System overview
Shortcuts and keys
A. Patient category
N. The Quick access key leads to functions such as suction support, open lung tool and stress index.
B. Mode of ventilation C. Automode On/Off* D. Trigger indicators E. Nebulizer On/Off* F. System status G. Measured values and alarms H. Fixed keys I.
Main Rotary Dial
J. Ventilation settings K. Start/Standby key L. Waveforms M. On/Off Switch (rear side)
O. The Menu access key leads to functions such as Alarm history, Biomed and Panel lock. P. Additional values shows measured and calculated numerical values. Total PEEP and Static compliance on page three (expiratory/inspiratory hold required). Q. C means Patient circuit compliance compensation is activated. R. Direct access knobs-used for
*This feature is an option and may be unavailable on your Servo-i
This document is intended for an international audience outside of US
T
immediate adjustment of ventilation parameters. S. Additional settings T. Special Function Keys 1. Start breath will initiate a new breath cycle with the current ventilator settings. 2. O2 Breaths will deliver 100% oxygen for 1 minute. 3. Inspiratory hold. 4. Expiratory hold. CAUTION: Suction support is not intended to be used together with closed-suction systems.
Alarm handling
1
Alarms can be displayed in one of three colors depending on priority: High
Medium
2
Low
1. Silence and pre-silence alarms Press the Audio pause fixed key for less than 2 seconds to silence the alarm for two minutes. To Pre-silence low and medium alarms for 2 minutes, tap when no alarms are active.
2. Adjust alarm settings Press the Alarm Profile fixed key to open the alarm window. Autoset of alarm limits are available in VC, PC and PRVC modes.
6
Setting up and starting ventilation 1.
Plug in the power cord.
2.
Switch ON the ventilator, at the back of the panel.
3.
Connect oxygen and air hoses.
4.
Lock cart wheels.
5.
Start Pre-use check (needs test tube) and follow onscreen instructions. Includes Patient circuit test.
6.
Choose Patient category: Adult or Infant*
7.
Choose Ventilation type: Invasive ventilation or NIV*
8.
Choose Ventilation mode and adjust the settings.
9.
Adjust alarm limits in the alarm window.
7
8
9
10
2
1
2
3
4
Changing ventilation mode 1. Press touchpad to access the mode menu. 2. Press the arrow and the available ventilation modes appear. Select the desired mode. 3. Change the settings by activating, changing and confirming each one with the help of the Main rotary dial or by pressing the relevant touchpad. 4. To activate all settings in the window press Accept, or to cancel the settings press Cancel.
Disconnect and stop ventilation
4
1. Physically disconnect the patient from the ventilator. 2. Press the Start/Standby key. 3. Press Yes to stop ventilation.
2
4. Turn the ventilator off using the On/Off switch on the lower back of the User Interface.
This document is intended for an international audience outside of US Manufacturer · Maquet Critical Care AB · Röntgenvägen 2 SE-171 54 Solna · Sweden · +46 (0)10 335 00 00 www.getinge.com
© Maquet Critical Care AB 2020 · Servo-i is a trademark of Maquet Critical Care AB · MX-7734 Rev01 v7.0 · English
10. Connect the ventilator to the patient and press Start ventilation.