Addendum Rad-5®/5v®: Alarms and Messages This addendum provides updates to the following:
Operator's Manual, Rad-5/5v Signal Extraction Pulse Oximeter® 33606/LAB-5724D 1209 and equivalent translations
Messages The device will indicate other data or system errors. Message conditions for the device follow: Display Message
Type
Solution
nO CbL
No cable connected
Connect cable
Inc CbL
Incompatible cable
Replace cable
rPL CbL
Defective or unrecognized cable
Replace cable
Temporarily blinking message: rPL CbL
Cable life expired
Replace cable as soon as possible
nO SEn
No sensor connected
Connect sensor
Inc SEn
Incompatible sensor
Replace sensor
rPL SEn
Unrecognized or defective sensor
Replace sensor
Temporarily blinking message: rPL SEn
Sensor life expired
Replace sensor as soon as possible
nO AdH
No adhesive sensor connected
Connect adhesive sensor
www.masimo.com
1
Masimo
Display Message
Type
Solution
Inc AdH
Incompatible adhesive sensor
Replace adhesive sensor
rPL AdH
Defective or unrecognized adhesive sensor
Replace adhesive sensor
Temporarily blinking message: rPL AdH
Adhesive sensor life expired
Replace adhesive sensor as soon as possible
rPL AdH
Defective or unrecognized adhesive sensor
Replace adhesive sensor
38043/LAB-9026A 1015