SERVO-U/n
Disinfection Washer-Disinfector Wiping and Discarding
Rinse
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Washer-Disinfector 85-95°C (185-203°F)
Drying alternatives Shake/Tilt
Run the cassette in a SERVO-U/n with a testlung
5-7 times
Important!
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
If rinsing before disinfection is not included in the washer disinfector program: Rinse the parts thoroughly in water (<35°C/95°F). Let the water flow through the parts.
Important!
Rinsing the cassette in water (<35°C/95°F) immediately after use may be a useful alternative to disinfection. Immediate rinsing can remove particles and reduces the risk of cross contamination between patients.
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Position and flow
Note. The water pressure in the washer-disinfector should not exceed 1.5 bar. The maximum water flow should not exceed 10 l/min. The water must have free passage through the cassette.
10 min
Drying cabinet
Room air
1 hour maximum 70°C (158°F).
12-24 hours
If available, connect a 22 mm hose to expiratory inlet.
Caution! Never dry the cassette by applying high pressure air as you may damage the internal tubing.
SERVO-U/n
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Disinfection Manually Wiping and Discarding
Rinse
Disinfectant
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Drying alternatives Shake/Tilt
Run the cassette in a SERVO-U/n with a testlung
Ethyl- or isopropyl alcohol 70% Cidex OPA Hexanios G+R, Aniosyme DD1, Gigazyme Plus or Anioxyde 1000.
Rinse the parts thoroughly in water (<35°C/95°F). Let the water flow through the parts.
Important! When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
Important! Rinsing the cassette in water (<35°C/95°F) immediately after use may be a useful alternative to disinfection. Immediate rinsing can remove particles and reduces the risk of cross contamination between patients.
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Important! Follow the disinfectant manufacturer’s recommendations and instructions.
5-7 times
10 min
Drying cabinet
Room air
1 hour maximum 70°C (158°F).
12-24 hours
Rinse
Important! Rinse the parts thoroughly in water to remove all traces of disinfectant. Let the water flow through the parts. Carefully shake and tilt the cassette, turn upside down and repeat, at least 5-7 times.
If available, connect a 22 mm hose to expiratory inlet.
Caution! Never dry the cassette by applying high pressure air as you may damage the internal tubing.
SERVO-U/n
Disinfection Washer-Disinfector
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
(Note. Autoclaving reduces the lifetime of the expiratory cassette)
Wiping and Discarding
Rinse
Autoclave
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Shake/Tilt 3 min 134°C (273°F)
Rinse the cassette in water (<35°C/95°F). Let the water flow through the parts
5-7 times
Shake/Tilt
Important!
When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts.
5-7 times
Drying cabinet
Room air
Run the cassette in a SERVO-U/n with a test lung
15 min 121°C (250°F)
Dry
Important!
Drying alternatives
Before placing the expiratory cassette in an autoclave, make sure that no water remains inside the cassette. Note. Sterilization is normally not necessary for the expiratory cassette but, when applied, use validated processes only. Instrument parts are typically autoclaved at a temperature of 134°C
Drying cabinet
10 min
Room air
1 hour 12-24 hours maximum 70°C (158°F). If available, connect a 22 mm hose to expiratory inlet.
(273°F) for 3 minutes. 1 hour 12-24 hours maximum 70°C (158°F). If available, connect a 22 mm hose to expiratory inlet.
Caution! All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Rubber parts are typically autoclaved at a temperature of 121°C (250°F) for 15 minutes.
Caution! Never dry the cassette by applying high pressure air as you may damage the internal tubing.
SERVO-U/n
Cleaning procedure when Servo Duo Guard filter is used
Refer to the Cleaning and Maintenance User’s Manual for detailed information.
Important! When Servo Duo Guard filter (or equivalent) is used, the only cleaning needed is to wipe the ventilator and removable parts with a soft lint-free cloth moistened in soap & water or detergent-based disinfectant. Note. In case of more contaminated surfaces use ethyl alcohol (70%) or isopropyl alcohol (70%). Avoid contact with electrical contacts
Wipe off/discard
Whenever possible, clean and disinfect the equipment immediately after use. Do not leave foreign particles, such as blood or saliva, to dry on the equipment.
Hazardous waste (infectious)
Caution!
All personnel should be aware of the risk of parts being infected when disassembling and cleaning the ventilator.
Hazardous waste (infectious)
Maquet has now merged with Getinge Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization and endoscopic reprocessing departments, endovascular procedures and for life science companies and institutions. Based on our first-hand experience and close partnerships with clinical experts, healthcare professionals and medical-technology specialists, we are improving the everyday life for people – today and tomorrow. © 2018 Getinge Australia Pty Ltd ABN 67 009 673 518 Australian Phone: 1800 438 464 • Email: [email protected] New Zealand Phone: 0800 1 438 4643 • Email: [email protected] www.getinge.com/anz ANZ_CC_002_V01_0718