Brief operating instructions ALPHAMAXX
1133.12B1/B3/F1/F3
Control of functions: The electrohydraulically driven operating table functions can be controlled by • cable-connected hand control or • optional infrared hand control • optional foot switch • emergency control panel (override system) BACKW. Motor drive for the base backward
FORW. Motor drive for the base forward
LOCK Operating table locked
UNLOCK Table top unlocked; can be displaced
LEFT
L.FOOT Table top – longitudinal shift towards foot end
RIGHT Right leg plate up or down, depending on the key which is pressed simultaneously – LEG UP/DOWN L.HEAD Table top – longitudinal shift towards head end
BACK DN. Back plate down
BACK UP Back plate up
LEG DN. Leg plate down
LEG UP Leg plate up
FLEX (Kidney / gall bladder elevation) controls the ”back plate” and ”inclination” functions alternatively until the flex position is achieved.
BEACH CHAIR controls the ”back plate” and ”inclination” functions alternatively until the reverse flex position is achieved, then the leg plates are lowered
TREND Table top – Trendelenburg
REV. TREND Reverse Trendelenburg position
TILT LEFT Lateral tilt left
TILT RIGHT Lateral tilt right
Left leg plate up or down, depending on the key which is pressed simultaneously LEG UP/DOWN
DOWN Table top down
LEVEL Table top – horizontal alignment
UP Table top up
GK113312EN02
The back plate, seat plate and leg plate adjustments are motor-driven. The head rest is adjusted manually. The NORMAL patient orientation is preferable to the REVERSE patient orientation.
Fig. 1:
The preferred patient orientation
Precautionary measures •
This operating table may not be used for a patient weight exceeding 450 kg (1000 lb.).
•
Surgical procedures are only permissible on the locked table in the patient orientation shown above.
•
Displacing (Unlock) the operating table with a patient weight between 135 kg (300 lb.) and 225 kg (500 lb.) is only permissible when the table is in its lowest position.
•
Accessories which themselves are quite heavy or which have long lever arms (retractor systems and the like) are particularly liable to cause damage to the side rails.
•
The user must ensure that the operating table is set up at a flat, level and solid location.
•
Some adjustment functions can cause collisions between accessories and the operating table and damages to the accessories.
Î Note: Do not use the override control panel to set up extreme combinations of Trendelenburg/reverse Trendelenburg and lateral tilt adjustments. In case of malfunctions or if the hand control is defective, the most important operating table functions can be controlled via the control panel at the upper column casing (override system).