to a separate wall clamp.
Compressor Mini (optional) See separate data sheet.
| 16 | SERVO-i UNIVERSAL | Critical Care |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
DIMENSIONAL DRAWINGS
SERVO-i on Mobile Cart
SERVO-i Holder
| Critical Care | SERVO-i UNIVERSAL | 17 |
TECHNICAL SPECIFICATIONS SERVO-i UNIVERSAL
SERVO-i (patient unit) on SERVO-i Holder
SERVO-i Shelf Base
| 18 | SERVO-i UNIVERSAL | Critical Care |
ORDERING INFORMATION SERVO-i UNIVERSAL
SERVO-i, Ventilator and accessories: See separate information: “SERVO-i, System Version 4.0 - System Flow Chart” (Order no: 66 70 102).
© Maquet Critical Care AB 2008. All rights reserved. • MAQUET reserves the right to modify the design and specifications contained herein without prior notice. Order No. 66 70 105 • Printed in Sweden • 080501 • 07 English • Data Sheet
Maquet Critical Care AB SE-171 54 Solna, Sweden Phone: +46 (0) 8 730 73 00 www.maquet.com
For local contact: Please visit our website www.maquet.com
GETINGE Group is a leading global provider of equipment and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. Equipment, service and technologies are supplied under the brands ArjoHuntleigh for patient handling and hygiene, disinfection, DVT prevention, medical beds, therapeutic surfaces and diagnostics, GETINGE for infection control and prevention within healthcare and life science and MAQUET for Surgical Workplaces, Cardiovascular and Critical Care.