Service Manual
40 Pages
![Table of Contents Page 1 Overview ...1-1 1.1 About this Manual ...1-1 1.2 Warnings, Cautions and Notes ...1-1 2 Maintenance ...2-1 2.1 Introduction ...2-1 2.2 Cleaning...2-1 3 Battery Operation and Maintenance ...3-1 3.1 Overview ...3-1 3.2 Battery Charging...3-1 3.3 Replacing the Handheld Battery...3-2 3.4 Replacing the Docking Station Battery ...3-2 3.5 Replacing the Fuses ...3-3 4 Software ...4-1 4.1 Overview ...4-1 4.2 Software Download Procedure...4-1 5 Performance Verification ...5-1 5.1 Overview ...5-1 5.2 Power-On Self-Test ...5-1 5.3 Key Press Button Test ...5-1 5.4 Alarm Limit Test...5-1 5.5 Display Contrast Test on Monochrome only...5-2 5.6 Testing with Masimo SET Tester...5-2 5.7 %SpCO (Carboxyhemoglobin) Testing (If Applicable) ...5-3 5.8 %SpMET (Methemoglobin) Testing (If Applicable) ...5-3 5.9 PVI (Pleth Variability Index) Testing (If Applicable) ...5-4 5.10 Nurse Call Test ...5-4 5.11 Analog Output Test ...5-5 5.12 Serial Port ...5-5 5.13 Battery Test...5-6 6 Repair ...6-1 6.1 Safety Precautions:...6-1 6.2 General Procedures...6-1 6.3 Troubleshooting ...6-1 6.4 SatShare Diagnostics ...6-7 7 Repair and Returns...7-1 7.1 Repair Policy...7-1 7.2 Return Procedure ...7-1 8 Analog Output/Nurse Call Test...8-1 9 Docking Station Power Supply Power Measurement ...9-1 Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual iii](https://public.bioclinicalservices.com.au/library/representations/proxy/eyJfcmFpbHMiOnsiZGF0YSI6MTExNjgsInB1ciI6ImJsb2JfaWQifX0=--fb2433ea7f4f39e736a12b942206ae4ddb23ae53/eyJfcmFpbHMiOnsiZGF0YSI6eyJyZXNpemUiOiIzMjB4In0sInB1ciI6InZhcmlhdGlvbiJ9fQ==--4214356078d1f110e449cda512a9432562da3e70/radical-7_signal_extraction_pulse_co-oximeter_servi.pdf.png)
Preview
Page 1
Signal Extraction Pulse CO-Oximeter Monochrome and Color Display
Service Manual
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
The Radical-7 Pulse CO-Oximeter Service Manual is intended to provide the necessary information for proper servicing of all models of the Radical-7 Pulse Oximetry and CO-Oximetry systems. This manual is structured to support troubleshooting to the assembly or module level. This manual does not provide instructions for troubleshooting to the printed circuit board component level. There may be information provided in this manual that is not relevant for your system. Do not service the Radical-7 pulse oximeter without completely reading and understanding these instructions. Service training provided by Masimo is required prior to utilization of this manual. NOTICE: Purchase or possession of this device does not carry any express or implied license to use this device with replacement parts which would, alone or in combination with this device, fall within the scope of one of the patents relating to this device. CAUTION: Federal law (U.S.) restricts this device to sale by or on the order of a physician. Masimo Corporation 40 Parker Irvine, CA 92618 USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com
EU Authorized Representative for Masimo Corporation:
EC REP MDSS GmbH Schiffgraben 41 30175 Hannover, Germany Tel.: +49-511-62 62 86 30 Fax.: +49-511-62 62 86 33 MEDICAL ELECTRICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 606011/CAN/CSA C22.2 No. 601.1 Covered by one or more of the following U.S. Patents: RE38,492, RE38,476, 7,221,971, 7,215,986, 7,215,984, 7,186,966, 6,979,812, 6,861,639, 6,850,787, 6,826,419, 6,816,741, 6,745,060, 6,699,194, 6,684,090, 6,654,624, 6,650,917, 6,643,530, 6,606,511, 6,515,273, 6,501,975, 6,463,311, 6,430,525, 6,388,240, 6,360,114, 6,263,222, 6,236,872, 6,229,856, 6,157,850, 6,067,462, 6,011,986, 6,002,952, 5,919,134, 5,769,785, 5,758,644, 5,685,299, 5,632,272, 5,490,505, 5,482,036, international equivalents, or one or more of the patents referenced at www.masimo.com/patents.htm Products containing Satshare® feature are also covered by U.S. Patent 6,770,028. Other patents pending. © 2007 Masimo Corporation. LNOP, Masimo, Radical, the Radical logo, Rainbow, Satshare, SET, Signal IQ, and SpCO are federally registered trademarks of Masimo Corporation. FlashApp, PVI, Radical-7, RDS, SpMet and Signal Extraction Pulse CO-Oximeter are trademarks of Masimo Corporation.
ii
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
Table of Contents Page 1
Overview ...1-1 1.1 About this Manual ...1-1 1.2 Warnings, Cautions and Notes ...1-1 2 Maintenance ...2-1 2.1 Introduction ...2-1 2.2 Cleaning...2-1 3 Battery Operation and Maintenance ...3-1 3.1 Overview ...3-1 3.2 Battery Charging...3-1 3.3 Replacing the Handheld Battery...3-2 3.4 Replacing the Docking Station Battery ...3-2 3.5 Replacing the Fuses ...3-3 4 Software ...4-1 4.1 Overview ...4-1 4.2 Software Download Procedure...4-1 5 Performance Verification ...5-1 5.1 Overview ...5-1 5.2 Power-On Self-Test ...5-1 5.3 Key Press Button Test ...5-1 5.4 Alarm Limit Test...5-1 5.5 Display Contrast Test on Monochrome only...5-2 5.6 Testing with Masimo SET Tester...5-2 5.7 %SpCO (Carboxyhemoglobin) Testing (If Applicable) ...5-3 5.8 %SpMET (Methemoglobin) Testing (If Applicable) ...5-3 5.9 PVI (Pleth Variability Index) Testing (If Applicable) ...5-4 5.10 Nurse Call Test ...5-4 5.11 Analog Output Test ...5-5 5.12 Serial Port ...5-5 5.13 Battery Test...5-6 6 Repair ...6-1 6.1 Safety Precautions:...6-1 6.2 General Procedures...6-1 6.3 Troubleshooting ...6-1 6.4 SatShare Diagnostics ...6-7 7 Repair and Returns...7-1 7.1 Repair Policy...7-1 7.2 Return Procedure ...7-1 8 Analog Output/Nurse Call Test...8-1 9
Docking Station Power Supply Power Measurement ...9-1 Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
iii
10 Portable Functional Block Diagram ...10-1 11 Docking Station Functional Block Diagram ...11-1 12 Radical-7 Isolation Block Diagram ...12-1 13 Portable Wiring Interconnect...13-1 14 Docking Station Wiring Interconnect...14-1 15 Radical-7 Color Parts List ...15-1 16 Radical-7 Monochrome Parts List ...16-1 17 Docking Station Parts List (RDS-1, RDS-1B and RDS-3) ...17-1 18 RDS2 Parts List ...18-1
iv
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
1
Overview 1.1
About this Manual
This manual explains how to service the Radical-7 Pulse CO-Oximeter. Important safety information relating to general use of the Radical-7 Pulse CO-Oximeter appears in the Radical-7 CO-Oximeter Operator’s Manual. Other important safety information is located throughout this manual where appropriate. 1.2
Warnings, Cautions and Notes
Please read and follow any warnings, cautions and notes presented throughout this manual. An explanation of these labels is as follows: A warning is provided when actions may result in a serious outcome (i.e., injury, serious adverse affect, death) to the patient or user. Look for text in a gray shaded box. Sample of Warning: WARNING:
THIS IS A SAMPLE OF A WARNING STATEMENT.
A caution is given when any special care is to be exercised by the patient or user to avoid injury to the patient, damage to this device or damage to other property. Sample of Caution: CAUTION: This is a sample of a Caution statement. A note is provided when extra general information is applicable. Sample of Note: Note: This is a sample of a Note statement.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
1-1
2
Maintenance 2.1
Introduction
2.1.1
This chapter covers how to test the operation of the Radical7 and SatShare interface (if applicable), how to properly clean the Radical-7 pulse oximeter, how to recharge and replace the batteries, how to replace the fuses, and how to obtain service under normal operation. No internal adjustment or recalibration is required. Service should be done by qualified personnel only. Safety checks should be performed at regular intervals or in accordance with local and governmental regulations.
WARNING: ELECTRICAL SHOCK AND FLAMMABILITY HAZARD. BEFORE CLEANING THE OXIMETER, ALWAYS TURN THE UNIT OFF AND DISCONNECT THE POWER CORD FROM THE BACK OF THE UNIT.
2.2
Cleaning
2.2.1
To clean the display panel use a cotton swab moistened with 70% isopropyl alcohol and gently wipe the panel.
2.2.2
To clean the outer surface of the oximeter, use a soft cloth moistened with a 70% isopropyl alcohol. Do not allow liquids to enter the interior of the instrument.
2.2.3
To decontaminate unit, refer to Section 7.2.2.
CAUTIONS: •
Do not autoclave, pressure sterilize, or gas sterilize this oximeter.
•
Do not soak or immerse the monitor in any liquid.
•
Use the cleaning solution sparingly. Allow unit to air dry. Excessive solution can flow into the monitor and cause damage to internal components.
•
Do not touch, press, or rub the display panels with abrasive cleaning compounds, instruments, brushes, rough-surface materials, or bring them into contact with anything that could scratch the panel.
•
Do not use petroleum-based or acetone solutions, or other harsh solvents, to clean the oximeter. These substances attack the device’s materials and device failure can result.
2-1
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3
Battery Operation and Maintenance 3.1
Overview
3.1.1
The Radical-7 Handheld includes a 2.0 Amp-hr Nickel Metal Hydride (NiMH) battery. The Radical-7 Docking Station may include the optional 6.5 Amp/hr NiMH battery.
3.1.2
Before using the Radical-7 as a handheld or transport monitor, the Handheld battery and the optional Docking Station battery need to be fully charged.
3.2
Battery Charging
3.2.1
To charge the battery(s), attach the Handheld unit to the Docking Station.
3.2.2
Ensure that AC power is attached to the Docking Station.
Note: Some older units are configured with an AC power switch. Turn the AC power switch to the ‘I’ position.
3.2.3
The battery charging LED indicator(s) on the Docking Station will initially flash, then remain illuminated while the battery(s) is (are) charging. A prolonged flashing battery charging LED indicates that the internal battery temperature exceeds recommended operating conditions for proper battery charging. Proper battery charging will proceed when the temperature returns to recommended operating conditions.
3.2.4
The Handheld battery requires approximately 2 to 3 hours for charging. The optional Docking Station battery requires approximately 6 hours for charging.
3.2.5
When the battery charging LED indicators turn off, additional trickle charging may occur to complete charging. Although battery charging can occur while the Handheld is docked and powered on, most efficient charge times are achieved with the Handheld unit turned off. It is advisable to completely discharge and fully recharge the batteries periodically to maintain optimum battery health. Refer to the Battery Deep Discharge function as described in the Radical-7 Operators Manual, Section 4, Service. See
3.2.6
CAUTIONS: on next page.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3-1
CAUTIONS: •
If the Radical-7 has not been charged for 1 month or more, then recharge the battery prior to use.
•
At low battery warning, connect the Radical-7 to AC power to prevent interruption of power.
•
Additional information on the Radical-7 batteries may be found in the Radical-7 Operators Manual. 3.3
Replacing the Handheld Battery
3.3.1
Turn the Radical-7 Handheld off and remove the patient cable connection. Detach the Radical-7 Handheld from the Docking Station (if docked).
3.3.2
Place the Handheld face down on a soft surface such that no cosmetic damage occurs to the unit.
3.3.3
Loosen the closure screw on the battery, or on color models, the battery compartment door, and remove the battery.
3.3.4
Position the replacement battery into the Handheld, bottom first. On color models, the battery needs to be plugged in / unplugged.
3.3.5
Replace the enclosure (if applicable) and tighten the battery screw until it is secure. If there is a grasp loop, make sure it is pointing towards the top of the Handheld and lays flat into the plastic housing.
3.3.6
Place Handheld into Docking Station. Turn on power and if necessary, charge battery according to Section 9 of the Radical-7 Operator’s Manual: Battery Operation and Maintenance.
3.4
Replacing the Docking Station Battery
CAUTION: Do not over-tighten screws. Tighten the screws throughout the Radical-7 Handheld and Docking Station to 80 in.- oz.
Note: Docking Station Battery can only be installed in RDS-1B or RDS3B type replacement units.
Note: It will be necessary to perform safety testing after replacement.
3-2
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3.4.1
With the AC power cord disconnected, remove the Handheld from the Docking Station.
3.4.2
Lay the Docking Station on its side on a soft surface, with the screw holes facing up
3.4.3
Remove the five (5) screws from the unit using a #1 Phillips screw driver.
3.4.4
Carefully turn the Docking Station over to allow the screws to fall out of the unit in a manner that allows the technician to catch the screws.
3.4.5
Lay the Docking Station on its feet and remove the top plastic cover.
3.4.6
For units with a battery installed, disconnect the battery connector (J21) from the Docking Station system board. Remove the screw that holds the Docking Station System board in place. Lift this board to allow access to the Battery. Remove the battery.
3.4.7
Carefully install the replacement battery into its compartment.
3.4.8
Reinstall the screw to secure the Docking Station system board. Attach the battery connector.
3.4.9
Carefully install the top plastic cover. Make sure no internal cables are pinched between top and bottom case.
3.4.10 Turn the unit over, and re-install the screws. 3.4.11 Attach the Handheld onto the Docking Station. 3.4.12 If necessary, charge the battery before using the Docking Station according to Section 3.2 of this manual. CAUTION: Follow local governing guidelines for proper disposal of internal batteries. Do not incinerate. 3.5
Replacing the Fuses
3.5.1
Disconnect unit from AC power.
3.5.2
Use a small flat-blade screwdriver and gently pry loose the fuse holder from the power entry module.
3.5.3
Note how the fuses are placed in the fuse holder for installation of the new fuses.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
3-3
3.5.4
To remove the fuses from the fuse holder, use the edge of the screwdriver blade to pry against the bottom of the metal portion of the fuse.
3.5.5
Place fuses as needed (1 Amp, Metric, fast acting, 5x20mm, 250V) in the fuse holder, observing the proper orientation.
3.5.6
Slide the fuse holder back into the power entry module and press firmly to make sure it is firmly inserted.
3.5.7
The unit is ready to be reconnected to AC power.
Note: If the fuses blow after replacement, troubleshoot the unit.
WARNING: FIRE HAZARD: TO PROTECT AGAINST FIRE HAZARD, REPLACE ONLY WITH FUSES OF THE SAME TYPE, CURRENT RATING, AND VOLTAGE RATING.
3-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4
Software 4.1
Overview
4.1.1
The Radical-7 product is configured to allow software downloads directly from a PC. The Radical-7 product has three different boards that utilize software that may require upgrade. The Pulse Oximeter Board, Handheld System Board and Docking Station System Board all operate with software that is independent of each other. The software must be compatible between the boards for the Radical-7 product to operate properly. The boards communicate with each other through on board connections - in the case of the Oximeter Board to Handheld System Board. For the Handheld to Docking Station communication, this is accomplished through a Contact Pin/Interface Cable.
4.1.2
When software upgrades are affected, it is imperative that FRM-1312 is completed and returned to Masimo to ensure each unit is tracked with the software version of each unit. This form can be found in this manual, in the contents of the software disk and can also be acquired from Masimo if necessary by contacting Masimo using the information in Section 7.2.4.
Note: The Pulse Oximeter Board (MX-1) is not upgradeable using the PC_FlashApp Program. Only the Docking Station and Portable System Board are upgradeable in this fashion. The upgrade of the MX board requires the use of a special field upgrade tool.
4.2
Software Download Procedure
Note: No software is installed in the RDS-2 and the unit does not require programming.
4.2.1
On the PC, open PC_FlashApp.exe from your hard drive. Masimo software can be obtained by Authorized Service providers by calling Masimo Technical Support Within the United States at (800) 326-4890, option #2, or From Outside the United States at country code (+1) 949-297-7498 option #2.
4.2.2
Set the default “Target”, “Port” and “Baud” parameters as shown in the table below.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4-1
Programming Software PC_FlashApp PC_FlashApp
1
Target
Port
Baud
Portable Docking Station
COM1 COM1
57600 57600
4.2.3
For programming the Radical-7 Handheld System Board, perform the following steps:
4.2.4
Connect a Docking Station to AC outlet and turn the AC Switch to ON (if applicable).
4.2.5
Connect a straight through serial cable between the PC and the Docking Station.
4.2.6
Install the Radical-7 Handheld to the Docking Station.
4.2.7
Turn off the Handheld, if it is on.
4.2.8
Start the FlashApp program. Click Browse and select the proper Image file to be loaded. It will be listed under the monitor’s PSA subdirectory.
•
If loading on a Radical-7C “Color” the proper file will be named PPSA.ROM. If loading on a Radical-7 “Monochrome” the proper file will be named PPSAMX.ROM. Select the proper Port (usually COM1), and Select Portable (Blue) as the Target.
• 4.2.9
4.2.10 Press the “Green” Program button. A ‘status message’ of “waiting for docking station” will be displayed. 4.2.11 Turn on the Radical-7 Handheld. 4.2.12 The green Program button will turn off, and the red Stop button will illuminate. Wait until the green button turns on again indicating a successful transfer. 4.2.13 If the message on the Status window is “Complete – Success”, continue with the next step. If programming is not successful, go back to 4.2.7 and repeat. If this cycle is repeated 3 times without success, troubleshooting is required. 4.2.14 Turn off the Handheld.
1
If a unit does not program with 57,600, set baud rate to 9,600.
4-2
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4.2.15 To program Radical-7 Docking Station, perform the following steps: 4.2.16 Connect a Docking Station to AC outlet and turn the switch to ON position (if applicable). 4.2.17 Connect a serial cable between computer and the Docking Station. 4.2.18 Install a known good Radical-7 Handheld to the Docking Station. 4.2.19 Turn off the Handheld, if it is on. 4.2.20 Start the FlashApp program. Click Browse and select the proper Image file to be loaded. It will usually be listed under a SDSSA subdirectory. 4.2.21 Select the proper Port (usually COM1), and Select Docking Station as the Target. 4.2.22 Press the “Green” Program button. A ‘status message’ of “waiting for docking station” will be displayed. 4.2.23 Turn on the Handheld. 4.2.24 The green Program button will turn off, and the red Stop button will illuminate. Wait until the green button turns on again indicating a successful transfer. . 4.2.25 If the message on the Status window is “Complete – Success”, continue with next step. If the message is “Complete – Fail”, go back to 4.2.19 and repeat. If this cycle is repeated 3 times without success, troubleshooting is required. 4.2.26 To check software version numbers, perform the following steps: 4.2.27 Turn off the Handheld. 4.2.28 Turn on the Handheld. 4.2.29 Press the Menu key. 4.2.30 Press the Down key and select “About”. 4.2.31 Verify that the Handheld (Handheld, Masimo SET) and Docking Station (D-Station) software version(s) are the intended version. The Handheld must be docked to the Docking Station in order to see the version of the Docking Station.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
4-3
4.2.32 Record software version in FRM-1312 and notify Masimo as per Section 4.1.2.
4-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5 Performance Verification Note: Under normal operation, no internal adjustment or recalibration is required.
5.1
Overview
5.1.1
To test the performance of the Radical-7 pulse oximeter following repairs or during routine maintenance, follow the procedure outlined in this section. If the Radical-7 fails any of the described tests, discontinue its use and correct the problem before returning the unit back to the user.
5.1.2
Before performing the following tests place the Radical-7 Handheld into the Docking Station, connect the Radical-7 to AC power and fully charge the Radical-7 Handheld battery (the AC mains switch must be on). Disconnect any patient cables or pulse oximetry sensors, as well as SatShare, serial or Analog Output/Nurse Call cables from the instrument.
5.2
Power-On Self-Test
5.2.1
Connect the unit to AC power and verify that the AC Power Indicator LED is illuminated. If it is not, verify that the AC mains switch is on.
5.2.2
Turn the unit on by depressing the Power/Standby Button. Within 5 seconds all available LEDs are momentarily illuminated, a 1-second beep tone sounds, and the Masimo SET logo is displayed.
5.2.3
The Docking Indicator LED, AC Power Indicator LED, Handheld Battery Charge LED and Docking Station Charge LED (if Docking Station is equipped with optional battery) are illuminated and the Radical-7 begins normal operation.
5.3
Key Press Button Test
5.3.1
5.4
With the exception of the Power/Standby Button, press each soft key button and verify that the Radical-7 acknowledges each key-press with an audible beep tone or by indicating an associated change on the display.
Alarm Limit Test
5.4.1
With the monitor turned on, select the Menu Access key and enter the Alarm menu. Change the desired parameter, below, to a value two points below the currently selected value, and accept the change. Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-1
5.4.2
Verify that the newly set parameter is shown on the Saturation Alarm Limit Display, next to the SpO2 or pulse rate measurement display.
5.4.3
Return the Alarm parameter to its original setting.
5.4.4
Perform steps 5.4.1 through 5.4.3 with the High Saturation Alarm parameter.
5.4.5
Perform steps 5.4.1 through 5.4.3 with the Low Saturation Alarm parameter.
5.4.6
Perform steps 5.4.1 through 5.4.3 with the High Pulse Rate Alarm parameter.
5.4.7
Perform steps 5.4.1 through 5.4.3 with the Low Pulse Rate Alarm parameter.
5.4.8
Reset the alarm limits again to the original settings.
5.5
Display Contrast Test on Monochrome only
5.5.1
With the monitor turned on, select the Menu key and enter the Display menu. Change the Contrast parameter by scrolling through the contrast settings.
5.5.2
Return the Contrast setting to the original value, or a value that allows maximum viewing contrast.
5.5.3
Exit the Menu system and press and hold down the Backlight/Contrast button for several seconds. The display will scroll again through all the contrast settings.
5.5.4
Release the Backlight/Contrast button again when the display shows maximum viewing contrast.
5.6
Testing with Masimo SET Tester
5.6.1
Turn the Radical-7 off and then on again.
5.6.2
Set the alarm limits to:
5.6.3
5-2
% SpO2
Pulse Rate
High
100
140
Low
90
50
Connect the Masimo tester to the Radical-7. A patient cable LNOP to Red PC cable adapter (4', 8' or 12') may be required to connect a modular tester to the Radical-7. Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5.6.4
Verify that within 20 seconds a plethysmographic and a Signal IQ waveform displays.
5.6.5
Verify that the SpO2 measurement is between 78% and 84%. (with tester values of 81% +/- 3%)
5.6.6
Verify that the pulse rate measurement is between 59 bpm and 63 bpm (with tester values of 61 +/- 1 bpm).
5.6.7
Verify that an audible alarm occurs and that the SpO2 measurement and the low SpO2 alarm are flashing
5.6.8
Press the Alarm Silence button once and verify that the alarm is silenced.
5.6.9
Wait 120 seconds and verify that the alarm silence times out and the audible alarm is activated again. Disable the Alarm before continuing.
5.6.10 Press the Increase Loudness button several times and verify that the loudness of the pulse beep tone increases. 5.6.11 Press the Decrease Loudness button and verify that the loudness of the pulse beep tone decreases. Press the Decrease Loudness button and verify that the loudness of the pulse beep tone can be turned off. 5.7 %SpCO (Carboxyhemoglobin) Testing (If Applicable) Note: This is performed as a functional test only. 5.7.1
Connect a Masimo SET Rainbow sensor to the Radical-7 Handheld.
5.7.2
Insert the non-dominate hand’s ring finger into the finger sensor.
5.7.3
Under %SpCO, a healthy, non-smoking individual would have a typical reading between 0 – 5 on the display. (These numbers are for reference only and should not be used for diagnostic purposes)
5.8 %SpMET (Methemoglobin) Testing (If Applicable) Note: This is performed as a functional test only. 5.8.1
Connect a Masimo SET Rainbow sensor to the Radical-7 Handheld.
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-3
5.8.2
Insert the non-dominant hand’s ring finger into the finger sensor.
5.8.3
Under %SpMET, a healthy, non-smoking individual would have a typical reading between 0-2% on the display. (These numbers are for reference only and should not be used for diagnostic purposes)
5.9 PVI (Pleth Variability Index) Testing (If Applicable) Note: This is performed as a functional test only. 5.9.1
Connect a Red Masimo SET Rainbow sensor to the Radical7 Handheld.
5.9.2
Insert the non-dominant hand’s ring finger into the finger sensor.
5.9.3
PVI will fluctuate depending on motion, perfusion, physiology and other factors, so there is no ‘normal’ value. If a number is displayed, then the PVI feature is functioning properly.
5.10 Nurse Call Test Note: Refer to Section 8 in this manual for connector pin outs noted in Sections 5.10 5.10.1 Disconnect the patient cable or the Masimo SET Tester from the Radical-7 and turn the instrument on. Ensure that there are no audible alarms and that the audible alarms are not silenced. 5.10.2 Connect the common lead of a digital multimeter to pin 12 (Nurse Call - Common) of the Analog Output/Nurse Call Connector (see Service Manual, Troubleshooting, Section 8, Analog Output/Nurse Call Test) on the Radical-7. Connect the positive lead of the multimeter to pin 6 (Nurse Call Normally Open) of the Analog Output/Nurse Call Connector and measure that the resistance is greater than 1 MΩ (open circuit). 5.10.3 Trigger an alarm on the monitor (e.g. by disconnecting a sensor after it was measuring data) and verify the resistance is < 20Ω.
5-4
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5.11 Analog Output Test 5.11.1 Disconnect all patient cables and sensors from the Radical7. Turn the Radical-7 off and then on again. 5.11.2 Connect the common lead of a multimeter to the pin 2 (Ground) of the Analog Output/Nurse Call Connector. Connect the positive lead of the multimeter to pin 9 (Analog 1) of Analog Output/Nurse Call Connector. 5.11.3 Enter the menu system and set the "Output", "Analog 1" to "0v Output". Verify that the voltmeter measures a voltage of approximately between –10mV and 100mV. 5.11.4 Set “Analog 1” to 1v Output”. Verify that the voltmeter measures a voltage between 0.90V to 1.10V. 5.11.5 Reset “Analog1” to “Spo2 0 – 100%”. 5.11.6 Repeat steps 5.8.3 and 5.8.4 with the positive lead of the multimeter connected to pin 15 (Analog2) and set “Analog2” to “0V Output” and “1V Output”. 5.11.7 Reset “Analog2” to “Pulse Rate”. 5.12
Serial Port
5.12.1 The serial port outputs ASCII data using standard RS232 protocol via a straight through serial cable. The Handheld must be docked with AC power applied. 5.12.2 Validation of the port can be done by connecting a serial printer or using a PC with Hyper Terminal and the following setup: 5.12.3 Communication Parameters baud rate data bits stop bit parity hand shaking
9600 bps 8 1 none none
5.12.4 With the Radical-7 connected as per the above setup, attach a patient cable and sensor to the Radical-7. From the Output menu, select ASCII 1 and the printer (or PC monitor) will print in the following format every second.
05/15/07 08:12:21 SN=0000070986 SPO2=---% BPM=--- PI=--.--% SPCO=--.-% SPMET=--.-% DESAT=-- PIDELTA=+-- PVI=--ALARM=0000 EXC=000000
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual
5-5
5.13 Battery Test 5.13.1 Fully charge the Radical-7 Handheld Battery by placing the Handheld into the Docking Station and connecting the AC power (the AC mains switch must be on). 5.13.2 Verify that the green Handheld Battery Indicator LED is illuminated. 5.13.3 When the Radical-7 is fully charged the green Handheld Battery Indicator turns off. 5.13.4 Turn the Radical-7 on and verify that the Handheld Battery indicator shows a full charge. 5.13.5 Fully charge the Radical-7 Docking Station Battery by placing the Handheld into the Docking Station and connecting the AC power (the AC mains switch must be on). 5.13.6 Verify that the green Handheld Battery and blue Connection Indicator LEDs, on the Docking Station, are illuminated. 5.13.7 When the Radical-7 is fully charged the green Docking Station Battery Indicator turns off. 5.13.8 Turn the Radical-7 on and verify that the Docking Station Battery indicator shows a full charge.
5-6
Radical-7 Signal Extraction Pulse CO-Oximeter Service Manual