Quick Guide
1 Page
![EMMA® Waveform Capnograph Controls & Displays Power & Airway Adapters Battery Cover Release Alarm Mute Airway Adapter EtC02 Value a kP Battery Status On Button Respiratory Rate Alarm Status EtC02 Waveform Power On: Insert Airway Adapter 1 Press Power on button 2 device ready 3 Power Off: Automatic On removal of Adapter 1 or 2 mins after 2 NO Breath detected and Alarm silenced 4 How to Zero device Battery Cover Release 3 Infant Airway Adapter Battery types: Duracell Industrial 2 x AAA LR03 or Energizer Ultimate 2x AAA Lithium L92 [reduced deadspace 1 ml] 5 6 Note! Trained medical professional must determine correct airway adapter for each patient application. Close Battery Cover Alarms & Status Displays 2 kPa kPa kPa kPa 3 1 EtC02 Low Alarm [setting] EtC02 High Alarm [setting] Press & Hold Power On 1 Press & Hold Alarm Mute 2 button for 5 secs. for button for 5 secs. Display changes as shown. Display changes as shown. Press Power button to 1 Press Alarm button 2 decrease High alarm. to increase Low alarm. Press 2 to increase Press 1 to decrease setting, when set; wait setting, when set; wait 5 secs device returns 5 secs device returns to zero 3 ready to use. to zero 3 ready to use. Alarm settings revert to default after device power off. Alarm Priority: No Airway Adapter! 2 How to ZERO device NOTE! presence of ambient air [0% CO2] in the adapter is crucial for successful zero. Avoid breathing near device during zeroing process. 1 3 Clinical information kPa 2 4 EMMA should read 0.0 at power on. If it does NOT read 0.0 as shown 4 then insert NEW adapter. After inserting NEW adapter if device is NOT 0.0 but has offset 4 then the device needs re-zeroing. Insert new adapter Press Power On 1 & Alarm 2 buttons simultaneously for 10 secs, then device counts down 10 to 0. When device flashes ‘0’ release both buttons 3 ready to use. Adult/Paediatric Airway Adapter [deadspace 6 ml] How to Set Alarms 1 Replace batteries 4 5 6 0 Batteries 1 Dotted line = EtC02 reached/exceed 7kPa kPa Alarm Priority: Advisory! kPa Alarm Priority: Caution!!! kPa Normal C02 patient range 4.0 - 5.7 kPa Below 4.0 kPa is Hypocapnia Above 5.7 kPa is Hypercapnia EtC02 Inspiration Expiration Normal waveform Hyperventilation [decreasing C02] Hypoventilation [increasing C02] Alarm Priority: Advisory Clogged Adapter! 02392 469737 info@medacx.co.uk www.medacx.co.uk Alarm Priority: Advisory Replace batteries! Alarm Priority: Advisory Silenced! Alarm Priority: Advisory Slienced Alarm has Cleared! Apnea [loss of waveform] 142-CL-R1.0-MAY-16 MEDACX LIMITED • 1-3 ALEXANDER HOUSE • 60-62 STATION ROAD • HAYLING ISLAND • HAMPSHIRE • PO11 0EL](https://images.bioclinicalservices.com.au/80glns1iad1es8ljbik09qp6xp0v/320w/Masimo%20-%20142-CL-R1.0-May-16%20-%20EMMA%20MEDACX%20Waveform%20Capnograph%20Quick%20Guide%20May%202016%20-%202016-05.png)
Preview
Page 1
EMMA® Waveform Capnograph Controls & Displays
Power & Airway Adapters
Battery Cover Release
Alarm Mute
Airway Adapter
EtC02 Value a
kP
Battery Status
On Button
Respiratory Rate Alarm Status
EtC02 Waveform
Power On: Insert Airway Adapter 1 Press Power on button 2 device ready 3 Power Off: Automatic On removal of Adapter 1 or 2 mins after 2 NO Breath detected and Alarm silenced 4
How to Zero device Battery Cover Release
3
Infant Airway Adapter
Battery types: Duracell Industrial 2 x AAA LR03 or Energizer Ultimate 2x AAA Lithium L92
[reduced deadspace 1 ml]
5
6
Note! Trained medical professional must determine correct airway adapter for each patient application.
Close Battery Cover
Alarms & Status Displays
2
kPa
kPa
kPa
kPa
3
1
EtC02 Low Alarm [setting] EtC02 High Alarm [setting] Press & Hold Power On 1 Press & Hold Alarm Mute 2 button for 5 secs. for button for 5 secs. Display changes as shown. Display changes as shown. Press Power button to 1 Press Alarm button 2 decrease High alarm. to increase Low alarm. Press 2 to increase Press 1 to decrease setting, when set; wait setting, when set; wait 5 secs device returns 5 secs device returns to zero 3 ready to use. to zero 3 ready to use. Alarm settings revert to default after device power off.
Alarm Priority: No Airway Adapter!
2
How to ZERO device NOTE! presence of ambient air [0% CO2] in the adapter is crucial for successful zero. Avoid breathing near device during zeroing process.
1
3
Clinical information kPa
2
4
EMMA should read 0.0 at power on. If it does NOT read 0.0 as shown 4 then insert NEW adapter. After inserting NEW adapter if device is NOT 0.0 but has offset 4 then the device needs re-zeroing.
Insert new adapter Press Power On 1 & Alarm 2 buttons simultaneously for 10 secs, then device counts down 10 to 0. When device flashes ‘0’ release both buttons 3 ready to use.
Adult/Paediatric Airway Adapter [deadspace 6 ml]
How to Set Alarms
1
Replace batteries
4
5
6
0
Batteries 1
Dotted line = EtC02 reached/exceed 7kPa kPa
Alarm Priority: Advisory! kPa
Alarm Priority: Caution!!! kPa
Normal C02 patient range 4.0 - 5.7 kPa Below 4.0 kPa is Hypocapnia Above 5.7 kPa is Hypercapnia EtC02 Inspiration Expiration
Normal waveform
Hyperventilation [decreasing C02]
Hypoventilation [increasing C02]
Alarm Priority: Advisory Clogged Adapter!
02392 469737 info@medacx.co.uk www.medacx.co.uk
Alarm Priority: Advisory Replace batteries!
Alarm Priority: Advisory Silenced!
Alarm Priority: Advisory Slienced Alarm has Cleared!
Apnea [loss of waveform]
142-CL-R1.0-MAY-16
MEDACX LIMITED • 1-3 ALEXANDER HOUSE • 60-62 STATION ROAD • HAYLING ISLAND • HAMPSHIRE • PO11 0EL