Cleaning Posters
4 Pages
Preview
Page 1
SERVO-U/n Disinfection Washer-Disinfector Wiping and Discarding
Rinse
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Washer-Disinfector
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
85-95°C (185-203°F)
Drying alternatives Shake/Tilt
Run the cassette in a SERVO-U/n with a testlung
5-7 times
Important! When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
If rinsing before disinfection is not included in the washer disinfector program: Rinse the parts thoroughly in water (<35°C/95°F). Let the water flow through the parts.
Important! Rinsing the cassette in water (<35°C/95°F) immediately after use may be a useful alternative to disinfection. Immediate rinsing can remove particles and reduces the risk of cross contamination between patients.
Position and flow
Note. The water pressure in the washer-disinfector should not exceed 1.5 bar. The maximum water flow should not exceed 10 l/min. The water must have free passage through the cassette.
10 min
Drying cabinet
Room air
1 hour maximum 70°C (158°F).
12-24 hours
If available, connect a 22 mm hose to expiratory inlet.
Caution! Never dry the cassette by applying high pressure air as you may damage the internal tubing.
0123
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
© Maquet Critical Care AB 2015. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1512 • Rev. 10 English • Routine wall cleaning diagram English ANZ - MCC0009
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. ArjoHuntleigh focuses on patient mobility and wound management solutions. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.
SERVO-U/n Disinfection Manually Wiping and Discarding
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Disinfectant
Rinse
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Drying alternatives Shake/Tilt
Run the cassette in a SERVO-U/n with a testlung
Let the parts soak in a disinfectant: Ethyl- or isopropyl alcohol 70% Cidex OPA Hexanios G+R, Aniosyme DD1, Gigazyme Plus or Anioxyde 1000.
Rinse the parts thoroughly in water (<35°C/95°F). Let the water flow through the parts.
Important! Follow the disinfectant manufacturer’s recommendations and instructions.
5-7 times
10 min
Drying cabinet
Room air
1 hour maximum 70°C (158°F).
12-24 hours
Rinse Important!
Important!
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant.
Rinsing the cassette in water (<35°C/95°F) immediately after use may be a useful alternative to disinfection. Immediate rinsing can remove particles and reduces the risk of cross contamination between patients.
Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Important! Rinse the parts thoroughly in water to remove all traces of disinfectant. Let the water flow through the parts. Carefully shake and tilt the cassette, turn upside down and repeat, at least 5-7 times.
If available, connect a 22 mm hose to expiratory inlet.
Caution! Never dry the cassette by applying high pressure air as you may damage the internal tubing.
0123
© Maquet Critical Care AB 2015. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1512 • Rev. 10 English • Routine wall cleaning diagram ANZ - MCC0009
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. ArjoHuntleigh focuses on patient mobility and wound management solutions. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.
SERVO-U/n Disinfection Washer-Disinfector (Note. Autoclaving reduces the lifetime of the expiratory cassette)
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Wiping and Discarding
Autoclave
Rinse
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Drying alternatives Shake/Tilt
3 min 134°C (273°F)
Rinse the cassette in water (<35°C/95°F). Let the water flow through the parts
Run the cassette in a SERVO-U/n with a test lung
15 min 121°C (250°F)
Dry
5-7 times
Shake/Tilt
10 min
Important! Before placing the expiratory cassette in an autoclave, make sure that no water remains inside the cassette.
Important! When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
5-7 times
Drying cabinet
Room air
Note. Sterilization is normally not necessary for the expiratory cassette but, when applied, use validated processes only. Instrument parts are typically autoclaved at a temperature of 134°C
Drying cabinet
Room air
1 hour 12-24 hours maximum 70°C (158°F). If available, connect a 22 mm hose to expiratory inlet.
(273°F) for 3 minutes. 1 hour 12-24 hours maximum 70°C (158°F). If available, connect a 22 mm hose to expiratory inlet.
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Rubber parts are typically autoclaved at a temperature of 121°C (250°F) for 15 minutes.
Caution! Never dry the cassette by applying high pressure air as you may damage the internal tubing.
0123
© Maquet Critical Care AB 2015. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1512 • Rev. 10 English • Routine wall cleaning diagram ANZ - MCC0009
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. ArjoHuntleigh focuses on patient mobility and wound management solutions. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.
SERVO-U/n
Hazardous waste (infectious)
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Hazardous waste (infectious)
Wipe off/discard
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts
Important!
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Cleaning procedure when Servo Duo Guard filter is used
0123
© Maquet Critical Care AB 2015. All rights reserved. • Maquet reserves the right to modify the design and specifications contained herein without prior notice. Order No. 65 15 014 • Printed in Sweden • 1512• Rev. 10 English • Routine wall cleaning diagram ANZ - MCC0009
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost efficiency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. ArjoHuntleigh focuses on patient mobility and wound management solutions. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.